Expandable PXI bundle based on PXI-4065 Digital Multimeter, 6 1/2-Digit, 300V, 3A Specifications PXIe-1083 and PXI-4065

This document combines the PDFs of this bundle together. The page numbers in the table of contents correspond to the page number of PDF the component’s documentation begins. Test Workflow P/N: 788509-35 Test Workflow is a bundle of select NI software featuring engineering-specific tools that help test professionals accomplish anything from their day-to-day work to overcoming their most challenging obstacles. Test Workflow includes:

  • LabVIEW - a graphical programming environment engineers use to develop automated research, validation, and production test systems.
  • InstrumentStudio- an application software that provides an integrated approach to interactive PXI measurements.
  • TestStand- a test executive software that accelerates system development and deployment for engineers in validation and production.
  • And more NI Software! PXI-4065 (Digital Multimeter) PXIe-1083 (5-Slot PXIe Chassis) Accessories:
  • Thunderbolt cable
  • Power cable (varied by PN) o 867114-01 (US) o 867114-02 (EUR) o 867114-03 (Generic)

This document contains specifications for the PXIe-1083 chassis. Electrical The following section provides information about the PXIe-1083 AC input and DC output. AC Input Input rating 100 VAC to 240 VAC, 50 Hz/60 Hz, 6 A to 3 A Operating voltage range1 90 VAC to 264 VAC Nominal input frequency 50 Hz/60 Hz Operating frequency range1 47 Hz to 63 Hz Efficiency 78% typical Over-current protection Internal fuse in line Main power disconnect The AC power cable provides main power disconnect. Do not position the equipment so that it is difficult to disconnect the power cord. The front-panel power switch causes the internal chassis power supply to provide DC power to the PXI Express backplane. Caution Disconnect power cord to completely remove power. © National Instruments 3 PXIe-1083 Specifications

DC output characteristics of the PXIe-1083. Maximum total available power for the PXIe-1083 is 293 W. The maximum combined power available on +3.3 V and +5 V is 180 W. The maximum power available for each Thunderbolt port is 15 W (5 V/3 A). Table 1. Backplane Slot Current Capacity Note PCI V(I/O) pins in Hybrid Peripheral Slots are connected to +5 V. temperature considerations at 58 W.

Module cooling Forced air circulation (positive pressurization) through one 150 CFM fan Module slot airflow direction Bottom of module to top of module Module intake Bottom of chassis Module exhaust Top, right side of chassis Slot cooling capacity 58 W; slot 6 supports 58 W cooling with high fan mode Power supply cooling Forced air circulation through integrated fans Power supply intake Front and left side chassis Power supply exhaust Rear of chassis Minimum chassis cooling clearances Above 44.45 mm (1.75 in.) Rear 44.45 mm (1.75 in.) Sides 44.45 mm (1.75 in.) Below Rack 44.45 mm (1.75 in.) Desktop 25.4 mm (1.00 in.) © National Instruments 5 PXIe-1083 Specifications

Maximum altitude 2,000 m (6,560 ft.), 800 mbar (at 25 °C ambient, high fan mode) Pollution Degree 2 Indoor use only. Operating Environment Ambient temperature range When all peripheral modules require ≤38 W cooling capacity per slot 0 °C to 50 °C (IEC 60068-2-1 and IEC 60068-2-2.)2 Meets MIL-PRF-28800F Class 3 low temperature limit and high temperature limit. When any peripheral module requires >38 W cooling capacity per slot 0 °C to 40 °C (IEC 60068-2-1 and IEC 60068-2-2.)2 Meets MIL-PRF-28800F Class 3 low temperature limit and MIL- PRF-28800F Class 4 high temperature limit. Relative humidity range 20% to 80%, noncondensing Storage Environment Ambient temperature range –40 °C to 71 °C (IEC-60068-2-1 and IEC-60068-2-2.)[3] Meets MIL- PRF-28800F Class 3 limits. Relative humidity range 10% to 95%, noncondensing ni.com6 PXIe-1083 Specifications

Operational shock 30 g peak, half-sine, 11 ms pulse (IEC-60068-2-27.)3 Meets MIL- PRF-28800F Class 2 limits. Operational random vibration 5 to 500 Hz, 0.3 grms Non-operating vibration 5 to 500 Hz, 2.4 grms (IEC 60068-2-64.)3 Non-operating test profile exceeds the requirements of MIL-PRF-28800F, Class 3. Acoustic Emissions Sound Pressure Level (at Operator Position) (Tested in accordance with ISO 7779. Meets MIL-PRF-28800F requirements.)

