1 www.PDUKE.com 2015/6/10 MAD40 SINGLE -OUTPUT SERIES

2 X 3 Inch AC-DC POWER SUPPLIES

  • 0.11 WATT LOW STANDBY POWER CONSUMPTION
  • WIDE INPUT VOLTAGE RANGE 85 TO 264VAC, 47 TO 63HZ
  • BUILT-IN CLASS B EMI FILTER
  • OUTPUT VOLTAGE ADJUSTABLE (SINGLE OUTPUT ONLY)
  • 4000VAC INPUT TO OUTPUT 2MOPP INSULATION
  • PROTECTION TYPE CLASS I AND CLASS II
  • LOW LEAKAGE CURRENT UNDER 75 µA
  • OPERATING ALTITUDE 5000M
  • 5 YEAR WARRANTY
  • ANSI/AAMI ES60601-1, EN60601-1, & IEC60601-1 3 rd EDITION SAFETY APPROVALS
  • CE MARKED
  • COMPLIANT TO RoHS II & REACH
  • MEDICAL EQUIPMENT
  • WIRELESS NETWORK
  • TELECOM/DATACOM
  • INDUSTRY CONTROL SYSTEM
  • MEASUREMENT EQUIPMENT
  • SEMICONDUCTOR EQUIPMENT 4000VAC ISOLATION OCP SCP OVP LOW STANDBY POWER PART NUMBER STRUCTURE M A D 40 U S 12 ƶ Application Package Dimension Output Input Output Output Protection Code Code Power Voltage Quantity Voltage Type (W) (VAC) (VDC) Medical A: Open type U:Universal S: Single 05: 5 N o suffix: CLASS I Application U: U chassis type 85 ~ 264 7P5: 7.5 B: CLASS II E: Enclosed type 09: 9 D: Din rail type 12: 12 15: 15 24: 24 28: 28 36: 36 48: 48 53: 53

2 www.PDUKE.com 2015/6/10 TECHNICAL SPECIFICATION All specifications are typical at 230VAC input, full load and 25 ഒ otherwise noted Input Range Output Voltage Output Current @Convection cooled 73 ഒ Ta Max. Output Power Input Power @ No Load Efficiency Model Number VAC VDC A W W % M AD40US05 M U D40US05 M ED40US05 M D D40US05 85 ~ 264 5 8 40 0.11 90 M AD40US7P5 M U D40US7P5 M ED40US7P5 M D D40US7P5 85 ~ 264 7.5 5.34 40 0.11 90 M AD40US09 M U D40US09 M ED40US09 M D D40US09 85 ~ 264 9 4.45 40 0.11 91 M AD40US12 M U D40US12 M ED40US12 M D D40US12 85 ~ 264 12 3.34 40 0.11 92 M AD40US121 M U D40US121 M ED40US121 M D D40US121 85 ~ 264 12 3.34 40 0.11 90 M AD40US15 M U D40US15 M ED40US15 M D D40US15 85 ~ 264 15 2.67 40 0.11 92 M AD40US151 M U D40US151 M ED40US151 M D D40US151 85 ~ 264 15 2.67 40 0.11 90 M AD40US24 M U D40US24 M ED40US24 M D D40US24 85 ~ 264 24 1.67 40 0.11 92 M AD40US28 M U D40US28 M ED40US28 M D D40US28 85 ~ 264 28 1.43 40 0.11 91 M AD40US36 M U D40US36 M ED40US36 M D D40US36 85 ~ 264 36 1.12 40 0.11 92 M AD40US48 M U D40US48 M ED40US48 M D D40US48 85 ~ 264 48 0.84 40 0.11 93 M AD40US53 M U D40US53 M ED40US53 M D D40US53 85 ~ 264 53 0.77 40 0.11 92.5

3 www.PDUKE.com 2015/6/10 OUTPUT SPECIFICATIONS Parameter Conditions Min. Typ. Max. Unit Output power 40 Watts Initial set voltage accuracy 230VAC and Full Load -1.0 +1.0 % Line regulation Low Line to High Line at Full Load -0.2 +0.2 % Load regulation No Load to Full Load 5Vout Others -0.7 -0.5 +0.7 +0.5 % 10% Load to 90% Load 5Vout Others -0.6 -0.4 +0.6 +0.4 Voltage adjustability Single output 53Vout Others -20 -10 +10 +10 % Minimum load 0 % Ripple and noise Measured by 20MHz bandwidth With a 10 µF/25V 1206 X7R MLCC 5Vout, 7.5Vout, 9Vout 12Vout, 15Vout 75 mVp-p With a 1 µF/50V 1206 X7R MLCC 24Vout, 28Vout, 36Vout 75 mVp-p With a 0.1 µF/100V 1206 X7R MLCC 48Vout, 53Vout 150 mVp-p Temperature coefficient -0.02 +0.02 %/ ഒ Transient response Load step form 50 ~ 75% change at 2.5A/ µs Peak deviation Recovery time 600 %Vout µs Over voltage protection % of Vout(nom); Latch mode 125 140 % Over load protection % of Iout rated; Hiccup mode 145 % Short circuit protection Continuous, automatics recovery GENERAL SPECIFICATIONS Parameter Conditions Min. Typ. Max. Unit Isolation voltage 1 minute (2MOPP insulation) Input to Output Input (Output) to F.G. 4000 2500 VAC Isolation resistance 500VDC 0.1 G Ω Switching frequency 230VAC 5Vout Others 120 kHz Safety approvals ANSI/AAMI ES60601-1 EN60601-1 IEC60601-1 Weight MAD MUD MED MDD 114g (4.02oz) 154g (5.43oz) 169g (5.96oz) 190g (6.70oz) MTBF MIL-HDBK-217F, Full load 3.010 x 10 6 hrs ENVIRONMENTAL SPECIFICATIONS Parameter Conditions Min. Typ. Max. Unit Operating ambient temperature Natural convection With derating -40 +85 ഒ Storage temperature range -40 +85 ഒ Operating altitude 5000 m Shock IEC60068-2-27 Vibration IEC60068-2-6 Relative humidity Non-condensing 5% to 95% RH INPUT SPECIFICATIONS Parameter Conditions Min. Typ. Max. Unit Operating input voltage range AC input DC input 120 264 370 VAC VDC Input frequency AC input 47 63 Hz Input current 100VAC and Full Load 240VAC and Full Load 1.0 0.5 A No load input power 230VAC 0.11 Watts Leakage current 264VAC 75 µA Start up time 1000 ms Rise time 20 ms Hold up time 115VAC and Full Load 25 ms Input inrush current 230VAC 60 A Input protection Internal fuse in line and neutral T3.15A/250VAC

4 www.PDUKE.com 2015/6/10 EMC SPECIFICATIONS Parameter Conditions Level EMI EN55011, EN55022 and FCC Part 18 Conducted Radiated Class B Class B Harmonic currents EN61000-3-2 Full Load Class A Voltage flicker EN61000-3-3 ESD EN61000-4-2 Air ± 8kV and Contact ± 6kV Perf. Criteria A Radiated immunity EN61000-4-3 20 V/m Perf. Criteria A Fast transient EN61000-4-4 ± 2kV Perf. Criteria A Surge EN61000-4-5 DM ± 1kV and CM ± 2kV Perf. Criteria A Conducted immunity EN61000-4-6 20 Vr.m.s Perf. Criteria A Power frequency magnetic field EN61000-4-8 10 A/m Perf. Criteria A 230VAC 50Hz 30% 60% >95% >95% 500mS 100mS 10mS 5000mS Perf. Criteria A Perf. Criteria A Perf. Criteria A Perf. Criteria B Dip and interruptions EN60601-1-2 EN61000-4-11 100VAC 50Hz 30% 60% >95% >95% 500mS 100mS 10mS 5000mS Perf. Criteria A Perf. Criteria B Perf. Criteria A Perf. Criteria B CHARACTERISTIC CURVE Derating Curve vs. Ambient Temperature M □D40USxx xx=5/7P5/9/121/151/28 Derating Curve vs. Ambient Temperature M □D40USxx xx=12/15/24/36/48/53 Derating Curve vs. Input Voltage M ƶD40 Efficiency vs. Output Load MAD40US24 Efficiency vs. Input Voltage MAD40US24

5 www.PDUKE.com 2015/6/10 MECHANICAL DRAWING M AD Open type M D D Din rail type M ED Enclosed type M U D U chassis type 1. All dimensions in inch (mm) CONNECTORS CONNECTIONS CON1 – Input Connector CON2 – Output Connector Pin 1 Line Pin 1,2 -Vout Pin 3 Neutral Pin 3,4 +Vout Mates with JST housing : VHR-3N JST crimp terminals : SVH-21T-P1.1 Mounting holes marked with must be connected to safety earth for CLASS I application Mates with JST housing : VHR-4N JST crimp terminals : SVH-21T-P1.1