2650A_V01 BK | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 89
Technical content
Datasheet sections
- 1.1 PRODUCT OUTLINES --------------------------------------------------------------------------
- 1.2 STANDARD ACCESSORIES -------------------------------------------------------------------
- 1.3 OPTIONAL ACCESSORIES --------------------------------------------------------------------
- 1.4 OVERVIEW OF ALL FOUR MODELS ----------------------------------------------------------
- 2.1 PERFORMANCE --------------------------------------------------------------------------------
- 2.2 EXTERNAL VIEW -------------------------------------------------------------------------------
- 5.1 LIST OF FUNCTION MENU ------------------------------------------------------------------
- 5.2 MENU TREE ----------------------------------------------------------------------------------
- 6.1 STAND ----------------------------------------------------------------------------------------
- 6.2 CONNECTION TO POWER SUPPLY --------------------------------------------------------
- 6.3 BATTERY CHARGE --------------------------------------------------------------------------
- 6.4 INSTALLATION OF BATTERY ---------------------------------------------------------------
- 7.1 SETTING WITH STEP KEYS ( [F1], [F2] ) -------------------------------------------------
- 7.2 SETTING WITH ENCODER ------------------------------------------------------------------
- 7.3 SETTING WITH NUMERIC KEY --------------------------------------------------------------
- 7.4 ACCORDING TO MARKER POSITION -----------------------------------------------------
- 8.1 SWITCHING FREQUENCY BAND -----------------------------------------------------------
- 9.1 SETTING OF REFERENCE LEVEL ---------------------------------------------------------
- 9.2 CHANGE OF UNIT OF AMPLITUDE AXIS --------------------------------------------------
- 9.3 SETTING OF STEP SIZE OF REFERENCE LEVEL ----------------------------------------
- 9.4 ON-OFF SETTING OF OFFSET ------------------------------------------------------------
- 9.5 SETTING OF OFFSET LEVEL ---------------------------------------------------------------
- 9.6 SETTING OF INPUT IMPEDANCE -----------------------------------------------------------
- 9.7 REFERENCE LEVEL SETTING RANGE FOR EACH UNIT -------------------------------
- 9.8 RELATIONSHIP BETWEEN REFERENCE LEVEL AND ATT/AMP (AT DBM) ----------
- 10.1 SETTING THE FUNCTION KEY -------------------------------------------------------------
- 11.1 MANUAL MODE ----------------------------------------------------------------------------
- 11.2 AUTO MODE --------------------------------------------------------------------------------
- 11.3 ALL AUTO MODE --------------------------------------------------------------------------
3.3 GHz/8.5 GHz Spectrum Analyzer USER MANUAL
i Before Using the Product
- Please note the following described on the back panel.
- For safe use: 1) When abnormal sound, abnormal smell or smoke is found, stop using the product and remove the battery and AC adapter. 2) Never handle this product with wet hands to prevent risks of electric shock, fire, or damage. 3) Never use this product under lightning conditions to prevent risks of damage. 4) Never use an y AC adapter other than specified, or it will cause damage. Connect the power cable only to three-terminal outlet for protection from static electricity, otherwise this product or a device under test may be damaged. 5) Never use any battery other than specified, or it may damage this product. Turn off the power and disconnect the AC adapter before removing the battery out of the battery compartment. 6) Be sure to charge the battery as specified in the instructions in this manual. Moreover, an explosion, fire or smoke may occur if battery is handled improperly. Please read the notes for handling the battery.
- Set clock function The time information is set at Japan standard time. Set the year, month, day and time when the 2650A series is used for the first time. (Refer to “24.4 Setting the clock” for the details.)
- Quality assurance Warranty If the defect by our responsibility occurs within one year after delivered, it shall be repaired free of charge. However, this warranty does not cover such defect that : 1) is caused by a fire or natural disasters. 2) is caused by inappropriate handling such as dropping while moving the unit delivered. 3) is caused by handling in contradiction to usage or precautions described in the operating manual. 4) is caused by modification or misuse. We will not be responsible for direct or indirect damage caused by use or defect of this product. Refer to the end of this manual for details. Warm-up time Warm up the product for at least ten minutes after turning on the power in order to stabilize the internal circuit. Precautions for storage 1) Avoid direct sunlight or dust. 2) Store this unit in a place where temperature is between -20 °C to 60 °C, humidity less than 60 °C/70 %RH and variations in temperature and humidity are small. After service Please contact us without hesitation if you have any questions about this product: B&K PRECISION
22820 Savi Ranch Parkway
Yorba Linda CA, 92887 TEL. (714) 921 - 9095 FAX. (714) 921 - 6422 URL: http://www.bkprecision.com
v
- Outline
1.1 Product outlines
The 2650A series are authentic spectrum analyzers in a compact and lightweight form factor providing performance and functions comparable to large-size bench type instruments. The features are as follows. 1) Compact and lightweight 1.8 kg. The dimensions are 162(W)×71(H)×265(D) mm, and the weight is only 1.8 kg including the battery, making the instrument very convenient for field use. 2) Large color TFT display 5.7 inches, 640×480 color LCD 3) Four hours battery operation The included Lithium-ion battery, when fully charged, can provide power for up to four hours of operation with the backlight turned off. And about 3 ½ hours with the backlight set to the minimum level. 4) USB host adapter USB host port can be used with removable storage USB flash drives. The screen image is stored as BMP format, and the spectrum and the setting parameters are stored as CSV format. The port conforms to USB 1.1 standards and is backwards compatible with USB 2.0 devices. 5) USB device for PC connectivity USB interface with transfer rates up to 12 Mbps maximum (conform to USB 1.1 standards). 6) Accurate frequency measurement by PLL synthesizer The center frequency is accurately set by PLL (Phase Locked Loop) synthesizer. Moreover, the frequency counter (factory option) enables more accurate measurements of the frequency of a signal. 7) Average noise level -127 dBm The low average noise level of -127 dBm @ 1 GHz provides a wide dynamic range. 8) 100 dB display dynamic range Wide dynamic range display with a display scale of 100 dB/10div (at 10 dB/div) in the amplitude axis. 9) Easy operation with AUTO mode
- Auto range operation: RBW, VBW and sweep time are automatically selected based on the frequency span.
- Auto tuning operation: The center frequency is adjusted to the maximum level within full span, and the optimum RBW, VBW and sweep time are chosen.
10) Abundant functions
- Measuring functions : Channel power, Adjacent channel power, Occupied bandwidth, Electric field strength, Magnetic field strength and Frequency measurement.
