Features
/square4 Ultra Small size of 2” x 3” x 1” /square4 For 1U Applications /square4 60W convection cooled /square4 Universal Input 90-264Vac /square4 Approved to IEC60601-1, 3 rd Edition with 2 MOPP /square4 Level V Efficiency Compliant Models /square4 Less than 0.5W no-load Power Consumption /square4 3 Year Warranty /square4 Optional LED indicator for power-on /square4 RoHS Compliant
Description
The MB60S Series models provide a reliable power source in high power density in 2” x 3” x 1” package. Fully compliant to the applicable safety and EMC standards, these models will allow easy integration into many Medical applications. All 6 models are CE marked to low voltage directive and approved to Medical standards of IEC60601-1 3 rd edition with 2 MOPP. Model Selection Model Output Current Output Power Ripple & Total OVP Number Volts Convection Cooled Convection Cooled Noise* Regulation Threshold MB60S12K 12V 4.58A 55W 120mV pk-pk ±2% 14.4-18Vdc MB60S15K 15V 4.00A 60W 150mV pk-pk ±2% 18-22.5Vdc MB60S18K 18V 3.33A 60W 180mV pk-pk 2% 21-25.5Vdc MB60S24K 24V 2.50A 60W 240mV pk-pk ±2% 28.8-36Vdc MB60S36K** 36V 1.67A 60W 360mV pk-pk ±2% 42-47Vdc MB60S48K 48V 1.25A 60W 480mV pk-pk ±2% 57.6-72Vdc Notes: * At -20C, the noise and ripple is 2% of the output. ** For product availability, please contact the factory Input Specifications PARAMETER SPECIFICATION NOTES AC Input Voltage: 90-264Vac, single phase AC Input Frequency: 47-63Hz AC Input Current: 120Vac: 1.4A, 240Vac: 0.75A Turn-on Input Voltage: 75V Ramping Up Turn-off Input Voltage: 65V Ramping Down Inrush Current: 40A maximum @ 0C
www.slpower.com MB60S 6-July-15 2 Leakage Current (Input–Earth): <275µA@264Vac, 60 Hz input, NC IEC 60601-1 3 rd Ed – 8.7.3.c Leakage Current (Output–Earth): N/A Leakage Current (Input-Output): <90µA@264Vac, 60 Hz input, NC Input Fuses: F1, F2: 4A, 250VAC Fuses provided on all models Efficiency Typical Measured at 120Vac and full load MB60S12K 83% MB60S15K 85% MB60S18K 85% MB60S24K 88% 24V, 36V, and 48V Models meet Level V requirement MB60S36K 88% MB60S48K 88% No Load Input Power: <0.5W Meet Level V, standby Power Consumption Turn -on Time: <2 Seconds at 120Vac. Hold-up Time: 16mS minimum from loss of ac input at 120 Vac, full load. 55 Watts for 12V output DC Output Specifications PARAMETER SPECIFICATION NOTES Output Power: 60W continuous for operation from -10ºC to 50ºC 55 Watts for 12V output. Cooling: Convection Total Regulation: ±2% for all models Total regulation is the maximum deviation from nominal voltage for all loading conditions Overload Protection: 120% - 180% of rated output current value, Hiccup Mode Short Circuit Protection: Short across the output terminals will not cause damage to the unit. Hiccup Mode Overvoltage Protection: OVP firing reduces output voltage to <50% of nominal in <50mS. See chart for trip range Overtemperature Protection: Automatic Power Shutdown at Tc = 155ºC, Minimum Load: No minimum load is required Ripple and Noise: 0.5% RMS, 1% pk-pk for all models. 20 MHz Bandwidth, differential mode. Measured with noise probe directly across output terminals, and load terminated with 0.1 µF ceramic and 10 µF low ESR capacitors Transient Response: 500 µ s typ. response time for return to within 0.5% of final value for a 50% load change, ∆ i/ ∆ t< 0.2A/ µ s. Max. voltage deviation is 3.5%. Overshoot: 5% overshoot at turn-on, 5% overshoot at turn-off, under all conditions.
