F or medical electric equipment Inter nal dual fuses Lo w leakage current High po wer & peak power (option) Small and compact PCB constr uction Built-in inrush current, overcurrent and overvoltage protection circuits Har monic attenuator (Complies with IEC61000-3-2 class A) Univ ersal input (AC85-264V) P ower factor correction Saf ety agency approvals ANSI/AAMI ES60601, EN60601-1 3rd EMI Complies with FCC-B, CISPR22-B, EN55011-B, EN55022-B, VCCI-B 5-y ear warranty AC-DC Power Supplies Medical Type Power Factor Correction World wide Cost Effective Rugged PCB type Safety Approvals EMI Inrush current limiting OCP OVPMedical electric equipment CE marking Lo w Voltage Directive RoHS Directiv e EMS Compliance : EN61204-3, EN61000-6-2 IEC60601-1-2 (2014), EN60601-1-2 (2015) EN61000-4-2 EN61000-4-3 EN61000-4-4 EN61000-4-5 EN61000-4-6 EN61000-4-8 EN61000-4-11 June 29, 2020

AC-DC Power Supplies Medical Type LM A 100 F -O -O

Ordering information

R 1 2 3 4 5 6

1 Series name

2 Single output

3 Output wattage

4 Universal input

5 Output voltage

6 Optional *1

C : with Coating G : Low leakage current H : with the function to be acceptable to output peak current J1 : VH(J.S.T.)connector type R : with Remote ON/OFF R2 : with Remote ON/OFF S : with Chassis SN : with Chassis & cover P : Setting in the overcurrent protection ratingThis power supply is manufactured by SMD technology. The stress to P .C.B like twisting or bending causes the defect of the unit, so handle the unit with care. *Make sure necessary tests will be carried out on your end equipment with the power supply installed in accordance with any required EMC/EMI regulations. MODEL LMA100F-24-Y LMA100F-24-HY MAX OUTPUT WATTAGE[W] 103.2 103.2 (206.4) *2 DC OUTPUT 24V 4.3A 24V 4.3A (8.6A) *2 SPECIFICATIONS MODEL LMA100F-24-Y LMA100F-24-HY INPUT VOLTAGE[V] AC85 - 264 1f (Refer to “Derating” ,Instruction Manual 1 and 3) CURRENT[A] ACIN 100V 1.4typ (Io=100%) ACIN 200V 0.7typ (Io=100%) FREQUENCY[Hz] 50 / 60 (47 - 63) EFFICIENCY[%] ACIN 100V 84.0typ (Io=100%) 84.0typ (Io=100%) ACIN 200V 86.0typ (Io=100%) 86.0typ (Io=100%) POWER FACTOR ACIN 100V 0.99typ (Io=100%) ACIN 200V 0.95typ (Io=100%) INRUSH CURRENT[A] ACIN 100V 15typ (Io=100%) (At cold start) (Ta=25C) ACIN 200V 30typ (Io=100%) (At cold start) (Ta=25C) LEAKAGE CURRENT[mA] 0.10 / 0.25max (ACIN 100V / 240V 60Hz, Io=100%, According to IEC60601-1) OUTPUT VOLTAGE[V] 24 24 CURRENT[A] 4.3 4.3 (Peak 8.6) *2 LINE REGULATION[mV] *7 96max 96max LOAD REGULATION[mV] *7 150max 150max RIPPLE[mVp-p] *3 0 to +50C 120max 120max -10 - 0C 160max 160max RIPPLE NOISE[mVp-p]*3 0 to +50C 150max 150max -10 - 0C 180max 180max TEMPERATURE REGULATION[mV] 0 to +50C 240max 240max -10 to +50C 290max 290max DRIFT[mV] *4 96max 96max START-UP TIME[ms] 350typ (ACIN 100V, Io=100%) HOLD-UP TIME[ms] 20typ (ACIN 100V, Io=100%) OUTPUT VOLTAGE ADJUSTMENT RANGE[V] 19.20 to 27.50 19.20 to 27.50 OUTPUT VOLTAGE SETTING[V] 24.00 to 24.96 24.00 to 24.96 PROTECTION CIRCUIT AND OTHERS OVERCURRENT PROTECTION Works over 105% of rating (works over 101% of peak current at option -H) and recovers automatically OVERVOLTAGE PROTECTION[V] 27.60 to 33.60 27.60 to 33.60 OPERATING INDICATION Not provided REMOTE SENSING Not provided REMOTE ON/OFF Option (Required external power source.) ISOLATION