Technical content
from: 2014-05-01 Zhengzhou Winsen Electronic s Technology Co., Ltd
Co., LTD. W ithout the written permission, any part of this manual shall not be cop ied translat ed stored in database or retrieval system, also can t spread through electronic, copying record ways. Thanks for purchasing our product. In order to let customers use it better and reduce the faults caused by misuse, please read the manual carefully and operate it correctly in accordance with the instructions. If users disobey the terms or remove, disassemble, change the c omponents inside of the sensor, we shall not be responsible for the loss. T h e specific such as color, appearance sizes &etc, please in kind prevail. We are devoting ourselves to products develop ment and technical innovation so we r eserve the right to improve the products without notice. Please confirm it is the valid version before using this manual. At the same time, users comments on optimized using way are welcome. Please keep the manual properly, in order to get help if you have questions during the usage in the future. Zhengzhou Winsen Electronics Technology CO., LTD
Co., Ltd www.winsen -sen sor.com T el: 86-371-67169097 /67169670 Fax: 86-371-60932988 Email: sales@winsensor.com ME4-Cl C hlorine Gas Sensor Profile ME4- Cl chlorine gas sensor is constant potential electrolysis type. Oxidation-reduction reaction with chlorine gas and oxygen take place respectively on the working electrode and on the counter electrode. The process releases electric charge and generates current. The current is in direct proportion to the concentration of chlorine gas and conform to faraday's law. So the concentration of the target gas could be got by measuring the value of current.
Features
consumption, high precision, high sensitivity, wide linear range, g oo d anti-interference ability, excellent repeatability and stability. Main
applications
field. Technical Parameters Stable1. Detection Gas Chlorine Gas Cl Detection Range 20ppm Max range 250ppm Sensitivity 0.25 µ A/ppm Resolution 0.1ppm Response Time(T 60S Bias Voltage 0mV Load Resistance Ω (recommended) Repeatability Output value Stability(/month) Output Linearity Linear Zero drift(-20 0.2ppm Temperature range -20 Humidity range RH No condensation Pressure range standard atmospheric pressure Lifespan years Fig1.Sensor S tructure
Co., Ltd www.winsen -sen sor.com T el: 86-371-67169097 /67169670 Fax: 86-371-60932988 Email: sales@winsensor.com Basic Circuit Fig2: ME4-Cl test circuit
Description
Fig3.Responce and Resume Fig4.Linearity Fig5. O utput of sensor at different temperature Fig 6.Zero output of sensor at different temperature
Co., Ltd www.winsen -sen sor.com T el: 86-371-67169097 /67169670 Fax: 86-371-60932988 Email: sales@winsensor.com Cross Interference ME Cl sensor also can respon d to other gases besides target gas Cl Following data are the response characteristics of the sensor to interferential gases at certain concentration for your reference. Stable2. C ross interference Interferential Gas Concentration/ppm ME4-Cl /ppm H S CO 200 NO SO HCN H 400 0.1 C H 400 0.1 HCl CO NH 0.1 Cautions! Tin soldering is prohibited. Before using, power on to aging for more than hours is necessary. Pins can t be broke off or bent. Don t disassemble the sensor to avoid the damage caused by electrolyte leakage. Avoid contacting organic solvent (including Silicone rubber and other adhesive coatings, medicine, oil and high concentration gases. All the electrochemical sensors shall not be encapsulated completely by resin materials, and shall not immerse in pure oxygen environment, otherwise, it will damage the function of sensor All electrochemical sensors shall not be applied in corrosive gas environment, or the sensor will be damaged Zero calibration should be finished in clean air. During test and usage, sensors should avoid the gas inflow vertically The side for inflow can t be choked and polluted. The laminating film above the sensor surface can t be uncovered and damage. Excessive impact or vibration should be avoided It takes some time for the sensor to return to normal state after it is applied in high
Co., Ltd www.winsen -sen sor.com T el: 86-371-67169097 /67169670 Fax: 86-371-60932988 Email: sales@winsensor.com concentration gas Working electrode and reference electrode of the sensor shall be in short circuit when stored Prohibit to use the hot cement or sealant of which the curing temperature is higher than80 to make the capsulation for the sensor. Prohibit storage and usage for long time in alkaline gases with high concentration. Do not use the sensor when the shell is damaged Zhengzhou Winsen Electronics Technology Co., Ltd Add.: NO.299 Jin Suo Road, National Hi-Tech Zone, Zhengzhou, 450001 China Tel.: 371- 67169097 67169670 Fax :0371-60932988 E-mail: sales@winsensor.com