Long Island, NY 11590 IET LABS, INC. in the Gen Rad Tradition www.ietlabs.com TEL: (516) 334-5959 • (800) 899-8438 • FAX: (516) 334-5988 Electrometer Calibration Standard SCR-5156-IET Page 1 of 1 SCR-5156-IET Datasheet Sept. 2018 High-performance, electrometer calibra on standard for use with the Keithley 6517 and others. Resistances and capacitance values as specifi ed in the Keithley Service Manual for current range and cou- lombs verifi ca on. Replacement for the Keithley 5156. SPECIFICATIONS SCR-5156 Nominal Value Initial adjustment to nominal Stability (ppm/yr) Voltage coeffi cient (ppm/V) Temperature coeffi cient (ppm/°C) Resistor type 100 MW 100 1 ±35 Thick-fi lm 1 GW 100 1 ±35
10 GW 500 2 ±50
100 GW 500 5 ±50
1 nF 300 5 ±50 Polystyrene 100 nF 300 5 ±50 MPP Calibrated Values: 23°C; traceable to SI with measurement uncertainties Capacitance is calibrated at 1 kHz, 15 Vac Maximum Input Voltage:
500 Vpk
Features:
- 4 resistance standards
100 MΩ to 100 GΩ
- 1 nF and 100 nF capacitors for charge calibra on
- Accuracy: 1%
- Accredited calibra on cer fi cate
ORDERING INFORMATION
SCR-5156-IET Electrometer Calibration Standard Includes: Instruction Manual SCR-5156-200 Cable, 3 Lug Triaxial to 3 Lug Triaxial, 5' SCR-5156-300 Short, BNC Accredited Calibration Certifi cate Traceable to SI The SCR-5156-IET contains 4 resistors and 2 capacitors as recommended in the Keithley
6517 Service Manual for current and coulombs
verifi cation. The metal housing provides shielding from noise and a shorting link is provided to connect guard and ground when required. The output of the SCR-5156-IET is a triaxial connector for direct connection to the Keithley 6517. There are 6 bnc connectors for each of the internal standards. The calibrated value of each standard is located on the front panel. Terminals: 6 bnc connectors for standard values One triaxial output which is common for all 6 standards One fi ve-way binding post for guard with shorting link for connection to ground, and one ground post electrically connected to case Environmental: Operating 23°C ± 3°C Humidity: 30 to 60 % RH non-condensing Dimensions: 20 cm W x 11 cm H x 5.7 cm D (7.55" W x 4.38" H x 2.24" D) Weight: 2.4 kg (3.5 lbs), nominal Custom resistance values are available upon request