INPUT-OUTPUT.RC *6 AC4,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature) 2MOOP INPUT-FG AC2,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature) 1MOOP OUTPUT.RC-FG *6 AC500V 1minute, Cutoff current = 25mA, DC500V 50MW min (At Room Temperature) OUTPUT-RC *6 AC100V 1minute, Cutoff current = 25mA, DC100V 10MW min (At Room Temperature) ENVIRONMENT OPERATING TEMP.,HUMID.AND ALTITUDE *5 -10 to +70C, 20 - 90%RH (Non condensing), (Refer to “Derating”, Instruction Manual 3) 3,000m (10,000feet) max STORAGE TEMP.,HUMID.AND ALTITUDE -20 to +75C, 20 - 90%RH (Non condensing), 9,000m (30,000feet) max VIBRATION 10 - 55Hz, 19.6m/s2 (2G), 3minutes period, 60minutes each along X, Y and Z axis IMPACT 196.1m/s2 (20G), 11ms, once each X, Y and Z axis SAFETY AND NOISE REGULATIONS AGENCY APPROVALS (AT ONIY AC input) ANSI/AAMI ES60601-1, EN60601-1 3rd, Complies with IEC60601-1-2 4th Ed. CONDUCTED NOISE Complies with FCC-B, VCCI-B, CISPR22-B, EN55011-B, EN55022-B HARMONIC ATTENUATOR Complies with IEC61000-3-2 (Class A) *8 OTHERS CASE SIZE/WEIGHT 62X33X155mm [2.44X1.30X6.10 inches] (WXHXD) / 290g max (with chassis & cover : 470g max) COOLING METHOD Convection (Refer to “Derating”, Instruction Manual 3) *5 *1 Specification is changed at option, ref er to Instruction Manual. P eak loading for 10sec. And Duty 40% max. ( ) means peak current. There is a possibility that an internal device is damaged when the specification is exceeded. This is the v alue that measured on measuring board with capacitor of 22mF at 150mm from output terminal. Measured b y 20MHz oscilloscope or Ripple-Noise meter (Equivalent to KEISOKU-GIKEN: RM103). *4 Dr ift is the change in DC output for an eight hour period after a half-hour warm-up at 25C , with the input voltage held constant at the rated input/output. Der ating is required. Applicab le when remote control (optional) is added. Please contact us about dynamic load and input response Please contact us about another class * T o meet the specifications. Do not operate over-loaded condition. P arallel operation is not possible. Der ating is required when operated with chassis and cover. Sound noise ma y be generated by power supply in case of pulse load. Example recommended EMI/EMC filter NAM-04-101 * A higher current rating EMI/EMC filter may be recommended in view of the other devices that could be connected in parallel with the power supply. High voltage pulse noise type : NAP series Low leakage current type : NAM series June 29, 2020

¶ Tolerance : ±1 [±0.04] ¶ Optional chassis and cover material : Electric galvanizing steel board. ¶ Weight : 290g max (with chassis & cover : 470g max) ¶ PCB material : CEM3 ¶ Dimensions in mm, [ ]=inches Standard type Chassis and cover type ¶ 4 Mounting holes are existing. ¶ The back side of P.C.B. of the power supply is assembled some SMDs. Be attention not to bump against the attached area by vibration. ¶ Use the spacer of 8mm length or more regarding insulation. And do not use press-fitting bush. ¶ Point A, Point B are thermometry points. Please refer to Instruction Manual 3. Barrier strip type Model B2B-XH-A Mating Connector (Terminal) RC(-) RC(+) ContentsPIN No. or SXH-001T-P0.6 BXH-001T-P0.6 XHP-2 CN4 Option (Mfr:J.S.T) Connector type <PIN CONNECTION> CN1 CN2 Pin No. Input

