Features

  • Designed for Medical equipment (meet MOOP and MOPP)
  • High efficiency up to 90% at 230Vac full load
  • Universal AC input voltage range: 90-264 VAC
  • Active power factor correction
  • Over temperature, output over voltage, over current and short circuit protection
  • Input over voltage, under voltage protection
  • Low conducted and radiated EMI (EN 55011 class A 6db margin)
  • Safety approval to UL 60601-1 3rd,CSA60601-1 3rd and IEC60950-1 3rd
  • High power density design: 11.2 W/in3
  • Compact size: 3.0 (W) x 1.5 (H) x 5.0 (L) inch
  • ROHS compliant

Description

This is a compact and efficient series of AC -DC power supplies suited for medical and many o ther applications. This family meets the international information technology safety standards. Their high efficiency allows a very minimal power loss in end equipment, resulting in higher reliability, ease of thermal management, and regulatory approvals for an environmentally frien dly end product. With approval to Means of Patient Protection (MOPP), the MBC250 is suitable for use in body floating (BF) applications wherein equipment may come in direct contact with the patient, such as blood pressure monitors or ultrasound equipment. Content

Rev. AA2, 26-Mar-2012 Page 2 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

1 ORDERING INFORMATION

Dash # Outputs Standard: 0 S-Type: 1-9 Vo-Voltage RoHS compliant MBC250-10XX 250W - 1 0 12VDC / 24VDC / 48VDC G

2 OVERVIEW

The MBC250-10XXG is a high efficiency and high power density AC to DC power supply . It use single phase transition mode PFC converter and well proven two -FET forward converter with output diode rectifiers , providing increased system reliability and high efficiency. With around 1 m/s system air cooling, MBC250-1012G can delivery up to 228W continuous output power and 252W for MBC250-1024G and MBC250-1048G . Figure 1: MBC250-10XXG block diagram

Rev. AA2, 26-Mar-2012 Page 3 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

3 ABSOLUTE MAXIMUM RATINGS

Stresses in excess of the absolute maximum ratings may cause performance degradation, adversely affect long -term reliability, and cause permanent damage to the supply. Parameter Conditions / Description Min Nom Max Unit Vi maxc Max continuous input Continuous 264 VAC

4 ENVIRONMENTAL AND MECHANICAL

Parameter Conditions / Description Min Nom Max Unit TA Ambient temperature Vi min to Vi max, I1 nom, -10 +50 °C TAext Extended temp range Derated output +50 +70 °C TS Storage temperature Non-operational -40 +85 °C Dimensions Width 76.2 mm Height 38.1 mm Depth 127.0 mm M Weight 0.5 kg

5 INPUT SPECIFICATIONS

General Condition: TA = -10… 50 °C unless otherwise noted. Active fan air cooling required: 200 LFM (1 m/s) Parameter Conditions / Description Min Nom Max Unit Vi nom Nominal input voltage 100 240 VAC Vi Input voltage ranges Normal operating (Vi min to Vi max) 90 264 VAC Ii max Max input current 3.5 Arms Ii p Inrush current limitation Vi nom=115VAC, T= 25°C (see Figure 2) Vi nom=230VAC, T= 25°C (see Figure 3) 30 60 Ap Fi Input frequency 47 50/60 63 Hz PF Power factor Vi nom==264VAC, > 0.5 Io nom 0.9 W/VA Vi on Turn-on input voltage1) Ramping up 82 89 VAC Vi off Turn-off input voltage1) Ramping down 78 85 VAC η Efficiency Vi nom=230VAC, 0.5∙Io nom, Vo nom, TA = 25 °C 882) % Vi nom=230VAC, 1.0∙Io nom, Vo nom, TA = 25 °C 90 Thold Hold-up Time After last AC zero point, Vo within regulation, Vi = 115 VAC, Po nom 12 18 ms Li Earth leakage current 264Vac at 60Hz 300 uA 1) The power supply is provided with a minimum hysteresis of 2 V during turn-on and turn-off within the ranges. 2) For MBC250-1012G/1048G half load minimum efficiency is 87.5%.

Rev. AA2, 26-Mar-2012 Page 4 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC INPUT FUSE Double fusing with 5A slow-blow type (5 × 20 mm), in series with Live and neutral line inside of the power supply protects against severe defects. The fuses and a VDR form together with the input filter an effective protection against high input transients.

5.1 INRUSH CURRENT

The AC-DC power supply exhibits an X-capacitance of only 2.0 μF, resulting in a low and short peak current, when the supply is connected to the mains. The internal bulk capacitor will be charged through an power resistor which will limit the inrush current (see Figure 2 and 3). Figure 2: Inrush current, Vin = 115VAC, Phase=90° CH1: Vin (200V/div), CH2: Iin (10A/div) Test data: the inrush current is 13A at 115VAC input. Figure 3: Inrush current, Vin = 230VAC,Phase=90° CH1: Vin (200V/div), CH2: Iin (10A/div) Test data: the inrush current is 30A at 230VAC input.