38 W Profile

Auto fan (up to 30 °C ambient) 33.7 dBA High fan 50.8 dBA

58 W Profile

Auto fan (up to 30 °C ambient) 54.7 dBA High fan 55.3 dBA Sound Power Level © National Instruments 7 PXIe-1083 Specifications

Auto fan (up to 30 °C ambient) 44.9 dBA High fan 60.3 dBA Auto fan (up to 30 °C ambient) 63.4 dBA High fan 64.2 dBA Note The protection provided by the PXIe-1083 can be impaired if it is used in a manner not described in this document. Safety Compliance Standards This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: ■ IEC 61010-1, EN 61010-1 ■ UL 61010-1, CSA C22.2 No. 61010-1 Note For safety certifications, refer to the product label or the Product Certifications and Declarations section. EMC Guidelines This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic environment. ni.com8 PXIe-1083 Specifications

This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a peripheral device or test object, or if the product is used in residential areas. To minimize interference with radio and television reception and prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation. Furthermore, any changes or modifications to the product not expressly approved by NI could void your authority to operate it under your local regulatory rules. EMC Notices Refer to the following notices for cables, accessories, and prevention measures necessary to ensure the specified EMC performance. Notice For EMC declarations and certifications, and additional information, refer to the Product Certifications and Declarations section. Notice Changes or modifications to the product not expressly approved by NI could void your authority to operate the product under your local regulatory rules. Notice Operate this product only with shielded cables and accessories. Electromagnetic Compatibility Standards This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: ■ EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity ■ EN 55011 (CISPR 11): Group 1, Class A emissions © National Instruments 9 PXIe-1083 Specifications

■ AS/NZS CISPR 11: Group 1, Class A emissions Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note In Europe, Canada, Australia, and New Zealand (per CISPR 11) Class A equipment is intended for use in nonresidential locations. CE Compliance This product meets the essential requirements of applicable European Directives, as follows: ■ 2014/35/EU; Low-Voltage Directive (safety) ■ 2014/30/EU; Electromagnetic Compatibility Directive (EMC) ■ 2011/65/EU; Restriction of Hazardous Substances (RoHS) Product Certifications and Declarations Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, search by model number, and click the appropriate link. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This page contains the environmental ni.com10 PXIe-1083 Specifications

regulations and directives with which NI complies, as well as other environmental inf ormation not included in this document. EU and UK Customers

  • Waste Electrical and Electronic Equipment (WEEE)—A t the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/ environment/weee. 电子信息产品污染控制管理办法(中国 R oHS)
  • 5 40e 中国 RoHS— NI 符合中国电子信息产品中限制使用某些有害物 质指令(R oHS)。关于 NI 中国 RoHS 合规性信息,请登录 ni.com/environment/ rohs_china。(For information about China RoHS compliance, go to ni.com/ environment/rohs_china.) Backplane Size 3U-sized; 5 peripheral slots. Compliant with IEEE 1101.10 mechanical p ackaging. PXI Express Specification compliant. Accepts both PXI Express and CompactPCI (PICMG 2.0 R 3.0) 3U modules. Backplane bare-board mat erial UL 94 V-0 Recognized Backplane connectors Conforms to IEC 917 and IEC 1076-4-101, UL 94 V-0 rated System Synchronization Clocks

10 MHz Sys

tem Reference Clock: PXI_CLK10 © National Instruments 11 PXIe-1083 Specifications

Maximum slot-to-slot skew 250 ps Accuracy ±25 ppm max (guaranteed over the operating temperature range) Maximum jitter 5 ps RMS phase-jitter (10 Hz–1 MHz range) Duty-factor 45% to 55% Unloaded signal swing 3.3 V ±0.3 V Note For other specifications, refer to the PXI-1 Hardware Specification.

100 MHz System Reference Clock: PXIe_CLK100 and PXIe_SYNC100

Maximum slot-to-slot skew 100 ps Accuracy ±25 ppm max (guaranteed over the operating temperature range) Maximum jitter 3 ps RMS phase-jitter (10 Hz to 12 kHz range), 2 ps RMS phase-jitter (12 kHz to 20 MHz range) Duty-factor for PXIe_CLK100 45% to 55% Absolute differential voltage (When terminated with a 50 Ω load to 1.30 V or Thévenin equivalent) 400 mV to 1000 mV Note For other specifications, refer to the PXI-5 PXI Express Hardware Specification. ni.com12 PXIe-1083 Specifications

for installing the optional rack mount kits. Figure 1. PXIe-1083 Chassis Dimensions (Front)

Figure 2. PXIe-1083 Chassis Dimensions (Side) Figure 3. PXIe-1083 Chassis Dimensions (Bottom) 1 The operating range is guaranteed by design.

2 This product meets the requirements of the environmental standards for electrical

equipment for measurement, control, and laboratory use.