1 AC(L)

3 AC(N)

Pin No. Output 1 to 4- V 5 to 8+ V ¶ Keep drawing current per pin below 5A for CN2. I/O Connector Mating connector CN1 1-1123724-3 1-1123722-5 Chain Loose CN2 1-1123723-8 1-1123722-8 Terminal Chain Loose 1123721-1 1318912-1 1123721-1 1318912-1 (Mfr:Tyco Electronics) ¶ I/O Connector is Mfr. Tyco Electronics ¶ Option:-J1:VH(J.S.T) connector type. ¶ Mounting torque (Mounting hole of chassis) :1.5N m (16kgf cm) max ¶ External size of option is different from standard model. Output(-) Output(+) FG Input(N) Input(L) 4-M4 Mounting Hole 2- 4.5 2-M4 Mounting Hole 4.5 CN2 CN1 PCB t=1.6 Name plate Connector for Remote ON/OFF (Optional) CN4 Voltage adjust Mounting Hole 3- 3.5 Mounting Hole Point A Point B

1 Input(L)

3 Input(N)

Output(+) Output(-) 4.5 185 4.5 303max 52±0.55 5 145±0.5 16.5 155 3.5 6173±0.5 2325±0.5 1345±0.5 6173±0.5 20 15 FG ¶1 Surface mount device [0.2] [0.65] [0.16] [0.2] [2.44] [1.18][0.12] [0.14] [0.91] [0.51] [0.18] [0.18] [0.59][0.79] [1.77] [1.42] [2.83] [6.1] [0.24] [0.24] [7.28][5.71] [6.81] [6.81] [2.05] [0.98] [1.77] FG DC OUT AC IN 85 - 264V RC E 250V3.15A×2 INRUSH CURRENT LIMIT RECTIFIER AND FILTER CONTROL TRANSFORMER CONTROL OVER VOLTAGE PROTECTION NOISE FILTER RECTIFIER AND FILTER INVERTER FUSE RECTIFIER CURRENT SENSING BOOSTER INDUCTOR INVERTER CONTROL CURRENT SENSING EXTERNAL SOURCE OPTION Photocoupler Photocoupler Photocoupler June 29, 2020

