MVAC400 MURATA-PS | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 10
Technical content
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 1 of 10 ORDERING GUIDE Model Number Natural Convection Cooling Forced Air Cooling Main Output (V1) Fan Output (V2) Aux Output (V3) MVAC400-12AF 250W 400W @ 250LFM 12V 12V MVAC400-24AF 24V MVAC400-48AF 50V MVAC400-12AFD 12V MVAC400-24AFD 24V MVAC400-48AFD 50V MVAC400-24AFT* 24V MVAC400-4AFJT*# 24V MVAC400-12AFR* 12V MVAC400-12AFT* 12V MVAC400-54! 54 Not Available PQC-COVER Optional cover kit assembly; see PQC-COVER datasheet for details. Refer to page 2 for current sharing details for MVAC400-xxAFD and MVAC400-xxAFR models. *CCC Certification is not available for these models. #JST: B2P3-VH Series AC Input Connector Variant ! See Isolation requirements
DESCRIPTION
The MVAC400 series switching power supplies utilize advanced component and circuit technologies to deliver high efficiency. Designed for medical, computing, communications, telecom and other OEM applications to satisfy 1U height design considerations, the MVAC400 Series measures only 3.0" x 5.0" x 1.40". All models offer universal AC input with active power factor correction (PFC) and compliance to worldwide safety and EMC standards Available now at: www.murata -ps.com/en/3d/acdc.html
FEATURES
IEC60601 Ed 3 Medical (2 X MOPP Pri-Sec) EN60950 ITE safety approved Designed to comply with IEC60601-1-2 4th Edition EMC Standard Requirements1 400W compact high density 3" x 5" standard footprint High efficiency up to 94% Remote sense Remote On/Off, Power OK Universal AC input with active PFC Less than 1U high – 1.4" Convection cooled operation up to 250W Isolated 12V@1A fan output Isolated 5V@2A standby output RoHS compliant Active inrush protection Current sharing MVAC400-54 main output is PoE Compatible 1When deployed in the End User equipment INPUT CHARACTERISTICS Parameter Conditions Min. Typ. Max. Units Input Voltage Operating Range Single phase 90 115/230 264 Vac DC 127 300 Vdc Input Frequency 47 50/60 63 Hz Turn-on Input Voltage Input rising 80 90 Vac Turn-off Input Voltage Input falling 70 80 Input Current 90Vac input, full load all outputs 5.5 A No Load Input Power7 (PS_ON = OFF, 5V_Aux = 0A) 1.5 2.0 W Inrush Current At 264Vac, at 25°C cold start 15 Apk Power Factor At 230Vac, full load 0.98 OUTPUT CHARACTERISTICS Model Number Main Output Voltage (V1) Load Current Maximum Load Capacitance Line, Load, Cross Regulation6 Typical Efficiency @230Vac MVAC400-12AFx 12V 0 to 33.3A 0 to 2200µF ± 1% 93% MVAC400-24AFxx 24V 0 to 16.7A 0 to 470µF ± 1% 93% MVAC400-48AFx 50V 0 to 8.0A 0 to 150µF ± 1% 94% MVAC400-54 54V 0-7.4A 0 to 150µF ± 1% 94% MAIN OUTPUT CHARACTERISTICS (ALL MODELS) Parameter Conditions Typ. Max. Units Transient Response9 50% load step, 1A/µsec slew rate ± 5 % Settling Time to 1% of Nominal 500 µsec Turn On Delay After application of input power 3 sec Output Voltage Rise Monotonic5 50 msec Output Holdup 120Vac/60Hz, full load 20 Temperature Coefficient 0.02 %/°C Ripple Voltage & Noise1 1 % Remote SenseNB Compensates for up to 0.5V of lead drop with remote sense connected. Protected against short circuit and reverse connection. 500 mV Hot Swap Transients10 All outputs remain in regulation ± 10 % NB Not available on MVAC400-54 variant AUXILIARY OUTPUT CHARACTERISTICS (ALL MODELS) Auxiliary Output Aux Output Voltage8 Load Current Load Capacitance Line, Load, Cross Regulation3 Ripple & Noise1 Fan (V2) 12V 0 to 1A 0 to 220µF ±10% 2% Aux (V3) 5V 0 to 2A 0 to 220µF ±5% 1% Certificate and Test Report For full details go to www.murata -ps.com/rohs
