FB5201B PF | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 3

Technical content

< 1-channel < Input Ex ia < Installation in suitable enclosures in Zone 1 < Module can be exchanged under voltage (hot swap) < Converter for 2-, 3- and 4-wire Pt100, slide wire sensors < Simulation mode for service operations (forcing) < Line fault detection (LFD) < Permanently self-monitoring Function The RTD converter accepts 2-, 3-, 4-wire RTD signals (Pt100) from the hazardous area. Open and short-circuit line faults are detected. The intrinsically safe input is galvanically isolated from the bus and the power supply. Connection Zone 1Zone 0, 20 COM mV Technical Data Slots Occupied slots 1 Supply Connection backplane bus Rated voltage Ur 12 V DC , only in connection with the power supplies FB92** Power dissipation 0.4 W Power consumption 0.4 W Internal bus Connection backplane bus Interface manufacturer-specific bus to standard com unit temperature input Number of channels 1 Release date: 2021-11--

16 Date of

Release date: 2021-11-16 Date of issue: 2021-11-16 Filename: 542104_eng.pdf Germany: +49 621 776 2222 Pepperl+Fuchs Group Refer to "General Notes Relating to Pepperl+Fuchs Product Information". USA: +1 330 486 0002 Singapore: +65 6779 9091

Field device resistance thermometer Field device [3] slide-wire sensors Field device interface Connection 2-wire sensor Connection [2] 3-wire sensor Connection [3] 4-wire sensor Connection 2-wire connection: 5, 6 3-wire connection: 1, 5, 6 4-wire connection: 1, 2, 5, 6 Measurement range 10 ... 400 Ω (500 Ω incl. line resistance) Slide-wire sensor 10 ... 400 Ω Measuring current 200 µA Smallest span 20 Ω for 0.1 % accuracy Linearity error 0.1 % Conversion time max. 20 ms without LFD max. 150 ms with LFD Lead resistance max. 50 Ω per strand Line fault detection can be switched on/off for each channel via configuration tool Short-circuit < 10 Ω Open-circuit > 1 kΩ Transfer characteristics Deviation Influence of ambient temperature max. 0,1 %/10 K Indicators/settings LED indication LED green: supply LED red: line fault Coding optional mechanical coding via front socket Directive conformity Electromagnetic compatibility Directive 2014/30/EU EN 61326-1:2013 Conformity Electromagnetic compatibility NE 21 Degree of protection IEC 60529 Environmental test EN 60068-2-14 Shock resistance EN 60068-2-27 Vibration resistance EN 60068-2-6 Damaging gas EN 60068-2-42 Relative humidity EN 60068-2-56 Ambient conditions Relative humidity 95 % non-condensing Shock resistance shock type I, shock duration 11 ms, shock amplitude 15 g, number of shocks 18 Vibration resistance frequency range 10 ... 150 Hz; transition frequency: 57.56 Hz, amplitude/acceleration ± 0.075 mm/1 g; 10 cycles frequency range 5 ... 100 Hz; transition frequency: 13.2 Hz amplitude/acceleration ± 1 mm/0.7 g; 90 minutes at each resonance Damaging gas designed for operation in environmental conditions acc. to ISA-S71.04-1985, severity level G3 Mechanical specifications Degree of protection IP20 (module) , a separate housing is required acc. to the system description Connection removable front connector with screw flange (accessory) Mass approx. 350 g Dimensions 28 x 107 x 132 mm (1.1 x 4.2 x 5.2 inch) RTD Converter FB5201B Release date: 2021-11-16 Date of issue: 2021-11-16 Filename: 542104_eng.pdf Germany: +49 621 776 2222 Pepperl+Fuchs Group Refer to "General Notes Relating to Pepperl+Fuchs Product Information". USA: +1 330 486 0002 Singapore: +65 6779 9091

Data for application in connection with hazardous areas EU-type examination certificate PTB 97 ATEX 1074 U Marking 1 II 2(1) G Ex d [ia Ga] IIC Gb

1 II (1) D [Ex ia Da] IIIC

Voltage Uo 2.7 V Current Io 43 mA Power Po 93 mW (trapezoid characteristic curve) Galvanic isolation Input/power supply, internal bus safe electrical isolation acc. to EN 60079-11, voltage peak value 375 V Directive conformity Directive 2014/34/EU EN IEC 60079-0:2018+AC:2020 EN 60079-1:2014 EN 60079-11:2012 International approvals ATEX approval PTB 97 ATEX 1075 ; PTB 97 ATEX 1074 U General information System information The module has to be mounted in appropriate backplanes and housings (FB92**) in Zone 1, 2, 21, 22 or outside hazardous areas (gas or dust). Here, observe the corresponding EC-type examination certificate. Supplementary information EC-Type Examination Certificate, Statement of Conformity, Declaration of Conformity, Attestation of Conformity and instructions have to be observed where applicable. For information see www.pepperl-fuchs.com. Assembly Front view LED green: Power supply Space for labelling LED red: Line fault Socket for removable plug blue (accessory) Accessories FB9224* Field Unit FB9225* FB9248* RTD Converter FB5201B Release date: 2021-11-16 Date of issue: 2021-11-16 Filename: 542104_eng.pdf Germany: +49 621 776 2222 Pepperl+Fuchs Group Refer to "General Notes Relating to Pepperl+Fuchs Product Information". USA: +1 330 486 0002 Singapore: +65 6779 9091