- Calculation functions : Max hold, Min hold, Averaging, Over write
- Marker measurement and peak search function
- Save/Load function
- Hard copy with printer 11) PC Software The included PC software controls all four models of the 2650A/2651 series via PC. The software emulates the front panel keys and supports transfer of 1001 point trace data from the analyzer to the PC. Screen images can be stored in BMP format and the spectrum data in CSV format (stores frequency and level). 12) Options There are many options available, such as dipole antennas, magnetic field probes and USB printer
1.2 Standard accessories
- AC adaptor MA400 2. Carrying case LC2650A 3. Accessory pouch 4. User manual 5. PC software and USB cable 6. Lithium Ion battery MB400
1.3 Optional accessories
- Dipole antennas M401, M402, M403, M404, M405, M406 2. Magnetic field probe PR 26M with a dedicated double shielded coaxial cable 3. Printer PT2650A with AC adaptor, 4pcs of AA batteries, one roll paper 4. Roll paper PX2650A for PT2650A printer (with 10 rolls)
1.4 Overview of all four models
50 kHz to 3.3 GHz Applications: Cellular phone, 2.4 GHz wireless LAN, 2.5 GHz WiMAX, RF-ID, Broadcasting 2658A 50 kHz to 8.5 GHz Covering most of the wireless communication frequency spectrum Applications: 5 GHz wireless LAN, 3.5/5.8 GHz WiMAX, Maintenance of wireless base station 2652A 50 kHz to 3.3 GHz With 5 MHz to 3.3 GHz tracking generator Applications: Frequency characteristics measurement of electronic component/circuit and return loss measurement 2651 50 kHz to 3.3 GHz For EMI test Applications: Radiated and conducted emission measurement
- Specifications
2.1 Performance
Frequency section 2650A, 2652A, 2651 2658A Frequency range 50 kHz to 3.3 GHz 50 kHz to 8.5 GHz ――― Frequency range Frequency band Harmonic order 50 k to 3.5 GHz Base band 1 3.3 G to 6.3 GHz Band 1- 1 6.1 G to 8.5 GHz Band 1+ 1 Center frequency Setting resolution 20 kHz Allows rotary encoder, numeric key and function key Accuracy within ± (30+20T) kHz±1 dot @frequency span: 200 kHz to 10 MHz, RBW 3 kHz, 23±5°C within ± (60+300T) kHz±1dot @frequency span: 20 MHz to 3.3 GHz, RBW 100 kHz、23±5°C T: sweep time (s) within ±(30+20T) kHz ± 1 dot @frequency span: 200 kHz to 10 MHz, RBW 3 kHz, 23±5°C within ± (60+300T) kHz±1dot @frequency span: 20 MHz to 8.5 GHz, RBW 100 kHz, 23±5°C T: sweep time (s) RBW frequency error within ±4 kHz @ 3 kHz, 10 kHz, 30 kHz within 20% of RBW @ RBW: 100 kHz, 300 kHz within 10% of RBW @ RBW: 1 MHz, 3 MHz Frequency span Setting range 0 Hz (zero span), 200 kHz to 2 GHz (1-2-5 step) and
3.3 GHz (full span)
0 Hz (zero span),
200 kHz to 5 GHz (1-2-5 step) and
8.5 GHz (full span)
Accuracy within ±3% ±1dot @ sweep time of 0.3s, 23±5 °C Display resolution # of points per trace 501dots on LCD screen, 1001dots readout via PC (501 dots are visible on the display, 1001 dots of trace data are captured internally and can be transferred to a PC via USB device interface.) Resolution bandwidth 3 dB bandwidth (6 dB for 2651 @ 9 kHz, 120 kHz) Setting range 3 kHz to 3 MHz (1-3 step) and AUTO (2651: 3 kHz, 9 kHz, 30 kHz, 120 kHz, 300 kHz, 1 MHz, 3 MHz) Accuracy ±20% Selectivity 1:12 (typical, 3 dB : 60 dB) Video bandwidth 100 Hz to 1 MHz (1-3 step) and AUTO SSB phase noise -90 dBc/Hz (typical) @100 kHz offset, RBW : 3 kHz, VBW : 100 Hz, sweep time: 1 s Spurious response less than -60 dBc Harmonics less than -40 dBc @ ≧100 MHz T:掃引時間
Amplitude section 2650A, 2652A, 2651 2658A Reference level Setting range +10 to -60 dBm (1dB step) Accuracy within ± 0.8 dB ± 1 dot @ center frequency : 100 MHz, RBW : 3 MHz, VBW : 1 MHz, REF : -15 dBm, 23±5 °C Unit dBm, dBV, dBmV, dBμV, dBμV/m, dBμA/m (dBμV/m and dBμA/m are used for measurement functions) Average noise level -127 dBm (typical) @ CF : 1 GHz, RBW : 3 kHz, VBW : 100 Hz, Ref. level < - 40dBm ( preamp automatically ON) Frequency characteristics within ±2.0 dB ±1dot @50 kHz to 100 MHz within ±1.0 dB ±1dot @ 100 MHz to 3.3 GHz within ±2.0 dB ±1dot @50 kHz to 100 MHz within ±1.0 dB ±1dot @100 MHz to 8.5 GHz Input impedance 50 Ω Input VSWR less than 2.0 Input attenuator Operating range 0 to 25 dB (1 dB step), coupled with reference level Switching error ±0.6 dB @100 MHz RBW switching error ±0.6 dB Display resolution (vertical) 381 dots/10div Display scale Scale 10 dB/div, 5 dB/div, 2 dB/div Accuracy ±(0.2 dB+1dot)/2 dB ±(0.4 dB+1dot)/5 dB ±(0.8 dB+1dot)/10 dB ±(1.8 dB+1dot)/83 dB Input damage level +27 dBm (CW average power), 25 VDC Sweep section 2650A, 2652A, 2651 2658A Sweep time Setting range 10 ms to 30 s and AUTO @frequency span : 0 to 2 GHz 30 ms to 30 s and AUTO @frequency span : full span 1-3 step 10 ms to 30 s and AUTO @frequency span : 0 to 2 GHz 30 ms to 30 s and AUTO @frequency span : 5 GHz, full span 1-3 step Accuracy Within ±0.1 %±1dot @frequency span : 0 to 2GHz Within ±0.1 %±1dot @frequency span : 0 to 5GHz within ±1.5 %±1dot @ full span within ±2.5 %±1dot @ full span Trigger Trigger mode AUTO (Available only for zero span) Trigger source Internal and External
