S10 ASHCROFT | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 7

Technical content

Ashcroft Series: S10 Sheath Diameter: 3 mm, 4.5 mm, 6 mm, 8 mm Stem Length: Minimum: 50 mm/2 in Maximum: 3 m/120 in Sensor Type & Measuring Range RTDs Platinum 385 Pt 100 –200 to +600°C Pt 1000 –40 to +600°C Thermocouples Type J –40 to +750°C Type E –200 to +800°C Type K –200 to +1000°C Type N –200 to +1000°C Wiring Configuration: RTDs single or dual

2 Wire

3 Wire

4 Wire

Accuracy Class (IEC 60751) KEY BENEFITS

  • Flexible designs to work in most applications
  • Designs for hazardous locations

DESCRIPTION

The Ashcroft S10 temperature sensor assemblies provide accurate temperature measurements for most applications. Each temperature sensor assembly consists of a spring loaded temperature sensor insert, a connection head and lag extension. The assembly may also include an optional terminal block for wiring and/or transmitters. Thermocouple assemblies are manufactured to either to IEC 60584-2 or ANSI MC 96.1 and RTD assemblies are manufactured to IEC 60751. TYPICAL USES n Process temperature measurements for industrial, process and power generation. n Exhaust gas temperature measurements for diesel engines. n Oven temperature measurements for industrial drying ovens. n Special designs for intrinsically safe and non-incendive application. With DIN lag extension (1) Absolute temperature in °C Interior View Thermocouples (IEC 60584-2) Type J Type K Type E Type N Class 1 ±1.5°C or ±0.0040*|t|(1) ±1.5°C or ±0.0040*|t|(1) ±1.5°C or ±0.0040*|t|(1) ±1.5°C or ±0.0040*|t|(1) Class 2 ±2.5°C or ±0.0075*|t|(1) ±2.5°C or ±0.0075*|t|(1) ±2.5°C or ±0.0075*|t|(1) ±2.5°C or ±0.0040*|t|(1) Class 3 N/A ±2.5°C or ±0.0040*|t|(1) ±2.5°C or ±0.0150*|t|(1) ±2.5°C or ±0.0150*|t|(1) Thermocouples (ANSI MC 96.1) Type J Type K Type E Type N Standard ±2.2°C or ±0.0075*|t|(1) ±2.2°C or ±0.0075*|t|(1) ±1.7°C or ±0.0050*|t|(1) ±2.2°C or ±0.0040*|t|(1) Special ±1.1°C or ±0.0040*|t|(1) ±1.1°C or ±0.0040*|t|(1) ±1.0°C or ±0.0075*|t|(1) ±1.1°C or ±0.0040*|t|(1) Data Sheet S10 RTD's & Thermocouples, Metric Connections of 71 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258

FM Intrinsically safe: Class I, Division 1, Groups A, B, C, D T4 for –55°C ≤ Ta ≤ +80°C T5 for –55°C ≤ Ta ≤ +55°C T6 for –55°C ≤ Ta ≤ +40°C FM Nonincedive: Class I, Division 2, Groups A, B, C, D T4 for –55°C ≤ Ta ≤ +80°C T5 for –55°C ≤ Ta ≤ +55°C T6 for –55°C ≤ Ta ≤ +40°C ATEX or IECEx: ATEX or IECEx II 1 G Ex ia IIC T6 Ga –50°C to +60°C II 2 G Ex ib IIC T6 Gb –50°C to +60°C II 2 G Ex e IIC T6 Gb –55°C to +60°C Optional S10 Heads BUZH-AL Type E BUZ-AL Type D DIN B Type B BBK-PA Type C Data Sheet S10 RTD's & Thermocouples, Metric Connections of 72 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258

