Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) \\EG-2K1A Seri Dimensions MEG-1K2A4 88.9 x 254 x 40.5mm 3.5 x 10 x 1.59 inch MEG-2K1A6 127 x 254 x 40.5mm 5 x 10 x 1.59 inch Input Input Voltage 90 VAC ~ 264 VAC Input Frequency 47 Hz ~ 63 Hz Input Current <15 A Inrush Current <40 A Power Factor >0.95 @ rated load Efficiency Up to 93% 1) Patient Leakage Current <100 uA normal, <500 uA SFC Earth Leakage Current <300 uA normal, <1 mA SFC Output Module Output Number Single Output Dual Output Consuming Slots Single Slot Triple Slot Single Slot Output Voltage 2 V ~ 60 V 8 V ~ 60 V 3.3 V ~ 30 V Output Power 300 W Max 1200 W Max 240 W Max Ripple & Noise <1% Vrated pk-pk or 100 mV, which is larger Standby Power 5 V / 1 A (No minimum load required) Environmental MTBF 500 KHrs Operation Temperature -20°C ~ 70°C 2) Operation Altitude 5000 m or 50 kPa 1) Exclude fan power 2) Power de-rating with temperature above 50°C, refer to power de-rating curve for detail Highlights & Features Up to 21.6 W/inch3 Power Density for MEG-1K2A Up to 26.4 W/inch3 Power Density for MEG-2K1A6 Full Power up to 50°C Ambient Up to 500 kHrs MTBF 2xMOPP Isolation for Medical Application Output selectable from 2 V to 60 V Current sharing Class B Conducted and Radiated EMI IEC 60601-1-2 4th edition immunity compliance Normal and Reversed Option for Global Remote On/Off & Air Flow Direction without power de-rating Analog and Digital Voltage Trimming PMBus Ver 1.3 Supported Intelligent Fan Speed Control Optional RS485/RS232/USB Communication Adapters PC GUI for easy parameter setting and monitoring Conformal coating on PCBAs to protect against common dust and pollutants Safety Certifications IEC 60601-1 CB report UL 60601-1+CAN/CSA 60601-1 IEC 62368-1 CB report UL 62368-1+CAN/CSA 62368-1 SEMI F47 CCC GB 17625.1; GB 4943.1; GB/T 9254.1 (MEG-1K2A4) CQC GB 17625.1; GB 4943.1; GB/T 9254.1 (MEG-2K1A6)
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) Model Numbering MEG - 2K1 A 6 X ̺ Y-Y-Y-Y ̺ zzzzz ME: Delta Medical Power Supply Max Wattage in Product Series Family Code Slot Number Inlet Type Output Configurations PSU Configuration G: Configurable 2K1: 2,100W 1K2: 1,200W A series 4: 4 Slots 6: 6 Slots T: US Terminal E: EU Terminal C: C14 See below See below Model Information: Model Number Output Slots Input Voltage Maximum Power MEG-1K2A4 4 90-160 Vac Input Current Limit to 8.5 A 160-264 Vac 1200 W MEG-2K1A6 6 90-160 Vac Input Current Limit to 15 A 160-264 Vac 2100 W Output Configurations: Y-Y-Y-Y Slot1-Slot2-Slot3-Slot4 Output Configurations See below For single output module, output module code combined with a voltage code and a current code. For dual output module, output module code is combined with two voltage code. Please check Table 1 for all available combinations. For example: J1: 12 V, 25 A, single slot, single output module. O2: 24 V, 50 A, triple slot, single output module. OJ: Dual output module, one 24 V/4 A output, one 12 V/5 A output. Split the modules with a “-”. If any slot to be left empty, use code “NU”.