www.slpower.com MB60S 6-July-15 3 Safety Standard Compliance Agency CONDITIONS UL ANSI/AAMI ES60101:2005, 3 rd Edition CSA CAN/CSA-C22.2 No. 60601-1 (2008) Demko EN 60601-1:2006 CB Report IEC 60601-1 (3rd Edition) Isolation Type: B rated Isolation Specifications PARAMETER CONDITIONS Rating NOTES Insulation Safety Rating: Input to Ground 1 MOPP Input to Output 2 MOPP Output to Ground Functional Electric Strength Test Voltage: Input to Ground 1800Vac Input to Output 4000Vac Output to Ground 500Vac . Environmental Specifications PARAMETER SPECIFICATION NOTES Operating Temperature: -10° C to +80° C -40° C Startup guaranteed Temperature Derating: For Output Voltage ≥24V, derate output power to 50W @ 60C, 40 Watt @ 70C, and 20 Watts for 80C <24V will derate to 40W at 60C, 30W at 70C, and 20W at 80C Cooling: Convection Storage Temperature: -40° C to +85° C Altitude: Operating: -500 to 3,000 meter Non-operating: -500 to 40,000 ft. Relative Humidity: 5% to 95%, non-condensing Shock: Non-Operating: Half-sine, 40 gpk, 10mS, 3 axes, 6 shocks total Vibration: Random vibration per MIL-STD-810E, Method 514.4, Cat. 1, Figure 514.4-1, 1 hr in each of three axes Reliability Specifications PARAMETER SPECIFICATION NOTES MTBF: 700,000 hours, 25° C ambient, full load Calculation is done based on Telcordia. Reports for each model is available Warranty: 3 Years Limited HALT Data: Per SL Power Halt procedure Report is av ailable
www.slpower.com MB60S 6-July-15 4 EMI/EMC Compliance PARAMETER SPECIFICATION NOTES Conducted Emissions: EN55011/22 Class B; FCC Part 15 Radiated Emissions: EN55011/22 Class A; FCC Part 15 Harmonic Current Emissions EN61000-3-2, Class A Voltage Fluctuations & Flicker EN61000-3-3 Static Discharge Immunity: EN61000-4-2 6kV contact, 8kV air, Criteria A Performance criteria are defined as following: A – Normal performance during and after the test B – Temporary degradation, self-recoverable C – Temporary degradation, operator intervention required to recover the operation RF Field Susceptability EN61000-4-3 (3V/m), Criteria A Fast Transients/Bursts EN61000-4-4 (PS: 2kV-40A, other lines 1kV- 20A), Criteria B Surge Susceptability EN61000-4-5, Installation Class 3 (1kV diff. mode, 2kV common mode), Criteria A Conducted RF Susceptability EN61000-4-6 (3Vrms ), Criteria A Power Frequency Magnetic Field Test EN61000-4-8 (3A/m), Criteria A Voltage Sags & Surges EN61000-4-11, 95% dip/0.5 cycle (Criteria A), 60%/5cycles (Criteria B ), 30%/25 cycles (Criteria A). Notes: 1. Specifications subject to change without notice. 2. Specifications are for convection rating at factory settings with 115Vac input and 25° C ambient unless otherwise stated.
www.slpower.com MB60S 6-July-15 5 Mechanical Drawing Connector Information Input Connector J100 DC Output Connector Ground (FG) PIN 1) AC LINE PIN 2) EMPTY PIN 3) AC NEUTRAL PIN 1) +Vout PIN 3) -Vout PIN 2) +Vout PIN 4) -Vout 19-30258-0187 (Keystone 1285) (Zierick 895)(.187*0.020) Mating Connector: Tyco/AMP 640250-3 Pins = 770461-1 Mating Connector: AMP 640250-4 Pins = 770461-1 Mating Connector Molex 19002-0005 1. Mounting holes should be connected together for EMI purpose 2. FG is safety ground connection 3. This power supply requires mounting on metal sta ndoffs 0.20” (5mm) in height
www.slpower.com MB60S 6-July-15 6 Characteristic Curves Output vs. Temperature -40C start up. At -20C, the supply meet its full spec except ripple & noise might be increased from 1% to 2% of the output voltage 55W convection cooled, derating output power to 30W at 70 °C for outputs 12V and 15V 60W convection cooled, derating output power to 50W at 60 °C and 40W at 70 °C for Output Voltages ≥ 24V 20W convection cooled at 80C Fig.1 Efficiency vs. Loading Fig.2 -10 0 10 20 30 40 50 60 70 80 Output Power(W) Temperature ( °°°°C) Output Power vs . Temperature Convection for 12V Model Convection for 24V and above 70.0% 80.0% 90.0% 100.0% Efficiency (%) Output Power (W) Efficiency vs. Output Power 120Vac 240Vac
www.slpower.com MB60S 6-July-15 7 Ripple & Noise To verify that the output ripple and noise does not exceed the level specified in the product specification, measured using a scope probe socket with 0.1uF ceramic and a 10uF electrolytic capacitor connected in parallel across it, 20MHz BW. 24V OUT , NO LOAD , 90V AC , 60H Z 24V OUT , FULL LOAD , 90V AC , 60H Z 24V OUT , NO LOAD , 264VAC , 50H Z 24V OUT , FULL LOAD , 264VAC , 50H Z
www.slpower.com MB60S 6-July-15 8 Output Transient Response 50% load step within the regulation limits of minimum and maximum load, dI/dt< 0.2A/ µSec. Recovery time not specified as there is no laps in regulation with a 50% Load Step. Maximum voltage deviation is 3.5% 24V OUT , 120V AC , 25% TO 75% LOAD STEP 24V OUT , 240V AC , 25% TO 75% LOAD STEP Output Overload Characteristic 24V OUT , 90V AC 24V OUT , 264V AC Short Circuit Protection Supply shall protect itself against Short Circuit conditions. No damage will occur if the output is shorted. 24V OUT , 90V AC 24V OUT , 264V AC
www.slpower.com MB60S 6-July-15 9 Overvoltage Protection OVP firing reduces output voltage to <50% of nominal in <50ms. See models chart for trip ranges. 24V OUT , FULL LOAD , 90VAC , 60H Z 24V OUT , FULL LOAD , 264VAC , 50H Z Turn On Time 24V OUT , FULL LOAD , 90VAC , 60H Z 24V OUT , FULL LOAD , 264VAC , 50H Z Hold Up Time 24V OUT , FULL LOAD , 120VAC , 60H Z Data Sheet © 2012 SL Power Electronics Corp. The information and specifications contained herein are believed to be correct at the time of publication. However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice .