MBC120 BEL | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 5

Technical content

The MBC120 Series of open frame medical power supplies feature a wide universal AC input range of 85 V – 264 VAC, offering 120 W of output power in a compact footprint, with a variety of isolated single output voltages. The MBC series is designed and approved to the latest Medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I & Class II applications. These power supplies are ideal for medical, telecom, datacom, industrial equipment and other applications.

  • 3 x 2 Inch Footprint
  • 120 Watts with Forced Air Cooling
  • Approved to EN/IEC 60601-1
  • Efficiencies up to 93%
  • -40 To 70°C Operating Temperature
  • Dual Fusing
  • Thermal Shut-Down Feature
  • >3.00 Million Hours, Telcordia -SR332-Issue 3
  • Standby Power < 0.3 W
  • Class II Option Available
  • RoHS Compliant
  • CE Marked
  • Diagnostic
  • Drug Pump
  • Dialysis
  • Home Health Care
  • Monitoring
  • Portable Equipment

2 MBC120 Series

tech.support@psbel.com MBC120-1T12L MBC120-1012L Screw Terminal Molex Header 12 V 8.33 A 10.0 A 0.0 A 1% MBC120-1T15L MBC120-1015L Screw Terminal Molex Header 15 V 6.66 A 8.0 A 0.0 A 1% MBC120-1T24L MBC120-1024L Screw Terminal Molex Header 24 V 4.16 A 5.0 A 0.0 A 1% MBC120-1T30L MBC120-1030L Screw Terminal Molex Header 30 V 3.33 A 4.0 A 0.0 A 1% MBC120-1T48L MBC120-1048L Screw Terminal Molex Header 48 V 2.08 A 2.5 A 0.0 A 1% MBC120-1T58L MBC120-1058L Screw Terminal Molex Header 58 V 1.72 A 2.07 A 0.0 A 1% COVER-120-XBC3 metal cover kit accessory Specifications are for nominal input voltage, 25°C unless otherwise stated. Input Voltage Universal (see derating under output power) 85-264 VAC / 390 VDC4 Input Frequency 47-63 Hz Input Current 115 VAC:

230 VAC:

1.2 A max. 0.65 A max. No Load Power Typical < 0.3 W Inrush Current

115 VAC:

264 VAC:

Leakage Current Typical (N.A. For Class II Option) Touch current 300 uA <100 uA Power Factor @ Full Load, Active PFC > 0.95 Switching Frequency Typical 60 KHz 1 For Class II Option (without input Earth pin) add suffix: -2 (e.g.: MBC120-1012L-2). 2 Ripple is peak to peak with 20 MHz bandwidth and 10 μF (Tantalum capacitor) in parallel with a 0.1 μF capacitor at rated line voltage and load ranges.

3 When used in Cover Kit, de-rate output power to 70 % under all operating conditions

4 Functional, not approved.

+86 755 298 85888 Europe, Middle East +353 61 225 977 North America +1 408 785 5200 © 2017 Bel Power Solutions & Protection BCD.00667_AC Output Power Forced cooling (with 200 LFM): Convection cooling (for input 100-264 VAC, de-rate linearly to 80 W @ 85 VAC): 120 W 100 W Efficiency

48 V, 58 V:

24 V, 30 V:

12 V, 15 V:

93% 91% 90% Hold-up Time Typical >10 ms Line Regulation +/-0.5% Load Regulation +/-1% Transient Response 25% step load change, at 0.1A/uS slew rate, 50% duty cycle, 50 Hz = 4% recovery time < 5 ms Voltage Adjustment +/-3% Rise Time Typical 55 ms Set Point Tolerance +/-1% Over Current Protection > 110% Over Voltage Protection Latch type (AC recycling required) 110 to 140% Short Circuit Protection Hiccup mode Operating Temperature5 Startup guaranteed (derate linearly above 50°C to 70°C as per Fig 1) -40 to +70°C -40 to 0°C Storage Temperature -40 to +85°C Cooling Forced: with 300 LFM (Refer to Mech. Drawing) Convection: for input 100-264 VAC (derate linearly to 80 W @ 85 VAC) 120 W 100 W Relative Humidity Noncondensing 5% to 95% Altitude Operating: Nonoperating: 16,000 ft 40,000 ft. Reliability MTBF according to Telcordia –SR332-Issue 3 3.00 million hours Conducted Emissions EN 55011-B, CISPR22-B, FCC PART15-B Pass Radiated Emissions EN 55011 A; with external core (King core K5B RC 25x12x15-M in input cable) Pass Level B Input Current Harmonics EN 61000-3-2 Class D Voltage Fluctuation and Flicker EN 61000-3-3 Pass ESD Immunity EN 61000-4-2 Level 4, Criterion A Radiated Field Immunity EN 61000-4-3 Level 3, Criterion A Electrical Fast Transient Immunity EN 61000-4-4 Level 3, Criterion A Surge Immunity EN 61000-4-5 Level 3, Criterion A Conducted Immunity EN 61000-4-6 Level 3, Criterion A Magnetic Field Immunity EN 61000-4-8 Level 4, Criterion A Voltage Dips, Interruptions EN 61000-4-11 Criterion B 5 Output ripple can be more than 10% of the output voltage.

4 MBC120 Series

4000 VAC

1500 VAC

500 VAC

2 MOPP

1 MOPP

Figure 1. Derating Curve