C100 CHAUVINARNOUX | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 18
Technical content
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.00 (1/2) C100 seriesCurrent clamps for AC current "C 100" SERIES The "C100" series is a range of thirteen transformer clamps with all the advantages of our old "C30" series clamps whilst incorporating considerable improvements, particularly in the fi eld of safety, ergonomics and performance:
- 1,000 A measurement, excellent metrology, high accuracy, high level of linearity, symmetrical coil windings for minimum phase shift, pendular adjusting system for magnetic elements, maximum conductor diameter Ø 52 mm and also some models with /uni03BC metal core specially made for wattmeter use.
- Innovative design: excellent ergonomics, handle with fi nger grips, assisted opening system for jaws (patented system), IEC 1010 600 V cat. III safety (industry and services), antislip protection, conductor anti- pinching system,... All this technology and manufacturing quality has been combined to provide the best measurement possible without any complications. A "C100" series clamp is compatible with any instrument (multimeter, wattmeter, recorder, oscilloscope...) for safe measurement of AC currents without shutting down the installation.
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.00 (2/2) C100 seriesCurrent clamps for AC current 101 111 43.5
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.01 (1/1) C100 seriesCurrent clamp for AC current ELECTRICAL SPECIFICATIONS
- Current range: 0.1 A/uni00A0AC .. 1,200 A/uni00A0AC
- Current transformation ratio: 1000:1
- Output signal: 1 mA/uni00A0AC / A/uni00A0AC (1 A to 1,000 A)
- Accuracy and phase shift (1): Primary current 0.1 A .. 10 A 10 A 50 A (2) 200 A( 2) 1,000 A (2) 1,200 A (2) Phase shift not specified ≤/uni00A03°≤/uni00A01.5°≤/uni00A00.75 °≤/uni00A00.5 °≤/uni00A00.5 °
- Bandwidth: 30 Hz .. 10 kHz (-3 dB)
- Crest factor: ≤/uni00A06 for a current ≤/uni00A03,000 A peak (500 ARMS)
- Maximum currents: 1,000 A continuous for a frequency ≤ 1 kHz (limitation proportional to the inverse of frequency beyond) 1,200 A for 40 minutes max (interval between measurements > 20 minutes)
- Load impedance: ≤/uni00A015 /uni03A9
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
600 V category III and pollution degree 2
- Infl uence of adjacent conductor: ≤/uni00A01 mA / A at 50 Hz
- Infl uence of conductor position in jaws: ≤/uni00A00.1 % of output signal for frequencies ≤/uni00A0400 Hz
- Load infl uence: From 5 Ω to 15 Ω </uni00A00.5 % on measurement </uni00A00.5 ° on phase
- Infl uence of frequency (3): </uni00A01 % of output signal from 30 Hz .. 48/uni00A0Hz </uni00A00.5 % of output signal from 65/uni00A0Hz .. 1 kHz </uni00A01 % of output signal from 1 kHz .. 5 kHz
- Infl uence of crest factor: </uni00A01 % of output signal for crest factor ≤/uni00A06 with current ≤/uni00A03,000 A peak (500/uni00A0A RMS)
- Influence of DC current superimposed on rated current: </uni00A01 % of output signal for a current ≤/uni00A030 A DC MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C to +50 °C
- Storage temperature: -40 °C to +70 °C
- Infl uence of temperature: ≤/uni00A00.1 % of output signal per 10 °K
- Relative humidity for operation: 0 to 85/uni00A0% RH decreasing linearly above 35 °C
- Infl uence of relative humidity: </uni00A00.1 % of output signal from 10/uni00A0% to 85/uni00A0% RH
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm Patented progressive opening system
- Clamping capacity: Cable: Ø max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm / 4 busbars of 30/uni00A0x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm 15/25/uni00A0Hz 1/uni00A0mm 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: Casing and jaws: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: Safety sockets (4/uni00A0mm) SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, no DC components, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement, load impedance 5 Ω (5 VA) (2) Accuracy class in accordance with IEC/uni00A0185: 5 VA - class 0.5 - 48 Hz .. 65/uni00A0Hz (3) Out of frequency domain Current 1,000 A Ratio 1000 / 1 Output 1 mA / A To order Reference AC current clamp model C100 with operating manual P01120301 Current clamp for AC current Model C100
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.02 (1/1) C100 series ELECTRICAL SPECIFICATIONS
- Current range: 0.1 A/uni00A0AC .. 1,200 A/uni00A0AC
- Current transformation ratio: 1000:1
- Output signal: 1 mA/uni00A0AC / A/uni00A0AC (1 A to 1,000 A)
- Accuracy and phase shift (1): Primary current 0.1 A .. 10 A 10 A 50 A (2) 200 A ( 2) 1,000 A (2) 1,200 A (2) Phase shift not specified ≤/uni00A03°≤/uni00A01.5°≤/uni00A00.75 °≤/uni00A00.5 °≤/uni00A00.5 °
- Bandwidth: 30 Hz .. 10 kHz (-3 dB)
- Crest factor: ≤/uni00A06 for a current ≤/uni00A03,000 A peak (500 ARMS)
- Maximum currents: 1,000 A continuous for a frequency ≤ 1 kHz (limitation proportional to the inverse of frequency beyond) 1,200 A for 40 minutes max (interval between measurements > 20 minutes)
