The MBC450 Series of open -frame medical power supplies, with its wide universal 90-264 VAC input range, is available at 450 W of output power and a variety of single output voltages. The MBC series is designed and approved to the latest Medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I applications. These medical power supplies are ide al for monitoring, home health equipment as well as surgical devices.

  • 4 x 6.5 x 1.61 inches
  • Approved to EN/IEC 60601-1
  • Dual Fusing
  • Peak Power Capability
  • Class B EMI & Medical (BF) Safety Approvals
  • Side Fan or Top Fan Mounting Option
  • Current Sharing Option
  • Class B EMI & Medical (BF) Safety Approvals
  • RoHS Compliant
  • CE marked
  • Diagnostic
  • Drug Pump
  • Dialysis
  • Hospital Beds
  • Home Health Care
  • Monitoring
  • Imaging
  • Therapy Devices

2 MBC450 Series

tech.support@psbel.com MODEL 1 OUTPUT VOLTAGE MAX LOAD MINIMUM LOAD RIPPLE & NOISE 2 POWER CONVECTION 420 LFM MBC450-1T05G 5 VDC 31.0 A 55.0 A 0.0 A 2% 275 W MBC450-1T12G 12 VDC 20.83 A 37.5 A 0.0 A 2% 450 W MBC450-1T15G 15 VDC 16.66 A 30.0 A 0.0 A 2% 450 W MBC450-1T24G 24 VDC 12.30 A 18.75 A 0.0 A 2% 450 W MBC450-1T30G 30 VDC 10.0 A 15.0 A 0.0 A 2% 450 W MBC450-1T48G 48 VDC 6.25 A 9.37 A 0.0 A 2% 450 W 1 For Side Fan Mounting option, add suffix -S to the part number (e.g.: MBC450-1T12G-S) For Top Fan Mounting option add suffix -T to the part number (e.g.: MBC450-1T24G-T) For Current Sharing option, add suffix -I to the part number (e.g.: MBC450-1T48G-I or MBC450-1T48G-I-T or MBC450-1T48G-I-S) 2 Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Electrolytic capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges. Please contact factory/ sales representative for minimum load required for ripple to be within specification. Specifications are for nominal input voltage, 25°C unless otherwise stated. PARAMETER DESCRIPTION / CONDITION SPECIFICATIONS Input Voltage Universal 90 – 264 VAC / 120 – 390 VDC Input Frequency 47 – 63 Hz Input Current 120 VAC:

230 VAC:

4.5 A max. 2.3 A max. No Load Power 120 VAC: 0.4 W 0.8 W Inrush Current 120 VAC: 40 A max. 75 A max. Leakage Current Earth Leakage Current Touch Leakage Current 270 µA 45 µA @ 120 VAC / 63 Hz Input Protection Dual fusing, in Live & Neutral T8A / 250 V No Load Power 120 VAC: 0.4 0.3 Switching Frequency PFC converter: Variable Resonant converter: Variable 45 – 160 kHz typical 35 – 250 kHz, 90 kHz typical

+86 755 298 85888 Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2021 Bel Power Solutions & Protection BCD.00335_AG PARAMETER DESCRIPTION / CONDITION SPECIFICATIONS Output Voltage 5 to 48 V Output Power 3, 4 475 W for 24 V, 30 V models & 500 W for 48 V model only for 5 seconds max. 155 to 450 W Standby Output 5 5 VDC Fan Output 6 12 V Efficiency (Full Load)

120 VAC

24, 30 & 48 V models 12 & 15 V models

5 V model

88% 86% 83% typical

230 VAC 24, 30 & 48 V models 90%

Hold Up Time 120 / 230 VAC 10 ms Power Factor 120 VAC

230 VAC

0.98 0.95 Line Regulation ± 0.5% Load Regulation ± 3% Transient Response <10%, 50% to 100% load change, 50 Hz, 50% duty cycle, 0.1 A/µs Recovery time < 5 ms Rise Time < 100 ms Set Point Tolerance ± 1% Output Voltage Adjustment V1 ± 3 % Over Voltage Protection Latch Type >114% Over Current Protection Hiccup type 120 to 150% Short Circuit Protection Short term, auto recovery Over Temperature Protection Automatic recovery 130°C primary heat sink Current Share Up to 2 supplies connected in parallel (optional) Cooling Convection 12 & 15 V models 24, 30 & 48 V models 155 W 250 W 300 W With 420 LFM 12 & 15 V models 24, 30 & 48 V models 275 W 450 W 450 W 3 Combined output power of main output, fan supply and standby supply shall not exceed max. power rating. 4 Derate output power linearly to 80% from 90 VAC to 80 VAC input. 5 Standby output voltage 5 V / 1.5 A (convection) / 2 A (420 LFM) with tolerance including set point accuracy, line & load regulation is +/-10%. Ripple and noise is less than 5%. 6 Fan supply output voltage 12V / 500 mA with tolerance including set point accuracy, line and load regulation is +/-30% and needs min. 1% load on main output to be within regulation band. Ripple and noise is less than 10%. PARAMETER DESCRIPTION / CONDITION Power Good Signal TTL signal goes high after main output is within regulation band, delay is 0.1 to 0.3 s Remote Sense Compensates for 200 mV drop Remote on/off To turn on PSU short remote pin to ground