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 2 of 10 ENVIRONMENTAL CHARACTERISTICS Parameter Conditions Min. Typ. Max. Units Storage Temperature Range -40 85 °C Operating Temperature Range See power rating curves -10 70 Start up -20 Operating Humidity Non-condensing 10 95 % Operating Altitude -200 5000 m MTBF Telcordia SR-332 M1C3 @25°C 474K Hours Shock Operating, MIL-HBK-810E Complies Non-operating, MIL-HBK-810E Complies Operational Vibration IEC-68-2-27 standard Complies to levels of IEC721-3-2 Safety – Medical Standards 2 x MOPP (Primary-Secondary) IEC60601-1 (Ed. 3) – CB Cert and Report ANSI/AAMI ES60601-1 (2005+C1:09+A2:10) CAN/CSA 22.2 No. 60601-1 (2008) 3rd Edition EN60601-1:2006+CORR:2010 Safety – ITE Standards UL 62368-1, 2nd Ed, 2014-12-01 CAN/CSA C22.2 No. 62368-1-14, 2nd Ed IEC 62368-1:2014 EN 62368-1:2014/A11:2017, EN 62368-1:2014 CSA22.2 No.60950-1-07, 2nd Edition, 2001-12. CE Marking per LVD UKCA Marking per Electrical Equipment (Safety) Regulations 2016 Warranty 2 years Weight (typ.) 0.8lbs (362.87g) RESIDUAL RISK (PER ISO 14971 & IEC60601-1) FOR USER CONSIDERATION Fault Condition Residual Risk Complies Contact your Murata salesperson for details PROTECTION CHARACTERISTICS Parameter Conditions Min. Typ. Max. Units Over Voltage Protection4 V1 (main output) latching 110 125 % V3 (aux output) latching 5.5 7.5 V Over Current Protection4 V1, hiccup mode 110 130 %A max V3, auto-recovery 110 150 Over Temperature Protection Auto-recovery Complies Remote Sense Short Circuit Protection Complies Remote Sense Reverse Connection Protection Complies ISOLATION CHARACTERISTICS Parameter Conditions Min. Typ. Max. Units Isolation Primary to Chassis 1500 Vac Primary to Secondary (2xMOPP) 4000 Secondary to Chassis 500 Output to Output 500 Isolation; PoE Variant It shall be required that the 54VDC Main V1 output is provided with the necessary isolation to chassis ground (frame) and other outputs/signals (not associated with the output) to allow compliance with IEEE 802-3at: a) 1500 VRMS steady-state at 50-60 Hz for 60 seconds, applied as specified in sub clause 6.2 of IEC 60950-1:2001. b) An impulse test consisting of a 1500V, 10/700μs waveform, applied 10 times, with a 60 second interval between pulses, applied as specified in sub clause 6.2 of IEC 60950-1:2001. Earth Leakage Current (under single fault condition) 264Vac, 60Hz, 25°C 300 µA Earth Leakage Current (under normal conditions) 264Vac, 60Hz, 25°C 150 µA CURRENT SHARING VARIANTS – MVAC400-xxAFD AND MVAC400-xxAFR Model Number Description MVAC400-12AFD MVAC400-24AFD MVAC400-48AFD MVAC400-12AFR Main Output: Current share is achieved using the droop method. Nominal output voltage is achieved at 50% load and output voltage increases/drops at a rate of:
- 30mv per amp for the 12V output
- 120mV per amp for the 24V output
- 500mV per amp for the 50V output. Startup of parallel power supplies is not internally synchronized. If more than 400W combined power is needed, start-up synchronization must be provided by using a common PS_ON signal. To account for ±10% full load current sharing accuracy and the reduction in full load output voltage due to droop, the available output power must be derated by 15% when units are operated in parallel. Current sharing can be achieved with or without remote sense connected to the common load. If ORing protection is desired use the AFR model or if the AFD model is selected an external ORing FET is recommended (see Applications Note ACAN-42). Aux (V3) output can be tied together for redundancy, but total combined output power must not exceed 10W, external ORing devices must be used. Fan (V2) can be tied together for redundancy, but total com load must not exceed 12W, external ORing devices must be used.
