DATASHEET SEARCH SITE | WWW.ALLDATASHEET.COM
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 2
Technical content
7” LCD GPT-12000 Series AC/DC/IR/GB Electrical Safety Analyzer ACOutput Capacity DC IR GB GPT-12002 GPT-12003 GPT-12004 GPT-12001 Function 200VA 200VA 200VA 200VA Rear OutputContinuityModel Intelligent Safety Analyzer 200VA AC Test Capacity Comply with IEC 61010-2-034 7” TFT LCD & Manual/Auto Mode True RMS Current Measurement Zero Crossing Turn-on Operation Controllable Ramp-up & Ramp-down Time Capacitive Load Testing Capability up to 47 Fμ Statistics Function Sweep Function for DUT Characteristic Analysis USB Storage Available & Rear Panel Output Available Interface:RS-232C,USB Host/Device,Signal I/O and GPIB(Opt.) Universal Power Input
ORDERING INFORMATION
GPT-12004 AC/DC/IR/GB Electrical Safety Analyzer AC/DC/IR Electrical Safety Analyzer AC/DC Electrical Safety Analyzer Specifications subject to change without notice. REMOTE, USB host Rear Output, RS-232C, USB device, Signal I/O, GPIB 0.050kV~5.000kV ±(1% of setting + 5V) [no load]
200 VA (5kV/40mA)
40mA (0.5kV< V 5kV); 10mA (0.05kV V 0.5kV)≦≦ ≦ Sine wave
50 Hz / 60 Hz selectable
±(1% + 5V) [maximum rated load -- no load] ±(1% of reading + 5V) 1 A~40.00mAμ
1 A / 10 Aμμ
±(1.5% of reading + 30 A)μ Yes Yes 0.1s~999.9s 0.0s~999.9s OFF, 0.3s~999.9s 0.0s~999.9s ON/OFF 0.050kV~6.000kV ±(1% of setting + 5V) [no load] 50W (5kV/10mA) 10mA (0.5kV< V 6kV);≦ 2mA (0.05kV V 0.5kV)≦≦ ±(1% + 5V) [maximum rated load -- no load] ±(1% of reading + 5V) 1 A~10.00mAμ
0.1 A /1 A /10 Aμμ μ
±(1.5% of reading + 3 A) when I Reading < 1mA ;μ ±(1.5% of reading + 30 A) when I Reading 1mAμ Yes Yes 0.1s~999.9s 0.0s~999.9s OFF, 0.3s~999.9s 0.0s~999.9s ON/OFF 50V~1200V dc 50V ±(1% of setting + 5V) [no load] Display Range 0.1M ~Ω 10.00GΩ 0.1M ~Ω 20.00GΩ 0.1M ~Ω 50.00GΩ Measurement Range / Accuracy 0.1M 1M 5% of reading + 3 count);Ω~ Ω:±( 1 M 50M 5% of reading + 1 count);Ω~ Ω:±( 51M 2G 10% of reading + 1 count)Ω~ Ω:±( 0.1M 1M 5% of reading + 3 count);Ω~ Ω:±( 1 M 500M 5% of reading + 1 count);Ω~ Ω:±( 501M 9.999G 10% of reading + 1 count);Ω~ Ω:±( 10G 50G 20% of reading + 1 count)Ω~ Ω:±( ±(1% + 5V) [maximum rated load -- no load] ±(1% of reading + 5V) 10mA max. 2kΩ Yes 0.1s~999.9s 0.0s~999.9s 0.3s~999.9s 0.0s~999.9s ON/OFF 03.00A~32.00A ac 0.01A 3A I 8A : ±(1% of reading + 0.2A);≦≦ 8A<I 32A : ±(1% of reading + 0.05A)≦ 8Vac max (open circuit) 50Hz/60Hz selectable 1m ~ 650mΩΩ 0.1mΩ ±(1% of reading + 2 m )Ω Yes 0.3s~999.9s Four Terminal ON/OFF 100mA dc (fixed) 0.10 ~ 70.00ΩΩ 0.01Ω ±(10% of reading + 2 )Ω Yes 0.3s~999.9s MANU : 100 blocks AUTO : 100 blocks, Manu per auto : 10 7” color LCD AC 100V~240V ± 10%, 50Hz/60Hz; Power consumption : Max. 400VA 380(W) x 148(H) x 454(D) mm; Approx. 15kg GPT-12003 GPT-12002 GPT-12001 OPTIONAL ASSESSORIES OPTION Opt.1 GPIB card Quick Start Guide x 1, Power cord x 1, CDx1(complete user manual), Interlock Key x 1, Remote terminal Cable GHT-119 x 1, Test lead GHT-115 x 1 for GPT- 12001/12002/12003, Test lead GHT-115 x 1, GTL-215 x 1 for GPT-12004 ACCESSORIES AC WITHSTANDING Output-Voltage Range Output-Voltage Resolution Output-Voltage Accuracy Maximum Rated Load Maximum Rated Current Output-Voltage Waveform Output-Voltage Frequency Voltage Regulation Voltmeter Accuracy Current Measurement Range Current Best Resolution Current Measurement Accuracy Window Comparator Method ARC Detect RAMP UP (Rise Time) RAMP DOWN (Fall Time) TIMER (Test Time)* WAIT TIME GND DC WITHSTANDING Output-Voltage Range Output-Voltage Resolution Output-Voltage Accuracy Maximum Rated Load Maximum Rated Current Voltage Regulation Voltmeter Accuracy Current Measurement Range Current Best Resolution Current Measurement Accuracy Window Comparator Method ARC Detect RAMP UP (Rise Time) RAMP DOWN (Fall Time) TIMER (Test Time)* WAIT TIME GND INSULATION RESISTANCE Output Voltage Output-Voltage Resolution Output-Voltage Accuracy Resistance Measurement Test Voltage 50V V 100V≦≦ 150V V 450V≦≦ 500V V 1200V≦≦ Voltage Regulation Voltmeter Accuracy Short-Circuit Current Output Impedance Window Comparator Method RAMP UP (Rise Time) RAMP DOWN (Fall Time) TIMER (Test Time)* WAIT TIME GND GROUND BOND Output-Current Output-Current Resolution Output-Current Accuracy Test-Voltage Test-Voltage Frequency Ohmmeter Measurement Range Ohmmeter Measurement Resolution Ohmmeter Measurement Accuracy Window Comparator Method TIMER (Test Time)* Test Method GND CONTINUITY TEST Output-Current Ohmmeter Measurement Range Ohmmeter Measurement Resolution Ohmmeter Measurement Accuracy Window Comparator Method TIMER (Test Time)* MEMORY Single Step Memory Automatic Testing Memory INTERFACE DISPLAY POWER SOURCE DIMENSIONS & WEIGHT Standard (Front) Standard (Rear) Option SPECIFICATIONS GHT-113 GHT-117 GHT-118 GHT-205 GTL-232 GTL-246 GTL-248 GRA-440High Voltage Test Pistol High Voltage Adapter Box High Voltage / Ground Bond Adapter Box High Voltage Test Probe RS232C Cable, 9-pin Female to 9-pin, null Modem for Computer USB Cable, A-B type, approx. 1.2m GPIB Cable, approx. 2m Rack Adapter Panel (19', 4U) Note : * TIMER Accuracy: +/-(100ppm+20ms)