- Load impedance: ≤/uni00A015 /uni03A9
- Max. voltage output: Electronic protection circuit limiting voltage to 30/uni00A0V/uni00A0peak max
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
- Infl uence of adjacent conductor: ≤/uni00A01 mA / A at 50 Hz
- Infl uence of conductor position in jaws: ≤/uni00A00.1 % of output signal for frequencies ≤/uni00A0400 Hz
- Load infl uence: from 5 Ω to 15 Ω </uni00A00.5 % on measurement </uni00A00.5 ° on phase
- Infl uence of frequency (3): </uni00A01 % of output signal from 30 Hz .. 48/uni00A0Hz </uni00A00.5 % of output signal from 65/uni00A0Hz .. 1 kHz </uni00A01 % of output signal from 1 kHz .. 5/uni00A0kHz
- Infl uence of crest factor: </uni00A0 1 % of output signal for crest factor ≤/uni00A0 6 with current ≤/uni00A03,000 A peak (500/uni00A0ARMS)
- Infl uence of DC current superimposed on rated current: </uni00A01 % of output signal for a current ≤/uni00A030/uni00A0A/uni00A0DC MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C to +50 °C
- Storage temperature: -40 °C to +70 °C
- Infl uence of temperature: ≤ 0.1 % of output signal per 10 °K
- Relative humidity for operation: 0 to 85/uni00A0% RH with a linear decrease above 35 °C
- Infl uence of relative humidity: </uni00A00.1 % of output signal from 10/uni00A0% to 85/uni00A0% RH
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm, patented progressive opening system
- Clamping capacity: Cable: Ø max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm / 4 busbars of 30 x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm -15/25/uni00A0Hz 1/uni00A0mm - 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: Casing and jaws: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: C102: Safety sockets (4/uni00A0mm) C103: two-wire cable with reinforced insulation or double insulation, length 1.5 m, terminated by 2/uni00A0insulated elbowed male banana plugs (4/uni00A0mm) SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, no DC components, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement, load impedance 5 Ω (5 VA). (2) Accuracy class in accordance with IEC/uni00A0185: 5 VA - class 0.5 - 48 .. 65/uni00A0Hz. (3) Out of reference domain.
DESCRIPTION
An electronic voltage limiter protects the output of the clamp, if the secondary circuit is opened accidentally. Current 1,000 A Ratio 1000 / 1 Output 1 mA / A To order Reference AC current clamp model C102 with operating manual P01120302 AC current clamp model C103 with operating manual P01120303 Current clamps for AC current Models C102 and C103
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.03 (1/1) C100 series ELECTRICAL SPECIFICATIONS
- Current range: 0.1 A/uni00A0AC .. 1,200 A/uni00A0AC
- Output signal: 1 mV AC / A/uni00A0AC (1/uni00A0V for 1,000 A)
- Accuracy and phase shift (1): Primary current 0.1 A .. 10 A 10 A 50 A 200 A 1,000 A 1,200 A Phase shift not specified ≤/uni00A03°≤/uni00A01.5°≤/uni00A00.75 °≤/uni00A00.5 °≤/uni00A00.5 °
- Bandwidth: 30 Hz .. 10 kHz
- Crest factor: ≤/uni00A06 for a current ≤/uni00A03,000 A peak (500 ARMS)
- Maximum currents: 1,000 A continuous for a frequency ≤ 1 kHz (limitation proportional to the inverse frequency beyond) 1,200 A for 40 minutes max (interval between measurements > 20 minutes)
- Output impedance: 1 /uni03A9 ± 1 %
- Load impedance: ≥ 1 M/uni03A9 and ≤/uni00A0100 pF
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
- Infl uence of adjacent conductor: ≤/uni00A01 µV / A at 50 Hz
- Infl uence of conductor position in jaws: ≤/uni00A00.1 % of output signal for frequencies ≤/uni00A0400 Hz
- Load infl uence: On receiver, for an input impedance of 100/uni00A0 /uni03A9: ≤/uni00A01 % on measurement, no measurement on phase On receiver, for an input impedance of 1/uni00A0 k/uni03A9: ≤/uni00A0 0.1 % on measurement, no measurement on phase
- Infl uence of frequency (2): </uni00A01 % of output signal from 30 Hz .. 48/uni00A0Hz </uni00A00.5 % of output signal from 65/uni00A0Hz .. 1 kHz </uni00A01 % of output signal from 1 kHz .. 5 kHz
- Infl uence of crest factor: </uni00A01 % of output signal for crest factor ≤/uni00A06 with current ≤/uni00A03,000 A peak (500/uni00A0A RMS)
- Infl uence of DC current superimposed on rated current: </uni00A01 % of output signal for a current ≤/uni00A030/uni00A0A DC MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C to +50 °C
- Storage temperature: -40 °C to +70 °C
- Infl uence of temperature: ≤/uni00A00.1 % of output signal per 10 °K
- Relative humidity for operation: 0 to 85/uni00A0% RH decreasing linearly above 35 °C
- Infl uence of relative humidity: </uni00A00.1 % of output signal from 10/uni00A0% to 85/uni00A0% RH
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm Patented progressive opening system