4 MBC450 Series

tech.support@psbel.com Conducted Emissions EN 55011-B, CISPR22-B, FCC PART15-B Pass Radiated Emissions EN 55011 B Pass Input Current Harmonics EN 61000-3-2 Class D Voltage Fluctuation and Flicker EN 61000-3-3 Pass ESD Immunity EN 61000-4-2 Level 4, Criterion A Radiated Field Immunity EN 61000-4-3 Level 3, Criterion A Electrical Fast Transient Immunity EN 61000-4-4 Level 3, Criterion A Surge Immunity EN 61000-4-5 Level 3, Criterion A Conducted Immunity EN 61000-4-6 Level 3, Criterion A Magnetic Field Immunity EN 61000-4-8 Level 4, Criterion A Voltage Dips, Interruptions EN 61000-4-11 Criterion A & B PARAMETER DESCRIPTION / CONDITION SPECIFICATION Isolation Voltage Input to Output, 2MOPP Input to Earth,1MOPP Output to GDN for type BF

5940 VDC

2121 VDC

1500 VAC

Safety Standards EN 60601-1, IEC 60601-1 (ed.3), ANSI / AAMI ES 60601 - 1, CSA C22.2 No. 60601-1 Agency Approvals Nemko, UL, C-UL CE mark Complies with LVD Directive PARAMETER DESCRIPTION / CONDITION SPECIFICATIONS Operating Temperature Refer to derating curves -20 to 0°C, start-up is guaranteed 0 to +70°C Storage Temperature -40 to 85° C Humidity Non-Condensing 95% HR Altitude Operating: Non-Operating: 10,000 ft. 40,000 ft. Reliability MTBF according to Telcordia -SR332-Issue 3 1.28 million hours

Figure 1. Derating Curves

6 MBC450 Series

tech.support@psbel.com CONNECTOR PIN DESCRIPTION / CONDITION MANUFACTURER / PN AC Input Connector J1 Pin 1 Pin 3 Pin 5 AC line AC neutral Earth Tyco: 1-1123724-3 Mating: 1-1123722-5 DC Output Connector J2 Lug 1 Lug 2 +V1 RTN 6-32 inches Screw Pan HD Mating: Designed to accept Ring Tongue Terminal AMP: 8-31886-1, wherein one 16 AWG (max) wire can be crimped. Note: One Ring Tongue Terminal with 16 AWG is recommended for current up to 11 A only. Use multiple tongue terminals with wire for more current. Signals 7 J3 Pin 1 Pin 2 Pin 3 Pin 4 Pin 5 Pin 6 Pin 7 Pin 8 Pin 9 Pin 10 NC Power Fail Power Good DC Return +5Vstby +VE Remote Sense -VE Remote Sense CS DC Return Remote On/Off Molex: 22-23-2101 Mating: 22-01-2107; Pins 08-50-0113 Fan J4 Pin 1 Pin 2 +VE -VE Mating Connector: Molex 22-01-2025 Pins = 08-50-0113 Earth (Spade Connector) 8 J5 Molex: 19705-4301 Mating: 190030001 7 PSU is supplied with J3 housing, pin-9 and pin-10 shorted to enable main output without remote on/off feature. 8 The J5 (Earth) spade connector can be used for U-Channel option products only. When fan options are required the earth connection provided in the input AC connector should be used (Pin 5 – J1) PARAMETER DESCRIPTION / CONDITION Weight 900 g (1.98 lbs)

Figure 2. Mechanical Drawing (Without Fan Mounting)

8 MBC450 Series

Figure 3. Mechanical Drawing (With Top Fan Mounting)

Figure 4. Mechanical Drawing (With Side Fan Mounting)

10 MBC450 Series

tech.support@psbel.com During the installation and setup of parallel supplies in a system it is important that a single remote sense point be used for all the supplies. The remote sense voltage between the supplies must be adjusted to within 2% to ensure the supplies are inside the 3% capture window. If the supplies are not initially adjusted inside the capture window the supplies will not current share. NOTE: “CURRENT SHARING “facility is inclusive with the unit only with ordering of the “ CURRENT SHARING “ option by adding suffix -I to the part number, i.e. MBC450-1XXX-I or MBC450-1XXX-I. SET-UP PROCEDURE: 1 Connect load cables to the outputs of each supply. 2 Connect the remote sense lines to the load in twisted style. (A common remote sense point must be used for all the supplies in parallel). 3 Connect all the “current share” pins on the J3 connector between the supplies. 4 Adjust remote sense voltage of each supply to within 1% of rated output voltage or readjust to required set point. (Adjustment to be done with all other parallel supplies off). 5 Current sharing between the supplies can be verified by monitoring the output current of each supply with a hall effect DC current probe. The supplies should share to within 10% of the total load current. 6 The current share circuit has a capture window voltage of +/- 3% of the rated output voltage. If the output remote sense voltage of one of the supplies is adjusted outside the 3% window the supplies will not current share. NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice.