WEBSITE : www.jdsu.com/test COMMUNICATIONS TEST & MEASUREMENT SOLUTIONS Key Features SB and SC series

  • SB series can accommodate up to 48 channels and offer up to two input channels
  • SC series can accommodate up to 180 channels and offer up to four input channels
  • Low IL, 0.4 dB typical
  • Excellent repeatability, +/- 0.003 dB typical
  • High return loss (RL) > 65 dB typical
  • GPIB and RS-232 remote control SCG series
  • Offer up to 45 input channels and 90 output channels
  • Mass input reconfiguration possible
  • Low IL, 0.5 dB typical for D configuration
  • High RL > 65 dB typical
  • Excellent repeatability, +/- 0.005 dB typical
  • Replaces multiple switch elements with one switch instrument The JDSU SB, SC, and SCG series of Benchtop/Rackmount Programmable Switches can be controlled using the front panel keys and a numeric pad or via GPIB and serial RS-232 interface. The SCG series ganged input switches allow a single switch instrument to replace multiple switch elements while maintaining low loss. In this series of switches, the inputs are ganged together in a particular sequence and are thus able to offer three different modes of operation. The SB, SC, and SCG series switches are available in four basic configurations:
  • C configuration - is a single common input model (SB, SC)
  • D configuration - provides simultaneous connection of a bank of input fibers to output fibers (SB, SC, SCG)
  • E configuration - allows any input to be connected to any output while other inputs/outputs are aligned to subsequent/adjacent channels. The switch is non-blocking in this mode and other inputs/outputs are aligned (SB, SC, SCG)
  • F configuration - enables one of the inputs to be aligned with an output in a blocking sense, with a result in reduction of available output channels and a low-loss M x N blocking switch. (SB, SC, SCG) Operation of these switches is based upon JDSU's proven expanded beam lens technology, which utilizes a precision stepper-motor to align optical channels. The use of collimating lenses minimizes insertion loss (IL) and improves repeatability and performance. Internal temperature control of the switching mechanism ensures excellent operational stability.

Applications

  • Fiberoptic component testing and measurement
  • System testing
  • Research and development (R&D)
  • Mass reconfiguration of large numbers of inputs/outputs with SCG series (D configuration)
  • Connection of multiple wavelength sources to any one of a number of devices with SCG series (F configuration)
  • Network monitoring Safety Information
  • Complies to CE requirements plus UL3101-1 and CAN/CSA-C22.2 No. 1010.1 Benchtop/Rackmount Programmable Switches SB/SC/SCG Series

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES Both single-mode (SM) and multimode (MM) versions of the SB, SC, and SCG series switches are available. The series features the high level of performance required for multi-unit testing in R&D and in manufacturing environments. The compact, portable SB switch and the standard rackmount enclosure SC and SCG switches are highly suited for applications in telecommunications, manufacturing, and test environments. JDSU's SB, SC and SCG switches are known in the fiberoptic industry for their low IL and excellent repeatability. In addition to the many standard options available, we also customize switches in this series to meet your specific applica- tion needs. Continued 1x N Switching: C Configuration (SB and SC) The 1xN configuration allows a single common input to be switched to any of the outputs. Ganged Switching: D Configuration (SB/SC/SCG) The MxN configuration allows for mass reconfiguration of optical paths. It provides simultaneous connections of a bank of inputs to outputs, and replaces several 1xN with single switch element. Sequential Switching: E Configuration (SB/SC/SCG) The MxN configuration aligns any input with any output, while other inputs are aligned to adjacent outputs. Blocking Switching: F Configuration (SB/SC/SCG) The MxN configuration allows any one of a bank of inputs to connect with any output with only one active connection at a time. Configurations

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES Parameter1 Single Common Multiple Common Multiple Common C Configuration D Configuration E and F Configurations Typical Maximum Typical Maximum Typical Maximum Insertion loss (IL) Return loss (RL)2 SM standard/analog > 65 dB 60/65 dB > 65 dB 60/65 dB 65 dB 60 dB MM3 standard/analog 25/35 dB 20/30 dB 25/35 dB 20/30 dB > 25 dB 20 dB Repeatability5 Crosstalk (maximum) SM -80 dB Maximum input power (optical) 300 mW Lifetime > 80 million cycles Switching time One channel 300 ms Each additional channel 12 ms Power supply 100 to 240 V , 50 to 60 Hz Power consumption 100 V A maximum Control Local and remote via GPIB and serial RS-232 interfaces Drivers for external switch modules Four open collector drivers with maximum 100 mA sink current Operation temperature 0 to 55 °C Storage temperature - 40 to 70 °C Humidity maximum 95 % RH from 0 to 55 °C non-condensing Dimensions (W x H x D) 21.2 x 8.9 x 35.5 cm with rackmount kit (optional)6 48.3 x 8.9 x 35.5 cm Weight 3.75 kg 1. Excluding connectors. All optical measurements taken after temp erature has been stabilized for one hour, at ambient (room) conditions. 2. RL specification based on 1 m pigtail length. 3. Values shown for 62.5 µm diameter maximum fiber core. 4. Drift of any channel relative to reference channel at ± 3 °C deviation of ambient temperature over a seven-day period. 5. Measured between two consecutive readings over 100 cycles. 6. ED000899-A-00 standard rackmount kit, ED000899-A-01 Japan rackmount kit. Requires two kits to mount two units side-by-side. Please specify part number when ordering (if needed). SB Model Specifications

