MBC180 BEL | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 7
Technical content
The MBC180 Series of ultra low open frame medical power supplies feature a wide universal AC input range of 85 V – 264 VAC, offering output power 180 W with 13 CFM of forced air cooling or up to 120 W with natural convection cooling . They are available in a variety of isolated single output voltages. The MBC Series is designed and approved to the latest medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I & Class II applications. These power supplies are ideal for medical, telecom, datacom, industrial equipment and other applications.
- 4 x 2 x 0.75 Inches Form factor
- 180 Watts with Forced Air Cooling
- Approved to EN/IEC 60601-1
- Efficiencies up to 92%
- -40 to 70°C degree operating temperature
- Dual Fusing
- 12 V / 0.5 A Fan Output, Thermal Shut-Down feature
- 3.37 million Hours, Telcordia -SR332-issue 3 MTBF
- Standby Power < 0.5 W
- Medical (BF) Safety Approvals
- Diagnostic
- Drug Pump
- Monitoring
- Dialysis
- Home Health Care
- Portable Equipment
2 MBC180 Series
tech.support@psbel.com MBC180-1T12L Screw Terminal 12 V 9.37 A 10 A 15 A 0.0 A 2% MBC180-1012L Molex Connector MBC180-1T15L Screw Terminal 15 V 7.5 A 8 A 12 A 0.0 A 2% MBC180-1015L Molex Connector MBC180-1T24L Screw Terminal 24 V 4.68 A 5 A 7.5 A 0.0 A 1% MBC180-1024L Molex Connector MBC180-1T30L Screw Terminal 30 V 3.75 A 4 A 6 A 0.0 A 1% MBC180-1030L Molex Connector MBC180-1T48L Screw Terminal 48 V 2.34 A 2.5 A 3.75 A 0.0 A 1% MBC180-1048L Molex Connector MBC180-1T58L Screw Terminal 58 V 1.94 A 2.07 A 3.1 A 0.0 A 1% MBC180-1058L Molex Connector COVER-180-XBC3 metal cover kit accessory Specifications are for nominal input voltage, 25°C unless otherwise stated. Input Voltage Universal (Derate from 100% at 100 VAC to 77% at 80 VAC) 80 - 264 VAC / 390 VDC Input Frequency 47 - 63 Hz Input Current 115 VAC:
230 VAC:
2.2 A max. 1.1 A max. No Load Power Typical for MBC180-1XXX Typical for MBC180-1XXX-PGPF < 0.5 W < 0.85 W Inrush Current
115 VAC:
264 VAC:
Leakage Current Typical (N.A. For Class II Option- without input Earth pin) Touch current 300 µA < 100 µA Power Factor 115 VAC:
0.95 0.90 Switching Frequency PFC PWM 70 to 130 kHz 50 to 80 kHz 1 For Class II Option (without input Earth pin) add suffix: -2 (e.g.: MBC180-1012L-2). 2 Ripple is peak to peak with 20 MHz bandwidth and 10 µF (Tantalum capacitor) in parallel with a 0.1 µF capacitor at rated line voltage and load ranges 3 When used in Cover Kit, de-rate output power to 70 % under all operating conditions.
