DC/DC Medical Converter TRV 2M Series, 2 Watt
- I/O isolation 5000 VAC (reinforced)
- Wide 1.5:1 input voltage range
- Operating temperature range -40 to +80 °C without derating
- Certification according to IEC/EN/ES 60601-1 3rd edition for 2xMOPP and IEC/EN/UL 62368-1
- Short circuit protection
- Regulated outputs
- Low leakage current < 2 μA
- Efficiency up to 84 %
- Operation up to 5000 m altitude
- 5-year product warranty IEC 60601-1 IEC 62368-1 ES 60601-1 UL 62368-1 The TRV 2M is a series of 2 Watt DC/DC converters in a compact SIP-9 pack- age with reinforced isolation of 5000 VAC for medical and industrial applica- tions. The series offers a 1.5:1 input voltage range with a nominal input ranging between 5 and 24 VDC. With a continuous short circuit protection and a low leakage current of less than 2 µA, this converter is especially suited to protect any connected interfaces or applied parts to patients. Featuring almost fully regulated outputs this series provides great level of regulation without affecting the cost efficiency. It is an ideal solution for applications where an unregulated DC/DC converter would not meet your regulation requirements but cost still is a critical factor. With an operating temperature range from -40 to +80°C without derating and certifications acc. IEC/EN/ES 60601-1 3rd ed. for 2xMOPP and IEC/EN/UL 62368-1 this series is suitable for many different applications where a medical isolation system and short circuit protection is needed. Models Order Code Input Voltage Output 1 Output 2 Efficiency Range Vnom Imax Vnom Imax typ. TRV 2-0510M 4.5 - 7 VDC (5 VDC nom.)
3.3 VDC 600 mA 78 %
TRV 2-0511M 5 VDC 400 mA 81 % TRV 2-0512M 12 VDC 167 mA 83 % TRV 2-0513M 15 VDC 134 mA 83 % TRV 2-0521M +5 VDC 200 mA –5 VDC 200 mA 82 % TRV 2-0522M +12 VDC 83 mA –12 VDC 83 mA 83 % TRV 2-0523M +15 VDC 67 mA –15 VDC 67 mA 81 % TRV 2-1210M 9.6 - 14.4 VDC (12 VDC nom.)
3.3 VDC 600 mA 79 %
TRV 2-1211M 5 VDC 400 mA 81 % TRV 2-1212M 12 VDC 167 mA 84 % TRV 2-1213M 15 VDC 134 mA 83 % TRV 2-1221M +5 VDC 200 mA –5 VDC 200 mA 81 % TRV 2-1222M +12 VDC 83 mA –12 VDC 83 mA 83 % TRV 2-1223M +15 VDC 67 mA –15 VDC 67 mA 82 % TRV 2-1510M 12 - 18 VDC (15 VDC nom.) TRV 2-1511M 5 VDC 400 mA 81 % TRV 2-1512M 12 VDC 167 mA 84 % TRV 2-1513M 15 VDC 134 mA 83 % TRV 2-1521M +5 VDC 200 mA –5 VDC 200 mA 81 % TRV 2-1522M +12 VDC 83 mA –12 VDC 83 mA 83 % TRV 2-1523M +15 VDC 67 mA –15 VDC 67 mA 80 % TRV 2-2410M 19.2 - 28.8 VDC (24 VDC nom.) TRV 2-2411M 5 VDC 400 mA 80 % TRV 2-2412M 12 VDC 167 mA 82 % TRV 2-2413M 15 VDC 134 mA 82 % TRV 2-2421M +5 VDC 200 mA –5 VDC 200 mA 81 % TRV 2-2422M +12 VDC 83 mA –12 VDC 83 mA 81 % TRV 2-2423M +15 VDC 67 mA –15 VDC 67 mA 80 % Note - If the input will be switched electromechanically, use an external 47 μF/63 V E/C. to avoid voltage transient. www.tracopower.com November 30, 2022 Page 1 / 5