AC-DC Power Supplies Medical Type LM A 150 F -O -O R 1 2 3 4 5 6 C : with Coating G : Low leakage current H : with the function to be acceptable to output peak current J1 : VH(J.S.T.)connector type R : with Remote ON/OFF R2 : with Remote ON/OFF S : with Chassis SN : with Chassis & cover P : Setting in the overcurrent protection rating This power supply is manufactured by SMD technology. The stress to P .C.B like twisting or bending causes the defect of the unit, so handle the unit with care. *Make sure necessary tests will be carried out on your end equipment with the power supply installed in accordance with any required EMC/EMI regulations. MODEL LMA150F-24-Y LMA150F-24-HY MAX OUTPUT WATTAGE[W] 151.2 151.2 (302.4) *2 DC OUTPUT 24V 6.3A 24V 6.3A (12.6A) *2 SPECIFICATIONS MODEL LMA150F-24-Y LMA150F-24-HY INPUT VOLTAGE[V] AC85 - 264 1f (Refer to “Derating” ,Instruction Manual 1 and 3) CURRENT[A] ACIN 100V 2.0typ (Io=100%) ACIN 200V 1.0typ (Io=100%) FREQUENCY[Hz] 50 / 60 (47 - 63) EFFICIENCY[%] ACIN 100V 85.0typ (Io=100%) 85.0typ (Io=100%) ACIN 200V 87.0typ (Io=100%) 87.0typ (Io=100%) POWER FACTOR ACIN 100V 0.99typ (Io=100%) ACIN 200V 0.95typ (Io=100%) INRUSH CURRENT[A] ACIN 100V 15typ (Io=100%) (At cold start) (Ta=25C) ACIN 200V 30typ (Io=100%) (At cold start) (Ta=25C) LEAKAGE CURRENT[mA] 0.10 / 0.25max (ACIN 100V / 240V 60Hz, Io=100%, According to IEC60601-1) OUTPUT VOLTAGE[V] 24 24 CURRENT[A] 6.3 6.3 (Peak 12.6) *2 LINE REGULATION[mV] *7 96max 96max LOAD REGULATION[mV] *7 150max 150max RIPPLE[mVp-p] *3 0 to +50C 120max 120max -10 - 0C 160max 160max RIPPLE NOISE[mVp-p]*3 0 to +50C 150max 150max -10 - 0C 180max 180max TEMPERATURE REGULATION[mV] 0 to +50C 240max 240max -10 to +50C 290max 290max DRIFT[mV] *4 96max 96max START-UP TIME[ms] 350typ (ACIN 100V, Io=100%) HOLD-UP TIME[ms] 20typ (ACIN 100V, Io=100%) OUTPUT VOLTAGE ADJUSTMENT RANGE[V] 19.20 to 27.50 19.20 to 27.50 OUTPUT VOLTAGE SETTING[V] 24.00 to 24.96 24.00 to 24.96 PROTECTION CIRCUIT AND OTHERS OVERCURRENT PROTECTION Works over 105% of rating (works over 101% of peak current at option -H) and recovers automatically OVERVOLTAGE PROTECTION[V] 27.60 to 33.60 27.60 to 33.60 OPERATING INDICATION Not provided REMOTE SENSING Not provided REMOTE ON/OFF Option (Required external power source.) ISOLATION INPUT-OUTPUT.RC *6 AC4,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature) 2MOOP INPUT-FG AC2,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature) 1MOOP OUTPUT.RC-FG *6 AC500V 1minute, Cutoff current = 25mA, DC500V 50MW min (At Room Temperature) OUTPUT-RC *6 AC100V 1minute, Cutoff current = 25mA, DC100V 10MW min (At Room Temperature) ENVIRONMENT OPERATING TEMP.,HUMID.AND ALTITUDE *5 -10 to +70C, 20 - 90%RH (Non condensing), (Refer to “Derating”, Instruction Manual 3) 3,000m (10,000feet) max STORAGE TEMP.,HUMID.AND ALTITUDE -20 to +75C, 20 - 90%RH (Non condensing), 9,000m (30,000feet) max VIBRATION 10 - 55Hz, 19.6m/s2 (2G), 3minutes period, 60minutes each along X, Y and Z axis IMPACT 196.1m/s2 (20G), 11ms, once each X, Y and Z axis SAFETY AND NOISE REGULATIONS AGENCY APPROVALS (AT ONIY AC input) ANSI/AAMI ES60601-1, EN60601-1 3rd, Complies with IEC60601-1-2 4th Ed. CONDUCTED NOISE Complies with FCC-B, VCCI-B, CISPR22-B, EN55011-B, EN55022-B HARMONIC ATTENUATOR Complies with IEC61000-3-2 (Class A) *8 OTHERS CASE SIZE/WEIGHT 75X36.5X160mm [2.95X1.44X6.30 inches] (WXHXD) / 370g max (with chassis & cover : 600g max) COOLING METHOD Convection (Refer to “Derating”, Instruction Manual 3) *5 *1 Specification is changed at option, ref er to Instruction Manual. P eak loading for 10sec. And Duty 40% max. ( ) means peak current. There is a possibility that an internal device is damaged when the specification is exceeded. This is the v alue that measured on measuring board with capacitor of 22mF at 150mm from output terminal. Measured b y 20MHz oscilloscope or Ripple-Noise meter (Equivalent to KEISOKU-GIKEN: RM103). *4 Dr ift is the change in DC output for an eight hour period after a half-hour warm-up at 25C , with the input voltage held constant at the rated input/output. Der ating is required. Applicab le when remote control (optional) is added. Please contact us about dynamic load and input response Please contact us about another class * T o meet the specifications. Do not operate over-loaded condition. P arallel operation is not possible. Der ating is required when operated with chassis and cover. Sound noise ma y be generated by power supply in case of pulse load. Example recommended EMI/EMC filter NAM-04-101 * A higher current rating EMI/EMC filter may be recommended in view of the other devices that could be connected in parallel with the power supply. High voltage pulse noise type : NAP series Low leakage current type : NAM series June 29, 2020