- Clamping capacity: Cable: Ø/uni00A0max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm / 4 busbars of 30 x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm 15/25/uni00A0Hz 1/uni00A0mm 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: Casing and jaws: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: C106: Safety sockets (4/uni00A0mm) C107: two-wire cable with reinforced insulation or double insulation, length 1.5 m, terminated by 2/uni00A0insulated elbowed male banana plugs (4/uni00A0mm) SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, no DC components, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement. (2) Out of reference domain. Current 1,000 A Output 1 mV / A To order Reference AC current clamp model C106 with operating manual P01120304 AC current clamp model C107 with operating manual P01120305 Current clamps for AC current Models C106 and C107
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.04 (1/1) C100 series (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, no DC components, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement, load impedance 1 Ω (1 VA) (2) Out of reference domain. ELECTRICAL SPECIFICATIONS
- Current range: 0.001 A/uni00A0AC .. 1,200 A/uni00A0AC
- Current transformation ratio: 1000:1
- Output signal: 1 mA/uni00A0AC / A/uni00A0AC (1/uni00A0A for 1,000 A)
- Accuracy and phase shift (1): % Accuracy of output signal ≤/uni00A03 % + 5 µA≤/uni00A02 % + 3 µA ≤/uni00A01 % ≤/uni00A00.5 % ≤/uni00A00.3 % Phase shift not specified not specified ≤/uni00A02° ≤/uni00A01° ≤/uni00A00.7°
- Bandwidth: 30 Hz .. 10 kHz
- Crest factor: ≤/uni00A06 for a current ≤/uni00A02,000 A peak (300 ARMS)
- Maximum currents: 1,000 A continuous for a frequency ≤ 1 kHz (limitation proportional to the inverse of frequency beyond) 1,200 A for 40 minutes max (interval between measurements > 20 minutes)
- Load impedance: ≤/uni00A01 /uni03A9
- Max. voltage output: Electronic protection circuit limiting voltage to 30/uni00A0V peak max
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
- Infl uence of adjacent conductor: ≤/uni00A00.5 mA / A at 50 Hz
- Infl uence of conductor position in jaws: ≤/uni00A00.1 % of output signal for frequencies ≤/uni00A0400 Hz
- Load infl uence: From 1 Ω to 5 Ω </uni00A00.1 % on measurement </uni00A00.2° on phase
- Infl uence of frequency (2): </uni00A00.5 % of output signal from 30 Hz .. 48/uni00A0Hz </uni00A01 % of output signal from 65/uni00A0Hz .. 1 kHz </uni00A02 % of output signal from 1 kHz .. 5 kHz
- Infl uence of crest factor: </uni00A01 % of output signal for crest factor ≤/uni00A06 with current ≤/uni00A02,000 A peak (300/uni00A0A RMS)
- Infl uence of DC current superimposed on rated current: </uni00A01 % of output signal for a current ≤/uni00A015 A DC MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C to +50 °C
- Storage temperature: -40 °C to +70 °C
- Infl uence of temperature: ≤/uni00A00.2 % of output signal per 10 °K
- Relative humidity for operation: 0 to 85/uni00A0% RH with a linear decrease above 35 °C
- Infl uence of relative humidity: </uni00A00.1 % of output signal from 10/uni00A0% to 85/uni00A0% RH
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm, patented progressive opening system
- Clamping capacity: Cable: Ø max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm / 4 busbars of 30/uni00A0x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm, 15/25/uni00A0Hz 1/uni00A0mm, 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: Casing and jaws: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: C112: Safety sockets (4/uni00A0mm) C113: two-wire cable with reinforced insulation or double insulation, length 1.5 m, terminated by 2/uni00A0insulated elbowed male banana plugs (4/uni00A0mm) SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8
Thanks to their excellent technical performance (phase shift and linearity), these /uni03BC-metal core clamps are highly recommended for wattmeter use. These clamps are protected at output against overvoltages. Current 1,000 A Ratio 1000 / 1 Output 1 mA / A To order Reference AC current clamp model C112 with operating manual P01120314 AC current clamp model C113 with operating manual P01120315 Current clamps for AC current Models C112 and C113
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.05 (1/1) C100 series
- Bandwidth: 30 Hz .. 10 kHz
- Crest factor: ≤/uni00A06 for a current ≤/uni00A02,000 A peak (300 ARMS)
- Maximum currents: 1,000 A continuous for a frequency ≤ 1 kHz (limitation proportional to the inverse of frequency beyond) 1,200 A for 40 minutes max (interval between measurements > 20 minutes)
- Output impedance: 1 /uni03A9 ± 1 %
- Load impedance: ≥ 1 M/uni03A9 and ≤/uni00A0100 pF
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
- Infl uence of adjacent conductor: ≤/uni00A00.5 mA / A at 50 Hz
- Infl uence of conductor position in jaws: ≤/uni00A00.1 % of output signal for frequencies ≤/uni00A0400 Hz
- Load infl uence: On receiver, for an input impedance of 100/uni00A0 /uni03A9: ≤/uni00A01 % on measurement, no measurement on phase. On receiver, for an input impedance of 1/uni00A0 k/uni03A9: ≤/uni00A0 0.1 % on measurement, no measurement on phase.