Input Current - At no load 5 Vin models: 5 mA typ. (3.3 Vout model) 5 mA typ. (5 Vout model) 20 mA typ. (12 Vout model) 20 mA typ. (15 Vout model) 5 mA typ. (5 / -5 Vout model) 20 mA typ. (12 / -12 Vout model) 25 mA typ. (15 / -15 Vout model) 12 Vin models: 4 mA typ. (3.3 Vout model) 4 mA typ. (5 Vout model) 10 mA typ. (12 Vout model) 10 mA typ. (15 Vout model) 4 mA typ. (5 / -5 Vout model) 10 mA typ. (12 / -12 Vout model) 10 mA typ. (15 / -15 Vout model) 15 Vin models: 4 mA typ. (3.3 Vout model) 4 mA typ. (5 Vout model) 8 mA typ. (12 Vout model) 8 mA typ. (15 Vout model) 4 mA typ. (5 / -5 Vout model) 8 mA typ. (12 / -12 Vout model) 8 mA typ. (15 / -15 Vout model) 24 Vin models: 3 mA typ. (3.3 Vout model) 3 mA typ. (5 Vout model) 6 mA typ. (12 Vout model) 6 mA typ. (15 Vout model) 3 mA typ. (5 / -5 Vout model) 6 mA typ. (12 / -12 Vout model) 6 mA typ. (15 / -15 Vout model) - At full load 5 Vin models: 510 mA typ. 12 Vin models: 210 mA typ. 15 Vin models: 170 mA typ. 24 Vin models: 105 mA typ. Surge Voltage 5 Vin models: 15 VDC max. (1 s max.) 12 Vin models: 25 VDC max. (1 s max.) 15 Vin models: 25 VDC max. (1 s max.) 24 Vin models: 35 VDC max. (1 s max.) Recommended Input Fuse 5 Vin models: 1'000 mA (slow blow)
12 Vin models: 500 mA (slow blow)
15 Vin models: 500 mA (slow blow)
24 Vin models: 315 mA (slow blow)
(The need of an external fuse has to be assessed in the final application.) Input Filter Internal Capacitor Input Specifications Voltage Set Accuracy ±3% max. (60% load: 3.3, 5, ±5 Vout models) ±3% max. (90% load: other models) Regulation - Input Variation (Vmin - Vmax) single output models: 2% max. dual output models: 2% max. - Load Variation See application note: www.tracopower.com/overview/trv2m - Voltage Balance (symmetrical load) dual output models: 4% max. - Cross Regulation (25% / 100% asym. load) dual output models: 4% max. Output Specifications TRV 2M Series, 2 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 2 / 5
(20 MHz Bandwidth) - single output 3.3 Vout models: 100 mVp-p typ. 5 Vout models: 100 mVp-p typ. 12 Vout models: 125 mVp-p typ. 15 Vout models: 125 mVp-p typ. - dual output 5 / -5 Vout models: 100 / 100 mVp-p typ. 12 / -12 Vout models: 125 / 125 mVp-p typ. 15 / -15 Vout models: 125 / 125 mVp-p typ. Capacitive Load - single output 3.3 Vout models: 2'000 µF max. 5 Vout models: 820 µF max. 12 Vout models: 470 µF max. 15 Vout models: 470 µF max. - dual output 5 / -5 Vout models: 470 / 470 µF max. 12 / -12 Vout models: 220 / 220 µF max. 15 / -15 Vout models: 220 / 220 µF max. Minimum Load Not required Temperature Coefficient ±0.03 %/K max. Start-up Time 20 ms max. Short Circuit Protection Continuous, Automatic recovery Safety Standards - IT / Multimedia Equipment EN 62368-1 IEC 62368-1 UL 62368-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Certification Documents www.tracopower.com/overview/trv2m Pollution Degree PD 2 Over Voltage Category Not mains connected Safety Specifications EMI Emissions EN 60601-1-2 edition 4 (Medical Devices) - Conducted Emissions EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter) - Radiated Emissions EN 55011 class A (with external filter) EN 55011 class B (with external filter) EN 55032 class A (with external filter) EN 55032 class B (with external filter) External filter proposal: www.tracopower.com/overview/trv2m EMS Immunity EN 55024 (IT Equipment) EN 60601-1-2 edition 4 (Medical Devices) - Electrostatic Discharge Air: EN 61000-4-2, ±15 kV, perf. criteria A Contact: EN 61000-4-2, ±8 kV, perf. criteria A - RF Electromagnetic Field EN 61000-4-3, 10 