Single Output DC Power Supplies Data Sheet

standards — at an affordable price. power supplies family to provide a wider range of power source alternatives. Figure 1. The U8001A 90 W and U8002A 150 W single output DC power supplies

enabling brighter data display Highly visible LCD displaying actual values of output voltage and current in meter mode and limit values of voltage and current in limit mode Three storage locations (M1, M2, and M3) for user-defi ned operating states with save-and-recall capability Fully integrated over voltage and over current protection to prevent damages to the DUT Keypad Lock/Unlock options to prevent accidental front panel usage Selects the knob control function for voltage and current adjustment Control for enabling and disabling the output of the power supply Setting and display of the voltage and current limit values Versatile binding posts offer fl exibility to use safety test leads, banana plugs or stripped wires Rotary knob for quick and analog-like control of voltage and current Features of the U8000 Series

Performance Specifi cations Electrical Specifi cations Parameter U8001A U8002A Output Ratings (at 0 °C to 40 °C) 0 to +30 V 0 to 3 A 0 to +30 V 0 to 5 A Line and Load Regulation CV: <0.01% +2 mV CC: <0.02% +2 mA Ripple and Noise (25 °C ±5 °C) CV: 12 mVp-p, <1 mVrms; CC: 3 mArms Load Transient Response Time (within 15 mV from full load to half load and from half load to full load) <50 μs Programming Accuracy <0.35% +20 mA Readback Accuracy1 (25 °C ±5 ° C) <0.35% +20 mV <0.35% +20 mA Meter Resolution Voltage: 10 mV Current: 10 mA Maximum Output Float Voltage ±240 Vdc 1 Specifi cations are based on one hour warm-up period.

Supplemental Characteristics Supplemental Characteristics Parameter U8001A U8002A Temperature Coeffi cient (for 12 months) CV: <100 ppm/ °C CC: <380 ppm/ °C CC: <300 ppm/ °C Output Voltage Overshoot (during turn-on or turn-off of AC power state with the output control set to less than 1 V) <1 V Voltage Programming Speed, to within 1% of total excursion Up Full Load No Load 150 ms 100 ms Down Full Load No Load 30 ms 450 ms Last Memory Setting Enabled Yes Over Voltage Protection Response Time <1.5 ms when the trip voltage is equal to or greater than 3 V and <10 ms when the trip voltage is less than 3 V (average time for output to drop from 90% of output voltage to 1 V after OVP condition occurred) Protection Features Parameter U8001A U8002A Over Voltage Protection Accuracy ± (% of output + offset) <0.5% +0.5 V Over Voltage Protection Programmable Range

1 V to 33 V

Over Current Protection Accuracy ± (% of output + offset) <0.5% +0.5 A Over Current Protection Programmable Range 1 A to 3.3 A 1 A to 5.5 A AC Power Input Specifi cations Parameter U8001A U8002A Input Power Option (selectable) 100 Vac ± 10%, 47 to 63 Hz

115 Vac ± 10%, 47 to 63 Hz

230 Vac ± 10%, 47 to 63 Hz

Fuse External, customer assessable Physical Specifi cations Parameter U8001A U8002A Dimensions (H x W x L) 88.1 mm x 212.3 mm x 394.3 mm Weight 7.3 kg 8.3 kg