approx. 0.56 V(fix) approx. 10 kΩ / less than 15 pF ±50 V(DC+AC peak) SMA(J) Detection mode Positive peak, Negative peak, Sample ( QP and AV for model 2651 only.) Function 2650A, 2652A, 2658A, 2651 common Marker measurement NORM:displays frequency (8digits max) and level (4digits max) at marker point. DELTA:displays frequency difference and level difference between two markers. Peak search function Searches for peak level with all of 10 div (NORM mode) or within ZONE specified (ZONE mode) and displays frequency and level at peak level, and moreover NEXT peak at NORM mode. Calculation function NORM, MAX HOLD, MIN HOLD, AVERAGE, OVER WRITE Number of sweeps is 2 to 1024 (power of 2 ) and infinite. Measuring function Channel power, Adjacent channel power, Occupied bandwidth, Electric field strength (needs optional dipole antenna), Magnetic field strength (needs optional magnetic field probe) and Frequency counter Auto tuning When pressing AUTO TUNE key, the spectrum of maximum level within full span is adjusted to center, and reference level, RBW, VBW and sweep time are set to optimum parameters. Save/ Load Save Saves 200 spectrums and 200 setting parameters Load Loads one spectrum and one setting parameter General 2650A, 2652A, 2658A, 2651 common Input connector N(J) connector USB Communication Protocol USB version 1.1 (will also work with USB 2.0 devices) Device Connector B plug Transfer rate 12 Mbps Hard copy USB printer (option) connected to A plug (host) enables hard copy of screen. host connector A plug Display Display 5.7 inches and color LCD Backlight LED backlight Number of dots 640(H) x 480(V) dots Power supply Dedicated AC adaptor Input : 100 to 240 VAC Output : 9 VDC/2.6 A Lithium-ion battery Charge function Remainder indication MB400 : 7.4 V/5000 mAh Li-Ion battery Capable of charging during power-off. Indicates 4 conditions with two colors LED (red and green). 5 level indicator
Other 2650A, 2652A, 2658A, 2651 common Operating temperature 0 to 50° C (guaranteed at 23±10 °C, without soft carrying case) Operating humidity less than 40 °C/80 %RH (guaranteed at less than 33 °C/70 %RH, without soft carrying case) Storage temperature -20 to 60 °C, less than 60 °C/70 %RH Dimensions 162(W) x 71(H) x 265(D) mm , 6.38 (W) x 2.80 (H) x 10.43 (D) inch (excluding projections, protection bumper and stand) Weight approx. 1.8 kg or 4 lbs (including battery) * Refer to section 22) Tracking Generator Mode for the specifications of the tracking generator, model 2652A.
2.2 External view
- B&K Precision reserves the right to make changes in design, specification and other information without prior notice. [Unit : mm]
- Explanation of Panel 1) TFT display This is a large liquid crystal display with 640 (H) x 480 (V) dots. It simultaneously displays spectrum (10div x 10 div), various setting parameters and measured values. 2) Function key (F1 to F6) The function can change according to the operation key. 3) Center frequency key
The center frequency is set with this key. The setting range is 0 to 3.3 GHz (for 2650A, 2652A, and 2651), and 0 to 8.5 GHz (for 2658A). The setting resolution is 20 kHz. 4) Frequency span key The frequency span is set with this key. For 2650A, 2652A, and 2651, it is set in the range from 200 kHz to 2 GHz, ZERO SPAN or FULL SPAN (3.3 GHz). For 2658A, it is set in the range from 200 kHz to 5 GHz, ZERO SPAN or FULL SPAN (8.5 GHz). 5) Reference level key The reference level is set with this key. It can be set in the range from +10 dBm to –60 dBm by 1 dB step. 6) Resolution bandwidth key The resolution bandwidth is set with this key. It can be set in the range from 3 kHz to 3 MHz and to AUTO. 7) Video bandwidth key The video bandwidth is set with this key. It can be set in the range from 100 Hz to 1 MHz and to AUTO. 8) AUTO tuning key When pushing this key, the spectrum with the maximum level is searched within full span (3.3 GHz @ 2650A/2652A/2651 and 8.5 GHz @ 2658A), and then it is adjusted to the center of the screen, and the optimum setting parameters are set. In the case when in zero span, full span, input signal level lower than –40 dBm and input frequency lower than 50 MHz, this function will not work correctly. 9) Measuring function key Available for Channel power, Adjacent channel leakage power, Occupied frequency bandwidth, Electric field strength, and Magnetic field strength measurement. 10) Calculation function key Available for Max hold, Min hold, Average and Over write. 11) Display scale key The display scale of amplitude axis can be selected as 2 dB/div, 5 dB/div or 10 dB/div with this key. 12) Sweep key The sweep time can be set in the range from 10 ms to 30 s or AUTO. This key can also select the detection mode. 13) Hold/Run key Used to hold or run measurements.