V/m, perf. criteria A - EFT (Burst) / Surge EN 61000-4-4, ±2 kV, perf. criteria A EN 61000-4-5, ±2 kV, perf. criteria A External filter proposal: www.tracopower.com/overview/trv2m - Conducted RF Disturbances EN 61000-4-6, 10 Vrms, perf. criteria A - PF Magnetic Field Continuous: EN 61000-4-8, 100 A/m, perf. criteria A 1 s: EN 61000-4-8, 1000 A/m, perf. criteria A EMC Specifications Relative Humidity 95% max. (non condensing) General Specifications TRV 2M Series, 2 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 3 / 5
Temperature Ranges - Operating Temperature –40°C to +95°C - Case Temperature +105°C max. - Storage Temperature –55°C to +125°C Power Derating - High Temperature Depending on model See application note: www.tracopower.com/overview/trv2m Cooling System Natural convection (20 LFM) Altitude During Operation 5'000 m max. Switching Frequency 180 - 360 kHz (PFM) Insulation System Reinforced Insulation Isolation Test Voltage - Input to Output, 60 s 5'000 VAC - Input to Output, 1 s 7'000 VDC Creepage - Input to Output 8 mm min. Clearance - Input to Output 8 mm min. Isolation Resistance - Input to Output, 500 VDC 10'000 MΩ min. Isolation Capacitance - Input to Output, 100 kHz, 1 V 10 pF typ. 20 pF max. Leakage Current - Touch Current 2 µA max. (240 VAC, 60 Hz) Reliability - Calculated MTBF 10'410'000 h (MIL-HDBK-217F, ground benign) Washing Process Allowed (hermetical product) See Cleaning Guideline: www.tracopower.com/info/cleaning.pdf Environment - Vibration MIL-STD-810F 7.6 g, 3 axis, 60 min, 20-2000 Hz - Mechanical Shock MIL-STD-810F 50 g, 3 axis, 11 ms - Thermal Shock MIL-STD-810F -55°C to +125°C, 72 cycles, 30 min each Housing Material Non-conductive Plastic (UL 94 V-0 rated) Potting Material Silicone (UL 94 V-0 rated) Pin Material Tinned Copper Pin Foundation Plating Nickel (2 - 3 µm) Pin Surface Plating Tin (3 - 5 µm), matte Housing Type Plastic Case Mounting Type PCB Mount Connection Type THD (Through-Hole Device) Footprint Type SIP9 Soldering Profile Wave Soldering 260°C / 6 s max. Weight 4.8 g Environmental Compliance - REACH Declaration www.tracopower.com/info/reach-declaration.pdf REACH SVHC list compliant REACH Annex XVII compliant - RoHS Declaration www.tracopower.com/info/rohs-declaration.pdf Exemptions: 7a, 7c-I (RoHS exemptions refer to the component concentration only, not to the overall concentration in the product (O5A rule). The SCIP number is provided on request.) Supporting Documents Overview Link (for additional Documents) www.tracopower.com/overview/trv2m TRV 2M Series, 2 Watt All specifications valid at nominal voltage, resistive full load and +25°C after warm-up time, unless otherwise stated. www.tracopower.com November 30, 2022 Page 4 / 5
TRV 2M Series, 2 Watt Bottom view 0.5x0.25 (0.02x0.01) 19.6 (0.77) 2.0 (0.079) 1.8 (0.07) 2.0 (0.079) 10.0 (0.394) 2.0 (0.079) 1 2 97 8 16.0 (0.63) 4.1 (0.16) 2.3 (0.09) 9.9 (0.39) 12.5 (0.49) Dimensions in mm (inch) Tolerances: x.x ±0.5 (±0.02) x.xx ±0.25 (±0.01) Pin diameter ±0.1 (±0.004) Pinout Pin Single Dual 1 +Vin +Vin 2 in in 7 out out
8 No Pin Common
© Copyright 2022 Traco Electronic AG Specification s can be changed without notice. www.tracopower.com Rev. November 30, 2022 Page 5 / 5