Liquid Crystal Display (LCD) Operating Environment

  • Operating temperature from 0 °C to 40 °C (for full rated output) and 40 °C to 55 °C (for derated output)
  • Relative humidity up to 95%
  • Altitude up to 2000 meters Storage Compliance -20 °C to 70 °C Safety and EMC Compliance Certifi ed with:
  • IEC 61326:2002/EN61326:1997+ A1:1998+A3:2003
  • CISPR 11:1990/EN55011:1990
  • Canada: ICES-001:2004
  • Australia/New Zealand: AS/NZS CISPR11:2004
  • IEC 61010-1:2001/EN61010-1:2001 nd Edition)
  • Canada: CAN/CSA-C22.2 No. 61010-1-04
  • USA: ANSI/UL 61010-1:2004 Memory Operations Up to three operating states Programming Language
  • Standard Commands for
  • Programmable Instruments (SCPI) Warranty
  • One year for U8000 series power supplies
  • Three months for standard shipped accessories U8001A DC Power Supply, 30 V, 3 A U8002A DC Power Supply, 30 V, 5 A Included Documentation: U8001A-0B0 No documentation package Additional Documentation:
  • U8001A-AB0 Traditional Chinese User’s and Service Guide
  • U8001A-AB2 China – Simplifi ed Chinese printed User’s and Service Guide
  • U8001A-ABA English User’s and Service Guide
  • U8001A-ACF Japanese User’s and Service Guide Calibration Document: U8001A-UK6 Commercial calibration with test result data Power Options Opt 0E9 90 to 110 Vac, 47 to 63 Hz Opt 0EM 104 to 126 Vac, 47 to 63 Hz Opt 0E3 207 to 253 Vac, 47 to 63 Hz Other Options
  • Option U800xA-AUT Custom Power On
  • E3600A-100 Test lead kit
  • Option 1CM Rack-mount kit Rack-mount Kits To rack-mount a single instrument: Adapter kit (P/N 5063-9240) To rack-mount two instruments side-by- side: Lock-link kit (P/N 5061-9694) Flange kit (P/N 5063-9212) To rack-mount two instruments in a slid- ing support shelf: Support shelf (P/N 5063-9255) Slide kit (P/N 1494-0015) General Characteristics

Ordering Information

www.agilent.com/fi nd/emailupdates Get the latest information on the products and applica- tions you select. www.agilent.com/fi nd/agilentdirect Quickly choose and use your test equipment solutions with confi dence. www.agilent.com/fi nd/open Agilent Open simplifi es the process of connecting and programming test systems to help engineers design, validate and manufacture electronic products. Agilent offers open connectivity for a broad range of system- ready instruments, open industry software, PC-standard I/O and global support, which are combined to more easily integrate test system development. Agilent Email Updates Agilent Direct Agilent Open Remove all doubt Our repair and calibration services will get your equipment back to you, performing like new, when prom- ised. You will get full value out of your Agilent equipment through- out its lifetime. Y our equipment will be serviced by Agilent-trained technicians using the latest factory calibration procedures, automated repair diagnostics and genuine parts. You will always have the utmost confi dence in your measurements. Agilent offers a wide range of ad- ditional expert test and measure- ment services for your equipment, including initial start-up assistance onsite education and training, as well as design, system integration, and project management. For more information on repair and calibration services, go to www.agilent.com/fi nd/removealldoubt For more information on Agilent T echnologies’ products, applications or services, please contact your local Agilent office. The com- plete list is available at: www.agilent.com/fi nd/contactus Phone or Fax Americas Canada Latin America United States (877) 894-4414 305 269 7500 (800) 829-4444 Asia Pacifi c Australia China Hong Kong India Japan Korea Malaysia Singapore T aiwan Thailand 1 800 629 485 800 810 0189 800 938 693 1 800 112 929 0120 (421) 345 080 769 0800 1 800 888 848 1 800 375 8100 0800 047 866 1 800 226 008 Europe & Middle East Austria Belgium Denmark Finland France Germany Ireland Israel Italy Netherlands Spain Sweden Switzerland United Kingdom Other European Countries: www.agilent.com/fi nd/contactus Revised: October 6, 2008 01 36027 71571 32 (0) 2 404 93 40 45 70 13 15 15 358 (0) 10 855 2100 0825 010 700* *0.125 /minute 07031 464 6333 1890 924 204 972-3-9288-504/544 39 02 92 60 8484 31 (0) 20 547 2111 34 (91) 631 3300 0200-88 22 55 0800 80 53 53 44 (0) 118 9276201 © Agilent T echnologies, Inc. 2009 Printed in USA, June 16, 2009 5989-7182EN Product specifi cations and descriptions in this document subject to change without notice. www.agilent.com