MBC800 BEL | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 13

Technical content

The MBC800 Series of open-frame medical power supplies, with its wide universal 85 - 264 VAC input range, is available at 800 W of output power and a variety of single output voltages. The M BC series is designed and approved to the latest Medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I & Class II applications. These medical power supplies are ide al for monitoring, home health equipment as well as surgical devices.

  • 5 x 8.5 x 1.61 Inch Form Factor (127 x 216 x 41 mm)
  • Convection or Forced Air Cooling
  • Dual Fusing
  • Current Sharing Option
  • Peak Power Capability
  • Standard IEC 60601-1-2: 2014 (4th Edition)
  • 5 VDC Stand by output
  • 12 V fan output
  • Power Good / Power Fail Signal
  • Suitable for BF application
  • Lesser than 1U high
  • High voltage output range up to 58 VDC
  • N+1 redundant power supply
  • Single wire current sharing
  • Built in OR-ing diode / FET (- R suffix)
  • Diagnostic
  • Drug Pump
  • Dialysis
  • Hospital Beds
  • Home Health Care
  • Monitoring
  • Imaging
  • Therapy Devices

2 MBC800 Series

tech.support@psbel.com MBC800-1T12 MBC800-1T12-H MBC800-1T12-P 12 V U-Channel U-Channel + Slotted Cover U-Channel + Plain Cover 25 A 17.5 A 15 A 33.33 A 31.25 A 31.25 A

0.0 A 2%

17.5 A 15 A 33.33 A 31.25 A 31.25 A 17.5 A 15 A 33.33 A 31.25 A 31.25 A 26.66 A 25 A 25 A 12.5 A 8.75 A 7.5 A 16.66 A 15.625 A 15.625 A 10.34 A 7.25 A 6.2 A 13.78 A 12.93 A 12.93 A 1 To order product without the redundancy diode option please add the suffix-Sxxx to your required part number. Please contact the factory for availability. 2 For Ripple measurement minimum output power requirement is 25 W. Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Electrolytic capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges. Specifications are for nominal input voltage, 25°C unless otherwise stated. Input Voltage Universal 85 – 264 VAC / 120 – 390 VDC Input Frequency 47 – 63 Hz Input Current 120 VAC:

240 VAC:

8.00 A max. 3.64 A max. Input Protection In Live & Neutral both F16 A / 250 V No Load Power Over entire input range with main output kept OFF using Remote ON/OFF 3 W typ. Inrush Current 240 VAC: 25 A max. Leakage Current 240 VAC / 50 Hz 400 μA Touch Current < 100 μA Power Factor 120 VAC: 0.98 0.95 Switching Frequency PFC converter: Variable Resonant converter: Variable 85 kHz typical 100 kHz typical

+86 755 298 85888 Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2020 Bel Power Solutions & Protection BCD.01019_B Output Voltage Refer to Model Selection table for details 12 V to 58 V Standby Output Voltage 3 5 V Output Power 4 Forced Air Cooling U-Channel U-Channel + Slotted Cover / Plain Cover 5 up to 800 W up to 750 W Efficiency 120 VAC: 88% Typical 93% Hold-up Time 120 VAC / 240 VAC: 8 ms Line Regulation ± 0.5% Load Regulation ± 1.0% Transient Response 50% to 100% load change, 50 Hz, 50% duty cycle, 0.1 A/μs < 10%, recovery time < 5 ms Voltage Adjustment ± 3% Set Point Tolerance ± 1% Rise Time <100 ms Over Current Protection Hic-Up Type, autorecovery 110% Over Voltage Protection Latch Type, AC Power to be recycled for recovery 114% Short Circuit Protection Latch Type, AC Power to be recycled for recovery Over Temperature Protection Autorecovery 130 - 140°C primary heat sink Current Share Up to 3 supplies connected in parallel (optional) Cooling Forced Air Cooling U-Channel U-Channel + Slotted Cover U-Channel + Plain Cover up to 800 W up to 560 W up to 480 W 3 Standby output voltage 5 V / 1.5 A (convection) with tolerance including set point accuracy, line and load regulation is +/-10%. Ripple and noise is less than 5%. 4 Combined output power of main output, fan supply and standby supply shall not exceed max. power rating.