- Infl uence of frequency (2): </uni00A00.5 % of output signal from 30 Hz .. 48 Hz </uni00A01 % of output signal from 65/uni00A0Hz .. 1 kHz </uni00A02 % of output signal from 1 kHz .. 5 kHz
- Infl uence of crest factor: </uni00A0 1 % of output signal for crest factor ≤/uni00A0 6 with current ≤/uni00A02,000 A peak
- Infl uence of DC current superimposed on rated current: </uni00A01 % of output signal for a current ≤/uni00A015 A DC MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C to +50 °C
- Storage temperature: 40 °C to +70 °C
- Infl uence of temperature: ≤/uni00A00.2 % of output signal per 10 °K
- Relative humidity for operation: 0 to 85/uni00A0% RH decreasing linearly above 35 °C
- Infl uence of relative humidity: </uni00A00.1 % of output signal from 10/uni00A0% to 85/uni00A0% RH
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm Patented progressive opening system
- Clamping capacity: Cable: Ø max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm / 4 busbars of 30 x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm 15/25/uni00A0Hz 1/uni00A0mm 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: Casing and jaws: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: C116: Safety sockets (4/uni00A0mm) C117: two-wire cable with reinforced insulation or double insulation, length 1.5 m, terminated by 2/uni00A0insulated elbowed male banana plugs, Ø 4 mm SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, no DC components, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement, load impedance ≥ 1 MΩ and ≤/uni00A0100 pF (2) Out of reference domain
Thanks to their excellent technical performance (phase shift and linearity), these /uni03BC-metal core clamps are highly recommended for wattmeter use. ELECTRICAL SPECIFICATIONS
- Current range: 0.001 A/uni00A0AC .. 1,200 A/uni00A0AC
- Output signal: 1 mV AC / A/uni00A0AC (1 V for 1,000 A)
- Accuracy and phase shift (1): % Accuracy of output signal ≤/uni00A03 % + 5 µA≤/uni00A02 % + 3 µA ≤/uni00A01 % ≤/uni00A00.5 % ≤/uni00A00.3 % Phase shift not specified not specified ≤/uni00A02° ≤/uni00A01° ≤/uni00A00.7° Current 1,000 A Output 1 mV / A To order Reference AC current clamp model C116 with operating manual P01120316 AC current clamp model C117 with operating manual P01120317 Current clamps for AC current Models C116 and C117
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.06 (1/1) C100 seriesCurrent clamp for AC current Model C122 An electronic voltage-limiting system protects output of clamp when operating, if the secondary circuit is opened accidentally. ELECTRICAL SPECIFICATIONS
- Current range: 1 A AC .. 1,200 A/uni00A0AC
- Current transformation ratio: 1000:5
- Output signal: 5 mA/uni00A0AC / A/uni00A0AC (5 A for 1,000 A)
- Accuracy and phase shift (1): Primary current 1 A .. 20 A 20 A 50 A (2) 200 A (2) 1,000 A (2) 1,200 A (2) Accuracy en % ≤/uni00A06 % + 0.5 mA ≤/uni00A05 % ≤/uni00A03 %≤/uni00A01.5 % ≤/uni00A01 %≤/uni00A01 % Phase shift not specified ≤/uni00A03° ≤/uni00A03°≤/uni00A01.5°≤/uni00A01° ≤/uni00A01°
- Bandwidth: 30 Hz .. 10 kHz
- Crest factor: ≤/uni00A06 for a current ≤/uni00A03,000 A peak (500 ARMS)
- Maximum currents: 1,000 A continuous for a frequency ≤ 1 kHz (limitation proportional to the inverse of frequency beyond) 1,200 A for 30 minutes max (interval between measurements > 15 minutes)
- Load impedance: ≤/uni00A00.6 /uni03A9
- Impedance of connection leads: ≤/uni00A040 m/uni03A9
- Maximum output voltage (secondary open): Electronic protection circuit limiting voltage to 30/uni00A0V peak max
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
- Infl uence of adjacent conductor: ≤/uni00A01 mA / A at 50 Hz
- Infl uence of conductor position in jaws: ≤/uni00A00.2 % of output signal for frequencies ≤/uni00A0400 Hz
- Load infl uence: From 0.2 Ω to 0.6 Ω </uni00A00.5 % on measurement </uni00A00.5 ° on phase
- Infl uence of frequency (3): </uni00A01 % of output signal from 30 Hz .. 48/uni00A0Hz </uni00A00.5 % of output signal from 65/uni00A0Hz .. 1 kHz </uni00A01 % of output signal from 1 kHz .. 5 kHz
- Infl uence of crest factor: </uni00A01 % of output signal for crest factor ≤/uni00A06 with current ≤/uni00A03,000 A peak (500/uni00A0ARMS)
- Infl uence of DC current superimposed on rated current: </uni00A01 % of output signal for a current ≤/uni00A030 A DC MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C to +50 °C
- Storage temperature: -40 °C to +70 °C
- Infl uence of temperature: ≤/uni00A00.1 % of output signal per 10 °K
- Relative humidity for operation: 0 to 85/uni00A0% RH with a linear decrease above 35 °C
- Infl uence of relative humidity: </uni00A00.2 % of output signal from 10/uni00A0% to 85/uni00A0% RH
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm Patented progressive opening system