14) Marker & Peak search key This key is used for setting marker or peak search. 15) Save/Load key Used for saving or loading a spectrum or settings parameters. 16) Copy key This key allows the screen image to be printed on USB printer (option) or stored in USB memory. 17) UTIL key Used for setting label entry, clock, and buzzer. 18) Display control key Used for setting color, backlight ON/OFF or brightness of backlight. 19) Rotary encoder This is used for adjusting settings and parameters. 20) Power switch This is for power ON or OFF. 21) Input connector N(J) connector 22) Indicator for charging condition Two color LED indicates the charging conditions of battery. 23) Input connector for DC power source Connect AC adaptor MA400. 24) USB A plug For connecting USB printer (option) or USB memory. 25) USB B plug For interfacing with a PC
- Explanation of Screen Function Menu UNCAL **UNCAL Display Remaining Battery Display Spectrum Display 10div x 10 div Active Area Display Center frequency, Frequency span, Resolution bandwidth, Video bandwidth Reference level, Input attenuator, Display scale Sweep time, Detection mode Calculation function display Measuring function display Operating Information display Loaded spectrum information display * This is also used as a display for set values when the measuring function is used. **UNCAL is displayed when a normal measurement cannot be done due to sweep rate being too fast. In such cases, slowing down the sweep rate will change this indicator. Measured Values display Sweep Axis Setting Values Display Amplitude Axis Setting Values Display Frequency Axis Setting Values Display Upper: Label Display Lower: Clock Display
- Function Menu
5.1 List of Function Menu
The function menu is shown in the table below. For descriptions of each function, see their respective page. For the sequence of selection for the function menu, refer to “5.2 Menu tree”. 1 2651 only 2 2658A only 3 2652A only Function menu Key Sequence Page A ACP OFFSET MEAS → (F2)→ F2 42 ACP MEAS → F2 41 ACP WIDTH MEAS → (F2) → F3 42 ANT MEAS → (F4)→ F1 44 AVG CALC → F4 32 B BACK LT DSPL → F2 53 BACK SPACE FREQ → F6 → F5 21 BAND CNTR MEAS → (F1) → (F1) → F2 BAND WIDTH MEAS → (F1) → (F1) → F3 BRIGHT DSPL → F3 53 BUZZER UTIL → F3 59 C CENTER FREQ ← FREQ → F1 20 CENTER FREQ → FREQ → F2 20 CH POWER MEAS → F1 41 CLEAR FREQ → F6 → F4 21 CLOCK CONFIG UTIL → F4 59 COLOR DSPL → F1 53 CONV MKR → F6 33 D DELETE SAVE/LOAD → F3 36 DEVICE MEM SAVE/LOAD → F4 36 DET SWEEP → F4 29 DISP CLEAR SAVE/LOAD → F2 → E E/F ANT MEAS → F4 43 EncST FREQ → F4 20 EMI-C 1 SAVE/LOAD → F6 → EMI-R 1 SAVE/LOAD → F6 → F FREQ COUNT MEAS → F6 49 I IMP REFER → F6 25 K KeyST FREQ → F3 20 L LABEL UTIL → F1 58 LOAD SAVE/LOAD → F2 38 M M/F PROBE MEAS → F5 47 MAX HLD CALC → F2 31 MEAS OFF MEAS → (F1~5) → F6 41 MIN HLD CALC → F3 31 MARKER DELTA MKR → F2 33 MARKER NORMAL MKR → F1 33 Function menu Key Sequence Page M MODE MEAS → (F1 → F3) → F1 41 to 43 N NEXT PEAK MKR → (F4) → F2 34 NORMAL CALC → F1 31 NORMAL 1 SAVE/LOAD → F6 → O OBW MEAS → F3 42 OFSdB REFER → F5 24 OFS STEP REFER → F3 24 OVR WR CALC → F5 32 P PEAK SEACH MKR → (F5)→ F1 34 PEAK SERCH NORM MKR → F4 34 PEAK SERCH ZONE MKR → F5 34 PRE SET SAVE/LOAD → F6 36 PROBE MEAS → (F5) → F1 48 R RATIO MEAS → (F3) → F2 43 RBW ALL AUTO RBW → F3 28 RBW AUTO RBW → F2 28 RBW MANUAL RBW → F1 27 REFERENCE CENTER MEAS → (F2) → F4 42 REFERENCE WIDTH MEAS → (F2) → F1→ S SAVE SAVE/LOAD → F1 36 SCALE 5dB SCALE → F3 27 SET MKR FREQ → F5 20 SPR. FR 2 CALC → F6 32 SWEEP AUTO SWEEP → F2 29 SWEEP MANUAL SWEEP → F1 29 T T.G. MODE 3 SWEEP → F6 55 TRIG SWEEP → F5 30 Ten Key MODE FREQ → F6 20 U UNIT REFER → F1 → 4 24 V VBW ALL AUTO VBW → F3 29 VBW AUTO VBW → F2 28 VBW MANUAL VBW → F1 28 Z ZONE CENTER MKR → (F5) → F1 35 ZONE WIDTH MKR → (F5) → F2 35
5.2 Menu Tree
The following is the sequence of selection for the function menu. The function menu corresponds to the function key of F1 to F6. “Function menu” * Refer to “8. Frequency Span” for details. NORMAL SPAN FULL SPAN ZERO SPAN SPAN
50 MHz 1 GHz 1 MHz
: Set the frequency span * Refer to “9. Reference Level” for details. UNIT dBm RL STEP 1dB OFS STEP 1dB REFER OFFSET OFF OFS dB 0dB IMP 50Ω : Set the reference level ANT E/F M/F PROBE FREQ COUNT CH POWER ACP OBW F2 F3 F4 F5 F6 : Set the center frequency CENTER FREQ KeyStep
100 MHz
10 MHz
GHz MHz CLEAR “.” BACK SPACE RETURN * Refer to “7. Center Frequency” for details. FREQ
- T.G. MODE is only for 2658A. * Refer to “13. Sweep Axis / Detection Mode” for details. SWEEP TIME AUTO MANUAL ALL AUTO SWEEP : Set the sweep time T.G. ON NORM ON DET SMPL TRG INT T.G. MODE * Refer to “12. Video Bandwidth” for details. VBW AUTO MANUAL ALL AUTO VBW : Set the VBW * Refer to “11. Resolution Bandwidth” for details. RBW AUTO MANUAL ALL AUTO RBW : Set the RBW * Refer to “10. Display scale” for details. SCALE 5dB/ 10dB/ 2dB/ SCALE : Set the display scale
- “SPR. FR” is only for 2658A. * Refer to “16. Calculation Function” for details. NORMAL MAX HLD 256 MIN HLD 256 CALC AVG 256 OVR WR 256 SPR. FR OFF : Set the number * Refer to “17. Marker & Peak Search” for details. MKR CONV dBm → W : Move the marker (NORM mode) PEAK SEARCH NEXT PEAK RETURN RETURN MARKER DELTA NORMAL OFF PEAK SEARCH NORMAL ZONE ZONE CENTER WIDTH : Move the marker (NORM mode) : Set the zone center frequency (ZONE mode) F4: Peak Search F5: Zone marker
- Refer to “18. Save / Load” for details. SAVE/LOAD PRE SET RETURN RETURN MODE SELECT LOAD SAVE DELETE DEVICE MEM USB : Set the address to store the spectrum or the parameter : Set the address to call the spectrum or the parameter OBJECT PARAM SPECT BOTH SAVE EXECUTE Save menu Load menu File Search Step 10 1 100 LOAD EXECUTE DISP CLEAR RETURN Delete menu File Search Step 10 1 100 DELETE EXECUTE : Set the address to eliminate the trace or parameter For 2651 Main menu NORMAL EMI - C PRESET EMI - R Preset menu for 2651
- Refer to “19. Measuring Function” for details MEAS MEAS OFF MEAS OFF MEAS OFF CH POWER ACP OBW E/F ANT M/F PROBE Freq COUNT MODE BAND BAND CENTER WIDTH ADJ CH OFFSET WIDTH REFERENCE CENTER WIDTH MODE BAND MEAS OFF MODE RATIO % 99.5% Enc Step ANT M401 PROBE CP-2S MEAS OFF Input the probe ID (first time only) *Refer to “19.5 Magnetic field strength measurement” for details. : Set the parameter : Set the parameter : Set the parameter
- Refer to “24. Utility Function” for details UTIL LABEL MENU OFF BUZZER ALARM CLOCK CONFIG 012 abc ABC @%* YEAR MONTH JUN DATE : Select the character : Select the numeric value BACK SP ENTER HOUR MIN ENTER * Refer to “23. Storage/Print Screen Image” for details COPY AREA WHOLE PRINT EXECUTE Bmp->USB COPY MEM->USB EXECUTE * Refer to “21. Screen Control” for details DSPL COLOR COLOR1 BACK LT ON BRIGHT
- Preparing for Operation
6.1 Stand
Utilize the stand on the back to use the screen in an easier-to-see angle on the desk.
