MODEL: M2XT2 M2XT2 SPECIFICATIONS ES-5062 Rev.16 Page 1/5 http://www.m-system.co.jp/ Super-mini Signal Conditioners Mini-M Series THERMOCOUPLE TRANSMITTER (PC programmable) Functions & Features
- Accepts direct input from a thermocouple and provides a linearized process signal
- PC programmable
- Wide selection of thermocouples
- User’s temperature table can be used
- Leadwire resistance compensation Typical Applications
- Small burnout sensing current enables long distance transmission with minimum offset drifts
- Electric furnace (isolation ensured even when 200 V AC power for heater leaks through furnace wall)
- No burnout type can connect to a single T/C in parallel with a recorder
- Ideal for quick spare part 23 (.91) (2.99) 124 (4.88) Model mm (inch) MODEL: M2XT2–[1][2]-[3][4]
ORDERING INFORMATION
- Code number: M2XT2-[1][2]-[3][4] Specify a code from below for each [1] through [4]. (e.g. M2XT2-2Z1-M2/BL/CE/Q) K thermocouple setting will be used if the input code is not specified.
- Temperature range (e.g. 0 – 800°C)
- Output range (e.g. 4 – 20 mA DC)
- Specify the specification for option code /Q (e.g. /C01/S01) [1] INPUT THERMOCOUPLE 1: (PR) 2: K (CA) 3: E (CRC) 4: J (IC) 5: T (CC) 6: B (RH) 7: R 8: S 9: C (WRe 5-26) N: N U: U L: L P: P (Platinel II) 0: Specify (Configurator software is used to change the input type and precise range.) [2] OUTPUT Current Z1: Range 0 – 20 mA DC Voltage V1: Range -2.5 – +2.5 V DC V2: Range -10 – +10 V DC (Configurator software is used to change output over the described range of the selected suffix code. For changing out of this range, set the Output Range Selectors inside the unit before software adjustment.) [3] POWER INPUT AC Power M2: 100 – 240 V AC (Operational voltage range 85 – 264 V, 47 – 66 Hz) (90 – 264 V for UL) DC Power R: 24 V DC (Operational voltage range 24 V ±10 %, ripple 10 %p-p max.) P: 110 V DC (Operational voltage range 85 – 150 V, ripple 10 %p-p max.) (110 V ±10 % for UL) [4] OPTIONS (multiple selections) Burnout blank: Upscale burnout /BL: Downscale burnout (UL not available) /BN: No burnout (UL not available) Standards & Approvals (must be specified) /N: Without CE or UL /CE: CE marking /UL: UL approval, CE marking Other Options blank: none /Q: Option other than the above (specify the specification)
MODEL: M2XT2 M2XT2 SPECIFICATIONS ES-5062 Rev.16 Page 2/5 http://www.m-system.co.jp/ SPECIFICATIONS OF OPTION: Q (multiple selections) COATING (For the detail, refer to M-System's web site.) /C01: Silicone coating /C02: Polyurethane coating /C03: Rubber coating (UL not available) TERMINAL SCREW MATERIAL /S01: Stainless steel (UL not available)
RELATED PRODUCTS
- PC configurator software (model: JXCON) Downloadable at M-System’s web site. A dedicated cable is required to connect the module to the PC. Please refer to the internet software download site or the users manual for the PC configurator for applicable cable types. GENERAL SPECIFICATIONS Construction: Plug-in Connection: M3 screw terminals (torque 0.8 N·m) Screw terminal: Chromated steel (standard) or stainless steel Housing material: Flame-resistant resin (black) Isolation: Input to output to power Overrange output: Approx. -15 to +115 % (Negative current output is not provided.) Manual zero/span adjustments: See Front View. Programming: Downloaded from PC; T/C type, temp. range, output range, zero and span, simulating output, etc. Burnout protection: Upscale standard; downscale or no burnout optional by programming (Set to "no burnout" when the unit is connected with another instrument in parallel.) Linearization: Standard Cold junction compensation: CJC sensor attached to the input terminals Status indicator LED: Blinking patterns indicate different operating status of the transmitter. Configurator connection: 2.5 dia. miniature jack; RS-232-C level INPUT SPECIFICATIONS Input resistance: 1 MΩ min. Burnout sensing: 45 nA ±10 % If not specified, the input range is shown below.
