MSRF ADVANCEDENERGY | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 4
Technical content
©2020 Advanced Energy Industries, Inc. The HiTek Power MSRF series consists of high-stability, reversible source modules that perform reliably even under short circuit or arc conditions. They offer fast reversible output voltage, medium ripple, and excellent repeatability. PRODUCT HIGHLIGHTS Output power: 3.2 W Output voltage: ±2.5 to ±10 kV Ripple: < 400 mV to < 1 V T emperature coefficient: 10 or 25 ppm per °C High stability: < 0.01% per hour R eversible outputs Fast switching:
- ±2.5 kV < 20 mS to 99%
- ±5 kV < 20 mS to 99%
- ±8 kV < 30 mS to 99%
- ±10 kV < 30 mS to 99% Scr eened case for low magnetic radiation High r eliability Diff erential programming input HITEK POWER MSRF SERIES MASS SPECTROMETRY POWER SUPPLY MODULES
2 advancedenergy.com HITEK POWER MSRF SERIES ELECTRICAL SPECIFICATIONS Output Power 3.2 W, max Output Voltage ±2.5 to ±10 kV Output Current 0.3 to 0.5 mA Input Voltage +24 VDC, ±10% Input Current 1 A max Line Regulation < 10 ppm for a 1 V input voltage change Load Regulation < 10 ppm for a 10 to 100% load change Ripple < 400 mV to < 1 V, depending on model Voltage Control 0 to 10 V = 0 to 100%, accuracy ±2% Current Control Fixed at approximately 110 to 130% of max Voltage Monitor ±10 V = +100 to -100%, accuracy ±2% Current Monitor ±10 V = +100 to -100%, accuracy ±2% Polarity Control Low < 0.8 V = Negative High > 3.5 V or open = Positive Inhibit Low < 0.8 V = Enabled High > 3.5 V or open = Inhibited Stability < 0.01% per hour, 0.05% in eight hours (after one hour warmup) Temperature Coefficient 10 or 25 ppm per °C at max output voltage (tested with external voltage control) Cooling Convection cooled Protection The units are fully protected against over voltage, short circuit and intermittent arcs to ground. Operating Temperature 10 to 50°C (50 to 122°F) Storage/Transport Temperature Operational Altitude Sea level to 2000 m (6500') Storage/Transport Altitude Sea level to 18,000 m (59,055') Reliability MTBF > 50,000 hours Humidity 80% max relative humidity up to 31°C (88°F), reducing linearly to 50% at 40°C (104°F); non-condensing (ref. EN61010-1) Safety Meets the requirements of the Low Voltage Directive, 2006/95/EC by complying with BS EN61010-1:2010 when installed as a component part of compliant equipment. Units are CE marked accordingly. RoHS Meets the requirements of EU Directive 2011/65/EU on the Restriction of use of certain Hazardous Substances in Electrical and Electronic Equipment (RoHS). Construction A fabricated aluminum alloy case is used for good heat dissipation and screening. Options A control option can be supplied with a bipolar input voltage program of ±10 V without the polarity control signal. Please consult our sales team for part numbering for this option. Switching Settle to 99%, ±2.5 kV < 20 mS, ±5 kV < 20 mS, ±8 kV < 30 mS, ±10 kV < 30 mS
advancedenergy.com 3 HITEK POWER MSRF SERIES MECHANICAL SPECIFICATIONS Dimensions 159 mm x 182 mm x 47 mm (6.25" x 7.16" x 1.85") Weight 1.5 kg (3.3 lb) Casing Aluminum, clear non-chrome passivate finish Output Cable Unterminated URM76; 1 m (3.3') of screened output cable Connectors Various options are available upon request. MOUNTING: 4 OFF M3 BLIND FASTENERS; POSITION AS SHOWN DIMENSIONS IN mm (in) .0 (1.85) 154.0 (6.06) 154.0 (6.06) 10.0 (0.39) 160 (6. 31) 182 77.0 (3.03) 28.0 (1.1 184 .4 (7 6) O VER FI XING S 158.7 (6.25) OVER FIXINGS 100 0 (3 37)
ENG-HV-MSRF-235-06 1.20 For international contact information, visit advancedenergy .com. sales.support@aei.com +1 970 221 0108 PRECISION | POWER | PERFORMANCE Read and understand all documentation before you install, operate, or maintain Advanced Energy high voltage power supplies. Follow all safety instructions and precautions to protect against property damage and serious or possibly fatal bodily injury. Never defeat safety interlocks or grounds. CAUTION: High Voltage Specifications are subject to change without notice. Not responsible for errors or omissions. ©2020 Advanced Energy Industries, Inc. All rights reserved. Advanced Energy®, AE®, and HiTek Power® are U.S. trademarks of Advanced Energy Industries, Inc. INTERFACE 20-Way IDC Connector + 24 VDC INPUT SUPPLY NOT CONNECTED +24VDC INPUT SUPPLY VOLTAGE MONITOR +24 VDC INPUT SUPPLY CURRENT MONITOR +24 VDC INPUT SUPPLY POSITIVE INPUT VOLTAGE CONTROL +24 VDC INPUT SUPPLY NEGATIVE INPUT VOLTAGE CONTROL