7615 BURKERT | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 10
Technical content
Visit product website DATA SHEET 1000185431 | Standard | EU | en | 202105 | 07 Micro Dosing Unit for precise dosing in microlitre-range
- Diaphragm pump for high chemical resistance
- Dosing volume 5 µl/stroke up to 8 ml/min. in both directions
- Dosing accuracy < +/- 3,5 %
- Self priming
- Integrated electronics, easy to use Bürkert’s Micro Dosing Unit has been designed for precise dosing applica- tions in the microliter range. It combines high dosing accuracy and precision with excellent chemical inertness. The unit is comprised of three valves which can be opened simultaneously for flushing. Active inlet and outlet valves en- able the device to pump liquid in two directions. This feature can be used to mix fluids inside a tube or channel or to constantly keep them in motion. Rea- gents can be sucked back from the dispensing tip to prevent drying out after dosing. The integrated heating function heats up the valves and the medium, and the unit comes with dry running capabilities. Product variants described in the data sheet may differ from the product presentation and description. Type description Type 7615 © Christian Bürkert GmbH & Co. KG, Subject to alteration
2 | 10 Visit product website Table of contents 1. Gener al Technical Data 3 P roduct versions Materials 4. Dimensions 5 4.2. Thr eaded version (UNF ¼"- 28) 5. Device/P rocess connections 7 6. Performance specifications 7 7. P roduct operation 7 8. P roduct accessories 8 9. Or dering information 8 9.2. Bürkert product filter 9.3. Or dering chart 9.4. Or dering chart accessories Connection cable, fittings and hoses
3 | 10 Visit product website 1. Gener al Technical Data Product properties Dimensions Detailed information can be found in chapter “4. Dimensions” on page 5. Material Seal FFKM, EPDM pump diaphragm on request Body PEEK Fluidics PEEK Cover PPS (not in contact with media) Base plate Stainless steel (not in contact with media) Dosing quantity Adjusted to µl/stroke 1.5 %1.), max. approx. 8 ml/min; at 40 Hz bidirectional Dosing accur acy ± 3.5 %2.) Repeatability ± 2 %1.) Max. suction Lift > 0.7 m (dry) 4 m (wet) Weight Ca. 120 g Typical product service life Approx. M cycles (at °C,
10 Hz,
water) Performance data Max. outlet Pressure 1.0 bar3.) Electrical data Operating voltage 12 V DC,
24 V DC
(short term), 5 W Duty cycle 100 Voltage tolerance ± Medium data Medium t emperature + °C…+ (FFKM) °C…+ (EPDM)4.) Operating medium Neutral and aggressive liquids See also chapter “3.1. Chemical Resistance Chart – Bürkert resistApp” on page 4. Viscosity (max.) < 250 mm2/s Process/Port connection & communication Port connection Sub-base connection, UNF ¼"- Electrical connection E.g. suitable for connection t o Molex plug no. 50 9404 See also chapter “9.4. Ordering chart accessories” on page 9. Environment and installation Installation Variable, unit with two holes for fixing screws Degree of protection IP40 Ambient temperature + °C…+ 55°C4.) 1.) At 20 °C, 5 Hz, medium deionised, degassed water without backpressure 2.) T ypical value under Burkert test conditions see 1.). The typical repeatability is This is a guiding value only! 3.) Overpressure above atmospheric pressure 4.) For lower temperatures the unit can be electrically preheated (heating mode without media feed).