¶ Tolerance : ±1 [±0.04] ¶ Optional chassis and cover material : Electric galvanizing steel board. ¶ Dimensions in mm, [ ]=inches Standard type Chassis and cover type ¶ 4 Mounting holes are existing. ¶ The back side of P.C.B. of the power supply is assembled some SMDs. Be attention not to bump against the attached area by vibration. ¶ Use the spacer of 8mm length or more regarding insulation. And do not use press-fitting bush. ¶ Point A, Point B are thermometry points. Please refer to Instruction Manual 3. <PIN CONNECTION> CN1 CN2 CN3 Pin No. Input ¶ Keep drawing current per pin below 5A for CN2,CN3. Pin No. Output 1 to 6+ V Pin No. Output 1 to 7- V I/O Connector Mating connector CN1 1-1123724-3 1-1123722-5 Chain Loose CN2 1-1123723-6 1-1123722-6 (Mfr:Tyco Electronics) Terminal Chain Loose 1123721-1 1318912-1 1123721-1 1318912-1 CN3 1-1123723-7 1-1123722-7 Chain Loose 1123721-1 1318912-1 ¶ I/O Connector is Mfr. Tyco Electronics ¶ Option:-J1:VH(J.S.T) connector type. ¶ Mounting torque (Mounting hole of chassis) :1.5N-m (16kgf-cm) max ¶ External size of option is different from standard model. Barrier strip type Model B2B-XH-A Mating Connector (Terminal) RC(-) RC(+) ContentsPIN No. or SXH-001T-P0.6 BXH-001T-P0.6 XHP-2 CN4 Option (Mfr:J.S.T) Connector type 2 1 2-M4 Mounting Hole f4.5 2-f4.5 Output(-) Output(+) CN3 CN2 FG Input(N) Input(L) CN1 Output(-) Output(+) CN3 CN2 PCB t=1.6 Name plate Mounting Hole 3-f3.5 CN1 FG5 Input(N)3 Input(L)1 Point A Point B Mounting Hole Connector for RemoteON/OFF (optional) Voltage adjust 2 1 CN4 4-M4 Mounting Hole 55±0.5 4.5 188 4.5 33.53max 4.5 160 5 65±0.5 150±0.5 3.5 20 15 6176±0.5 2535±0.5 6176±0.5 ¶ Weight : 370g max (with chassis & cover : 600g max) ¶ PCB material : CEM3 ¶1 Surface mount device FG [0.2] [0.71] [0.16] [0.2] [2.95] [1.32][0.12] [0.14] [0.18] [0.98] [0.59] [0.18] [0.18] [0.59][0.79] [1.85] [1.65] [3.35] [6.3] [0.24] [0.24] [7.4] [5.91] [6.93] [6.93] [2.6] [1.38] [2.17] FG DC OUT AC IN 85 - 264V RC E 250V5A×2 INRUSH CURRENT LIMIT RECTIFIER AND FILTER CONTROL TRANSFORMER CONTROL OVER VOLTAGE PROTECTION NOISE FILTER RECTIFIER AND FILTER INVERTER FUSE RECTIFIER CURRENT SENSING BOOSTER INDUCTOR INVERTER CONTROL CURRENT SENSING EXTERNAL SOURCE OPTION Photocoupler Photocoupler Photocoupler June 29, 2020

