© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com Ion Pro™ Benchtop Ionizer Installation, Operation and Maintenance December 2021
Description
The SCS Ion Pro™ Benchtop Ionizer is used for neutralizing electrostatic charges on insulators and ungrounded conductors. Its fast discharge times (< 1.5 seconds at 12 inches) and ±3 volt offset voltage exceed the required limits of ANSI/ESD S20.20 and ESD TR53. The Ion Pro Benchtop Ionizer monitors both its offset voltage (balance) and discharge time. The ionizer will alarm should its offset voltage exceed the required limits defined by ANSI/ESD S20.20 or discharge time take longer than 4.5 seconds. Steady-state DC emitters are spaced farther apart to prevent ion recombination, increase decay efficiency, and provide a broader coverage area of ionization. The Ion Pro™ Benchtop Ionizer’s dual-speed fan produces consistent airflow performance. The Ion Pro™ Benchtop Ionizer features a stainless steel enclosure and multi-mount stand to minimize contamination and corrosion when used in clean or sensitive areas. The SCS Ion Pro™ Benchtop Ionizer operates on Steady-state DC. Steady-state DC systems consist of separate negative and positive ion emitters connected by a pair of high-voltage cables to their respective high-voltage power supplies. DC power is constantly applied to the emitter points. The ionizer utilizes patented capacitance-based sampling from its internal sensor to continuously adjust the output, maintain its offset voltage, and minimize maintenance. Made in the United States of America USER GUIDE TB-9107 “The primary method of static charge control is direct connection to ground for conductors, static dissipative materials, and personnel. A complete static control program must also deal with isolated conductors that cannot be grounded, insulating materials (e.g., most common plastics), and moving personnel who cannot use wrist or heel straps or ESD control flooring and footwear. Air ionization is not a replacement for grounding methods. It is one component of a complete static control program. Ionizers are used when it is not possible to properly ground everything and as backup to other static control methods. In clean rooms, air ionization may be one of the few methods of static control available.” [ESD Handbook ESD TR20.20 Ionization, section 5.3.6.1 Introduction and Purpose / General Information] The Ion Pro™ Benchtop Ionizer and its accessories are available as the following item numbers: Item Description
770116 Ion Pro™ Benchtop Ionizer, 120 VAC
770117 Ion Pro™ Benchtop Ionizer, 220 VAC
770000 Power Cord, North America Plug
770001 Power Cord, UK Plug
770002 Power Cord, Europe Plug
770003 Power Cord, China Plug
770118 Power Cord, Daisy Chain
770047 Benchtop Ionizer Boom Arm
770055 SMP Software
Figure 1. SCS Ion Pro™ Benchtop Ionizer into a database for on-going quality control purposes. balance (offset voltage) and discharge time statuses. SMP is sold separately. Click here to learn more.
balance exceeds the negative balance alarm threshold. when the discharge time exceeds the alarm threshold. settings for the audible alarm and balance alarm. while the ionizer is powered.
- The ionizer will toggle its
setting disables the buzzer.
- Push and hold the switch while
restoring power to the ionizer.
- The buzzer will chirp once to
- Push and hold the switch while
restoring power to the ionizer.
- The buzzer will chirp twice to
D. Power Inlet: Connect the power cord here. the 0 position to power the ionizer OFF.
1 Ion Pro™ Benchtop Ionizer
1 Power Cord with North America Plug (770116 only)
1 Emitter Point Cleaner
1 Certificate of Calibration
Figure 2. Ion Pro™ Benchtop Ionizer features and
ionizer’s power cord into an appropriate AC power source. EM Aware Monitor, and Ground Master Monitor. Figure 3. SMP system setup
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com DYNAMIC IP PROCEDURE 1. Verify that the Ethernet cable is securely connected to your network and Ion Pro™ Benchtop Ionizer. The LEDs on the Ethernet port will illuminate when a connection to the network is established. 2. Open SCS SMP Server Monitor. 3. The SMP Server icon will appear in the Windows taskbar. 4. Click the SMP Server icon located on the Windows taskbar. The SMP Server menu will appear. Click the Start button to start SMP Server.
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com 5. A message will appear and display “Starting SMP Server…” Click the SMP Server icon located in the Windows taskbar. 6. SMP Server will now indicate Active server status. 7. SMP Server will search and find the Ion Pro™ Benchtop Ionizer on the network. The Device and Active counts will increase to 1.
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com 8. Open SCS SMP Admin. 9. The window for SMP Server Admin will appear. “1 unassigned” is representative of the Ion Pro™ Benchtop Ionizer that was just found by SMP Server. Click the Add New Building button.
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com 10. Enter a building name, and then click the OK button. 11. The building name will appear on the navigation pane in SMP Server Admin. 12. Click on the building name in the navigation pane, and then click the Add New Floor button.
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com 13. Enter a floor name, and then click the OK button. 14. The floor name will appear on the navigation pane in SMP Server Admin. 15. Click on the floor name in the navigation pane, and then click the Add New Line button.
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com 16. Enter a line name, and then click the OK button. 17. The line name will appear on the navigation pane in SMP Server Admin. 18. Click on the line name in the navigation pane, and then click the Attachable Devices arrow located at the bottom of the window.
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com 19. The serial number for the Ion Pro™ Benchtop Ionizer will appear. This number will resemble the number found on the monitor's serial label with the exception of one zero. For example, if the serial number on the label says "XXXXX2025001", the serial number in SMP will read "XXXXXX2025 01". Click attach. 20. The Ion Pro™ Benchtop Ionizer will attach to the selected line.
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com 21. Open SCS SMP Client. 22. Verify that the Ion Pro™ Benchtop Ionizer was added to the appropriate building, floor and line.
STM3.1. Please refer to this standard for more complete information. than 1.5 seconds, and the balance should be ±3V or better. standard are shown in the figure below. were recorded in a factory ambient environment. Figure 7. Neutralization (Discharge) Times at 120VAC / 220VAC, 60Hz input
© 2021 DESCO INDUSTRIES INC Employee Owned SCS - 926 JR Industrial Drive, Sanford, NC 27332 East: (919) 718-0000 | West: (909) 627-9634 • Website: StaticControl.com Specifications Input Voltage and Frequency 100-240VAC, 50/60 Hz Neutralization (Discharge) Time at 12" < 1.5 seconds Offset Voltage (Balance) at 12" ±3 V typical Ion Emission Steady-state DC Airflow Dual speed, 77 CFM max. Ozone <0.05 ppm Discharge Time Alarm Threshold at 12" 4.5 s Offset Voltage Alarm Threshold (Standard Mode) ±35 V approx. Offset Voltage Alarm Threshold (CE Mode) ±7 V approx. Power Consumption 10 W Emitter Points Tungsten Dimensions (with stand) 9.1" H x 6.3" W x 3.1" D (231 mm x 160 mm x 79 mm) Weight 4.4 lbs (2.0 kg) U.S. Patents 9,404,945; 9,588,161 Country of Origin United States of America Limited Warranty, Warranty Exclusions, Limit of Liability and RMA Request Instructions See the SCS Warranty - StaticControl.com/Limited-Warranty.aspx