5.2 INPUT UNDER-VOLTAGE

If the sinusoidal input voltage stays below the input under voltage lockout threshold Vi on, the supply will be inhibited. Once the inp ut voltage returns within the normal operating range, the supply will return to normal operation again.

5.3 POWER FACTOR CORRECTION

Power factor correction (PFC) is achieved by controlling the input current waveform synchronously with the input voltage. A specified PFC controller is implemented in single phase transition mode topology giving outstanding PFC

Rev. AA2, 26-Mar-2012 Page 5 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC results over a wide input voltage and load ranges. The input current will follow the shape of the input voltage.

5.4 EFFICIENCY

The high efficiency ( see Figure 4, 5 and 6 ) is achieved by using state-of-the-art silicon power devices in conjunction with transition mode PFC topology minimizing switching losses. Synchronous rectifiers on the output reduce the losses in the high current output path for MBC250-1012G . Schottky diode and ultra fast di ode are used as rectifiers for MBC250-1024G and MBC250-1048G separately due to the high output voltage level. 0 50 100 150 200 250 Po(W) Efficiency (%) 65.00 68.00 71.00 74.00 77.00 80.00 83.00 86.00 89.00 92.00 0 50 100 150 200 250 300 Po(W) Efficiency(%) Figure 4: Efficiency vs. Output Power at 230VAC, MBC250-1012G Figure 5: Efficiency vs. Output Power at 230VAC, MBC250-1024G 65.00 68.00 71.00 74.00 77.00 80.00 83.00 86.00 89.00 92.00 0 50 100 150 200 250 300 Po(W) Efficiency(%) Figure 6: Efficiency vs. Output Power at 230VAC, MBC250-1048G

Rev. AA2, 26-Mar-2012 Page 6 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

6 OUTPUT SPECIFICATIONS

General Condition: TA = -10 … +50 °C unless otherwise noted. Active fan air cooling required: 200 LFM (1 m/s) Parameter Conditions / Description Min Nom Max Unit Main Output VO Vo Nominal output voltage 0.5 ∙IO, Tamb = 25 °C MBC250-1012G 12.0 VDC MBC250-1024G 24.0 VDC MBC250-1048G 48.0 VDC Vo set Output set point accuracy 0.5 ∙IO, Tamb = 25 °C MBC250-1012G -50 50 mV MBC250-1024G -75 75 mV MBC250-1048G -150 150 mV Po nom Nominal output power MBC250-1012G 0 228 W MBC250-1024G 0 252 W MBC250-1048G 0 252 W Io nom Nominal output current MBC250-1012G 0 19* A MBC250-1024G 0 10.5 A MBC250-1048G 0 5.25* A Vo p-p Output ripple voltage Vo nom , Io nom , 20 MHz BW MBC250-1012G 60 120 mVpp MBC250-1024G 80 240 mVpp MBC250-1048G 150 480 mVpp dVo Load Load regulation Vi = Vi nom, 0 - 100 % Io nom MBC250-1012G -360 50 360 mV MBC250-1024G -720 80 720 mV MBC250-1048G -1440 130 1440 mV dVo Line Line regulation Vi =Vi min…Vi max MBC250-1012G -120 80 120 mV MBC250-1024G -120 80 120 mV MBC250-1048G -120 80 120 mV dVdyn Dynamic load regulation ΔIo = 50% Io nom, Io = 50 … 100% Io nom, dIo/dt =1 A/μs, recovery within 5% of Vo nom MBC250-1012G -0.6 0.25 0.6 V MBC250-1024G -1.2 0.45 1.2 V MBC250-1048G -2.4 0.7 2.4 V trec Recovery time 2 ms tdelay Turn-on Delay Time required for output within regulation after intial application of AC input @90 Vac 0.3 1 2 Sec trise Turn on Rise time VO = 10…90% VO nom 10 20 ms CLoad Capacitive loading Tamb = 25 °C MBC250-1012G 4200 μF MBC250-1024G 2300 μF MBC250-1048G 1150 μF

Rev. AA2, 26-Mar-2012 Page 7 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC * For convection cooling condition, the maximum output current is 10A for MBC250-1012G; For convection cooling condition, the maximum output current is 5A for MBC250-1024G; * For convection cooling condition, the maximum output current is 2A for MBC250-1048G. CH1:Vo (10V/div) CH2:Vin (200V/div) Figure5:Turn on AC line 115VAC,full load (200ms/div) CH1:Vo (10V/div) CH2:Vin (200V/div) Figure6:Turn on AC line 230VAC,full load (200ms/div) CH1:Vo (10V/div) CH2:Vin (200V/div) Figure7:Turn off AC line at 115VAC,full load (20ms/div) CH1:Vo (10V/div) CH2:Vin (200V/div) Figure8:Turn off AC line at 230VAC,full load (20ms/div)

Rev. AA2, 26-Mar-2012 Page 10 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

7 PROTECTION

Parameter Conditions / Description Min Nom Max Unit F Input fuses (L and N) Not user accessible, time lag characteristic 5.0 A Vo ov OV threshold MBC250-1012G 14.0 16.0 VDC MBC250-1024G 28.0 32.0 VDC MBC250-1048G 53 57 VDC Io lim Current limit Vi > 90 VAC, Ta < 5 0 MBC250-1012G 21 22.5 25 A MBC250-1024G 11.5 12.6 13.7 A MBC250-1048G 5.8 6.4 6.8 A TSD Over temperature on heat sinks Automatic shut-down 91 96 101 °C

7.1 OVERVOLTAGE PROTECTION

The AC-DC power supply provides a fixed threshold overvoltage (OV) protection implemented with a HW comparator. Once an OV condition has been triggered, the supply will shut down and latch the fault condition. The latch can be unlocked by disconnecting the supply from the AC mains only.