5 Refer to Derating curves

Conducted Emissions EN 55011-B, CISPR22-B, FCC PART15-B Class B Radiated Emissions EN 55011 With External king core K5B RC 25x12x15-M or equivalent Class A Class B Input Current Harmonics EN 61000-3-2 Class A Voltage Fluctuation and Flicker EN 61000-3-3 Complies ESD Immunity EN 61000-4-2 A Radiated Field Immunity EN 61000-4-3 A Electrical Fast Transient Immunity EN 61000-4-4 A Surge Immunity EN 61000-4-5 A Conducted Immunity EN 61000-4-6 A Magnetic Field Immunity EN 61000-4-8 A Voltage Dips, Interruptions EN 61000-4-11 A & B

4 MBC800 Series

4245 VAC

1625 VAC

1500 VAC

Power Good: Is a TTL signal which goes high after main output reaches 90% of its set value. Pin 6 & Pin 7 of J3 can be used for Remote on/off. Note: - Provision of Inhibit Remote ON/OFF is available. +5V at Pin 7 will switch off the main output. Figure 1. Power Good / Fail Signal Diagram

+86 755 298 85888 Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2020 Bel Power Solutions & Protection BCD.01019_B AC Input Connector J1 Pin 1 Pin 2 Pin 3 AC Line Neutral Earth TE Connectivity: NC6-P107-03 DC Output Connector J2 J2-A J2-B +VE -VE 6-32 inches Screw Pan HD Mating: Designed to accept Ring Tongue Terminal AMP: 8-31886-1, wherein one 16 AWG (max) wire can be crimped. Note: One Ring Tongue Terminal with 16 AWG is recommended for current up to 11 A only. Use multiple tongue terminals with wire for more current. Signal Connector J3 6 Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9 Pin 10 GND 5V AUX PGPF VS – VS + GND RMT CL2 CL1 LS Molex: 22-23-2101 Mating: 22-01-2107; Pins: 08-50-0113 Fan Output 7 J10, J11 Pin 1 Pin 2 +VE -VE TE Connectivity AMP Connectors Description: CONN HEADER VERT 2 POS 2.54 mm, MPN: 640456-2 Mating: 3-641535-2 / TE Connectivity AMP Connectors or: 0022013027 / MOLEX with crimping 08-50-0114 / MOLEX 6 PSU is supplied with J3, pin-6 and pin-7 shorted to enable main output without remote on/off feature. 7 Fan supply output voltage is 12 V / 500 mA with regulation band+/-30 % and Ripple is less than 10%. To get 12 V Fan supply output voltage, minimum 10 % load on Main output voltage is required. Weight 1100 g Dimensions 127 x 216 x 41 mm (5.0 x 8.5 x 1.61 inches)

6 MBC800 Series

Figure 2. Mechanical drawing - U-Channel

Figure 3. Mechanical drawing - U-Channel + Slotted Cover

8 MBC800 Series

Figure 4. Mechanical drawing - U-Channel + Plain Cover

10 MBC800 Series

Figure 8. Power Derating Curve 12 V wi th Plain Cover Figure 9. Power Derating Curve 15 V Figure 10. Power Derating Curve 15 V with Slotted Cover

12 MBC800 Series

Figure 14. Power Derating Curve 24 V and above with Plain Cover

capture window. If the supplies are not initially adj usted inside the capture window the supplies will not current share satisfactorily.

  1. Connect load cables to the outputs of each supply.
  2. Connect the remote sense line s to the load in twisted style. (A common remote sense point must be use d for all the supplies
  3. Connect all the “LS” signal (Pin 10) on the J3 connector between the supplies.
  4. Adjust remote sense voltage of each supply to within 1% of rated output voltage or readjust to required set point. (Adjustmen t

to be done with all other parallel supplies off).

  1. Current sharing between the supplies can be verified by monitoring the output current of each supply with a hall effect DC

current probe. The supplies should share to within 10% of the total load current.

  1. The current share circuit has a capture window voltage of +/ - 1% of the rated output voltage. If the output remote sense

voltage of one of the supplies is adjusted outside the 1% window the supplies will not current share satisfactorily. Figure 7. Current Sharing Block Diagram equipment used in hazardous environments, or nuclear control systems. date manufactured. Specifications are subject to change without notice.