6.2 Connection to power supply
The AC adapter MA400 is used to power the instrument and for charging the MB400 battery. (Charging starts automatically if the AC adapter is connected and unit is power-off.) Connect the adapter as in the figure below and connect the AC plug to the power line (100-240 VAC, 50/60 Hz). For static electricity protection, ground the unit by connecting the three cores if possible. Not grounding the unit can damage the unit and/or device being measured. Do not use an AC adapter other than the MA400 supplied with the unit. Using an AC adapter other than the MA400 may cause damage to the unit. The battery indicator is divided into five levels and is displayed on the screen. When the mark of the battery remainder displays , the buzzer will sound even if buzzer is OFF under settings, and the power will turn off within a few minutes. Ensure to protect any data or measurements promptly by saving into memory to prevent data lost if the unit shuts down. Battery full charge time: approx. 4 hours Battery operating time: 4 hours maximum (with backlight set to minimum) *Tested at normal temperature with initial setting parameters. Connect to 100 to 240 VAC, 50/60 Hz Please make sure to connect to ground.
6.3 Battery Charge
When the unit is powered off and connected to AC adaptor MA400, the battery will be charged. The charging conditions are indicated by two colors LED , and they correspond to the table below. Charging condition Color of LED Good usage of battery On Charge Red Completion of Charge Green Remove the battery when unit is connected to the AC adaptor. Otherwise, battery life will be shortened. No battery Green Abnormal Blinking Red ※LED is turned off at power-on. The “abnormal” condition means that the charging time is more than the time expected, or the battery voltage is too high. Note: The battery charges only when the unit is turned off (and the AC adapter connected). It is not possible to charge the batteries while the unit is turned on. Please charge batteries with ambient temperature in the range of 0 ºC to 40 ºC.
6.4 Installation of Battery
To install the battery, please turn off the power and unplug the AC adaptor if in use. Remove the battery cover on the back of the unit by first pressing down with light pressure, then slide towards the direction of the arrow as locks in. Use only the MB400 battery as specified. Otherwise, the unit may cause problems or malfunction. Fig. 1 Fig. 2 Fig. 3 Caution Misuse of the battery may cause leaks, abnormal heat, fire or explosion. For safe use, please observe the following precautions fully. * Do not short-circuit the terminal of the removed battery. * Do not cause an impacted shock due to throwing, dropping or striking. * Do not disassemble or remodel the battery. * Do not throw the battery into a fire or any sources with heat. * Do not leave the battery in a place with high temperature. * Do not wet the terminal of the battery. * Do not cool the battery, and do not charge it in cold outdoor conditions. It decreases performance and battery life. * Do not charge the battery in any other way except according to specifications. * When you keep the battery, please keep it in a cool, dark environment where humidity is low. Moreover, please keep it out of reach from children. If the battery remains not used for a long time, it might decrease its charges. * Please exchange the battery for a new one (MB400) when operating time extremely shortens.
- Center Frequency <FREQ> When is pushed, the following function menu is displayed.
7.1 Setting with step keys ( [F1], [F2] )
- When is pushed, the center frequency decreases in the set step size. 2. When is pushed, the center frequency increases in the set step size. 3. Setting step size: When is pushed, the step size is changed according to the following. AUTO 100MHz 10MHz 1MHz 100kHz 20kHz AUTO : The step size is 1/10 of the frequency span. However, only for 500 kHz frequency span, the step size is set to 50 kHz. If the frequency span is less than 500 kHz, the step size is set to 20 kHz. AUTO is recommended for normal use.
7.2 Setting with encoder
- By rotating , the center frequency is changed in the set step size. 2. Setting step size: When is pushed, the step size is changed according to the following. AUTO 100MHz 10MHz 1MHz 100kHz 20kHz AUTO : The step size is 1/500 of the frequency span However, if the frequency span is less than 5 MHz, the step size is set to 20 kHz. AUTO is recommended for normal use. CENTER FREQ KeyStep 100M EncStep 0.1M SET MKR Ten Key Mode ← → F1 F2 F3 F4 F5 F6 FREQ * Center frequency setting range: 0 to 3.3 GHz for 2650A/2652A/2651, 0 to 8.5 GHz for 2658A * The center frequency may shift for a while (up to 10 sec.), after setting is changed.
7.3 Setting with numeric key
- When is pushed, the following function menu is displayed. 2. , and correspond to <GHz>, <MHz> and <. > and correspond to <CLEAR> and <BACK SPACE>. Changing the center frequency by encoder is available in Ten Key Mode. 3. The center frequency can be entered directly according to “Numeric Key Mapping Diagram” as follows. “Numeric Key Mapping Diagram” 4. The center frequency is determined by inputting the unit of frequency. To enter the unit, and are available. ( and are also available.) * Any figures below the setting resolution (20 kHz) will be truncated. 5. Change of setting The setting can be changed before pushing the unit key. : The setting values are cleared. : The last input digit is deleted. “.” CLEAR BACK SPACE RETURN GHz MHz F1 F2 F3 F4 F5 F6 HOLD RUN SCALE RBW VBW SWEEP AUTO TUNE MKR MEAS CALC SAVE LOAD COPY UTIL DSPL 1 2 3 4 5 6 7 8 9 0 . MHz GHz F1 F2 F3 F4 F5 UTIL DSPL F1 F2
- Cancel of Ten Key Mode By pushing or , unit will return to function menu.
