Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM Line Reg Load Reg Ripple Efficiency *Model Number Output V Output I (nom/max) (nom/max) & Noise (min/nom) SINGLE OUTPUT PMMK450S-12 U,E,D(-CS) 12 VDC 37.5 Amps 0.3% / 1.0% 0.3% / 1% 120mV pk-pk 77% / 85% *Available in a U-Channel Package (i.e. PMMK450S-12U), an Enclosed Package (i.e. PMMK450S-12E) or an Enclosed Package with a DIN Rail Clip (i.e. PMMK450S-12D) ** The PMMK450S Series power supplies are available with current-sharing (CS) option for N+1 paralleling and redundant power applications. To receive this option with your product simply add “-CS” to the end of the Model Number. For example PMMK450S-12U-CS. Please see the application note at the end of this specification for instructions on how to utilize.

450 Watts Output Power

5656VDC Input to Output Isolation Active Power Factor Correction

5 Volt Standby Power

**Optional Current Share (CS) Capability

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM INPUT SPECIFICATIONS Input Voltage Range 90-264VAC (100-240VAC Nom.) Frequency Range 47-63Hz (50-60 Hz Nom.) Inrush Current, typ: (cold start) 30A at FL, 115VAC 60A at FL, 230VAC Input Current (90Vin/264Vin) 7A / 2.5A max Power Factor (Nom I/P, FL) 0.95 min Leakage current Less than 0.3mA @ 240VAC, 50Hz OUTPUT SPECIFICATIONS Voltage and Current See Selection Chart 5 Volt Standby (+5Vsb) 0~2.0A (See Page 9) Preset Accuracy 1% Line Regulation (LL-HL) See Selection Chart Load Regulation (20%-FL) See Selection Chart DC Output Power 450W, Max (See Derating Curve) Ripple/Noise See Selection Chart 20MHz bandwidth with parallel 0.1uF ceramic capacitor 47uF eletrolytic capacitor Hold Up Time, typ. 16mS at nominal input voltage range Turn On Delay Time < 4S Rise Time < 25mS @110 Vin Short Curcuit Protection Auto-recovery* Over Voltage Protection 130% (max) Auto-recovery* Over Current Protection 150% (max) Auto-recovery* Efficiency (without Oring Diode) See Selection Chart PHYSICAL SPECIFICATIONS Size (PMMK450S-xxU) Millimeters / Inches 203.2x 127x 38 / 8”x 5”x 1.5” Size (PMMK450S-xxE, D) Millimeters / Inches 224.9x 127x 40 / 8.85”x 5”x 1.57” Weight U-Bracket (PMMK450S-xxU) 33.97 oz (963g) Enclosed (PMMK450S-xxE) 37.32 oz (1058g) Enclosed w. DIN Clip (PMMK450-xxD) 38 oz (1077g) ENVIRONMENTAL SPECIFICATIONS Oper. Temperature 0°C to 70°C (See Derate Curve) Storage Temperature -40°C to 85°C Oper. Humidity 10 - 90% for nominal input range Non-condensing Storage Humidity 5 - 95% MTBF @ 25°C 100K Hrs min. MIL-HDBK-217F Vibration* 4G’S pk, 50~500Hz, 3 AXES. 30 Min Drop Test * 70 cm no damage EMS Harmonics 61000-3-2 Class D Fluctuations 61000-3-3 ESD Contact 6KV Air 8KV 61000-4-2 RS FR: 80MHz-2.5GHz Field Strength: 3V / M 61000-4-3 EFT 2KV on AC power line 61000-4-4 Surge 1KV (L-N) & 2KV (L-PE, N-PE) 61000-4-5 CS 3V (EMF) 61000-4-6 DIPS 95% 250Cy , 70% 25 Cy 40% 5Cy , 5% 0.5Cy 61000-4-11 EMI CISPR/EN55011 Class B GENERAL SPECIFICATIONS Safety UL/cUL: ANSI/AAMI ES60601-1 3rd ed. CSA C22.2 No. 60601-1 3rd ed. UL-EU: EN60601-1 3rd ed. CB: IEC60601-1 3rd ed. CE: EN60601-1-2 Over-Shoot and under shoot Less than 10% of nominal output Remote sense 0.5V drop. max at terminals Hi-pot test (60S max) 5656VDC I/P - O/P

2121 VDC I/P - GND

Insulation Resistance (I/P-O/P) 500VDC, 1S, ≥ 20MΩ REMOTE SHUTDOWN 1. Open Collector, Logic “1”=On, “0”=Off. No termination with S/D pin to maintain constant “On” operation. * These are stress ratings. Exposure of the devices to any of these conditions may adversely affect long term reliability. Proper operation under conditions other than the standard operating conditions is neither warranteed nor implied. All specifications are typical at nominal input, full load, and 25°C unless other- wise noted

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM AAASSSTTTRRROOODDDYYYNNNEEE PPPAAACCCIIIFFFIIICCC CCCOOORRRPPP EEELLLEEECCCTTTRRRIIICCCAAALLL SSSPPPEEECCCIIIFFFIIICCCAAATTTIIIOOONNN