- Clamping capacity: - Cable: Ø max 52/uni00A0mm - Busbar: 1 busbar of 50 x 5/uni00A0mm / 4 busbars of 30 x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm 15/25/uni00A0Hz 1/uni00A0mm 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: Casing and jaws: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: Safety sockets (4/uni00A0mm) SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, no DC components, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement, load impedance 0.2 Ω (5 VA) (2) Accuracy class in accordance with IEC/uni00A0185: 5 VA - class 1 - 48 .. 65/uni00A0Hz (3) Out of reference domain Current 1,000 A Ratio 1000 / 5 Output 5 mA / A To order Reference AC current clamp model C122 with operating manual P01120306
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.07 (1/2) C100 seriesCurrent clamp for AC current An electronic voltage-limiting system protects output of clamp when operating if the secondary circuit is opened accidentally. ELECTRICAL SPECIFICATIONS
- Current range: 1 A/uni00A0AC .. 300 A/uni00A0AC 1 A/uni00A0AC .. 600 A/uni00A0AC 1 A/uni00A0AC .. 1,200 A/uni00A0AC
- Current transformation ratio 250:5 500:5 1000:5
- Output signal: 20 mA/uni00A0AC / A/uni00A0AC (5 A for 250 A) 10 mA/uni00A0AC / A/uni00A0AC (5 A for 500 A) 5 mA/uni00A0AC / A/uni00A0AC (5 A for 1,000 A)
- Accuracy and phase shift (1): I 250 A calibre Primary current 1 A .. 5 A 5 A 12,5 A (2) 50 A (2) 250 A (2) 300 A (2) Accuracy en % ≤/uni00A010 % + 2 mA ≤/uni00A010 % ≤/uni00A05 % ≤/uni00A02.5 % ≤/uni00A02 % ≤/uni00A02 % Phase shift not specified not specified ≤/uni00A010°≤/uni00A010°≤/uni00A010°≤/uni00A010° I 500 A calibre Primary current 1 A .. 10 A 10 A 25 A (3) 100 A (3) 500 A (3) 600 A (3) Accuracy en % ≤/uni00A06 % + 1 mA ≤/uni00A06 % ≤/uni00A03 % ≤/uni00A02 %≤/uni00A01 % ≤/uni00A01 % Phase shift not specified ≤/uni00A06° ≤/uni00A04°≤/uni00A03°≤/uni00A02.5 °≤/uni00A02.5 ° I 1,000 A calibre Primary current 1 A .. 20 A 20 A 50 A (4) 200 A (4) 1,000 A (4) 1,200 A (4) Accuracy en % ≤/uni00A06 % + 0.5 mA ≤/uni00A05 % ≤/uni00A03 % ≤/uni00A01.5 % ≤/uni00A01 % ≤/uni00A01 % Phase shift not specified ≤/uni00A05 ° ≤/uni00A03°≤/uni00A01.5°≤/uni00A01°≤/uni00A01°
- Bandwidth: 48/uni00A0Hz .. 1 kHz
- Crest factor: I 250 A calibre: ≤/uni00A06 with current ≤/uni00A0750 A peak I 500 A calibre: ≤/uni00A06 with current ≤/uni00A01,500 A peak I 1,000 A calibre: ≤/uni00A06 with current ≤/uni00A03,000 A peak
- Maximum currents: 1,200 A for frequencies ≤/uni00A01 kHz for 30/uni00A0minutes max (interval between measurements >/uni00A015/uni00A0minutes)
- Load impedance: I 250 A calibre: ≤/uni00A00.2 Ω I 500 A calibre: ≤/uni00A00.4 Ω I 1,000 A calibre: ≤/uni00A00.4 Ω
- Impedance of connection leads: ≤/uni00A040 m/uni03A9
- Maximum output voltage (secondary open): Electronic protection circuit limiting voltage to 30/uni00A0V peak max
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
- Infl uence of adjacent conductor: I 250 A calibre: ≤/uni00A015 mA / A at 50 Hz I 500 A calibre: ≤/uni00A010 mA / A at 50 Hz I 1,000 A calibre: ≤/uni00A01 mA / A at 50 Hz
- Infl uence of conductor position in jaws: For frequencies ≤/uni00A0400 Hz I 250 A calibre: ≤/uni00A00.6 % of output signal I 500 A calibre: ≤/uni00A00.4 % of output signal I 1,000 A calibre: ≤/uni00A00.2 % of output signal
- Load infl uence: I 250 A calibre: from 25 mΩ to 0.2 Ω </uni00A02 % on measurement </uni00A04° on phase I 500 A calibre: from 50 mΩ to 0.4 Ω </uni00A01 % on measurement </uni00A02° on phase I 1,000 A calibre: from 50 mΩ to 0.4 Ω </uni00A00.5 % on measurement </uni00A00.5 ° on phase
- Infl uence of frequency (5): I 250 A calibre: </uni00A01 % of output signal from 65/uni00A0Hz .. 100 Hz </uni00A05 % of output signal from 100 Hz .. 1 kHz I 500 A calibre: </uni00A01 % of output signal from 65/uni00A0Hz .. 1 kHz I 1,000 A calibre: </uni00A00.5 % of output signal from 65/uni00A0Hz .. 100 Hz </uni00A01 % of output signal from 100 Hz .. 1 kHz
- Infl uence of crest factor: </uni00A01 % of output signal for crest factor ≤/uni00A06 with current: ≤/uni00A0750 A peak (calibre 250 A) ≤/uni00A01,500 A peak (calibre 500 A) ≤/uni00A03,000 A peak (1,000 A calibre)
- Infl uence of DC current superimposed on rated current: </uni00A01 % of output signal for a current ≤/uni00A030 A DC Current 250 A/uni00A0AC500 A/uni00A0AC1,000 A/uni00A0AC Ratio 250:5 500:5 1000:5 Output 20 mA / A 10 mA / A 5 mA / A Current clamp for AC current Model C148
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.07 (2/2) C100 seriesCurrent clamp for AC current (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement, load impedance: - 250 A calibre: 0.1 Ω (2.5 VA) - 500 A calibre: 0.2 Ω (5 VA) - 1,000 A calibre: 0.2 Ω (5 VA) (2) Accuracy class in accordance with IEC/uni00A0185: 2.5 VA - class 3 - 48-65/uni00A0Hz (3) Accuracy class in accordance with IEC/uni00A0185: 5 VA - class 3 - 48-65/uni00A0Hz (4) Accuracy class in accordance with IEC/uni00A0185: 5 VA - class 1 - 48-65/uni00A0Hz (5) Out of reference domain MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C to +50 °C