AC-DC Power Supplies Medical Type LM A 240 F -O -O R 1 2 3 4 5 6 C : with Coating G : Low leakage current H : with the function to be acceptable to output peak current J1 : VH(J.S.T.)connector type R : with Remote ON/OFF R2 : with Remote ON/OFF S : with Chassis SN : with Chassis & cover P : Setting in the overcurrent protection rating This power supply is manufactured by SMD technology. The stress to P .C.B like twisting or bending causes the defect of the unit, so handle the unit with care. *Make sure necessary tests will be carried out on your end equipment with the power supply installed in accordance with any required EMC/EMI regulations. MODEL LMA240F-24-Y LMA240F-24-HY MAX OUTPUT WATTAGE[W] 300 300 (480) *2 DC OUTPUT Convection 24V 10A 24V 10A (20A) *2 Forced air 24V 12.5A 24V 12.5A (20A) *2 SPECIFICATIONS MODEL LMA240F-24-Y LMA240F-24-HY INPUT VOLTAGE[V] AC85 - 264 1f (Refer to “Derating” ,Instruction Manual 1 and 3) CURRENT[A] ACIN 100V 3.9typ (Io=100%) ACIN 200V 1.8typ (Io=100%) FREQUENCY[Hz] 50 / 60 (47 - 63) EFFICIENCY[%] ACIN 100V 86.0typ (Io=100%) 86.0typ (Io=100%) ACIN 200V 88.0typ (Io=100%) 88.0typ (Io=100%) POWER FACTOR ACIN 100V 0.99typ (Io=100%) ACIN 200V 0.95typ (Io=100%) INRUSH CURRENT[A] ACIN 100V 15 / 30typ (Io=100%) (Primary inrush current /Secondary inrush current) (More than 3 sec. to re-start) ACIN 200V 30 / 30typ (Io=100%) (Primary inrush current /Secondary inrush current) (More than 3 sec. to re-start) LEAKAGE CURRENT[mA] 0.15 / 0.40max (ACIN 100V / 240V 60Hz, Io=100%, According to IEC60601-1) OUTPUT VOLTAGE[V] 24 24 CURRENT[A] Convection 10 10 (Peak 20) *2 Forced air 12.5 12.5 (Peak 20) *2 LINE REGULATION[mV] *7 96max 96max LOAD REGULATION[mV] *7 150max 150max RIPPLE[mVp-p] *3 0 to +50C 120max 120max -10 - 0C 160max 160max RIPPLE NOISE[mVp-p]*3 0 to +50C 150max 150max -10 - 0C 180max 180max TEMPERATURE REGULATION[mV] 0 to +50C 240max 240max -10 to +50C 290max 290max DRIFT[mV] *4 96max 96max START-UP TIME[ms] 350typ (ACIN 100V, Io=100%) HOLD-UP TIME[ms] *9 20typ (ACIN 100V, Io=100%) OUTPUT VOLTAGE ADJUSTMENT RANGE[V] 19.20 to 27.50 19.20 to 27.50 OUTPUT VOLTAGE SETTING[V] 24.00 to 24.96 24.00 to 24.96 PROTECTION CIRCUIT AND OTHERS OVERCURRENT PROTECTION Works over 105% of rating (works over 101% of peak current at option -H) and recovers automatically OVERVOLTAGE PROTECTION[V] 27.60 to 33.60 27.60 to 33.60 OPERATING INDICATION Not provided REMOTE SENSING Not provided REMOTE ON/OFF Option (Required external power source.) ISOLATION INPUT-OUTPUT.RC *6 AC4,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature) 2MOOP INPUT-FG AC2,000V 1minute, Cutoff current = 10mA, DC500V 50MW min (At Room Temperature) 1MOOP OUTPUT.RC-FG *6 AC500V 1minute, Cutoff current = 25mA, DC500V 50MW min (At Room Temperature) OUTPUT-RC *6 AC100V 1minute, Cutoff current = 25mA, DC100V 10MW min (At Room Temperature) ENVIRONMENT OPERATING TEMP.,HUMID.AND ALTITUDE *5 -10 to +70C, 20 - 90%RH (Non condensing), (Refer to “Derating”, Instruction Manual 3) 3,000m (10,000feet) max STORAGE TEMP.,HUMID.AND ALTITUDE -20 to +75C, 20 - 90%RH (Non condensing), 9,000m (30,000feet) max VIBRATION 10 - 55Hz, 19.6m/s2 (2G), 3minutes period, 60minutes each along X, Y and Z axis IMPACT 196.1m/s2 (20G), 11ms, once each X, Y and Z axis SAFETY AND NOISE REGULATIONS AGENCY APPROVALS (AT ONIY AC input) ANSI/AAMI ES60601-1, EN60601-1 3rd, Complies with IEC60601-1-2 4th Ed. CONDUCTED NOISE Complies with FCC-B, VCCI-B, CISPR22-B, EN55011-B, EN55022-B HARMONIC ATTENUATOR Complies with IEC61000-3-2 (Class A) *8 OTHERS CASE SIZE/WEIGHT 84X46X180mm [3.31X1.81X7.09 inches] (WXHXD) / 540g max (with chassis & cover : 860g max) COOLING METHOD Convection / Forced air (Refer to “Derating”, Instruction Manual 3) *5 *1 Specification is changed at option, ref er to Instruction Manual. P eak loading for 10sec. And Duty 40% max. ( ) means peak current. There is a possibility that an internal device is damaged when the specification is exceeded. This is the v alue that measured on measuring board with capacitor of 22mF at 150mm from output terminal. Measured b y 20MHz oscilloscope or Ripple-Noise meter (Equivalent to KEISOKU-GIKEN: RM103). Dr ift is the change in DC output for an eight hour period after a half-hour warm-up at 25C , with the input voltage held constant at the rated input/output. Der ating is required. Applicab le when remote control (optional) is added. Please contact us about dynamic load and input response *8 Please contact us about another class . T o meet the specifications. Do not operate over-loaded condition. P arallel operation is not possible. Der ating is required when operated with chassis and cover. Sound noise ma y be generated by power supply in case of pulse load. Example recommended EMI/EMC filter NAM-06-101 * A higher current rating EMI/EMC filter may be recommended in view of the other devices that could be connected in parallel with the power supply. High voltage pulse noise type : NAP series Low leakage current type : NAM series June 29, 2020