1 PR: 0–1600°C
2 K: 0–1000°C
3 E: 0–500°C
4 J: 0–500°C
5 T: 0–300°C
6 B: 0–1800°C
7 R: 0–1600°C
8 S: 0–1600°C
9 C: 0–2000°C
N N: 0–1000°C U U: 0–300°C L L: 0–500°C P P: 0–1200°C OUTPUT SPECIFICATIONS ■ DC Current Operational range: 0 – 24 mA DC Output range: 0 – 20 mA DC Minimum span: 1 mA Offset: Lower range can be any specific value within the output range provided that the minimum span is maintained. Load resistance: Output drive 15 V max. (e.g. 4 – 20 mA: 750 Ω [15 V ÷ 20 mA]) If not specified, the output range is 4 – 20 mA DC. ■ DC Voltage Code V1 (narrow spans) Operational range: -3 – +3 V DC Output range: -2.5 – +2.5 V DC Minimum span: 250 mV Code V2 (wide spans) Operational range: -11.5 – +11.5 V DC Output range: -10 – +10 V DC Minimum span: 1 V Offset: Lower range can be any specific value within the output range provided that the minimum span is maintained. Load resistance: Output drive 1 mA max. If not specified, the output range is shown below. V1: 0 – 1 V DC V2: 1 – 5 V DC INSTALLATION Power Consumption
- AC: Approx. 3 VA at 100 V Approx. 4 VA at 200 V Approx. 5 VA at 264 V
- DC: Approx. 2 W Operating temperature: -30 to +60°C (-22 to +140°F) Operating humidity: 30 to 90 %RH (non-condensing) Mounting: Surface or DIN rail Weight: 120 g (0.26 lb)
MODEL: M2XT2 M2XT2 SPECIFICATIONS ES-5062 Rev.16 Page 3/5 http://www.m-system.co.jp/ PERFORMANCE Accuracy: [Table 1 Accuracy value + Cold Junction Compensation Error 0.5°C (0.9°F)] or ±0.1 % of span, whichever is greater. Add 0.2 % when the output span equals 1/10 or narrower of the max. span. Cold junction compensation error: ±0.5°C at 25 ±10°C ±0.9°F at 77 ±18°F Temp. coefficient: ±0.015 %/°C (±0.008 %/°F) of max. span at -5 to +55°C [23 to 131°F] Response time: ≤ 0.9 sec. (0 – 90 %) Burnout response: ≤ 10 sec. Line voltage effect: ±0.1 % over voltage range Insulation resistance: ≥ 100 MΩ with 500 V DC Dielectric strength: 2000 V AC @1 minute (input to output to power to ground) CALCULATION EXAMPLES OF OVERALL ACCURACY [Example 1] Input: K thermocouple, 0 – 1000°C, Output: 4 – 20 mA DC 1) Absolute value accuracy (Table 1): 0.25°C 2) CJC error (0.5°C) added: 0.75°C 5) Output span 16 mA (= 20 mA – 4 mA) Max. span 20 mA Output span ≥ 1/10 of max. span ➠ No need of adding 0.2 %. 6) Overall accuracy equals ±0.1 % of span [Example 2] Input: K thermocouple, 50 – 150°C, Output: 2.0 – 2.5 V DC 1) Absolute value accuracy (Table 1): 0.25°C 2) CJC error (0.5°C) added: 0.75°C 5) Output span 0.5 V (= 2.5 V – 2.0 V) Max. span 5 V Output span ≤ 1/10 of max. span ➠ Add 0.2 %. 