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES Parameter1 Single Common Multiple Common Multiple Common C Configuration D Configuration E and F Configurations (SC model only) Typical Maximum Typical Maximum Typical Maximum IL SC with 3 and 4 inputs and SCG models - - 0.5 dB 1.0 dB 0.7 dB 1.5 dB SC with 3 and 4 inputs and SCG models - - 0.5 dB 1.0 dB 0.7 dB 1.5 dB RL2 SM standard/analog > 65 dB 60/65 dB > 65 dB 60/65 dB 3 > 65 dB 60 dB MM4 standard/analog 25/35 dB 20/30 dB 25/35 dB 3 20/30 dB3 > 25 dB 20 dB Repeatability7 Crosstalk (maximum) SM -80 dB Maximum input power (optical) 300 mW Lifetime > 80 million cycles (> 10 million cycles on SCG) Switching time One channel (SCG model) 300 ms (420 ms) Each additional channel (SCG model) 12 ms (20 ms) Power supply 100 to 240 V , 50 to 60 Hz Power consumption 100 V A maximum Control Local and remote via GPIB and serial RS-232 interfaces Drivers for external switch modules Four open collector drivers with maximum 100 mA sink current Operation temperature 0 to 55 °C Storage temperature -40 to 70 °C Humidity Maximum 95 % RH from 0 to 55 °C non-condensing Dimensions (W x H x D) single (double height 6) 48 x 13 x 37 cm (48 x 26.6 x 37 cm) excluding handles Weight single (double height6) 9 kg (14 kg) 1. Excluding connectors. All optical measurements taken after temp erature has been stabilized for one hour, at ambient (room) conditions. 2. RL specification based on 1 m pigtail length. 3. Analog version available on one and two input SC model switches (C and D configurations). 4. Values shown for 62.5 µm diameter maximum fiber core. 5. Drift of any channel relative to reference channel at ± 3 °C deviation of ambient temperature over a seven-day period. 6. Applies to SC model only. 7. Measured between two consecutive readings over 100 cycles. SC and SCG Models Specifications

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES SB Switch Configuration Configuration C D E F 1xN 2xN 2xN 2xN Output Channel 2 to 48 4 to 44 2 to 44 2 to 20 Counts (N) SCG Switch Configurations Configuration Restrictions D: Up to 45 x 90 such that 'number of outputs' [N] is divisible by 'number of inputs' [M] E: Up to 45 inputs [M] and up to 84 outputs [N], such that M + N is not more than 90 F: Up to 13 inputs [M] and up to 14 outputs [N], such that M x (N + 1) is not more than 93 SC Switch Configuration C o n f i g u r a t i o n CD D D EEEFFF 1xN 2xN 3xN 4xN 2xN 3xN 4xN 2xN 3xN 4xN Output Channel 2 to 180 4 to 180 6 to 180 8 to 180 2 to 180 3 to 180 4 to 180 2 to 90 3 to 60 4 to 45 Counts (N)

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES (Specifications in inches [mm] unless otherwise noted.)SB Model Dimensions Diagram

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES

Ordering Information

SB Model Sample: SB2E10141+27XF000FP Code Connector Type FP FC/PC (bulkhead max. 24) FA FC/APC (bulkhead max. 24) SC SC/PC (bulkhead max. 24) SU SC/APC (bulkhead max. 24) NC No connector 1. Bulkheads and pigtails cannot be mixed in the same panel unless custom ordered. 2. For reverse direction, use bidirectional. SB Switch Package Information Connector Type Chassis Size Switch Configuration Max Channel Counts (input + output) Bulkhead 2U bench top C, D, E 24 Bulkhead 2U bench top F 22 Pigtail 2U bench top C, D 49 Pigtail 2U bench top F 22 If the configurations available do not meet your performance requirements, please contact our global sales and customer service team to discuss the potential for specialized solutions. SB +2 Code Wavelength Range (nm) F 1270 to 1670 Q 850 to 1350 (MM only) B 750 to 940 (MM only) Code Cable Length (3mm diameter) 001 1 m 003 3 m 009 9 m

000 Not applicable (bulkheads only)

Code Number of Input Channels 1C 1 input channel, C config. 2D 2 input channels, D config. 2E 2 input channels, E config. 2F 2 input channels, F config. Code Input Port Type 1

1 Bulkheads on front

4 Pigtails on back

Code Number of Output Channels 002 2 output channels 020 20 output channels 024 24 output channels 048 48 output channels Code Output Port Type 1 B Bidirectional, standard RL2 C Bidirectional analog RL2 Code Fiber Type (µm) 7 9/125 1 50/125 2 62.5/125