¶ Tolerance : ±1 [±0.04] ¶ Optional chassis and cover material : Electric galvanizing steel board. ¶ Dimensions in mm, [ ]=inches Standard type Chassis and cover type ¶ 5 Mounting holes are existing. ¶ The back side of P.C.B. of the power supply is assembled some SMDs. Be attention not to bump against the attached area by vibration. ¶ Use the spacer of 8mm length or more regarding insulation. And do not use press-fitting bush. ¶ Point A, Point B, Point C are thermometry points. Please refer to Instruction Manual 3. <PIN CONNECTION> CN1 CN2 CN3 Pin No. Input ¶ Keep drawing current per pin below 5A for CN2,CN3. Pin No. Output 1 to 6+ V Pin No. Output 1 to 7- V I/O Connector Mating connector CN1 1-1123724-3 1-1123722-5 Chain Loose CN2 1-1123723-6 1-1123722-6 (Mfr:Tyco Electronics) Terminal Chain Loose 1123721-1 1318912-1 1123721-1 1318912-1 CN3 1-1123723-7 1-1123722-7 Chain Loose 1123721-1 1318912-1 Barrier strip type Model B2B-XH-A Mating Connector (Terminal) RC(-) RC(+) ContentsPIN No. or SXH-001T-P0.6 BXH-001T-P0.6 XHP-2 CN4 Option (Mfr:J.S.T) Connector type ¶ I/O Connector is Mfr. Tyco Electronics ¶ Option:-J1:VH(J.S.T) connector type. ¶ Mounting torque (Mounting hole of chassis) :1.5N-m (16kgf-cm) max ¶ External size of option is different from standard model. ¶ Weight : 540g max (with chassis & cover : 860g max) ¶ PCB material : CEM3 [0.12] [1.69] ˞1 Surface mount device 433.0max PCB ̓=1.6 Point BPoint A [0.14] [0.16] [0.2] [7.09] [6.69] [0.2] [3.5] Mounting Hole3.5489 ±0.5 5 170 ±0.5 180 [2.91] [3.31] 74 ±0.5 Point C FG [0.65] Name plate 4-П3.5 Mounting Hole Voltage adjust Connector for Remote ON/OFF(Optional) 16.5 [0.83] CN3 CN2 CN4 Output(+) Output(-) CN1 Input(N) Input(L) FG [0.65] [1.04] [2.56] [1.77] f4.5 2-П4.5 FG Input(N) Input(L) 2-M4 Mounting Hole 16.565 ±0.5 26.545 ±0.5 200 ±0.5 [7.87] [0.24] 4-M4 Mounting Hole [0.18] [3.86] [1.93] CN3 CN2 Output(+) Output(-) 4.5 [0.24] 212 6200 ±0.5 [1.18] [0.18] [0.59] [2.32] 4.5 30 15 [8.35] [7.87] CN1 FG DC OUT AC IN 85 - 264V RC E INRUSH CURRENT LIMIT RECTIFIER AND FILTER CONTROL TRANSFORMER CONTROL OVER VOLTAGE PROTECTION NOISE FILTER RECTIFIER AND FILTER INVERTER RECTIFIER CURRENT SENSING BOOSTER INDUCTOR INVERTER CONTROL CURRENT SENSING EXTERNAL SOURCE OPTION Photocoupler Photocoupler Photocoupler 250V10A×2 FUSE June 29, 2020

