30W Medical and Industrial DC-DC Converters PXD-M Series 1 Features Benefi ts The PXD-M series of medical and industrial DC-DC converters feature a wide 4:1 input range in a compact 2 x 1” (50.8 x 25mm) industry standard pack- age. With 5kV input to output isolation, 2.5uA touch current and a low off-load power draw, these encapsulated modules are ideal for fi xed and portable applications.

  • Wide 4:1 Input Range
  • Compact 2” x 1” Industry Package
  • Certifi ed to IEC 60601-1 and IEC 62368-1
  • 2 x MOPP Isolation
  • Low No Load Power Consumption
  • Supports Dual 12/24V or 24/48V System Voltages
  • Less Board Area Needed
  • Suitable For Medical and Industrial Applications
  • High Input to Output Isolation
  • Longer Battery Life https://product.tdk.com/en/power/pxd-m www.emea.tdk-lambda.com/pxd-m Model Selector Model Input Output Maximum Maximum No Load Input Effi ciency Maximum Load Voltage (V) Voltage (V) Current (A) Power (W) Current (mA) (%) Capacitance (uF) Single Outputs PXD-M30-24WS05 9 - 36 5 6 30 9 88.5 7200 PXD-M30-24WS12 9 - 36 12 2.5 30 10 89 1200 PXD-M30-24WS15 9 - 36 15 2 30 10 90.5 1000 PXD-M30-24WS24 9 - 36 24 1.25 30 10 89.5 375 PXD-M30-48WS05 18 - 75 5 6 30 8 89 7200 PXD-M30-48WS12 18 - 75 12 2.5 30 9 89 1200 PXD-M30-48WS15 18 - 75 15 2 30 8 90 1000 PXD-M30-48WS24 18 - 75 24 1.25 30 9 89 375 Dual Outputs PXD-M30-24WD05 9 - 36 ±5 ±3.0 30 10 86 ±3600 PXD-M30-24WD12 9 - 36 ±12 ±1.25 30 10 89.5 ±750 PXD-M30-24WD15 9 - 36 ±15 ±1.0 30 9 90 ±500 PXD-M30-48WD05 18 - 75 ±5 ±3.0 30 9 86.5 ±3600 PXD-M30-48WD12 18 - 75 ±12 ±1.25 30 9 90 ±750 PXD-M30-48WD15 18 - 75 ±15 ±1.0 30 8 89.5 ±500 PXD-M- 30 -24W S 05 -N S Single Output D Dual Output 30 30W Output Power -24W 9 - 36V input -48W 18 - 75V input Blank No Remote On/Off -N Negative Logic Remote On/Off -P Positive Logic Remote On/Off 05 5V Output Voltage 12 12V Output Voltage 15 15V Output Voltage 24 24V Output VoltagePreferred model

Specifi cations Model PXD-M30 Input Input Voltage Range Vdc See model selector table Input Surge Voltage Vdc -24W models: 50, -48W models: 100. (3s maximum) Input Shutdown Voltage Vdc -24W models: 7.8 - 8.6, -48W models: 15.8 - 17.4 Start-up Time ms 60 max Touch Current uA 2.5 maximum No Load Power Consumption W See model selector table Effi ciency - See model selector table Conducted & Radiated EMI - EN55011-A, EN55032-A without external components. Class B with external components Immunity - See immunity section Safety Agency Certifi cations - IEC/UL/CSA/EN62368-1, IEC/ANSI/AAMI ES/CSA/EN60601-1, CE Mark (LVD and RoHS) Test Standard Test Level Criteria Notes ESD EN61000-4-2 Air ± 15kV and Contact ± 8kV A - Radiated Susceptibility EN61000-4-3 10V/m A - Surge EN61000-4-5 ± 2kV A capacitors and a TVS (SMDJ58A for PXG-M24W or SMDJ120A for PXG-M48W) Conducted Susceptibility EN61000-4-6 10 Vrms A - Magnetic Fields EN61000-4-8 100A/m continuous; 1000A/m 1s A - Immunity

See website for detailed specifi cations, test methods and installation manual (1): Measured with a 20MHz bandwidth oscilloscope across a 10uF 25V X7R MLCC (4.7uF 50V for 24V models) (2). The case temperature must be confi rmed in end application. The product rating may be affected by airfl ow direction and physical obstructions near the module. Specifi cations Model PXD-M30 Output Output Voltage Tolerance % ±1 Switching Frequency kHz 225 - 285 Line Regulation % Single output: ±0.2. Dual output: ±0.5 Load Regulation % Single output: ±0.2. Dual output: ±1 Cross Regulation % Dual output: ±5 (Asymmetrical 25% to 100% load change) External Load Capacitance uF See model selector table Ripple & Noise (1) mV Single output: 5V outputs: 50, 12V outputs: 75, 15/24V outputs: 100 Dual output: ±5V outputs: 50, ±12/15V outputs: 75 Temperature Coeffi cient %/°C ±0.02 Minimum Load - No minimum load required Transient Loading - 250us recovery time for a 25% load change Overcurrent Protection - Hiccup mode. 150 - 185% Overvoltage Protection V Zener clamp method. 5V: 6.2V, 12V: 15V, 15V: 20V, 24V: 30V (typical) Overtemperature Protection - Shutdown, typically 115°C Remote Sense No remote sense Remote On/Off Positive Logic (-P): ON: Open or 3.5-12V, OFF Short or 0-1.2V Negative Logic (-N): ON: Short or 0-1.2V, OFF: Open or 3.5-12V Parallel Operation - Not possible Series Operation - Possible Environmental Operating Temperature (2) °C Convection:-40 to +105, derate linearly to zero load from +40 to 105 Forced air (200LFM):-40 to +105, derate linearly to zero load from +73 to 105. (Model dependent, confi rm case temperatures in end system) Maximum Case Temperature °C 105. Measured centrally on the upper case (module top view) Thermal Impedance °C/W 12.85 Storage Temperature °C -55 to +125 Humidity (non condensing) %RH 5 - 95 (Operating & Storage) Cooling - Convection or forced air Altitude m 5,000 (operating) Withstand Voltage (For 1 minute) VAC Input to output 5,000 (2xMOPPs) Isolation Capacitance pF 20 Vibration (Operating) - MIL-STD-810F Thermal Shock - MIL-STD-810F Other Weight (Typ) g 32 Size (LxWxH) mm 50.8 x 25.4 x 10.2 Size (LxWxH) Inches 2 x 1 x 0.4 Case Material - Non-conductive black plastic. (Silicone (UL94 V-0) potting) MTBF - MIL-HDBK-217F, Full Load Hours 1,137,000 Warranty yrs 5

Outline Drawing Recommended Hole Pattern PIN Function Single Dual 1 +Vin 2 -Vin

3 Remote On/Off (Option)

6 Trim -Vout

Hole diameter: 0.05” [1.30mm) Top side pad diameter: 0.064” [1.63mm] Bottom side pad diameter: 0.102” [2.60mm] Pinout

PXD-M_Datasheet_260663 August 5th, 2020 v1.5 PXD-M Series 5 For additional information, please visit https://product.tdk.com/en/power/