Specifications are subject to change without notice 1
- 3-dgt µP-based frequency indicator
- 1.0 to 999 Hz
- Autoranging capability
- 72 x 72 mm
- Degree of protection (front): IP 50 (IP 65 on request) Product Description Model Range code Measurement Power supply Setpoints Options 3-dgt µP-based frequency meter with autoranging capa- bility. The housing is for panel mounting and ensures a degree of protection (front) of IP 50 (IP 65 on request). Panel Meters and Controllers Autoranging Frequency Meter Type DI3-72 F1KType Selection Power supply 50/60 Hz 1) 50/60 Hz 1) 50/60 Hz 1) 50/60 Hz (standard) Options XX: None (standard) EX: Degree of protection IP 65 Rated input 500 VAC, 1 to 1000 Hz, autoranging Overload protection Continuous 1.2 x rated input For 1 s 2 x rated input Accuracy Temperature drift ± 100 ppm/ºC Display 7-segment LED, h 14.2 mm, 3 digits Sampling rate 1 time/s Indication Max. 999 Min. 0.00 Over range EEE Input voltage 9 to 500 VAC Input impedance 500 kΩ Input Specifications 1) Power supply on request AC supply 230 VAC, -15% +10%, 50/60 Hz (standard)
24 VAC, 48 VAC, 115 VAC,
(on request) Power consumption 3.2 VA Supply Specifications Operating temperature 0 to 50°C (32 to 122°F) (R.H. < 90% non-condensing) Storage temperature -10 to 60°C (14 to 140°F) (R.H. < 90% non-condensing) Insulation reference voltage 300 Vrms to ground Dielectric strength 4000 Vrms for 1 minute Noise rejection CMRR 100 dB, 40 to 60 Hz EMC IEC 60801-2, IEC 60801-3, IEC 60801-4 (level 2), EN 50 081-1, EN 50 082-1 Safety standards EN 61010-1, IEC 61010-1, VDE 0411 Connector Screw-type Housing Dimensions 72 x 72 x 75 mm Material ABS, self-extinguishing: UL 94 V-0 Degree of protection Front: IP 50 (IP 65 on request) Weight 250 g approx. Approvals CE, CSA General Specifications
2 Specifications are subject to change without notice
66.5 x 66.5 68 x 68 Dimensions (mm) 1 2345 67 8 INNC INA INV/HzPower supply IN + 60 mV Wiring Diagram Hz Front Panel Description 1. Display 3-dgt (maximum read-out 999). Alphanumeric indication by means of 7-segment display for: - Displaying of the measured value, over-range. Range Table Measurement Dec. point Resolution position 1.00 to 9.99 Hz 9.99 0.01 Hz 10.0 to 99.9 Hz 99.9 0.1 Hz 100 to 999 Hz 999 1 Hz (Automatic selection)