7.4 According to Marker Position
When is pushed, the center frequency is set according to the frequency of current marker position. * Any figures below the setting resolution (20 kHz) will be truncated. * When the marker is not displayed, this operation is invalid. (The menu option disappears.) 8. Frequency Span <SPAN> When is pushed, the following function menu is displayed. By operating , the frequency span is set. 2650A / 2652A / 2651 By rotating , the frequency span is changed in the specified step size as follows. ZERO 200k 500k 1M 2M 5M 10M 20M 50M 100M 200M 500M 1G 2G FULL (3.3 GHz) 2658A 1. By rotating , the frequency span is changed in the specified step size as follows. ZERO 200k 500k 1M 2M 5M 10M 20M 50M 100M 200M 500M 1G 2G 5G FULL (8.5 GHz) 2. By pushing , the frequency span is set to FULL SPAN. Under this condition, the instrument will return to the former span when the rotary knob is rotated. 3. By pushing , the frequency span is set to ZERO SPAN. NORMAL SPAN FULL SPAN ZERO SPAN SPAN 1GHz 50MHz 1MHz F1 F2 F3 F4 F5 F6 FREQ F6 SPAN SPAN
- When is pushed and then the setting is FULL or ZERO span, the frequency span is returned to the last setting. 5. By pushing ~ , the frequency span in function menu is set.
8.1 Switching Frequency Band
2658A has three frequency bands. The frequency band is automatically set to the appropriate band based on the center frequency and span. (At the span less than 200 MHz, only one band is used.) When the setting frequency range belongs to two bands, the lower band has priority. The frequency connection point of two bands is fixed as follows. Two bands Frequency connection point Base band and Band 1- 3.4 GHz Band 1- and Band 1+ 6.2 GHz Note: The spectrum may change a little at the frequency connection point. For accurate measurement, center frequency and span should be set as the measured frequency range in one band. Frequency band Measured frequency range Base band 50 kHz to 3.5 GHz Band 1- 3.3 GHz to 6.3 GHz Band 1+ 6.1 GHz to 8.5 GHz F4 F6
- Reference Level <REFER> When is pushed, the following function menu is displayed.
9.1 Setting of Reference Level
By rotating , the reference level is changed. (Refer to “9.7 Reference Level Setting Range for Each Unit” for details.)
9.2 Change of Unit of Amplitude Axis
When is pushed, the unit is set as follows. dBm dBμV dBmV dBV
9.3 Setting of Step Size of Reference Level
By pushing , the step size is changed to 10 dB or 1 dB.
9.4 On-Off Setting of Offset
By pushing , the offer set ON/OFF is changed.
9.5 Setting of Offset Level
- By operating , the offset of reference level is set. When external amplifier or attenuator is used, the display level can be matched by the offset. The setting range is -50.0 to 50.0 dB. The reference level is displayed including the offset. * If the offset is set, “OFST” is displayed in Amplitude axis setting values display area. Furthermore, the level at the marker point is displayed including the offset. OFS STEP 1dB OFFSET OFF OFSdB 0.0 IMP 50Ω F1 F2 F3 F4 F5 F6 UNIT dBm RL STEP 1dB REFER
- If the unit is changed to dBμV, dBmV, dBV or dBm, the offset is automatically changed. 2. By pushing , the step size of offset is changed. (10 dB, 1 dB, 0.1 dB)
9.6 Setting of Input Impedance
By pushing , the input compensation can be set to 50 Ω (no offset) or 75 Ω. (5.6 dB offset compensation) The reference level is adjusted automatically, based on the input impedance selection. When the input impedance is set to 75 Ω, the reference level is displayed including offset and conversion of 75Ω. For the amplitude reading to be correct, a coaxial 50 Ω /75 Ω impedance conversion adapter must be connected to the RF input. * When “75 Ω” is selected, values in Amplitude axis setting values display area will change according to 75 Ω, and the offset is set to 5.7 dB (insertion loss of 50 Ω /75 Ω adapter ). Moreover, the offset can be changed. When the unit at marker point is set to W, V, V/m or other, it is converted correctly from dBm. * Be sure to attach a coaxial 50 Ω /75 Ω impedance conversion adapter when selecting “75 Ω”
9.7 Reference Level Setting Range for Each Unit
Unit dBm dBμV dBmV dBV Maximum 10 117 57 -3 Minimum -40 67 7 -53 Minimum (shifted spectrum data) -60 47 -13 -73 “Available unit in measuring function” Unit dBμV/m (Electric field strength measurement) dBμA/m (Magnetic field strength measurement) Setting M401 M402 M403 M404 M405 M406 CP-2S Maximum 143 146 149 150 137 159 160 ~ 203 Minimum 93 96 99 100 87 109 110 ~ 153 Minimum (shifted spectrum data) 73 76 79 80 67 89 90 ~ 133 * When the reference level is set between “Minimum” and “Minimum (shifted spectrum data)”, the spectrum of “Minimum” is shifted and displayed on the screen.
- When the reference level is set below “Minimum”, “*S/W AMP” is displayed in Amplitude axis setting values display area on the screen. Calculation formula (conversion from dBm)
- A[dBμV] = 107 + X [dBm] ● B[dBmV] = 47 + X [dBm] ● C[dBV] = -13 + X [dBm]
- D[dBμV/m] = 68.8 / λ×√(X/Gar) [dBm] λ : Wavelength[m] Gar : Antenna absolute gain[times]
- E[dBμA/m] = 107 + X + F[dBm] F : Probe calibration coefficient[dB] **changes depending on the frequency.
9.8 Relationship Between Reference Level and ATT/AMP (at dBm)
Internal input attenuator (ATT) and IF amplifier (AMP) are automatically set to the optimum values based on the reference level (REFER). (The input attenuator cannot be set independently.) REFER (dBm) ATT (dB) AMP (dB) REFER (dBm) ATT (dB) AMP (dB) REFER (dBm) ATT (dB) AMP (dB) REFER (dBm) ATT (dB) AMP (dB) 10 25 0 -3 12 0 -16 20 21 -29 7 21 9 24 0 -4 11 0 -17 19 21 -30 6 21 8 23 0 -5 10 0 -18 18 21 -31 5 21 7 22 0 -6 9 0 -19 17 21 -32 4 21 6 21 0 -7 8 0 -20 16 21 -33 3 21
- When the input signal level is higher than the proper level for 1st mixer’s terminal, harmonics distortion and spurious are generated. This product is designed so that the input signal level of 1st mixer is determined to proper level based on the reference level. 5 20 0 -8 7 0 -21 15 21 -34 2 21 4 19 0 -9 6 0 -22 14 21 -35 1 21 3 18 0 -10 5 0 -23 13 21 -36 5 26 2 17 0 -11 4 0 -24 12 21 -37 4 26 1 16 0 -12 3 0 -25 11 21 -38 3 26 0 15 0 -13 2 0 -26 10 21 -39 2 26 -1 14 0 -14 1 0 -27 9 21 -40 1 26 -2 13 0 -15 0 0 -28 8 21
- Display Scale <SCALE> When is pushed, the following function menu is displayed.
10.1 Setting the Function Key
- By pushing , 10 dB/div display scale is set. 2. By pushing , 5 dB/div display scale is set. 3. By pushing , 2 dB/div display scale is set. 11. Resolution Bandwidth <RBW> When is pushed, the following function menu is displayed.
11.1 MANUAL Mode
By pushing or rotating , MANUAL mode is set. By rotating , RBW is set as follows. 2650A/2652A/2658A: 3kHz 10kHz 30kHz 100kHz 300kHz 1MHz 3MHz 2651: 3kHz 9kHz 30kHz 120kHz 300kHz 1MHz 3MHz MANUAL RBW AUTO ALL AUTO F1 F2 F3 F4 F5 F6 10div/ SCALE 5div/ 2div/ F1 F2 F3 F4 F5 F6 SCALE RBW
11.2 AUTO Mode
By pushing , the optimum RBW is set based on the frequency span and sweep time. * When AUTO mode is set, “*” is displayed on the right end of RBW setting value display.
11.3 ALL AUTO Mode
By pushing , the optimum RBW, VBW and sweep time are set based on the frequency span. * When ALL AUTO mode is set, “*” is displayed on the right end of each setting value display. * When RBW is set to 3 kHz or 10 kHz, the selectivity at 60 dB becomes larger than an actual value due to SSB phase noise. 12. Video Bandwidth <VBW> When is pushed, the following function menu is displayed.
12.1 MANUAL Mode
By pushing or rotating , MANUAL mode is set. By rotating , VBW is set as follows. 100Hz 300Hz 1kHz 3kHz 10kHz 30kHz 100kHz 300kHz 1MHz
12.2 AUTO Mode
By pushing , the optimum VBW is set based on the frequency span and sweep time. When AUTO mode is set, “*” is displayed on the right end of VBW setting value display. MANUAL VBW AUTO ALL AUTO F1 F2 F3 F4 F5 F6 VBW
12.3 ALL AUTO Mode
By pushing , the optimum RBW, VBW and sweep time are set based on the frequency span. * When ALL AUTO mode is set, “*” is displayed on the right end of each setting value display. 13. Sweep Axis / Detection Mode <SWEEP> When is pushed, the following function menu is displayed. * T.G. MODE [F6] is only for 2652A. This menu is not displayed on 2650A, 2651 and 2658A. For the details, refer to “22. Tracking Generator Mode”. [F5] is only displayed when an external trigger signal is detected at the trigger port.
13.1 MANUAL Mode
By pushing or rotating , MANUAL mode is set. By rotating , sweep time is set as follows. 10ms 30ms 0.1s 0.3s 1s 3s 10s 30s * For 2650A, 2651 and 2652A, when the setting is FULLSPAN, it cannot be set to 10 ms. * For 2658A, when the setting is 5 GHz SPAN or FULLSPAN, it cannot be set to 10 ms.
13.2 AUTO Mode
By pushing , the optimum sweep time is set based on the frequency span and RBW. * When AUTO mode is set, “*” is displayed on the right end of SWEEP setting value display.
13.3 ALL AUTO Mode
By pushing , the optimum RBW, VBW and sweep time are set based on the frequency span. * When ALL AUTO mode is set, “*” is displayed on the right end of each setting value display.
13.4 Setting of Detection Mode
(For 2651, refer to “20. EMI test”) F1 F2 F3 F4 F5 F6 DET SMPL TRIG INT T.G. MODE MANUAL SWEEP AUTO ALL AUTO SWEEP
By pushing , the detection mode is set. POS SMPL NEG POS (Positive Peak): The maximum value of the sample points is detected. SMPL (Sample): The momentary value of the sample points is detected. NEG (Negative Peak): The minimum value of the sample points is detected.
13.5 Setting of Trigger Source
By pushing , the trigger source is set. INT EXT INT: The sweep is automatically repeated. This setting is normally used. EXT: When the signal over the trigger level (0.56 V) is input to external trigger input, the sweep starts. The sweep automatically starts without an input. (Available only for zero span) 14. AUTO Tuning <AUTO TUNE> When is pushed, the spectrum with the maximum level is searched within full span, and then it is adjusted to the center of the screen, and the optimum setting parameters are set. * Set the frequency span before setting AUTO tuning. * There is no function menu displayed for this mode. Auto tuning will execute upon pushing the key. * The auto tuning does not operate normally in the following conditions. 1) Zero span 2) Full span 3) The signal level is –40 dBm or lower. 4) The signal frequency is 50 MHz or lower. 15. Hold / Run <HOLD/RUN> By pushing , the sweep is switched to halt and continuance. * There is no function menu displayed for this mode. AUTO TUNE HOLD / RUN
- Calculation Function <CALC> When is pushed, the following function menu is displayed. * By pushing , the sweep is restarted. * By pushing ~ , the calculation function is selected. By rotating , the number of times of the sweep can be set.
16.1 NORMAL Mode
- When is pushed, the normal mode is set. The calculation is not performed in this mode. The number of times of the sweep is infinite. This mode is normally used. * When this mode is set, “NORMAL” is displayed in the CALC area on the screen. (Refer to “4. Explanation of Screen” for the details.)
16.2 MAX HOLD Mode
- By the operation of , the number of times of the sweep in the MAX HOLD mode is set. 2. The update spectrum data is compared with the data left last time at each point, and the larger one is retained and displayed. 2 4 8 16 32 64 128 256 512 1024 * * (infinite) * When this mode is set, “MAX --- (number of times)” is displayed in the CALC area on the screen. (Refer to “4. Explanation of Screen” for the details.)