All parameters are specified at 25°C ambient unless otherwise noted. Table 1. Output Modules
45.0 A 90 W - - -
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) PSU Configurations: Use following definition for PSU configurations z z z zz Parallel Code Control Code Communication Code CC code See Table 2 0: Normal Logic & Normal Fan Direction 1: Reversed Logic & Normal Fan Direction 2: Normal Logic & Reversed Fan Direction 3: Reversed Logic & Reversed Fan Direction 0: Default PMBus 1: RS232 adapter 2. USB adapter 3: RS485 adapter Use AA for default Parallel Code: Parallel feature is available for the same output modules. Select parallel code, Delta will parallel the outputs before shipping to customer. Parallel feature is designed for singe slot modules and only two modules can be paralleled. Triple slot modules and dual output module cannot support this option. Please notes that trimming the parallelled output voltage through potentiometer is forbidden on customer side. If customer needs to adjust output voltage after paralleling modules, please consult Delta for support. Table 2 Parallel Code Code 1 2 3 4 5 6
0 No Parallel
A B C D E F G H I J K L Examples: MEG-1K2A4T J1-J1-O1-C1 A00AA
4 Slots, US Terminal type input, two 12 V modules in parallel, one 24 V module, one 3 V module
MEG-2K1A6C J1-J1-J1-O1-O1-KP 011AA 6 Slots, C14 inlet, three 12 V module, two 24V module, one dual output module with 14 V and 28 V output, no parallel, reversed logic, with optional RS232 control module MEG-1K2A4E O2-KP 000AA 4 Slots, European Terminal type input, one triple slot 24 V module, one dual output module with 14V and 28 V output MEG-2K1A6C J2-O1-O1-NU D00AA 6 Slots, C14 inlet, one triple slot 12 V module, two 24 V modules in parallel, Slot 6 left empty.
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) Specifications Input Ratings / Characteristics Nominal Input Voltage 100-240 Vac Input Voltage Range 90-264 Vac Nominal Input Frequency 50-60 Hz Input Frequency Range 47-63 Hz Input Current (max) MEG-1K2A: 8.5 A MEG-2K1A: 15 A Input Surge Voltage (max) 300 Vac for 100 ms Full load Efficiency (typ.) 90% @ 115 Vac/60 Hz 93% @ 230 Vac/50 Hz Inrush Current (max) 40A @ 230 Vac, cold start Power Factor (min) >0.95 @ 115 V/50 Hz, 230 V/50 Hz, full load Output Ratings / Characteristics Total Regulation ±3% Output Power Up to 300 W per single slot module, 1200 W per triple slot module Output Voltage Trimming range ±10% of module rated output voltage Line Regulation (max) ±0.5% Load Regulation (max) ±1% Ripple & Noise (typ.) 1% pk-pk Vrated or 100 mV, which is greater Start-up Time (max) 3000 ms @ 115 Vac Hold-up Time (min) 12 ms @ rated load, with nominal input range 20 ms @ 1200 W load, with nominal input range for 6 slot series 20 ms @ 640 W load, with nominal input range for 4 slot series Half rated load @200 Vac/240 Vac for SEMI F47 Dynamic Response (Overshoot & Undershoot O/P Voltage) ±5% @ with 50-100% load change ±6% @ with 50-100% load change for H2/I2/J2 Capacitive load (max) Single Slot Single Output Module:1500 uF on each load Triple Slot Single Output Module:2800 uF on each load Single Slot Dual Output Module:1000 uF on each load Rise time (max) 100 ms Remote Sense Up to 500 mV compensation for voltage drop across external wire connections to load. Short and reverse connection protected. Inhibit Default ON, see detail in description Power Good Open collector signal when output is in regulation. See application note for detail
All parameters are specified at 25°C ambient unless otherwise noted. AC Fail Open collector signal. Pulled low when AC input is removed. Figure 1. Ripple & Noise Measurement Circuit Surrounding Air Temperature Operating Absolute Maximum/Minimum Rating.