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES SC Model Sample: SC2D30043+22XB009FP SC +2 Code Wavelength Range (nm) F 1270 to1670 Q 850 to 1350 (MM only) B 750 to 940 (MM only) Code Cable Length (3mm diameter) 001 1 m 003 3 m 009 9 m FP FC/PC (bulkheads max. 60/120) FA FC/APC (bulkheads max. 60/120) SC SC/PC (bulkheads max. 60/120) SU SC/APC (bulkheads max. 60/120) NC No connector Code Number of Input Channels 1C 1 input channel, C config. 2D 2 input channels, D config. 2E 2 input channels, E config. 2F 2 input channels, F config. 3D 3 input channels, D config. 3E 3 input channels, E config. 3F 3 input channels, F config. 4D 4 input channels, D config. 4E 4 input channels, E config. 4F 4 input channels, F config. Code Input Port Type 1

2 Bulkheads on back

3 Pigtails on front

Code Number of Output Channels 2 002 2 output channels 084 84 output channels 180 180 output channels Code Output Port Type 1 B Bidirectional, standard RL2 C Bidirectional analog RL2 Code Fiber Type (µm) 7 9/125 1 50/125 2 62.5/125 1. Bulkheads and pigtails cannot be mixed in the same panel unless custom ordered. 2. Single height: 84 output channel maximum. Double height: 180 output channel maximum. 3. For reverse direction, use bidirectional. If the configurations available do not meet your performance requirements, please contact our global sales and customer service team to discuss the potential for specialized solutions. SC Switch Package Information Connector Type Chassis Size Switch Configuration Max Channel Counts (input + output) Bulkhead 3U 19” rack mount C, D, E 60 Pigtail 3U 19” rack mount C, D, E 85 Bulkhead / Pigtail 3U 19” rack mount F 2xN 46 Bulkhead / Pigtail 3U 19” rack mount F 3xN 29 Bulkhead / Pigtail 3U 19” rack mount F 4xN 24 Bulkheads 6U 19” rack mount C, D, E 120 Pigtails 6U 19” rack mount C, D, E 181 Bulkhead / Pigtail 6U 19” rack mount F 2xN 92 Bulkhead / Pigtail 6U 19” rack mount F 3xN 63 Bulkhead / Pigtail 6U 19” rack mount F 4xN 49 Continued

BENCHTOP/RACKMOUNT PROGRAMMABLE SWITCHES UL is a registered trademark of Underwriters Laboratories Inc. LabVIEW is a registered trademark of National Instruments Corporation. All statements, technical information and recommendations related to the products herein are based upon information believed to be reliable or accurate. However, the accuracy or completeness thereof is not guaranteed, and no responsibility is assumed for any inaccuracies. The user assumes all risks and liability whatsoever in connection with the use of a product or its application. JDSU reserves the right to change at any time without notice the design, specifications, function, fit or form of its products described herein, including withdrawal at any time of a product offered for sale herein. JDSU makes no representations that the products herein are free from any intellectual property claims of others. Please contact JDSU fo r more information. JDSU and the JDSU logo are trademarks of JDS Uniphase Corporation. Other trademarks are the property of th eir respective holders. ©2006 JDS Uniphase Corporation. All rights SCG Model Sample: SCG06D20241+27XF000FP If the configurations available do not meet your performance requirements, please contact our global sales and customer service team to discuss the potential for specialized solutions. SCG Switch Package Information Connector Type Chassis Size Switch Configuration Max Channel Counts (input + output) Bulkhead 3U 19” rack mount C, D, E, F 60 front panel; 60 rear panel Pigtail 3U 19” rack mount C, D, E, F 90 front panel; 90 rear panel Continued SCG +2 Code Wavelength Range (nm) F 1270 to 1670 Q 850 to 1350 (MM only) B 750 to 940 (MM only) Code Cable Length (3mm diameter) 001 1 m 003 3 m 009 9 m

000 Not applicable

(bulkheads only) Code Connector Type FP FC/PC (maximum 60) FA FC/APC (maximum 60) SC SC/PC (maximum 60) SU SC/APC (maximum 60) NC No connector Code Number of Output Channels2 05 5 input channels 08 8 input channels 16 16 input channels 45 45 input channels Code Configuration D D configuration E E configuration F F configuration Code Number of Output Channels2 005 5 output channels 048 48 output channels 072 72 output channels 090 90 output channels Code Output Port Type 1

1 Bulkheads on front 2

2 Bulkheads on back 2

B Bidirectional, standard RL3 Code Fiber Type (µm) 7 9/125 1 50/125 2 62.5/125 1. The inputs and outputs must exit on opposite sides. (For example, if inputs exit from the front, then the outputs must exit from the rear.) 2. For exact layout of bulkheads and labeling, contact JDSU. 3. For reverse direction, use bidirectional. Code Input Port Type 1 TEL : 1 866 228 3762 FAX : +1 301 353 9216 Test & Measurement Regional Sales LA TIN AMERICA TEL : +55 11 5503 3800 FAX : +55 11 5505 1598 ASIA PACIFIC TEL : +852 2892 0990 FAX : +852 2892 0770 EMEA TEL : +49 7121 86 2222 FAX : +49 7121 86 1222 www.jdsu.com/test