Features

 Wide spectral range from 1 to 14 µm  High responsivity  Large dynamic range  Excellent long term stability and reliability  Quantity discounted price  Fast delivery

Applications

 FTIR spectroscopy and spectrometry Three-stage thermoelectric cooler parameters Parameter Value Tdet, K ~210 Vmax, V 3.6 Imax, A 0.45 Qmax, W 0.27 Thermistor characteristics Spectral transmission of wZnSeAR window (typical example) 1E+08 1E+09 1E+10 1 2 3 4 5 6 7 8 9 10 11 12 13 14 D*, cm·Hz1/2/W Wavelength, µm 100 1000 180 200 220 240 260 280 300 Thermistor resistance, kΩ Temperature, K 0 100 1 15 Transmission, % Wavelength, µm

PCI-3TE-12-1×1-TO8-wZnSeAR-36 detector datasheet P a g e | 2 25/09/2020 VIGO System S.A. reserves the right to change these specifications at any time without notification. www.vigo.com.pl Mechanical layout, mm Parameter Value Immersion microlens shape hyperhemisphere Optical area AO, mm×mm 1×1 R, mm 0.8 A, mm 4.8±0.35 Φ – acceptance angle R – hyperhemisphere microlens radius A – distance from the bottom of the 3TE-TO8 header to the focal plane Bottom view Function Pin number Detector 1, 3 Thermistor 7, 9 TE cooler supply 2(+), 8(–) Chassis ground 11 Not used 4, 5, 6, 10, 12 Precautions for use and storage  Heatsink with thermal resistance of ~2 K/W is necessary to dissipate heat generated by 3TE cooler.  Operation in 10% to 80% humidity and -20°C to 30°C ambient temperature.  Beam power limitations for optically immersed detector:  irradiance with CW or single pulse longer than 1 μs irradiance on the apparent optical active area must not exceed

2.5 W/cm2,

 irradiance of the pulse shorter than 1 μs must not exceed 10 kW/cm².  Storage in dark place with 10% to 90% humidity and -20°C to 50°C ambient temperature.