S10 RTD ORDERING CODE Example: S10 1 6 1 A A B 1 D 2 S Area Classification 1 - Standard 1 3 - Intrinsic Safety - ia B - Intrinsic Safety - ib E - Increased Safety N - Non-Incendive Sheath Diameter 3 - 3 mm 4 - 4.5 mm 6 - 6 mm 6 8 - 8 mm RTD Type 1 - PT 100 1 Accuracy or Class (IEC 60751) A - Class A A B - Class B C - 1/2 DIN D - Class AA - 1/3 DIN RTD Element/Range A - –50/+400°C A B - –200/+600°C D - vibrations-proof Electrical Circuit A - Single 2 wires B - Single 3 wires B C - Single 4 wires D - Dual 2 wires E - Dual 3 wires F - Dual 4 wires Sheath Material 1 - AISI 316L / 1.4404 1 Head Type B - DIN B Aluminum D - BUZ Aluminum D E - BUZH Aluminum C - BBK - Plastic Instrument Connection - 1/2 NPT Conduit Connection M - M20 x 1.5 P - Pg 16 2 - ½" NPT 2 Head Conduit Gland - - Without P - Polyamide PA, for unarmored cable L - Nickel plated brass, for unarmored cable M - Nickel plated brass, single seal for armoured cable N - Nickel plated brass, double seal for armoured cable S - Stainless steel, for unarmored cable S T - Stainless steel, single seal for armoured cable U - Stainless steel, double seal for armoured cable Data Sheet S10 Thermocouples, Probes of 73 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258 Data Sheet S10 RTD Probes of 73 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258 Continued on next page

S10 RTD ORDERING CODE Example: (Cont'd ) X C - 52 R3 - - 3P T N=27 LN=400 Inset Nominal Length X - LN=... (min=50, max=100000). (add actual length in mm LN=?? at the end of ordering code) X Lag Length - Lag length (add actual length in mm N=?? At the end of code) X - N=... (min=50, max=1000) N - N= 150mm F - N= 16mm C - N= 27mm C - - Without Lag Extension F5 - DIN Ø11/7 AISI 316/ 1.4401 N=150 F6 - DIN Ø14/11 AISI 316/ 1.4401 N=150 F8 - DIN Ø11/7 AISI 316/ 1.4401 N=non std F9 - DIN Ø14/11 AISI 316/ 1.4401 N=non std 51 -Threaded connection cyl. 52 - Threaded connection conical 52 -- - Without lag extension, without plug 4- - Without lag extension, with plug Process Connection Q3 - Thread G½ A Q4 - Thread G¾ A S6 - Thread M14 x 1.5 S7 - Thread M18 x 1.5 R3 - Thread ½ NPT R3 A3 - Compression fitting G ½ A, AISI 316 C3 - Compression fitting ½" NPT, AISI 316 -- - Without connection Electical Connection - - With terminal block - 1 - With transmitter Not available with FM IS or NI approval 3 - Without terminal block, with flying leads Certifications - - None required - F - FM A - ATEX X - IECEx S - SIL 2 + ATEX I - INMETRO D - ATEX + IECEX 2 - SIL 2 P - EAC (Gost R) + Metrological Russia Calibration Report -- - Without 3P 3P - 3 points 5P - 5 points 3D - 3 points 5D - 5 points Marking - - Without T - Label in stainless steel with tag T Prices subject to change without notice • All prices subject to escalation Nominal length in mm Extension length in mm Data Sheet S10 Thermocouples, Probes of 74 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258 Data Sheet S10 RTD Probes of 74 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258 mm = inches x 25.4

