PMP150 TTELEC | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 3

Technical content

© TT Electronics plc General Note All data sheets are subject to change without notice. TT Electronics | Power Partners, Inc 43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600 ISS.2 20/07/2020 Page: 1 Description: The PMP150 series of AC/DC switching power supplies are for 132-150 watts of continuous output power. They are enclosed in a 94 V-0 rated polyphenylene-oxide case with an IEC320/C14 or IEC320/C18 inlet to mate with interchangeable cord for world-wide use. All models meet EN55011 and FCC class B emission limits, and are designed for medical applications. Features:

  • Low Safety ground leakage current
  • Wide input range 90 to 264 VAC
  • Optional output connectors
  • 100% burn in
  • Overvoltage protection
  • Over-temperature protection
  • Overcurrent protection
  • Compliant with CEC and Energy Star Efficiency level V requirements (except PMP150-12 and PMP150-13 to level IV) *No Load power consumption *Average active efficiency ≥ 87%
  • Compliant with RoHS Requirements
  • IEC 60601-1-2-4th Edition EMC Compliant Model1 Output Average Active Efficiency (typical) @ 115/ 230 VAC Class I Class II V1 Min. Current Max. Current Tol. Ripple & Noise2 Max Power PMP150-12 PMP150F-12 12V 0A 11A ±5% 120mV 132W 87/86% PMP150-13 PMP150F-13 15V 0A 9A ±5% 150mV 135W 87/86% PMP150-13-2 PMP150F-13-2 19V 0A 7.9A ±5% 190mV 150W 88/88% PMP150-14 PMP150F-14 24V 0A 6.25A ±5% 240mV 150W 88/88% PMP150-15 PMP150F-15 27V 0A 5.56A ±5% 270mV 150W 89/88% PMP150-18 PMP150F-18 48V 0A 3.13A ±5% 480mV 150W 88/88% NOTES: 1. Class I models are equipped with IEC320/C14 inlet, and class II models with IEC320/C18 inlet. 2. Ripple and noise is maximum peak to peak voltage value measured at output within 20 MHz bandwidth, at rated line voltage and output load ranges, and with a 10 μF tantalum capacitor in parallel with a 0.1 μF ceramic capacitor across the output.

© TT Electronics plc General Note All data sheets are subject to change without notice. TT Electronics | Power Partners, Inc 43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600 ISS.2 20/07/2020 Page: 2 Specifications Safety Standards & EMC Specifications Safety Standard Approvals UL ES 60601-1, CSA C22.2 No. 60601-1 File No. E178020 TÜV EN 60601-1 EMI Standard EN55011, FCC & VCCI Class B conducted and radiated EMC Performance EN61000-3-2: Harmonic distortion, class A and D EN61000-3-3: Line flicker EN61000-4-2: ESD, ±15KV air and ±8KV contact EN61000-4-3: Radiated immunity, 10 V/m EN61000-4-4: Fast transient/burst, ±2KV EN61000-4-5: Surge, ±1KV diff., ±2KV com EN61000-4-6: Conducted immunity, 10 Vrms EN61000-4-8: Magnetic field immunity, 30 A/m EN61000-4-11: Voltage dip immunity, 30% reduction for 500ms, 100% reduction for 10ms *Consult with TT Electronics for information on additional country safety approvals Input Specifications Input Voltage Range 90-264 VAC Input Frequency Range 47-63 Hz Input Current 2A (rms) for 115 VAC, 1A (rms) for 230 VAC Earth Leakage Current 220 μA max. @ 264 VAC, 63 Hz Touch Current 100 μA max. @ 264 VAC, 63 Hz Power Factor >0.9 Output Specifications Ripple & Noise 1% peak to peak maximum at full load Overvoltage Protection Provided and set at 112-140% of its nominal output voltage Overcurrent Protection Protected to short circuit conditions Temperature Coefficient ±0.04 %/˚ ˚ ˚ ˚ C maximum Transient Response Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change Environmental Specifications Operating Temperature 0˚ C to +60˚ C (See Derating) Storage Temperature -40˚ C to + 85˚ C Relative Humidity 5% to 95% non-condensing Transient response Maximum excursion of 4% or better on all models, recovering to 1% of final value within 500 us after a 25% step load change General Specifications Switching Frequency 30-110KHz Power Factor 0.98 typical at 115VAC Efficiency Avg. 87% min. (except 95% min. for PMP150-12 and PMP150-13) Hold-up Time 15ms minimum at 110VAC Line Regulation ±0.5% maximum at full load Inrush Current 60A @ 115 VAC or 120A @ 230 VAC, at 25°C cold start Withstand Voltage

5600 VDC from input to output (2 MOPP)

2100VDC from input to ground (1 MOPP) 700VDC from output to ground (To verify AC strength, get correct test method to avoid power supply damage.) For Class II models, 4000 VAC from input to output MTBF 150,000 hours at full load at 25°C ambient , calculated per MIL- HDBK-217F

© TT Electronics plc General Note All data sheets are subject to change without notice. ISS.2 20/07/2020 Page: 3 TT Electronics | Power Partners, Inc 43 Broad Street Suite B206, Hudson, MA 01749, USA. t: +1 (978) 567-9600 Diagrams