- Storage temperature: -40 °C to +70 °C
- Infl uence of temperature: ≤/uni00A00.15 % of output signal per 10 °K
- Relative humidity for operation: 0 to 85/uni00A0% RH decreasing linearly above 35 °C
- Infl uence of relative humidity: From 10/uni00A0% to 85/uni00A0% RH I 250 A calibre: </uni00A00.6 % of output signal and </uni00A02° on phase I 500 A calibre: </uni00A00.4 % of output signal and </uni00A00.6° on phase I 1,000 A calibre: </uni00A00.2 % of output signal and </uni00A00.2° on phase
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm Patented progressive opening system
- Clamping capacity: Cable: Ø max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm / 4 busbars of 30 x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1/uni00A0m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm 15/25/uni00A0Hz 1/uni00A0mm 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: Safety jacks (4/uni00A0mm) SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 To order Reference AC current clamp model C148 with operating manual P01120307 Current clamp for AC current Model C148
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.08 (1/6) C100 series This 1,000 A AC clamp can be used for easy display and measurement of current curves. Equipped with a coaxial cable terminated by a BNC connector, it is ideal for use with any oscilloscope. It outputs a signal in mV directly proportional to the current. It offers 3 different sensitivities. ELECTRICAL SPECIFICATIONS
- Current range: 0.1 A AC .. 10 A AC (30 A peak) 1 A AC .. 100 A AC (300 A peak) 1 A AC .. 1,000 A AC (2,000 A peak)
- Output signal: 100 mV AC / A/uni00A0AC (1 V for 10 A) 10 mV AC / A/uni00A0AC (1 V for 100 A) 1 mA/uni00A0AC / A/uni00A0AC (1 V for 1,000 A)
- Accuracy and phase shift (1): I 10 A calibre % Accuracy of output signal ≤/uni00A03 % + 10 mV ≤/uni00A03 % + 10 mV ≤/uni00A03 % + 10 mV ≤/uni00A03 % + 10 mV Phase shift not specified not specified ≤/uni00A015 ° ≤/uni00A015 ° I 100 A calibre % Accuracy of output signal ≤/uni00A02 % + 5 mV ≤/uni00A02 % + 5 mV ≤/uni00A02 % + 5 mV ≤/uni00A02 % + 5 mV Phase shift not specified ≤/uni00A015 ° ≤/uni00A010° ≤/uni00A05 ° I 1,000 A calibre % Accuracy of output signal ≤/uni00A01 % + 1 mV ≤/uni00A01 % + 1 mV ≤/uni00A01 % + 1 mV ≤/uni00A01 % + 1 mV Phase shift not specified ≤/uni00A03° ≤/uni00A02° ≤/uni00A01°
- Bandwidth: 10 Hz .. 100 kHz (-3 dB) (depending on current value)
- Rise/fall time from 10 % to 90 %: 3.5 µs
- 10 % delay time: 0.5 µs
- Ampere second product: I 10 A calibre: 3,2 A.s I 100 A calibre: 26 A.s I 1,000 A calibre: 64 A.s
- Maximum currents: 1,000 A permanent 1,200 A for 40 minutes max. / > 20 minutes shutdown for a frequency ≤ 1 kHz (limitation proportional to the inverse of one third of the frequency beyond that)
- Insertion impedance (at 400 Hz / 10 kHz) I 10 A calibre: < 0.3 m/uni03A9 / < 6,6 m/uni03A9 I 100 A calibre: < 0.3 m/uni03A9 / < 2 m/uni03A9 I 1,000 A calibre: < 0.3 m/uni03A9 / < 1.6 m/uni03A9
- Output impedance at 1 kHz: I 10 A calibre: ≤ 515 /uni03A9 ± 10 % I 100 A calibre: ≤ 515 /uni03A9 ± 10 % I 1,000 A calibre: ≤ 515 /uni03A9 ± 10 %
- Infl uence of temperature: ≤ 150 ppm /k or 0.15 % of output signal per 10 °K
- Infl uence of relative humidity: < 0.1 % of output signal
- Infl uence of adjacent conductor: ≤/uni00A01/uni00A0mA / A at 50 Hz
- Infl uence of DC current ≤ 30 A superimposed on rated current: < 1 %
- Infl uence of conductor position in jaws: ≤/uni00A00.1 % of output signal for frequencies ≤/uni00A0400 Hz
- Infl uence of frequency (2): I 10 A calibre: </uni00A010 % of output signal from 10 Hz .. 1 kHz </uni00A05 % of output signal from 1 kHz .. 10 kHz </uni00A020 % of output signal from 10 kHz .. 50 kHz 3 dB of output signal from 50 kHz .. 100 kHz I 100 A calibre: </uni00A05 % of output signal from 10 Hz .. 1 kHz </uni00A03 % of output signal from 1 kHz .. 10 kHz </uni00A020 % of output signal from 10 kHz .. 50 kHz 3 dB of output signal from 50 kHz .. 100 kHz I 1,000 A calibre: </uni00A01 % of output signal from 10 Hz .. 1 kHz </uni00A02 % of output signal from 1 kHz .. 10 kHz </uni00A010 % of output signal from 10 kHz .. 50 kHz 3 dB of output signal from 50 kHz .. 100 kHz
- Infl uence of crest factor: </uni00A01 % of output signal for crest factor ≤/uni00A06 with current I 10 A calibre: ≤/uni00A030 A peak I 100 A calibre: ≤/uni00A0300 A peak I 1,000 A calibre: ≤/uni00A03,000 A peak Current 30 A peak 300 A peak 2,000 A peak Output 100 mV / A 10 mV / A 1 mV / A Oscilloscope clamp for AC current Model C160 (insulated AC current probe)