3 This product meets the requirements of the environmental standards for electrical

equipment for measurement, control, and laboratory use. © 2022 National Instruments Corporation.

ations apply to the PXI-4065. Definitions Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty. ■ Typical specifications describe the expected performance met by a majority of the models. ■ Nominal specifications describe parameters and attributes that may be useful in operation. Conditions Specifications are valid under the following conditions unless otherwise noted. ■ Ambient temperature of Tcal ± 5 °C.[1] ■ Calibr ation interval of 1 year ■ 30 minutes minutes warm-up time ■ niDMM Digits Resolution property or NIDMM_ATTR_RESOLUTION_DIGITS attribute set to 6.5 ■ niDMM Aperture Time Units property or NIDMM_ATTR_APERTURE_TIME_UNITS attribute set to Power Line Cycles ■ niDMM Aperture Time property or NIDMM_ATTR_APERTURE_TIME attribute set to 10 © National Instruments 16 PXI-4065

60 Hz (and 50 Hz) oper

Figure 1. Noise Performance[2]

1 V 1 μV >10 GΩ,

10 V 10 μV >10 GΩ,

100 V 100 μV 10 MΩ 20 + 8 75 + 12 110 + 12 9 + 1

300 V 1 mV 10 MΩ 20 + 24 75 + 40 110 + 40 9 + 1

Table 1. DC Voltage ± (ppm of Reading + ppm of Range)

24 Hr[4]

3 A 3 μA <1 V 1,000[7] + 200 1,200[7] + 200 1,200[7] + 200 65 + 20

Table 2. DC Current ± (ppm of Reading + ppm of Range)

24 Hr[9]

1 MΩ 1 Ω 5 μA 20 + 12 110 + 24 125 + 24 10 + 1

10 MΩ[11] 10 Ω 500 nA 150 + 12 400 + 24 500 + 24 30 + 2

100 MΩ[11] 100 Ω 500 nA||

Table 3. Resistance Table 4. Diode Test[12]

1 Normal 60 dB

2 Second-order >85 dB

Table 5. Normal Mode Rejection Ratio (NMRR)

10 Hz to 100 kHz 1 S/s 6½

100 Hz to 100 kHz 10 S/s 5½

500 Hz to 100 kHz 100 S/s 4½

300 V (± 425 V)

Table 6. AC Voltage (% of Reading + % of Range) Table 7. AC Current (% of Reading + % of Range) Table 8. High Crest Factor Additional Error

Table 9. Thermocouple Temperature Accuracy Specifications (°C) Table 10. RTD[21] T emperature Accuracy Specifications (°C) Table 11. Thermistor Temperature Accuracy Specifications (°C)

Maximum common-mode voltage 300 V ACrms or DC Measurement Category II Caution Do no t use this device for connection to signals or for measurements within Measurement Categories III or IV. Input Protection Characteristics DC I and AC I 3.15 Amp, fused F 3.15 A 250 V, fast-acting user-replaceable fuse Resistance, diode Up to 300 V DC DC V, AC V Up to 300 V DC, 300 V ACrms, 450 V A C peak Fuse When this fuse symbol is mark ed on a device, take proper precautions. Hazardous Voltage This icon denotes a warning advising you to take precautions to avoid electrical shock. Calibration Interval Calibration interval 1 year recommended 23 ni.com PXI-4065

Warm-Up Time Characteristics Warm-up time 30 minutes to rated accuracy Trigger Characteristics Input triggers Types Trigger, Sample Trigger (programmable edge) Sources Auxiliary connector (AUX I/O connector), PXI Trigger lines Minimum pulse width 200 ns Max samples per trigger 2.1 x 109 Trigger delay 0 to 149 s Logic level 5 V TTL, LVTTL Output triggers Types Measurement Complete (programmable edge) Destinations Auxiliary connector (AUX I/O connector), PXI Trigger lines Pulse width 1 μs Logic level 3.3 V Note The A UX I/O connector is not isolated. It is not referenced to your measurement circuit. The connector is referenced to the ground of your © National Instruments 24 PXI-4065

o 5.5 V of your chassis ground. The trigger signals are TTL-compatible. Table 12. PXI Power Consumption, Typical