6.3 Mechanical specification

U-Bracket (PMMK450S-xxU) Input connector:3P / 8.25 mm pitch terminal block Output connector:7P / 8.25 mm pitch terminal block Unit : mm MECHANICAL DIMENSIONS M3 hole size

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM AAASSSTTTRRROOODDDYYYNNNEEE PPPAAACCCIIIFFFIIICCC CCCOOORRRPPP EEELLLEEECCCTTTRRRIIICCCAAALLL SSSPPPEEECCCIIIFFFIIICCCAAATTTIIIOOONNN Enclosed with Fan (PMMK450S-xxE): Input connector:3P / 8.25 mm pitch terminal block Output connector:7P / 8.25 mm pitch terminal block Unit : mm M3 hole size

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM AAASSSTTTRRROOODDDYYYNNNEEE PPPAAACCCIIIFFFIIICCC CCCOOORRRPPP EEELLLEEECCCTTTRRRIIICCCAAALLL SSSPPPEEECCCIIIFFFIIICCCAAATTTIIIOOONNN Din-rail with Fan (PMMK450S-xxD): Input connector:3P / 8.25 mm pitch terminal block Output connector:7P / 8.25 mm pitch terminal block Unit : mm M3 hole size

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM AAASSSTTTRRROOODDDYYYNNNEEE PPPAAACCCIIIFFFIIICCC CCCOOORRRPPP EEELLLEEECCCTTTRRRIIICCCAAALLL SSSPPPEEECCCIIIFFFIIICCCAAATTTIIIOOONNN 7. Derating Curve U-Bracket (PMMK450S-xxU): Ambient Temperature Derating Curve 1. Convection air 2. Forced air cooling (30CFM air flow is required) Static Characteristics: Input voltage vs. Output Loading

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM AAASSSTTTRRROOODDDYYYNNNEEE PPPAAACCCIIIFFFIIICCC CCCOOORRRPPP EEELLLEEECCCTTTRRRIIICCCAAALLL SSSPPPEEECCCIIIFFFIIICCCAAATTTIIIOOONNN Diagram of external fan placement U-Bracket (PMMK450S-xxU):

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM AAASSSTTTRRROOODDDYYYNNNEEE PPPAAACCCIIIFFFIIICCC CCCOOORRRPPP EEELLLEEECCCTTTRRRIIICCCAAALLL SSSPPPEEECCCIIIFFFIIICCCAAATTTIIIOOONNN Enclosed & Din-rail (PMMK450S-xxE/D): Ambient Temperature Derating Curve Static Characteristics (Input voltage vs. Output Loading)

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM AAASSSTTTRRROOODDDYYYNNNEEE PPPAAACCCIIIFFFIIICCC CCCOOORRRPPP EEELLLEEECCCTTTRRRIIICCCAAALLL SSSPPPEEECCCIIIFFFIIICCCAAATTTIIIOOONNN *1. Test by dc loading side parallel with a 47uF/EC and 0.1u/ceramic capacitors and measured band-width with DC-20MHz. *2. Regulation included line, temperature drift etc tolerance variation. *3. Load regulation measured from 20% to Full load. *4. The efficiency is defined without Oring Diode. *5. All measurements should be made directly at the terminals of the power supply. *6. All specifications typical @ 25℃, unless otherwise noted, at nominal line and load. +5Vsb Standby Power: +5Vsb is a standby output voltage that is always active whenever AC power is present. +5Vsb supplies 1.0A current in normal operation and is capable of withstanding maximum 10W (2 Amps) peak power for 10 seconds. +5Vsb provides a power source for circuits that must remain operational when DC main output is in sleep/off state. +5Vsb over current protection is built internal to its controller. 5Vsb OCP trip point is set to be within 2.5A - 5A. 5Vsb OVP is clamp. PMMK450S-xx SPECIFICATION 24 36 48 Min Nom Max Min Nom Max Min Nom Max DC OUTPUT VOLTAGE (V) 24 36 48 PRESET ACCURACY (%) 1 1 1 OUTPUT CURRENT (A) 0 18.75 0 12.5 0 9.375 O/P NOISE (P-P , mV) *1 240 360 480 EFFICIENCY (typ@Full Load) (%) *4 78 86 79 86 81 88 DC OUTPUT POWER 450W, Max. PMMK450S-xx SPECIFICATION 5Vsb Min Nom Max DC OUTPUT VOLTAGE (V) 4.85 5 5.15 PRESET ACCURACY (%) 1 OUTPUT CURRENT (A) 0 1.0 2.0 O/P NOISE (P-P , mV) *1 75 REGULATION (%) *2 1 LOAD REGULATION (%) *3 1.0 2.0

Medical 450 Watt Enclosed & U-Channel Power Supplies with PFC PMMK450 series ASTRODYNE USA: 1-800-823-8082 ASTRODYNE PACIFIC: 886-2-26983458WWW.ASTRODYNETDI.COM Parallel Operation Instructions * 1.) Follow wiring diagram below 2.) No external oring diodes needed, oring diodes are built in with current share option 3.) Keep wires as short as possible to avoid noise pickup *Notes: Applications for increased power not to exceed 800 Watts Parallel operation of PMK450 not recommended for Medical Applications due to increased leakage current 2915