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) Protections Overvoltage Latch off. Reset by recycling AC or through I2C Min Nom Max 110% 117% 130% Overload / Overcurrent Hiccup Mode (Non-Latching, Auto-Recovery) Min Nom (default) Max 110% 115% 130% The OCP value can be programed from default 115% down to 0 through I2C for single slot modules and dual output modules. Over Temperature Latch off Short Circuit Hiccup Mode (Non-Latching, Auto-Recovery) Reliability Data MTBF (Minimum) at 160Vac, 35°C (1800W for MEG-2K1A6, 1200W for MEG-1K2A4) >500kHrs based on Telecordia SR-332 Safety Standards / Directives Medical Safety IEC 60601-1 CB report UL 60601-1+CAN/CSA 60601-1 ITE Safety IEC 62368-1 CB report UL 62368-1+CAN/CSA 62368-1 CCC GB 17625.1; GB 4943.1; GB/T 9254.1 (MEG-1K2A4) CQC GB 17625.1; GB 4943.1; GB/T 9254.1 (MEG-2K1A6) CE In conformance with EMC Directive 2014/30/EU and Low Voltage Directive 2014/35/EU EN 60601-1: 2006 + A11: 2011 + A1: 2013 + A12: 2014 & EN 60601-1-2: 2015 UKCA In conformance with Electrical Equipment (Safety) Regulations 2016, and Electromagnetic Compatibility Regulations 2016, Medical Devices Regulations 2002(UK MDR 2002) Galvanic Isolation Input to Output (2xMOPP) Input to Ground (1xMOPP) Output to Ground Output to Output
4000 Vac
1500 Vac
500 Vac (Type B application rated)
500 Vac
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) EMC EMC / Emissions EN/BS EN 55011, EN/BS EN 55032, FCC Title 47:Class B Harmonic Current Emissions IEC 61000-3-2 Meet Class A limit Voltage Flicker IEC 61000-3-3 Immunity to Electrostatic Discharge IEC 61000-4-2 Level 4 Criteria A1)5) Air Discharge: 15 kV Contact Discharge: 8 kV Radiated Field IEC 61000-4-3 IEC 60601-1-2 Level 3 Criteria A1)
80 MHz-1000 MHz, 10 V/m AM modulation
Criteria A1)5)
80 MHz-2700 MHz, 10 V/m AM modulation
385 MHz-5785 MHz, 28 V/m Pulse mode and other
Electrical Fast Transient / Burst IEC 61000-4-4 Level 3 Criteria A1):2 kV Surge IEC 61000-4-5 Level 3 Criteria A1)5) Common Mode3): 2 kV Differential Mode4): 1 kV Conducted IEC 61000-4-6 Level 2 Criteria A1)5) 150 kHz-80 MHz, 3 Vrms, 6 Vrms at ISM bands and Amateur radio bands Power Frequency Magnetic Fields IEC 61000-4-8 Criteria A1)5) Magnetic field strength 30 A/m Voltage Dips IEC 61000-4-11 30% 10 ms Criteria A1) 60% 100 ms Criteria B2) 100% 5000 ms Criteria B2) Voltage Dips5) Criteria A1) @ rated full load 0% UT, 0.5 cycle(10 ms) Criteria B 2), can meet Criteria A with 1200 W or lower load for 6 slot series and 64 0W or lower load for 4 slot series 0% UT,1 cycle (20 ms), 0° Criteria B2) 70% UT,25 cycle (500 ms), 0° Criteria B2) 0% UT,250 cycle (5000 ms), 0° 1) Criteria A: Normal performance within the specification limits 2) Criteria B: Output out of regulation, or shuts down during test. Automatically restored to normal operation after test. 3) Asymmetrical: Common mode (Line to earth) 4) Symmetrical: Differential mode (Line to line) 5) Compliant with IEC-60601-1-2 4th edition requirements.
All parameters are specified at 25°C ambient unless otherwise noted. Figure 2. Dimensional drawing MEG-2K1A
- Case Material: SGCC (conductive).
- Adjustable VR clockwise is to increase the output voltage.
- All dimensions are in millimeters and inches.
- Built-in cooling fan. Must prevent dust suction into power supply, or use natural convection power supply if any concerns.
All parameters are specified at 25°C ambient unless otherwise noted. Figure 3. Dimensional drawing MEG-1K2A
- Case Material: SGCC (conductive).
- Adjustable VR clockwise is to increase the output voltage.
- All dimensions are in millimeters and inches.