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.08 (2/6) C100 series (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz à 1,000 Hz, distortion factor </uni00A01 % with no DC component, external magnetic fi eld </uni00A040/uni00A0A/m, no DC components, no external conductor with circulating current, conductor centred for measurement, load impedance: ≥ 1 MΩ and </uni00A0100 pF (2) Out of reference domain MECHANICAL SPECIFICATIONS
- Max. jaw opening: 53/uni00A0mm
- Clamping capacity: Cable: Ø max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm/ 4 busbars of 30/uni00A0x/uni00A05/uni00A0mm
- Operating temperature: -10 °C to +55 °C
- Storage temperature: -40 °C to +70 °C
- Relative humidity for operation: 0 to 85/uni00A0% RH decreasing linearly above 35 °C
- Operating altitude: 0 to 2,000 m
- Casing protection rating: IP30 with clamp open (IEC 529) IP40 with clamp closed (IEC 529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g / 6 ms / half-periode (IEC 68-2-27)
- Protection against impacts: IK04 0.5 J (EN 50102)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm peak 15/25/uni00A0Hz 1/uni00A0mm peak 25/55/uni00A0Hz 0.25/uni00A0mm peak (IEC/uni00A068-2-6)
- Self-extinguishing capability: Casing and jaws: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: Via 2 m coaxial cable terminated by insulated BNC plug SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 without disturbance: 4 kV class 2 non-destructive: 15 kV class 4 - Radiated field: IEC/uni00A01000-4-3 without disturbance: 10 V/m performance criterion/uni00A0A - Fast transients: IEC/uni00A01000-4-4 without disturbance: 1 kV class 2 non-destructive: 2 kV class 3 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 field of 400 A/m at 50 Hz: < 1 A To order Reference AC current clamp model C160 with operating manual P01120308 Oscilloscope clamp for AC current Model C160 (insulated AC current probe)
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.08 (3/6) C100 seriesOscilloscope clamp for AC current Model C160 (insulated AC current probe) CURVES AT 50 Hz
10 A calibre
Error on measurement Phase shift Courant primaire en A AC Erreur en % -30.00 -25.00 -20.00 -15.00 -10.00 -5.00 0.00 5.00 Erreur typique Limite du FT Courant primaire en A AC Phase en ° 0.00 5.00 10.00 15.00 20.00 25.00 Phase typique Limite du FT
100 A calibre
Error on measurement Phase shift Courant primaire en A AC Phase en ° 0.00 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 Phase typique Limite du FT 1,000 A calibre Error on measurement Phase shift Error in % Phase in % Error in %Error in % Phase in %Phase in % PRIMARY CURRENT IN A AC PRIMARY CURRENT IN A AC PRIMARY CURRENT IN A AC Typical error FT limit Typical error FT limit Typical error FT limit Typical phase FT limit Typical phase FT limit Typical phase FT limit PRIMARY CURRENT IN A AC PRIMARY CURRENT IN A AC PRIMARY CURRENT IN A AC
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.08 (4/6) C100 series FREQUENCY RESPONSE (CONT.) 1,000 A calibre Fréquence en Hz Phase en ° -40.00 -35.00 -30.00 -25.00 -20.00 -15.00 -10.00 -5.00 0.00 5.00 à 10 A à 50 A Oscilloscope clamp for AC current Model C160 (insulated AC current probe) Error in %Error in %Error in %Error in % Phase in %Phase in % Phase in %Phase in % Frequency in Hz Frequency in Hz Frequency in Hz Frequency in Hz at 10 A at 50 A at 500 A at 1,000 A at 1 A at 5 A at 50 A at 100 A at 1 A at 5 A at 50 A at 100 A at 500 A at 1,000 A at 10 A at 50 A Frequency in Hz Frequency in Hz Frequency in Hz Frequency in Hz
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.08 (5/6) C100 seriesOscilloscope clamp for AC current Model C160 (insulated AC current probe) RESPONSE TO A SQUARE SIGNAL 1,000 A calibre
10 A à 1 Hz
10 A à 10 Hz
10 A à 1 kHz
10 A à 10 kHz
FREQUENCY RESPONSE (CONT.) Error in %Error in % Phase in %Phase in % Primary current Primary current Primary current Primary current Secondary voltage Secondary voltage Secondary voltage Secondary voltage
10 A at 1 Hz
10 A at 1 kHz
10 A at 10 Hz
10 A at 10 kHz
at 0.1 A at 0.5 A at 5 A at 10 A at 5 A at 10 A at 0.1 A at 0.5 A FREQUENCY IN HZ FREQUENCY IN HZ
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.08 (6/6) C100 seriesOscilloscope clamp for AC current Model C160 (insulated AC current probe) RESPONSE TO A SQUARE SIGNAL (CONT.)
10 A à 100 Hz
10 A at 100 Hz
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.09 (1/2) C100 seriesCurrent clamp for AC current The C173 clamp measures leakage or differential currents from 1 mA upwards and can also be used with multimeters equipped with a range in mV AC. The C173 clamp measures earth-loop currents and leakage currents. It/uni00A0 also locates faults in circuits of single and three-phase networks. For unearthed three-phase systems, use the optional Artifi cial Neutral. ELECTRICAL SPECIFICATIONS
- Current range: 0.001 A/uni00A0AC .. 1.2 A/uni00A0AC 0.01 A/uni00A0AC .. 12 A/uni00A0AC 0.1 A/uni00A0AC .. 120 A/uni00A0AC 1 A/uni00A0AC .. 1,200 A/uni00A0AC