7.2 CURRENT LIMITATION

The main output current limitation will decrease with linear derating to 50% at 70 ºC if the ambient (inlet) temperature increases beyond 50 °C (see Figure 17 ). Figure 17: Output Current Limitation Curve 100 120 -10 0 10 20 30 40 50 60 70 Ambient Temperature(°C) Io (%) Series1

Rev. AA2, 26-Mar-2012 Page 11 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

8 POWER GOOD SIGNAL

The Out -OK output gives a status indication of the converter and the output voltages. It can be used for control functions such as data protection, central syste m monitoring or as a part of a self -testing system. Connecting the Out -OK as shown in Figure 18, VOK <1.0 V indicates that the output voltage(s) of the converter are within the range. Note: Using the potentiometer, the monitor level tracks the programmed output voltage. In an error condition, if the output voltage is out of range due to overload or an external overvoltage, VOK will approach Vp. The output is formed by an NPN transistor. The signal is isolated from the output. Figure 18: Power Good signal

9 ELECTROMAGNETIC COMPATIBILITY

9.1 IMMUNITY

Note: Most of the immunity requirements are derived from EN 61000-6-1:2007. Test Standard / Description Criteria ESD Contact Discharge EN 61000-4-2, Level 2 A RF Susceptibility EN 61000-4-3, Level 3 A Fast Transient/Burst EN 61000-4-4, Level 3 B Surge EN 61000-4-5, Class 3 B RF Conducted Immunity EN 61000-4-6, Class 3 A Magnetic Fields EN 61000-4-8 A Voltage Dips and Interruptions EN 61000-4-11 C

9.2 EMISSION

Test Standard / Description Criteria Conducted Emission EN55011: 0.15 … 30 MHz, QP and AVG, single unit Class A 6 dB Margin Radiated Emission EN55011 : 30 MHz … 1 GHz, QP, single unit Class A 6 dB Margin Harmonic Emissions IEC61000-3-2, Vin = 100 VAC/ 60 Hz, 100% Load Class A IEC61000-3-2, Vin = 240 VAC/ 50 Hz, 100% Load Class A Voltage Fluctuation and Flicker EN61000-3-3 PASS Rules for Unintentional Radiations FCC Part 15, Sub Part-B PASS

Rev. AA2, 26-Mar-2012 Page 12 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

10 SAFETY / APPROVALS

Maximum electric strength testing is performed in the factory according to IEC 60950-1, Input-to-output electric strength tests should not be repeated in the field. Power-One will not honor any warranty claims resulting from electric strength field tests. Parameter Description / Conditions Min Nom Max Unit Agency Approvals UL60601-1 3rd CSA60601-1 3rd IEC 60950-1 3rd Approved by independent body (see CE Declaration) Insulation Safety Rating Input / Case Basic Input / Output Reinforced Output / Case Basic dC Creepage / clearance Primary (L/N) to protective earth (PE) According to safety standard mm Primary to secondary mm Electrical strength test Input to case 2.3 KVDC Input to output 4.0 KVAC Output and Signals to case 1.5 kVAC

Rev. AA2, 26-Mar-2012 Page 13 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

11 MECHANICAL

11.1 DIMENSIONS

19.05 115 M3x0.5P Thread Hole for customer system mounting; Mounting scerw insert into PSU chassis 5.0mm Max. (2holes on each side) Figure 19: Side view 1 127 ±0.5 J1-1 J1-5 J3-1 Outside Row Figure 20: Top view SN LabelI/O label 115 19.05 Figure 21: Side view 2 Note: A 3D step file of the power supply casing is available on request. Air Flow Direction Air Flow Direction

Rev. AA2, 26-Mar-2012 Page 15 / 15 www.power-one.com MBC250-10XXG Data Sheet 12VDC, 24 VDC, 48VDC

11.2 CONNECTIONS

Copyright © 2010 Power -One Inc. All rights reserved. Words and logos that are identified as trademarks and/or service marks are, unless noted otherw ise, the trademarks are protected under numerous U.S. and foreign patents and pending applic ations, maskwork rights, and copyrights. Power -One reserves the right to make changes to any products and services at any time without notice. Power -One assumes no responsibility or liability arising out of the application or use of any information, produc t, or service described herein except as expressly agreed to in writing by Power-One Inc. 1. NUCLEAR AND MEDICAL APPLICATIONS - Power-One products are not designed, intended for use in, or authorized for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems without the express written consent of the respective di visional president of Power -One, Inc. 2. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice.