MBC40 BEL | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 5
Technical content
The MBC40 Series of open-frame medical power supplies, with its wide universal 90-264 VAC input range, is available at 40 W of output power and a variety of single and multiple output voltages. The MBC series is designed and approved to the latest Medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I & Class II applications. These medical power supplies are ideal for, monitoring, home health equipment as well as surgical devices.
- 4 x 2 x 1.2 Inch Form Factor
- Single to Triple Outputs
- Ultra High Efficiency > 85%
- Approved to EN/IEC 60601-1
- Low leakage current < 250 µA
- No Load Power < 0.3 W
- IEC Protection Class Options: o Class I: Earth pin J4 (no suffix) o Class II: No Earth pin (-2 suffix)
- RoHS Compliant
- Cover Kit Accessory Available
- Dialysis
- Monitoring
- Pumps
- Surgical Devices
- Home Health
- Ultrasound
2 MBC40 Series
tech.support@psbel.com MODEL1 OUTPUT VOLTAGE (VDC) MAX LOAD (A)2 MINIMUM LOAD (A)3 RIPPLE & NOISE 4 Specifications are for nominal input voltage, 25°C unless otherwise stated. PARAMETER DESCRIPTION / CONDITION SPECIFICATION Input Voltage Universal 90 - 264 VAC Input Frequency6 47 to 400 Hz Input Current 120 VAC:
230 VAC:
0.85 A max. 0.45 A max. No Load Power Single output models Multi output models < 0.3 W < 0.5 W Inrush Current 120 VAC: 30 A max. 60 A max. Leakage Current 120 VAC: < 140 µA <250 µA Switching Frequency Typical 67 67 Hz 1 For Class II add suffix -2 (e.g.: MBC40-1012G-2). 2 Maximum current per output channel. Do not exceed total output power rating. 3 Minimum load specified to meet cross regulation. 4 Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges. 5 For MBC40-3003G efficiency is 75% typical.
6 Safety Approved: 47 to 63 Hz
Figure 1. Derating Curve
4 MBC40 Series
tech.support@psbel.com Conducted Emissions EN 55011-B, CISPR22-B, FCC PART15-B Pass Radiated Emissions EN 55011 B Pass Input Current Harmonics EN 61000-3-2 Class D Voltage Fluctuation and Flicker EN 61000-3-3 Pass ESD Immunity EN 61000-4-2 Level 4, Criterion A Radiated Field Immunity EN 61000-4-3 Level 3, Criterion A Electrical Fast Transient Immunity EN 61000-4-4 Level 3, Criterion A Surge Immunity EN 61000-4-5 Level 3, Criterion A Conducted Immunity EN 61000-4-6 Level 3, Criterion A Magnetic Field Immunity EN 61000-4-8 Level 4, Criterion A Voltage Dips, Interruptions EN 61000-4-11 Criterion A & B PARAMETER DESCRIPTION / CONDITION SPECIFICATION Isolation Voltage Input to Output, 2M0PP Input to Ground, 1M0PP Output to Ground Min. 5700 VDC
1500 VAC
500 VAC
Safety Standards Approved to the latest edition of the following standards: CSA/UL60601-1, EN60601-1 and IEC60601-1. Agency Approvals Nemko, UL, C-UL CE mark Complies with LVD Directive CONNECTOR PIN DESCRIPTION / CONDITION MANUFACTURER / PN AC Input Connector J1 Pin 1 Pin 2 AC Neutral AC Line Molex: 26-60-4030 or equivalent Mating: 09-50-3031; Pins: 08-50-0106 DC Output Connector J2 Pin 1,2 Pin 3,4 Pin 5 Pin 6 RTN Tyco: 640445-6 or equivalent Mating: 647402-6; Pins: 3-647409-1 Signal Connector J3 Pin 1 Pin 2 +V1 Sense -V1 Sense Molex: 22-23-2021 or equivalent Mating: 22-01-2021 Earth J4 Molex: 19705-4301 Mating: 190030001 PARAMETER DESCRIPTION / CONDITION Weight 150 g (0.33 lbs.) Dimensions 101.6 x 50.8 x 30.48 mm (4 x 2 x 1.2 inch)