User’s Guide 400A AC/DC Clamp Meter Model MA220

MA220 V1.3 10/06 2 Introduction Thank you for selecting the Extech MA200 AC/DC Clamp Meter. This meter measures AC/DC Current, AC/DC Voltage, Resistance, Capacitance, Frequency, Duty Cycle, Temperature, Diode Test, and Continuity. This professional meter, with proper care, will provide years of safe reliable service. Safety International Safety Symbols This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information. This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present Double insulation A UL mark does not indicate that this product has been evaluated for the accuracy of its readings. SAFETY NOTES

  • Do not exceed the maximum allowable input range of any function
  • Do not apply voltage to meter when resistance function is selected.
  • Set the function switch OFF when the meter is not in use.
  • Remove the battery if meter is to be stored for longer than 60 days. WARNINGS
  • Set function switch to the appropriate position before measuring.
  • When measuring volts do not switch to current/resistance modes.
  • Do not measure current on a ci rcuit whose voltage exceeds 240V.
  • When changing ranges using the selector sw itch always disconnect the test leads from the circuit under test.
  • Do not exceed the maximum rated input limits. OVERVOLTAGE CATEGORY III This meter meets the IEC 610-1-95 standard for OVERVOLTAGE CATEGORY III. Cat III meters are protected against overvoltage transients in fixed installation at the distribution level. Examples include switches in the fixed installation and some equipment for industrial use with permanent connection to the fixed installation.

MA220 V1.3 10/06 3 Specifications Function Range Accuracy (of reading) 0-20.00 ADC ± (2.5% + 6 digits) 40.00 ADC 20.00-40.00 ADC ± (3% + 6 digits) 0-300.0 ADC ± (2.5% + 6 digits) DC Current

400.0 ADC

300.0-400.0 ADC ± (3.5% + 6 digits) 0-20.00 AAC ± (3% + 10 digits) 40.00 AAC 20.00-40.00 AAC ± (5% + 10 digits) 0-300.0 AAC ± (3% + 10 digits) AC Current

400.0 AAC

300.0-400.0 AAC ± (5% + 10 digits) 400.0mV ± (0.8% + 3 digits) 4.000V 40.00V 400.0V ± (1.5% + 3 digits) DC Voltage 600V ± (2.0% + 3 digits) 400.0mV ± (1% + 10 digits) 4.000V 40.00V 400.0V ± (2% + 5 digits) AC Voltage 600V ± (2% + 5 digits) 400.0Ω ± (1.0% + 4 digits) 4.000kΩ 40.00kΩ 400.0kΩ ± (1.5% + 2 digits) 4.000 MΩ ± (2.5% + 3 digits) Resistance 40.00MΩ ± (3.5% + 5 digits) 40.00nF ± (5% + 30 digits) 400.0nF ± (3% + 5 digits) 4.000µF 40.00µF ± (3.5% + 5 digits) Capacitance 100.0µF ± (5% + 5 digits) 5.000Hz ± (1.5% + 5 digits) 50.00Hz 500.0Hz 5.000KHz 50.00KHz Frequency 150.0KHz ± (1.2% + 2 digits) Sensitivity: 5~5KHz: 10Vrms min. 5KHz~150KHz: 40Vrms min. 0.5% to 99.0% ± (1.2% + 2 digits) Duty Cycle Pulse Width: 100µs-100ms, Frequency: 5Hz to 150KHz -20.0 to 400.0°C 400 to 1000°C 400 to 1000°C ± (3% + 5 °C) -58.0 to 0°F ± 14°F -58.0 to 400.0°F 0 to 400.0°F ± (2.5% + 6 digits) Temperature 400 to 1832°F 400 to 1832°F ± (3% + 7°F)

MA220 V1.3 10/06 4 Jaw size 0.9" (23mm) approx. Display 4000 count LCD Continuity Audible tone < 150Ω approx. Diode Test Open circuit voltage < 1.5VDC; Test current 0.3mA (typical) AC V bandwidth 50Hz to 400Hz AC A bandwidth 50/60Hz Low battery indication “ ” is displayed Overrange indication “OL” is displayed Auto Power OFF After 30 minutes Measurement rate 2 per second, nominal Input Impedance 7.8MΩ (V DC and V AC) Operating Temperature 41ºF to 104ºF (5ºC to 40ºC) Storage Temperature -4 o F to 140 o F (-20 o C to 60 o Operating Humidity Max 80% up to 87ºF (31ºC) decreasing linearly to 50% at 113ºF(45ºC) Storage Humidity <80% Operating Altitude 6560ft. (2000meters) operating Batteries (2) 1.5V AAA batteries Weight 0.44lb (200g) Size 7.87” x 1.97” x 1.38” (200x50x35mm) Safety For indoor use and in accordance with the requirements for double insulation to IEC1010-1 (1995): EN61010-1 (1995) Overvoltage Category III, Pollution Degree 2.

MA220 V1.3 10/06 5 Meter Description 1. Conductor jaws 2. Jaw opening trigger 3. Function select switch 4. LCD Display 5. ZERO button 6. Data Hold and Backlight Button 7. Mode select button 8. Range select button 9. Hz/%/Duty Cycle button 10. COM input jack 11. V/ Ω/Temp jack 12. Battery cover (rear) AC AC (alternating current) DC DC (direct currrent) Minus sign AUTO AutoRange mode ZERO ZERO mode

  • ))) Audible Continuity HOLD Data Hold mode Low Battery icon Diode test mode m milli V Volts A Amps K kilo M Mega Ω Ohms °F Degrees Fahrenheit °C Degrees Centigrade

MA220 V1.3 10/06 6 Operation Notice: Read and understand all WARNING and CAUTION statements listed in the safety section of this operation manual prior to using this meter. Set the function select switch to the OFF position when the meter is not in use. DC/AC Current Measurements Warning: Disconnect the test leads from the meter before making current clamp measurements. 1. Set the Function switch to the 400ADC, 40ADC, 400AAC or 40AAC range. If the range of the measured is not known, select the higher range first then move to the lower range if necessary. 2. For DC current measurement, press the ZERO key to null the meter display. 3. Press the trigger to open jaw. Fully enclose one conductor to be measured. 4. The clamp meter LCD will display the reading. DC/AC Voltage Measurements 1. Set the rotary function switch to the Volts/Hz/% position. 2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into the positive (V/Ω/Temp) jack 3. Select AC or DC with the MODE button 4. Connect the test leads to the circuit under test 5. Read the voltage on the display. The disp lay will indicate the proper decimal point and value. Resistance Measurements 1. Set the function switch to the Ω •))) CAP position. 2. Insert the black test lead banana plug into the negative (COM) jack Insert the red test lead banana plug into the positive (VΩ Temp) jack. 3. Touch the test probe tips across the circui t or part under test. It is best to disconnect one side of the part under test so the rest of the circuit will not interfere with the resistance reading. 4. Read the resistance on the display. The disp lay will indicate the proper decimal point and value. Continuity Check 1. Set the function switch to the Ω •))) CAP position. 2. Push the mode button to indicate •))) on the display. 3. Insert the black lead banana plug into the negative (COM) jack Insert the red test lead banana plug into the positive (VΩ) jack. 4. Touch the test probe tips to the circuit or wire you wish to check. 5. If the resistance is less than approximately 150 Ω, the audible signal will sound. If the circuit is open, the display will indicate “OL.”.