4 | 10 Visit product website 2. P roduct versions Sub-base version Threaded version (UNF ¼"- 28) 3. Mat erials 3.1. Chemical Resistance Char t – Bürkert resistApp Bürkert resistApp – Chemical Resistance Chart You want to ensure the reliability and durability of the materials in your individual application case? Verify your combination of media and materials on our website or in our resistApp. Start Chemical Resistance Check
5 | 10 Visit product website 4. Dimensions 4.1. Sub-base version Note:
- Dimensions in mm
- Connection cable not included in delivery 11.8 34.5 4.1 min. t=6 (2.9) max. Ø 2.5 6.5 20.5 32.3 44.8 7.9 Ø 3.2 Ø 6 t=3 (76.1) (83.6) 7.1 Ø 3.3 11.8 28.4 34.5 70.5 Flange interface for manifold
6 | 10 Visit product website 4.2. Thr eaded version (UNF ¼"- 28) Note:
- Dimensions in mm
- Connection cable not included in the delivery 8.9 12.1 10.9 20.5 32.3 50.7 7.9 Ø 3.2 Ø 6 t=3 (76.1) (83.6) 7.2 UNF 1/4"-28 t=7 28.8 70.5 for M3 4.45 5.9 LED 11.8
7 | 10 Visit product website 5. Device/P rocess connections 5.1. Electrical connection PIN Forward dosing Reverse dosing Flush mode (all valves open) Heating Up mode 1 (Supply) + + not connected + (Reverse) not connected + + + 3 (Ground) – – – – 4 (Forward) + not connected + + 6. P erformance specifications 6.1. Dosing Function mode Dosing per stroke Dosing accuracy Max. possible dosing Impulse 120 ms 5 µl ± 3.5 %1.) 2.4 ml/min Frequency 40 Hz approx. 3.8 µl – > 8 ml/min Frequency 25 Hz 4.6 µl approx. % 7.0 ml/min Frequency 10 Hz 4.95 µl ± % 3.0 ml/min Frequency
5 Hz 5 µl ±
3.5 %1.) 1.5 ml/min Service mode 3.5 %1.) 500 µl Impulse 50 ms 4.8 µl ± % 5.8 ml/min 1.) The typical repeatability is ± 2 %. This is a guiding value only! 7. P roduct operation 7.1. F unction modes Note: Further information can be found in the operating instructions, see Type 7615 Different modes can be preselected via a rotary switch on the electronics:
- Impulse mode: An electrical power signal of min. ms (mode resp. 120 ms (mode length creates a dosing of µl at a time. Higher volumes (multiples of μl) can be dosed by repeating the signal.
- Frequency mode: The unit runs with a predefined frequency. Different frequencies can be selected (see following description). The standard setting is 5 Hz. Function modes Description Rotary switch to select mode Impulse mode 120 ms Frequency mode 40 Hz Frequency mode 25 Hz Frequency mode 10 Hz Frequency mode 5 Hz (Standard) Service mode 5 Hz (pump stops after 100 cycles) Quick impulse mode 50 ms
8 | 10Visit product website 8. P roduct accessories 8.1. Connection cable (15,2) 500 Pin 4: yellow Pin 3: blue Pin 2: green Pin 1: red 9. Or dering information 9.1. Bürk ert eShop – Easy ordering and quick delivery Bürkert eShop – Easy ordering and fast delivery You want to find your desired Bürkert product or spare part quickly and order dir ectly? Our online shop is available for you 24/7. Sign up and enjoy all the benefits. Order online now 9.2. Bürk ert product filter Bürkert product filter – Get quickly to the right product Y ou want to select products comfortably based on your technical requirements? Use the Bürkert product filter and find suitable articles for your application quickly and easily Try out our product filter 9.3. Or dering chart Note:
- Two stainless steel screws x are included in the scope of delivery of the flange version.
- The connection cable is not included in the scope of delivery and must be ordered separately, see “9.4. Ordering chart accesso- ries” on page 9. Version Voltage Connection Seal material Function mode 1.) Article no. Standard 24 V DC Sub-base FFKM 5 Hz 238190 Standard 24 V DC UNF ¼"-
28 FFKM 5 Hz 215793
Standard 24 V DC Sub-base FFKM/EPDM 5 Hz 238193 Standard 24 V DC UNF ¼"-
28 FFKM/EPDM 5 Hz 238194
Standard 12 V DC UNF ¼"-
28 FFKM/EPDM 5 Hz 238195
1.) Mode can be changed, please see “7.1. Function modes” on page 7 or Operating Instructions.
9 | 10Visit product website 9.4. Or dering chart accessories Connection cable, fittings and hoses Accessories Description Article no. Connection cable (4 pin) with 500 mm wire length 683613 Fittings and hoses1.) see data sheet TVU003
Bürkert – Close to You Austria Belgium Czech Republic Denmark Finland France Germany Italy Netherlands Norway Poland Spain Sweden Switzerland Turkey United Kingdom China Hong Kong India Japan Korea Malaysia Philippines Singapore Taiwan South Africa Russia United Arab Emirates Brazil Uruguay Canada USA Australia New Zealand Credits, © and concept: Christian Bürkert GmbH & Co. KG | Photographs: Marc Eggimann Fotografie - 4051 Basel | Scanner GmbH - Werbeagentur Künzelsau -
74653 Künzelsau