6) Overall accuracy equals ±0.95 % of span (0.75 + 0.2) Table 1. (PR) K (CA) E (CRC) J (IC) T (CC) B (RH) R S C ( WRe 5-26 ) N U L P ( Platinel II) 0 to 1760 -270 to +1370 -270 to +1000 -210 to +1200 -270 to +400 0 to 1820 -50 to +1760 -50 to +1760 0 to 2315 -270 to +1300 -200 to +600 -200 to +900 0 to 1395 ±1 .00 ±0.25 ±0.20 ±0.25 ±0.25 ±0.75 ±0.50 ±0.50 ±0.80 ±0.30 ±0.20 ±0.25 ±0.25 0 to 1760 -150 to +1370 -170 to +1000 -180 to +1200 -170 to +400 400 to 1760 200 to 1760 0 to 1760 0 to 2315 -130 to +1300 -200 to +600 -200 to +900 0 to 1395 THERMO- COUPLE MAXIMUM RANGE CONFORMANCE RANGEACCURACY (PR) K (CA) E (CRC) J (IC) T (CC) B (RH) R S C ( WRe 5-26 ) N U L P ( Platinel II) 32 to 3200 -454 to +2498 -454 to +1832 -346 to +2192 -454 to +752 32 to 3308 -58 to +3200 -58 to +3200 32 to 4199 -454 to +2372 -328 to +1112 -328 to +1652 32 to 2543 ±1 .80 ±0.45 ±0.36 ±0.45 ±0.45 ±1 .35 ±0.90 ±0.90 ±1 .44 ±0.54 ±0.36 ±0.45 ±0.45 32 to 3200 -238 to +2498 -274 to +1832 -292 to +2192 -274 to +752 752 to 3200 392 to 3200 32 to 3200 32 to 4199 -202 to +2372 -328 to +1112 -328 to +1652 32 to 2543 MAXIMUM RANGE ACCURACY THERMO- COUPLE MIN. SPAN MIN. SPAN CONFORMANCE RANGE STANDARDS & APPROVALS EU conformity: EMC Directive EMI EN 61000-6-4 EMS EN 61000-6-2 Low Voltage Directive EN 61010-1 Installation Category II Pollution Degree 2 Input or output to power: Reinforced insulation (300 V) Input to output: Basic insulation (300 V) RoHS Directive EN 50581 Approval: UL/C-UL nonincendive Class I, Division 2, Groups A, B, C, and D (ANSI/ISA-12.12.01, CAN/CSA-C22.2 No.213) UL/C-UL general safety requirements (UL 61010-1, CAN/CSA-C22.2 No.61010-1)
MODEL: M2XT2 M2XT2 SPECIFICATIONS ES-5062 Rev.16 Page 4/5 http://www.m-system.co.jp/ EXTERNAL VIEW UP/DOWN Switches The front cover cannot be turned open by 180 deg. when there is no extra space between units. Configurator Jack ■ FRONT VIEW (with cover open)Status Indicator LED Zero/Span Selector Top: SPAN Middle: OFF Bottom: ZERO ■ LEFT SIDE VIEW (with cover removed) Output Range Selectors [e.g.] Output code V1 O N 2 O SW2 ATT SW3 SW6 SW4 SW7 IV I V Refer to the instruction manual for detailed procedures. EXTERNAL DIMENSIONS unit: mm (inch) 21.5 (.85) 70.5 (2.78) 6 (.23) 59 (2.32) DIN RAIL 35mm wide [4 (.16)] 8–M3 SCREW 72 (2.83) 23 (.91) 2–4.2x5 (.17x.20) MTG HOLE 6 (.24) deep 116.7 (4.59)
- When mounting, no extra space is needed between units. 14 13 4 (.16) TERMINAL ASSIGNMENTS unit: mm (inch) 912 CJC SENSOR (model: CJM) 14 (.55) 1 14 redblack
MODEL: M2XT2 M2XT2 SPECIFICATIONS ES-5062 Rev.16 Page 5/5 http://www.m-system.co.jp/ SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM Isolation Base Socket Low Drift Amplifier & Burnout Output Driver OUTPUT Digital Computation ZERO/SPAN SELECTOR INPUT RANGE SELECTOR UP/DOWN SWITCHES CONFIGURATOR JACK U(+) V(–) POWER STATUS T / C CJC SENSOR metal leg model: CJM comp. leadwire red black OUTPUT RANGE SELECTORS Specifications are subject to change without notice.