MA220 V1.3 10/06 7 Diode Test 1. Turn the rotary switch to the Ω •))) CAP position. 2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into the positive (VΩ) jack. 3. Push the mode button to indicate on the display. 4. Touch the test probes to the diode under test. Typically for a normal diode, forward voltage will indicate 0.4V to 0.7V. Reverse voltage will indicate “OL”. Shorted devices will indicate near 0V and an open device will indicate “OL” in both polarities. Capacitance Measurements Warning: To avoid electrical shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords. 1. Set the function switch to the Ω •))) CAP position. 2. Push the mode button to indicate nF on the display. 3. Insert the black lead banana plug into the negative (COM) jack and insert the red test lead banana plug into the positive (VΩTemp) jack. 4. Press the ZERO key to null the meter display. 5. Touch the test probe tips to the capacitor you wish to check. 6. Read the capacitance value on the display. Frequency or % Duty Cycle Measurements 1. Turn the rotary switch to the Volts Hz % position. 2. Insert the black test lead banana plug into the negative (COM) jack and the red test lead banana plug into the positive (VΩ) jack. 3. Select Hz or % with the HZ/% button. 4. Touch the test probe tips to the circuit under test. 5. Read the frequency on the display.

MA220 V1.3 10/06 8 Temperature Measurements 1. Turn the rotary switch to the Temp position. 2. Insert the Temperature Probe into the negative (COM) and the (V ΩTemp) jacks, making sure to observe correct polarity. 3. Select °C or °F with the MODE button. 4. Touch the temperature probe head to the part whose temperature you wish to measure. Keep the probe touching the part under test until the reading stabilizes. 5. Read the temperature on the display. Warning: To avoid electrical shock, be sure the thermocouple has been removed before changing to another measurement function. Auto/Manual Ranging The meter turns on in Autoranging mode. Press the RANGE button to enter manual ranging. Each press of the range button will step to the next range as indicated by the units and decimal point location. Press and hold the RANGE button for two seconds to return to Autoranging mode. Note: Manual ranging does not function in AC Current or Diode and Continuity check functions. In Temperature function, it will change the resolution from 0.1° to 1°. Data Hold To freeze the LCD meter reading, press the HOLD button. While data hold is active, the HOLD display icon appears on the LCD. Press the HOLD button again to return to normal operation. Backlight Press and hold the HOLD button for >2 seconds to turn the backlight on/off. Note: The HOLD feature will activate when the backlight is turned on. Press the HOLD button again to exit the Hold feature. Zero Button Zeros Capacitance and DC Current measurements. Also allows the user to offset the meter by using the displayed value as the zero reference value. Press the ZERO key momentarily to activate and to exit Zero mode.

MA220 V1.3 10/06 9 Support Hotline (781) 890-7440 Tech support: Ext. 200; Email: support@extech.com Repair/Returns: Ext. 210; Email: repair@extech.com Website: www.extech.com Maintenance WARNING: To avoid electrical shock, disconnect the meter from any circuit, remove the test leads from the input terminals and turn OFF the meter before opening the case. Do not operate with open case. Cleaning and Storage Periodically wipe the case with a damp cloth and mild detergent; do not use abrasives or solvents. If the meter is not to be used for periods of longer than 60 days, remove the batteries and store them separately Battery Replacement 1. Remove the two rear Phillips head screws 2. Open the battery compartment 3. Replace the two 1.5V AAA batteries. 4. Re-assemble the meter Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment (a six month limited warranty applies on sensors and cables). If it should become necessary to return the instrument for service during or beyond the wa rranty period, contact the Customer Service Department at ( 781) 890-7440 ext. 210 for authorization. A Return Authorization (RA) number must be issued before any product is returned to Extech. The sender is responsible for shipping charges, freight, insurance and proper packaging to prevent damage in tr ansit. This warranty does not apply to defects resulting from action of the user such as mi suse, improper wiring, operati on outside of specification, improper maintenance or repair, or unauthorized modification. Extech specifically disclaims any implied warranties or merchantability or fitness for a s pecific purpose and will not be liable for any direct, indirect, incidental or consequential damages. Extech's total liab ility is limited to repair or replac ement of the produc t. The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or implied. Repair and Calibration Services Extech offers complete repair and calibration services for all Extech products. For periodic calibration, NIST certification or repair of any Extech product, call customer service for details on services available. Extech recommends that calibration be performed on an annual basis to insure calibration integrity. Copyright © 2004 Extech Instruments Corporation. All rights reserved including the right of reproduction in whole or in part in any form.