S10 TC ORDERING CODE Example: S10 S K 1 N 1 1 3 D M S Area Classification S - Standard - General Purpose S J - Intrinsic Safety - ia B - Intrinsic Safety - ib E - Increased Safety N - Non-Incendive Sheath Diameter 3 - 3 mm 4 - 4.5 mm 6 - 6 mm 8 - 8 mm Thermocouple Type E - temperature range: –200...+ 800°C J - temperature range: –40...+ 750°C K - temperature range: –200...+ 1000°C N - temperature range: –200...+ 1000°C Accuracy or Class N - ANSI MC 96.1: cl. standard N S - ANSI MC 96.1: cl. special 1 - IEC 60584-2 : class 1 2 - IEC 60584-2 : class 2 3 - IEC 60584-2 : class 3 Junction 1 - ungrounded 1 2 - grounded 3 - ungrounded, vibrations-proof 4 - grounded, vibrations-proof Electrical Circuit 1 - Single 1 2 - Dual Sheath Material 1 - AISI 316 / 1.4401 3 - Inconel 600/ 2.4816 3 Head Type B - DIN B Aluminum D- BUZ Aluminum D E- BUZH Aluminum C- BBK - Plastic Instrument Connection - 1/2 NPT Conduit Connection M M - M20 x 1.5 A - adapter M20x1.5 P - Pg 16 Head Conduit Gland - - Without P - Polyamide PA, for unarmored cable L - Nickel Plated Brass, for unarmored cable M - Nickel Plated Brass, single seal for armoured cable N - Nickel Plated Brass, double seal for armoured cable S - Stainless steel, for unarmored cable S T - Stainless steel, single seal for armoured cable U - Stainless steel, double seal for armoured cable Data Sheet S10 Thermocouples, Probes of 75 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258 Continued on next page

S10 RTD ORDERING CODE Example: (Cont'd ) X C - 52 R3 - - 3P T N=27 LN=400 Inset Nominal Length X - LN=... (min=50, max=100000) (add actual length in mm LN=?? at the end of ordering code) X Lag Length - - Lag length (add actual length in mm N=?? At the end of code) X - N= ... (min=40, max=1000) X N - N= 150mm F - N= 16mm C - N= 27mm - - Without - - Lag Extension F5 - DIN Ø11/7 AISI 316/ 1.4401 N=150 F6 - DIN Ø14/11 AISI 316/ 1.4401 N=150 F8 - DIN Ø11/7 AISI 316/ 1.4401 N=non std F8 F9 - DIN Ø14/11 AISI 316/ 1.4401 N=non std 51 - Threaded connection cyl. 52 - Threaded connection conical - - Without lag extension, without plug 4 - Without lag extension, with plug Process Connection Q3 - Thread G ½ A Q4 - Thread G ¾ A S6 - Thread M14 x 1.5 S7 - Thread M18 x 1.5 R3 - Thread ½ NPT R3 A3 - Compression fitting G ½ A, AISI 316 C3 - Compression fitting ½ NPT, AISI 316 -- - Without connection Electical Connection - - With terminal block - 1 - With transmitter 3 - Without terminal block, with flying leads Certifications - - None required - F - FM A - ATEX X - IECEx S - SIL 2 + ATEX I - INMETRO D - ATEX + IECEX 2 - SIL 2 P - EAC (Gost R) + Metrological Russia Calibration Report -- - Without 3P - 3 points 3P 5P - 5 points 3D - 3 points 5D - 5 points Tagging - - Without T - Label in stainless steel with tag T Prices subject to change without notice • All prices subject to escalation Nominal length in mm Extension length in mm Data Sheet S10 Thermocouples, Probes of 76 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258 mm = inches x 25.4

DIMENSIONS in [ ] are millimeters For reference only, consult Ashcroft for specific dimensional drawings HOW TO ORDER S10 TEMPERATURE PROBES: The ordering code is built by selecting the appropriate configuration for the various sections of the ordering code. The Insert nominal length LN is measured from base of the head to the tip of the probe. The lag extension length N is measured from the base of the head to the center of the threads on the lag extension. LN can be calculated by adding the lag extension length N to the probe insertion length L. The N length and the LN length are added to the end of the ordering code in millimeters. To convert inches to millimeters multiply by 25.4. mm = inches x 25.4 d = Stem diameter N = Lag Extension Length L = Insertion Length LN = Insert Nominal Length LN = N + L N LN d L Data Sheet S10 RTD's & Thermocouples, Metric Connections of 77 All specifications are subject to change without notice. All sales subject to standard terms and conditions. ©2022 Ashcroft Inc. S10_RTD_ds_RevB_04-05-22 ashcroft.com info@ashcroft.com 1.800.328.8258