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 3 of 10 EMISSIONS AND IMMUNITY Characteristic Standard Compliance Input Current Harmonics IEC/EN 61000-3-2 Class A Voltage Fluctuation and Flicker IEC/EN 61000-3-3 Complies Conducted Emissions EN 55022 Class B FCC Part 15 Class B ESD Immunity IEC/EN 61000-4-2 Level 4, Criterion 2 Radiated Field Immunity IEC/EN 61000-4-3 Level 3, Criterion A Electrical Fast Transient Immunity IEC/EN 61000-4-4 Level 4, Criterion A Surge Immunity IEC/EN 61000-4-5 Level 3, Criterion A Radiated Field Conducted Immunity IEC/EN 61000-4-6 Level 3, 10V/m, Criterion A Magnetic Field Immunity IEC/EN 61000-4-8 Level 3, Criterion A Voltage dips, interruptions IEC/EN 61000-4-11 Level 3, Criterion B EMI CONSIDERATIONS For optimum EMI performance, the power supply should be mounted to a metal plate grounded to all 4 mounting holes of the power supply. To comply with safety standards, this plate must be properly grounded to protective earth (see mechanical dimension notes). Pre-compliance testing has shown the stand-alone power supply to comply with EN55022 class A radiated emissions. Class B radiated emissions are achievable with a metal enclosure. Radiated emission results vary with system enclosure and cable routing paths. SAFETY CONSIDERATIONS 1. This power supply is a component level power supply intended for use in Class I or Class II applications. Secondary ground traces need to be suitably isolated from primary ground traces when used in Class II applications. 2. When the power supply is used in Class II equipment, all ground traces and components connected to the primary side are considered primary for spacing and insulation considerations. STATUS AND CONTROL SIGNALS Parameter Models Conditions PS_ON11 MVAC400-xxAF MVAC400-xxAFD MVAC400-xxAFR This pin must be pulled low (sink current >2mA) to +5V_AUX_RTN to turn on the main and Fan (V2) output. The +5V_AUX output is independent of the PS_ON signal, and comes up automatically when the input AC or input DC voltage is applied within their specified operating ranges. MVAC400-xxAFT MVAC400-xxAFJT This pin is pulled high internally and so all three outputs (main, Fan output and +5V_AUX) come up automatically when the input AC or input DC voltage is applied within their specified operating ranges. Pulling this pin low (sink current >2mA) to +5V_AUX_RTN will disable the main and fan outputs. PWR_OK All Models Open collector logic goes high 50-200ms after the main output is within regulation; it goes low at least 6msecs before loss of regulation. Internal 10K pull up to +5V_Aux is provided. Applications using the PWR_OK signal should maintain a minimum load of 5W on the main or fan output. 1. Noise and ripple is measured at an oscilloscope jack on the output, 20MHz bandwidth, and with 0.1µF ceramic and 10µF aluminum electrolytic capacitors across the output pins. 2. Unless otherwise specified all measurements are taken at 120Vac input and 25°C ambient temperature. 3. Fan (V2) regulation band applies from 0.1A to 1A load with a minimum of 10W load on the main (V1) output. 4. Fan (V2) has overvoltage protection (tracking V1) and short circuit protection. Overloading the Fan (V2) output can result in permanent damage to the unit. 5. 24V, 50V &54V models may exhibit up to 5% turn on overshoot for loads less than 4% of full load. 6. Load regulation for droop version models (MVAC400-xxAFD and MVAC400-xxAFR) is based the calculated droop voltage ±1.5% (see current sharing section for droop characteristics). 7. No load Input power varies by model and by input line. Measurement is difficult to make due to burst mode operation. Please contact Murata sales if additional information is required. 8. All three output returns are isolated from each other (see isolation characteristics section); the returns may be tied together externally. 9. Load steps beginning from combined loads on the main and fan outputs of less than 5W may result in a transient undershoot outside of the specification limit s. 10.For MVAC400-xxAFR models only: Measured with 220µF capacitance across main output. 11.The MVAC400 is not provided with a PS_ON signal; therefore the output is permanently enabled.