=8 mm min =4 mm min CN1 CN1 d2 Power supply Case (D)( E) (F) CN1 CN1 CN1 Standard Position CN1 (A)( B)( C) CN1 CN1 8 8 CN1 Unit [mm] 8 CN1, TB1 Unit [mm] LMA-8 Mounting screw Assembling and Installation Method Installation method ¡This power supply is manufactured by SMD technology.The stress to P.C.B like twisting or bending causes the defect of the unit, so han- dle the unit with care. ¡The mounting screw should be M3. The hatched area shows the allowance of metal parts for mounting. ¡If metallic fittings are used on the component side of the board,ensure there is no contact with surface mounted components. ¡This product uses SMD technology.Please avoid the PCB installation method which includes the twisting stress or the bending stress. *Recommendation to electrically connect FG to metal chassis for reducing noise. ¿ LMA100F, LMA150F ¿ LMA240F ¡In case of metal chassis, keep the distance between d1 & d2 for to insulate between lead of component and metal chassis, use the spacer of 8mm or more between d1. If it is less than d1 & d2, in- sert the insulation sheet between power supply and metal chassis. ¡There is a possibility that it is not possible to cool enough when the power supply is used by the sealing up space as showing in right figure.Please use it after confi rming the temperature of point A and point B of Instruction Manual 3. ¡(F) mounting is not possible when unit is with case cover, but if need to operate unit by (F) positioning with case cover, tempera- ture / load derating is necessary. For more details, please contact our sales or engineering departments. LMA-series June 29, 2020

[AC V] [%] 100 85 90 Load factor -10 7060 50400 3 35 45 2Forced air (0.5m3 /m in) 1Convection 100 20100 Loadf actor[ %] [C]Ambient temperature 1(A)mounting 1 1(B)mounting(D), (F) mounting 1(C)mounting 2(A) ~ (F) mounting 1(E)mounting -10 7060 50400 3 2Forced air (0.5m3 /m in) 1Convection 100 20100 Load factor [%] [C]Ambient temperature 1(A)mounting1(D), (E), (F) mounting 2(A) ~ (F) mounting 1(B), (C) mounting Output voltage Output power[W] 1Convection 2Forced air 24V 240.0 300.0 -10 7060 50400 3 2Forced air (0.5m3 /m in) 1Convection 100 20100 Load factor [%] [C]Ambient temperature 1 (A)mounting 11 (B)mounting 1 (F)mounting (C), (D), (E)mounting 2(A) ~ (F) mounting 35 45 LMA-9 Derating ¿ Derating curve for input voltage LMA100F Ambient temperature derating curve (Reference value) ¿ LMA240F Ambient temperature derating curve (Reference value) ¿ LMA150F Ambient temperature derating curve (Reference value) ¡The operative ambient temperature is different by with / without chassis cover or mounting position. Note: In the hatched area, the specification of Ripple, Ripple Noise is different from other area. ¡Make sure the temperature at point A and point B is less than the temperatures shown in Instruction Manual 3. ¡The ambient temperature should be measured 5 to 10 cm away from the power supply so that it won’t be influenced by the heat from the power supply. Please consult us for more details. LMA-series June 29, 2020

◆ It is neccessary to read the “Instruction Manual” and “Before using our product” before you use our product. Instruction Manual https://en.cosel.co.jp/product/powersupp ly/LMA/ Before using our product https://en.cosel.co.jp/technical/caution/in dex.html Instruction Manual NOTICE LMA Basic Characteristics Data LMA-series Model Circuit method Switching frequency [kHz] Input current *1 [A] Inrush current protection PCB/Pattern Series/Parallel operation availability Material Single sided Double sided Series operation Parallel operation LMA100F Active filter 60

1.4 Thermistor CEM-3 Ye s Ye s No

2.0 Thermistor CEM-3 Ye s Ye s No

3.9 SCR CEM-3 Ye s Ye s No

*1 The v alue of input current is at ACIN 100V and rated load. June 29, 2020