- Output signal:
1 V AC / A/uni00A0AC (1 V for 1 A)
100 mV AC / A/uni00A0AC (1 V for 10 A) 10 mV AC / A/uni00A0AC (1 V for 100 A) 1 mV AC / A/uni00A0AC (1 V for 1,000 A)
- Accuracy and phase shift (1): I 1 A calibre % Accuracy of output signal ≤/uni00A03 % + 1 mV ≤/uni00A03 % + 1 mV ≤/uni00A00.7 % + 1 mV ≤/uni00A00.7 % + 1 mV Phase shift not specified not specified ≤/uni00A010° ≤/uni00A010° I 10 A calibre Phase shift not specified ≤/uni00A05 ° ≤/uni00A02° ≤/uni00A02° I 100 A calibre Phase shift not specified ≤/uni00A02° ≤/uni00A01° ≤/uni00A01° I 1,000 A calibre Phase shift not specified ≤/uni00A02° ≤/uni00A01° ≤/uni00A01°
- Bandwidth: 10 Hz .. 3 kHz
- Crest factor: I 1 A calibre: ≤/uni00A03 for I ≤/uni00A03 A peak (1 ARMS) I 10 A calibre: ≤/uni00A03 for I ≤/uni00A030 A peak (10 A RMS) I 100 A calibre: ≤/uni00A03 for I ≤/uni00A0300 A peak (100 A RMS) I 1,000 A calibre: ≤/uni00A03 for I ≤/uni00A01700 A peak (500 A RMS)
- Maximum currents: 1,000 A continuous for a frequency ≤/uni00A0500 Hz (limitation proportional to the inverse of 1/2 of frequency beyond)
- Load impedance: ≥ 10 M/uni03A9 and ≤/uni00A047 pF
- Output impedance: I 1 A calibre: 10 k/uni03A9 ± 10 % I 10 A calibre: 1 k/uni03A9 ± 10 % I 100 A calibre: 100 /uni03A9 ± 10 % I 1,000 A calibre: 100 /uni03A9 ± 10 %
- Operating voltage: 600/uni00A0VRMS
- Common mode voltage:
- Infl uence of adjacent conductor: ≤/uni00A01 mA / A at 50 Hz
- Infl uence of conductor position in jaws: ≤/uni00A00.3 % of output signal for frequencies ≤/uni00A0400 Hz
- Infl uence of frequency (2): I 1 A calibre: </uni00A02 % of output signal 30 Hz .. 48/uni00A0Hz and 65/uni00A0Hz .. 1 kHz </uni00A010 % of output signal 1 kHz .. 3 kHz I 10 A calibre: </uni00A02 % of output signal 10 Hz .. 48/uni00A0Hz and 65/uni00A0Hz .. 3 kHz I 100 A calibre: </uni00A01.5 % of output signal 10 Hz .. 48/uni00A0Hz and 65/uni00A0Hz .. 3 kHz I 1,000 A calibre: </uni00A01 % of output signal 10 Hz .. 48/uni00A0Hz and 65/uni00A0Hz .. 1 kHz
- Infl uence of crest factor: ≤/uni00A00.5 % for crest factor limited to 3
- Infl uence of DC current superimposed on rated current: ≤/uni00A010 % at 1,000 A for a current DC from 10 A Current 1 A 10 A 100 A 1,000 A Output 1 V / A 100 mV / A 10 mV / A 1 mV / A Current clamp for AC current Model C173 (probe for leakage currents)
Serie MN Pinces ampèremetriques pour courant AC Non-contractual document 906131102D - Ed 1 4.09 (2/2) C100 seriesCurrent clamp for AC current MECHANICAL SPECIFICATIONS
- Operating temperature: -10 °C .. +50 °C
- Storage temperature: -40 °C .. +70 °C
- Infl uence of temperature: ≤/uni00A00.15 % of output signal per 10 °K from -10 °C .. +40 °C ≤/uni00A00.2 % of output signal per 10 °K from +40 °C .. +50 °C
- Relative humidity for operation: From 0 .. 85 % from RH decreasing linearly above 35 °C
- Infl uence of relative humidity: </uni00A00.1 % of output signal from 10 .. 85 % from RH
- Operating altitude: 0 to 2,000 m
- Max. jaw opening: 53/uni00A0mm Patented progressive opening system
- Clamping capacity: Cable: Ø/uni00A0max 52/uni00A0mm Busbar: 1 busbar of 50 x 5/uni00A0mm or 4 busbars of 30 x 5/uni00A0mm
- Casing protection rating: IP40 (IEC/uni00A0529)
- Drop test: 1 m (IEC/uni00A068-2-32)
- Shock resistance: 100 g (IEC/uni00A068-2-27)
- Vibration resistance: 5/15/uni00A0Hz 1.5/uni00A0mm 15/25/uni00A0Hz 1/uni00A0mm 25/55/uni00A0Hz 0.25/uni00A0mm (IEC/uni00A068-2-6)
- Self-extinguishing capability: UL94 V0
- Dimensions: 216 x 111 x 45/uni00A0mm
- Weight: 550 g
- Colours: Dark grey case with red jaws
- Output: 1.5 m two-wire lead with double or reinforced insulation terminated by 2 elbowed male safety plugs (4 mm) SAFETY SPECIFICATIONS
- Electrical safety: Instrument with double insulation or reinforced insulation between the primary the secondary and the grippable part located under the guard as per IEC/uni00A01010-1 & IEC/uni00A01010-2-032 - 600/uni00A0V category III, pollution degree 2 - 300/uni00A0V category IV, pollution degree 2
- Electromagnetic compatibility (EMC): EN/uni00A050081-1: class/uni00A0B EN/uni00A050082-2: - Electrostatic discharge: IEC/uni00A01000-4-2 - Radiated field: IEC/uni00A01000-4-3 - Fast transients: IEC/uni00A01000-4-4 - Magnetic field at 50/60 Hz: IEC/uni00A01000-4-8 (1) Conditions of reference: 23 °C ± 3 °K, 20 % to 75 % RH, signal sinus, frequency of 48/uni00A0Hz to 65/uni00A0Hz, distortion factor </uni00A01 %, no DC components, external magnetic fi eld </uni00A040/uni00A0A/m, no AC magnetic fi eld, conductor centred for measurement, load impedance: ≥/uni00A010/uni00A0MΩ and ≤/uni00A047 pF (2) Out of reference domain To order Reference AC current clamp model C173 with operating manual P01120309 Accessory: AN1 artificial neutral box (see capter 12) Bag n°11 P01197201 P01100120 Current clamp for AC current Model C173 (probe for leakage currents)