Indoor use only. Oper ating Environment Ambient temperature range 0 to 55 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) Relative humidity range 10% to 90%, noncondensing Storage Environment Ambient temperature range -40 °C to 70 °C (Tested in accordance with IEC 60068-2-1 and IEC 60068-2-2.) Relative humidity range 5% to 95%, noncondensing (Tested in accordance with IEC 60068-2-56.) Shock and Vibration Operational shock 30 g peak, half-sine, 11 ms pulse (Tested in accordance with IEC 60068-2-27. T est profile developed in accordance with MIL-PRF-28800F.) Random vibration Operating 5 Hz to 500 Hz, 0.3 grms (T ested in accordance with IEC 60068-2-64.) Nonoperating 5 Hz to 500 Hz, 2.4 grms (T ested in accordance with IEC 60068-2-64. Test profile exceeds the requirements of MIL-PRF-28800F, Class 3.) © National Instruments 26 PXI-4065

Compliance and Certifications Caution Y ou can impair the protection provided by the PXI-4065 if you use it in a manner not described in this document. Caution This product is intended for use in industrial locations. As a result, this product may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. Safety This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: ■ IEC 61010-1, EN 61010-1 ■ UL 61010-1, CSA C22.2 No. 61010-1 Note For UL and other safety certifications, refer to the product label or the Online Product Certification sec tion. Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: ■ EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity ■ EN 55011 (CISPR 11): Group 1, Class A emissions ■ EN 55022 (CISPR 22): Class A emissions ■ EN 55024 (CISPR 24): Immunity ■ AS/NZS CISPR 11: Group 1, Class A emissions ■ AS/NZS CISPR 22: Class A emissions 27 ni.com PXI-4065

■ F CC 47 CFR Part 15B: Class A emissions ■ ICES-001: Class A emissions Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations. Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For EMC declarations, certifications, and additional information, refer to the Online Product Certification sec tion. CE Compliance This product meets the essential requirements of applicable European Directives, as f ollows: ■ 2014/35/EU; Low-Voltage Directive (safety) ■ 2014/30/EU; Electromagnetic Compatibility Directive (EMC) Online Product Certification Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit ni.com/certification, se arch by model number or product line, and click the appropriate link in the Certification column. © National Instruments 28 PXI-4065

ommitted to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Minimize Our Environmental Impact web page at ni.com/environment. This p age contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. Waste Electrical and Electronic Equipment (WEEE) EU Customers At the end of the product life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/ environment/weee. 电子信息产品污染控制管理办法(中国 R oHS) 中国客户 National Instruments 符合中国电子信息产品中限制使用某 些有害物质指令(RoHS)。关于National Instruments 中国 RoHS 合规性信 息,请登录 ni.com/environment/rohs_china。(For information about China RoHS compliance, go to ni.com/environment/rohs_china.) 1 Tc al = temperature at which last external calibration was performed. NI factory calibration is 23 °C ± 1 °C. 2 Me asured on the 10 V range. 3 With offse t nulling. 4 R elative to external calibration source. DMM must remain powered on. 29 ni.com PXI-4065

5 Tc al is the temperature at which last external calibration was performed. NI factory calibration is 23°C ±1°C. 6 T empco is the temperature coefficient in ppm of range per degree Celsius. 7 Add 650 ppm/ A of reading for currents above 1.5 A. 8 Specific ations are for 4-wire measurements. For 2-wire measurements, perform offset nulling or add 200 mΩ to specification. For relative humidity >80%, add 100 ppm/MΩ. 9 R elative to external calibration source. DMM must remain powered on. 10 Tc al is the temperature at which last external calibration was performed. NI factory calibration is 23°C ±1°C. 11 2-wir e resistance measurement only. 12 Can be used t o test p-n junctions, LEDs, or zener diodes up to 10 V. 13 Tc al is the temperature at which last external calibration was performed. NI factory calibration is 23°C ±1 °C. 14 U se the 1 Year specification to calibrate on a 90-day cycle. 15 T empco is the temperature coefficient. Tempco values are valid within the device's ambient temperature range.

16 Applic able for non-sinewave signals up to the rated peak voltage, current, or

bandwidth. 17 Tc al = temperature at which last external calibration was performed. NI factory calibration is 23°C ±1 °C. For total measurement accuracy, add temperature probe error. 18 U sing simulated reference junction. © National Instruments 30 PXI-4065

19 Includes P XI 2527 with TB 2627 with a typical 0.5 °C CJC error and a typical thermal EMF offset of 2.5 μV for CJC temperatures between 15 °C and 35 °C. Add an additional 0.5 °C uncertainty when CJC is in the range 0 °C to 15 °C or 35 °C to 50 °C.

20 T empco = Temperature coefficient, expressed in degrees of measurement

uncertainty per degree change in DMM instrument operating temperature.

21 R TD with RO = 100 Ω Pt3851 RTD in a 4-wire configuration, using lowest possible

resistance range for each temperature.

22 T empco is the temperature coefficient, expressed in degrees of measurement

uncertainty per degree change in DMM instrument operating temperature. 31 ni.com PXI-4065 © 2022 National Instruments Corporation.