+86 755 298 85888 Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2017 Bel Power Solutions & Protection BCD.00880_AB Output Power4 With 13 CFM forced air cooling With natural convection cooling at 100 to 264 VAC 180 W up to 120 W Efficiency (typical @ 230 VAC full load)
48 V, 58 V:
24 V, 30 V:
12 V, 15 V:
92% 90% 88% Hold-up Time At 180 W: At 120 W: 10 ms 16 ms Line Regulation +/-0.5% Load Regulation +/-1% Transient Response 25% step load change, at 0.1 A/uS slew rate, 50% duty cycle, 50 Hz = 4% recovery time < 5 ms Voltage Adjustment5 +/-3% Rise Time Typical 55 ms Set Point Tolerance6 +/-1% Over Current Protection > 110% Over Voltage Protection 110 to 140% Short Circuit Protection Hiccup mode Operating Temperature7 Startup guaranteed with spec. deviation -40 to +70°C -40 to 0°C Storage Temperature -40 to +85°C Relative Humidity Non-condensing 5% to 95% Altitude Operating: Non-operating: 16,000 ft 40,000 ft. MTBF Telcordia -SR332-issue 3 3.37 million hours Conducted Emissions EN 55011-B, CISPR22-B, FCC PART15-B Pass Radiated Emissions EN 55011 A; with external core (King core K5B RC 25x12x15-M in input cable) Pass Level B Input Current Harmonics EN 61000-3-2 Class D Voltage Fluctuation and Flicker EN 61000-3-3 Pass ESD Immunity EN 61000-4-2 Level 4, Criterion A Radiated Field Immunity EN 61000-4-3 Level 3, Criterion A Electrical Fast Transient Immunity EN 61000-4-4 Level 3, Criterion A Surge Immunity EN 61000-4-5 Level 3, Criterion A Conducted Immunity EN 61000-4-6 Level 3, Criterion A Magnetic Field Immunity EN 61000-4-8 Level 4, Criterion A Voltage Dips, Interruptions EN 61000-4-11 Criterion B 4 Combined output power of main output, fan supply shall not exceed max. Power rating. 5 Adjustment potentiometer is located on the SMT side of the PCB. 6 Fan supply output voltage tolerance including set point accuracy, line & load regulation is +/-10% and Ripple & noise is less than 10%. 7 Output ripple can be more than 10% of the output voltage.
4 MBC180 Series
tech.support@psbel.com Isolation Voltage Input to Output: (Medical applications) Input to GND: (Not Applicable For Class II Option) Output to GND: for type BF (for type B (N/A for Class II Option)
4000 VAC
1500 VAC
500 VAC
Safety Standard(s) Approved to the latest edition of the following standards: CSA/UL60601-1, EN60601-1 and IEC60601-1. Agency Approvals Nemko, UL, C-UL CE mark Complies with LVD Directive AC Input Connector J1 Pin 1 Pin 2 Pin 3 AC Line Not Fitted AC Neutral Molex: 26-60-4030 Mating: 09-50-3031; Pins: 08-50-0106 DC Output Connector J2 Pin 1, 2, 3 Pin 4. 5. 6 V1 +VE V1 -VE Option 1 (Screw Terminal): Molex: 39357 Series or equivalent Option 2 (Molex Connector): Molex: 26-60-4060 Mating: 09-50-3061; Pins: 08-50-0106 Aux (Fan) Output J3 Pin 1 Pin 2 FAN +VE FAN -VE AMP: 640456-2 Mating: 640440-2 Signal Output8 J4 Pin 1 Pin 2 Pin 3 Vs PGPF GDN AMP :640456-3 Mating: 640440-3 Weight approx. 200 g Dimensions 101.6 x 50.8 x 19.05 mm (4 x 2 x 0.75 inches) Cooling9 180 W with 13 CFM forced air cooling (refer to Mechanical Drawing) Up to 120 W with natural convection cooling (refer to Derating Curve) Power good / AC fail signal specs 8 A TTL signal is available at pin 2 of J4 which goes high 100-500mS after output voltage reaches 90% of set value. It goes low a minimum of 1 ms before output falls below 90% of the set value, when input AC is switched off. 9 180 W with 13CFM forced air cooling and 120 W with natural convection cooling at 100 to 264 VAC.
+86 755 298 85888 Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2017 Bel Power Solutions & Protection BCD.00880_AB Mechanical Drawing – Option 1 without PGPF Mechanical Drawing – Option 2 without PGPF NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following: 1 Stand off, used to mount PCB has OD of 5.4 mm max. 2 Screws, used to fix PCB on stand off, have head dia of 6.0 mm max. 3 Washer, if used, to have dia of 6.5 mm max.
6 MBC180 Series
tech.support@psbel.com Mechanical Drawing – Option 1 with PGPF Mechanical Drawing – Option 2 with PGPF NOTES: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following: 1 Stand off, used to mount PCB has OD of 5.4 mm max. 2 Screws, used to fix PCB on stand off, have head dia of 6.0 mm max. 3 Washer, if used, to have dia of 6.5 mm max.
+86 755 298 85888 Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2017 Bel Power Solutions & Protection BCD.00880_AB NUCLEAR AND MEDICAL APPLICATIONS - Products are not designed or intended for use as critical components in life support systems, equipment used in hazardous environments, or nuclear control systems. TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date manufactured. Specifications are subject to change without notice.