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 4 of 10 PERFORMANCE DATA Inrush Current Time: 100 msec/div, Ch1: 500 V/div, Ch4: 20 A/div, Vin: 264 VAC, Ipk = 15.1A AC applied at peak of sine wave
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 5 of 10 THERMAL CONSIDERATIONS System thermal management is critical to the performance and reliability of the MVAC series power supplies. Performance derating curves are provided which can be used as a guideline for what can be achieved in a system configuration with controlled airflow at various input voltage conditions. The air flow curves are generated using an AMCA 210-99 and ASHRAE 51-1999 compliant wind tunnel with heated inlet air and a controlled CFM providing a duct test section having a calculated average LFM. A correlation between the test setup and the actual system environment is paramount to understanding what can be achieved in an actual system. In a power supply of this density, cooling air moving both through the unit as well as around the unit strongly influences local temperatures. The wind tunnel test setup was constructed to produce a flow with a slight back pressure to induce both flow conditions by providing a small gap between the power supply and duct walls of 0.5" (13mm). The optimal and characterized airflow direction is from the input connector to the output connector (see diagram below). The P-Q flow curve for this test setup is also shown below. The natural convection data is obtained from a horizontally mounted power supply with un -obstructed flow at room temperature. At elevated temperature the power supply data is taken while it is surrounded by a large vented enclosure to minimize forced cross flows inherent in the elevated temperature test system. DERATIONS
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 6 of 10 WIRING DIAGRAM FOR OUTPUT Note: For parallel (current share) operation it is required to connect the sharing power supplies in parallel (+DC out connected together and DC out Return connected together on sharing power supplies . Since each output has an identical “droop” share characteristic then each output will intrinsically share the total load current WIRING DIAGRAM FOR OUTPUT DATASHEET/APPLICATION NOTE Document Number Description Link ACAN-42 MVAC Series External ORING MOSFET Reference Circuit http://www.murata-ps.com/data/apnotes/acan-42.pdf PQC250 Optional cover kit https://power.murata.com/datasheet?/data/acdcsupplies/pqc250-cover.pdf
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 7 of 10 MECHANICAL DIMENSIONS – MVAC400-xxAF AND MVAC400-xxAFD SAFETY CONSIDERATION NOTES: 1. Protective bonding conductor from the end product protective earthing terminal must be tied to TB1. For optimum EMI performance, while maintaining Class I safety isolation all 4 mounting holes must be tied to the end product protective earthing terminal. To maintain Class II safety isolation mounting holes MTG1 and MTG2 need to be isolated from protective earth and should use standoffs of non-conductive material. 2. This power supply requires mounting standoffs of minimum 6mm in height. If there is risk of chassis deformation or shorter standoff height is required, an appropriate insulator must be used under the power supply with adequate extension beyond the outline of the power supply. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 3. The primary heatsink is considered a live primary circuit, and should not be touched. It is recommended that the primary heatsink be kept at least 3.5mm from chassis, and 7mm from secondary circuits. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 4. This product is subject to the following operating requirements and the Life and Safety Critical Application Sales Policy: Refer to: http://www.murata-ps.com/requirements/ 5. Used only in non-tropical conditions. 6. Double pole/neutral fusing. INPUT/OUTPUT CONNECTOR AND SIGNAL SPECIFICATION AND MATING CONNECTORS – MVAC400-xxAF and MVAC400-xxAFD Connector PIN Description Mating Housing Crimp terminal/pins Input Connector J1: Molex 26-62-4030 1 AC Neutral Molex 0009930300 Molex 0008500105 (18-24 AWG) Molex 0008500107 (22-26 AWG) 3 AC Line Output Connector J2: Molex 39-28-1123 1,2,3,4,5,6 +DC_OUT Molex 0039012125 Molex 0039000038 7,8,9,10,11,12 +DC_OUT_RTN Output Connector J3: Molex 90130-1108 1 +5V_AUX Molex 0901420008 Molex 0901190109
2 PWR_OK
3 +12V_FAN_RTN
4 PS_ON
7 +12V_FAN 8 +5V_AUX_RTN
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 8 of 10 MECHANICAL DIMENSIONS – MVAC400-xxAFT AND MVAC400-xxAFR SAFETY CONSIDERATION NOTES: 1. Protective bonding conductor from the end product protective earthing terminal must be tied to TB1. For optimum EMI performance, while maintaining Class I safety isolation all 4 mounting holes must be tied to the end product protective earthing terminal. To maintain Class II safety isolation mounting holes MTG1 and MTG2 need to be isolated from protective earth and should use standoffs of non-conductive material. 2. This power supply requires mounting standoffs of minimum 6mm in height. If there is risk of chassis deformation or shorter standoff height is required, an appropriate insulator must be used under the power supply with adequate extension beyond the outline of the power supply. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 3. The primary heatsink is considered a live primary circuit, and should not be touched. It is recommended that the primary heatsink be kept at least 3.5mm from chassis, and 7mm from secondary circuits. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 4. This product is subject to the following operating requirements and the Life and Safety Critical Application Sales Policy: Refer to: http://www.murata-ps.com/requirements/ 5. Used only in non-tropical conditions. 6. Double pole/neutral fusing. INPUT/OUTPUT CONNECTOR AND SIGNAL SPECIFICATION AND MATING CONNECTORS – MVAC400-xxAF and MVAC400-xxAFD Connector PIN Description Mating Housing Crimp terminal/pins Input Connector J1: Molex 26-62-4030 1 AC Neutral Molex 0009930300 Molex 0008500105 (18-24 AWG) Molex 0008500107 (22-26 AWG) 3 AC Line Output Connector J2: + +DC_OUT Molex 0039012125 6-32 machine screws - +DC_OUT_RTN Output Connector J3: Molex 90130-1108 1 +5V_AUX Molex 0901420008 Molex 0901190109 3 +12V_FAN_RTN 7 +12V_FAN 8 +5V_AUX_RTN
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 9 of 10 MECHANICAL OUTLINE; MVAC400-54 SAFETY CONSIDERATION NOTES: 1. Protective bonding conductor from the end product protective earthing terminal must be tied to TB1. For optimum EMI performance, while maintaining Class I safety isolation all 4 mounting holes must be tied to the end product protective earthing terminal. To maintain Class II safety isolation mounting holes MTG1 and MTG2 need to be isolated from protective earth and should use standoffs of non-conductive material. 2. This power supply requires mounting standoffs of minimum 6mm in height. If there is risk of chassis deformation or shorter standoff height is required, an appropriate insulator must be used under the power supply with adequate extension beyond the outline of the power supply. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 3. The primary heatsink is considered a live primary circuit, and should not be touched. It is recommended that the primary heatsink be kept at least 3.5mm from chassis, and 7mm from secondary circuits. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 4. This product is subject to the following operating requirements and the Life and Safety Critical Application Sales Policy: Refer to: http://www.murata-ps.com/requirements/ 5. Used only in non-tropical conditions. 6. Double pole/neutral fusing INPUT/OUTPUT CONNECTOR AND SIGNAL SPECIFICATION AND MATING CONNECTORS – MVAC400-54 Connector Pin Description Mating Housing Crimp terminal/pins Input Connector J1: Molex 26-62-4030
1 AC Neutral
Molex 0008500105 (18-24 AWG) Molex 0008500107 (22-26 AWG) 3 AC Line Output Connector J2: Molex 39-28-1123 1,2,3,4,5,6 +DC_OUT Molex 0039012125 Molex 0039000038 7,8,9,10,11,12 +DC_OUT_RTN Output Connector J3: Not Populated Not Applicable Not Applicable Not Applicable
PRELIMINARY MVAC400 Series 400W 3" x 5" High Density AC-DC Power Supply https://www.murata-ps.com/support MVAC400.B06 Page 10 of 10 MECHANICAL DIMENSIONS – MVAC400-xxAFJT SAFETY CONSIDERATION NOTES: 1. Protective bonding conductor from the end product protective earthing terminal must be tied to TB1. For optimum EMI performan ce, while maintaining Class I safety isolation all 4 mounting holes must be tied to the end product protective earthing terminal. To maintain Class II safety isolation mounting holes MTG1 and MTG2 need to be isolated from protective earth and should use standoffs of non-conductive material. 2. This power supply requires mounting standoffs of minimum 6mm in height. If there is risk of chassis deformation or shorter standoff height is required, an appropriate insulator must be used under the power supply with adequate extension beyond the outline of the power supply. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 3. The primary heatsink is considered a live primary circuit, and should not be touched. It is recommended that the primary heatsink be kept at least 3.5mm from chassis, and 7mm from secondary circuits. In all cases, the applicable safety standards must be applied to ensure proper creepage and clearance requirements are met. 4. This product is subject to the following operating requirements and the Life and Safety Critical Application Sales Policy: Refer to: http://www.murata-ps.com/requirements/ 5. Used only in non-tropical conditions. 6. Double pole/neutral fusing. INPUT/OUTPUT CONNECTOR AND SIGNAL SPECIFICATION AND MATING CONNECTORS – MVAC400-xxAFJT Connector PIN Description Mating Housing Crimp terminal/pins Input Connector J1: JST B2P3-VH JST SVT-41T-P1.1 (20~16 AWG) 3 AC Line Output Connector J2: + +DC_OUT N/A 6-32 machine screws – +DC_OUT_RTN Output Connector J3: Molex 90130-1108 1 +5V_AUX Molex 0901420008 Molex 0901190109 3 +12V_FAN_RTN 7 +12V_FAN 8 +5V_AUX_RTN Murata Power Solutions, Inc. 129 Flanders Rd. Westborough, Ma 01581, USA. ISO 9001 REGISTERED This product is subject to the following operating requirements and the Life and Safety Critical Application Sales Policy. Refer to: https://www.murata -ps.com/requirements/ Murata Power Solutions, Inc. (“Murata”) makes no representation that the use of its products in the circuits described herein, or the use of other technical information contained herein, will not infringe upon existing or future patent rights. The descriptions contained herein do not imply the granting of licenses to make, use, or sell equipment constructed in accordance therewith. Buyer represents and agrees that it has all the necessary expertise to create and implement safeguards that anticipate dangerous consequences of failures, monitor failures and their consequences, lessen the likelihood of failures that might cause harm, and take appropriate remedial actions. Buyer will fully indemnify Murata, its affiliated companies, and its representatives against any damages arising out of the use of any Murata products in safety-critical applications. Specifications are subject to change without notice. © 2022 Murata Power Solutions, Inc.