- Built-in cooling fan. Must prevent dust suction into power supply, or use natural convection power supply if any concerns.
All parameters are specified at 25°C ambient unless otherwise noted. Figure 8. AC Input Connector Figure 9. Global Control Signals and Communications Connector
1 AC Fail - “Collector” 1 SMBALERT’
2 AC Fail - “Emitter” 2 Address Bit 2 (A2)
7 Global Remote On_Off/Inhibit + 7 PMBus Clock (SCL)
8 Global Remote On_Off/Inhibit - 8 PMBus Data (SDA)
9 Reserve/No Connection 9 5V Communication Bus Return
10 Reserve/No Connection 10 5V Communication Bus +
Table 4. Global Control Signals and Communications Connector (CN503 & CN505) – Pin Table 3. AC Input Connector - pin assignment
All parameters are specified at 25°C ambient unless otherwise noted. Figure 10. -x1 Module Connector Table 5. DC output port - pin assignment Table 6. Control Signals Connector - pin assignment Screws are suitable for slotted and Phillips head screwdrivers.
1 Remote On_Off/Inhibit +
2 Remote On_Off/Inhibit -
3 Remote Sense +
4 Remote Sense -
5 Power Good- “Collector”
6 Power Good- “Emitter”
7 Current Share
8 Reserve/No Connection
Figure 11. -x2 Module Connector Screws are suitable for slotted and Phillips head screwdrivers. Table 7. DC output port - pin assignment Table 8. Control Signals Connector - pin assignment
7 Reserve/No Connection
All parameters are specified at 25°C ambient unless otherwise noted. Figure 12. -x3 Module Connector Table 9. DC output port - pin assignment
1 Remote Inhibit 2 +
2 Remote Inhibit 2 -
4 Remote Inhibit 1 +
5 Remote Inhibit 1 -
Table 10. Control Signals Connector - pin assignment steady state value, after the input voltage is applied. 90% of its final steady state value. falling to 90% of its steady state value. Figure 13. Time Sequence Figure 14. Dynamic from 50% load to 100% Load
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) Power Derating – MEG-2K1A Series and MEG-1K2A Series Power Derating – Triple Slot Modules No air flow direction power derating unless specifically identified.
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) Power Derating – Single Slot Modules & Dual Output Modules
Medical/Industrial AC-DC Configurable Power Supply Up to 2100 Watt / MEG-A Series TECHNICAL DATASHEET All parameters are specified at 25°C ambient unless otherwise noted. www.DeltaPSU.com (November 2023, Rev. S07.3) Certificate Delta has been certified as meeting the requirement of ISO 13485: 2003 and EN ISO 13485:2012 for the design and manufacture of switching power supply and adaptor for medical device. In addition to a UL Total Certification Program (TCP) approved client laboratory for IEC 62368-1. Delta also has participated UL Client Test Data Program (CDTP) for IEC 60601 Attention Delta provides all information in the datasheets on an “AS IS” basis and does not offer any kind of warranty through the info rmation for using the product. In the event of any discrepancy between the information in the catalog and datasheets, the datasheets shall prevail (please refer to www.DeltaPSU.com for the latest datasheets information). Delta shall have no liability of indemnification for any claim or action arising from any error for the provided information in the datasheets. Customer sha ll take its responsibility for evaluation of using the product before placing an order with Delta. Delta reserves the right to make changes to the information described in the datasheets without notice. Manufacturer and Authorized Representatives Information Manufacturer Thailand Delta Electronics (Thailand) PCL. 909 Pattana 1 Rd., Muang, Samutprakarn, 10280 Thailand Taiwan Delta Electronics, Inc.
3 Tungyuan Road, Chungli Industrial Zone, Taoyuan County
32063, Taiwan Authorized Representatives The Netherlands Delta Greentech (Netherlands) B.V. Zandsteen 15, 2132 MZ Hoofddorp, The Netherlands United Kingdom Delta Electronics Europe Limited
1 Redwood Court, Peel Park Campus,
East Kilbride, Glasgow, G74 5PF, United Kingdom