16.3 MIN HOLD Mode
- By the operation of , the number of times of the sweep in the MIN HOLD mode is set. MIN HLD * * AVG OVRWR * * SPRI FR OFF F1 F2 F3 F4 F5 F6 NORMAL MAX HLD * * CALC HOLD / RUN F1 F5
- The update spectrum data is compared with the data left last time at each point, and the smaller one is retained and displayed. 2 4 8 16 32 64 128 256 512 1024 * * (infinite) * When this mode is set, “MIN --- (number of times)” is displayed in the CALC area on the screen. (Refer to “4. Explanation of Screen” for the details.)
16.4 AVERAGE Mode
- By the operation of , the number of times of the sweep in the AVERAGE mode is set. 2. The simple averaging process is executed at each sweep. 2 4 8 16 32 64 128 256 512 1024 * When this mode is set, “AVG --- (number of times)” is displayed in the CALC area on the screen. (Refer to “4. Explanation of Screen” for the details.)
16.5 OVERWRITE Mode
- By the operation of , the number of times of the sweep in the OVER WRITE mode is set. 2. The image on the screen is not cleared at each sweep, and the overwriting display is executed. 2 4 8 16 32 64 128 256 512 1024 * * (infinite) * When this mode is set, “OVER WR” is displayed in the CALC area on the screen. (Refer to “4. Explanation of Screen” for the details.) * Only the last spectrum is saved.
16.6 SPURIOUS FREE Mode (2658A only)
- By pushing , the SPRIOUS FREE mode, by which the spurious response at band 1- and band 1+ is simply deleted, is selected. * “SPR” is displayed in CALC area on the screen. (Refer to “4. Explanation of Screen” for the details.)
- Differing from base band at which the up-conversion of input frequency is done with a frequency mixer, the spurious response peculiar to band 1- and band 1+ is generated because the down-conversion is done at those two bands. * About SPURIOUS FREE mode 1. SPURIOUS FREE mode is a mode by which the spurious response peculiarly caused at band 1+ is simply deleted. 2. SPURIOUS FREE mode is especially effective in the measurement of a stationary wave. 3. If SPURIOUS FREE mode is used in measurement of a signal with level change or frequency change, the phenomenon that the level goes down will occur. 4. The noise level goes down a little in case of a noise changing at random. In short, the same phenomenon as MIN HOLD function happens. * How to judge SPURIOUS response at band 1- and band 1+ The procedure for judging SPURIOUS response at band 1- and band 1+ is as follows. 1. Set the center frequency to < (current setting value) + f >. 2. The spectrum shifting left by f is a correct spectrum, but another spectrum shifting such as left by 2f, left by 3f, right by f, right by 2f or right by 3f is a spurious response. * “f ” should be changed according to SPAN for easy judgment. For reference; f = SPAN / (10 to 50) (corresponding to 0.2 to 1 div) Example: judging a spectrum displayed at 6.92 GHz (setting : center frequency 6.92 GHz and SPAN 500 MHz) 1. Set the center frequency to < 6.92 GHz + 500 MHz/25 > = 6.94GHz. right by 2f from 6.94 GHz) is spurious. * Additionally, a spurious response mentioned above shifting right by 2f at band 1+ is largest, and other spurious response are about 25 to 50 dB smaller than that. 17. Marker & Peak Search <MKR> When is pushed, the following function menu is displayed.
- Main menu of marker function F1 F2 F3 F4 F5 F6 CONV dBm →W NORMAL MARKER DELTA OFF CENTER FREQ NORMAL ZONE MKR
17.1 Movement and Basic Function of Marker
- By rotating , the marker moves. 2. By pushing , the normal marker mode is selected and the marker operates as an independent marker. The frequency and the level at the marker point are displayed in the lower right side. 3. When is pushed, the delta marker mode is selected, and the reference marker and the active marker are displayed. The frequency and the level at the marker points are displayed as follows; Frequency = (freq. at active marker) – (freq. at reference marker) Level = (level at active marker) – (level at reference marker) 4. By pushing , the marker is erased (marker off).
17.2 Normal Peak Search
- Menu of normal peak search When is pushed, the following function menu is displayed. 1. When is pushed, the marker moves to the maximum peak position of spectrums. 2. When is pushed, the marker moves to the next highest peak. Fifty or less peaks are available. 3. When is pushed, the function menu is returned to the main menu of marker function.
17.3 Zone Peak Search
- Menu of normal peak search When is pushed, the following function menu is displayed. RETURN F1 F2 F3 F4 F5 F6 PEAK SEARCH NEXT PEAK
At the Zone mode, the marker automatically moves to the maximum peak point within the range set beforehand. When entering this mode, the zone is displayed with thin blue shadow in the spectrum display area. The marker moves to maximum level within this zone at each sweep. The center of the zone can be adjusted by pressing The width of the zone can be adjusted by pressing Note: The marker does not move in HOLD.
17.4 Change Unit of Marker Level
When is pushed, the unit of the marker level is changed. If the unit of the reference level is dBm, it can be changed to [dBm] [W]. If the unit of the reference level is dBμV, dBmV or dBV, it can be changed to [dBμV, dBmV, dBV] [V]. If the unit of the reference level is dBμV/m, it can be changed to [dBμV/m] [V/m]. If the unit of the reference level is dBμA/m, it can be changed to [dBμA/m] [A/m]. Note: “V/M” is only displayed when the electric field strength measurement is selected “A/M” is only displayed when the magnetic field strength measurement is selected Each unit change as described above has their own sub-menu for more unit display options. They are shown below: [W] [W, mW, μW, nW, pW, fW] [V] [V, mV, μV, nV] [V/m] [V/m, mV/m, μV/m, nV/m] [A/m] [A/m, mA/m, μA/m, nA/m] 18. Save / Load <SAVE/LOAD> When is pushed, SAVE/LOAD menu is displayed as follows. F1 F2 F3 F4 F5 F6 RETURN ZONE CENTER WIDTH SAVE / LOAD
18.1 Selection of Storage Device
- When is pushed, the built-in flash memory is selected as a storage device. When is pushed, the external USB memory is selected as a storage device. * Install the USB memory in USB A- plug on the right side firmly. Refer to the figure below. * Once you disconnect, then reconnect the same or a different USB flash memory drive, the 2650A series will not automatically recognize the new USB drive. After connecting the drive, you must first press the Save/Load key, followed by F5 key (USB) in order for the 2650A to "see" the new drive., then proceed as described below. F1 F2 F3 F4 F5 F6 PRE SET SAVE MODE SELECT LOAD DELETE DEVICE MEM USB PRE SET
18.2 Save Function
When is pushed, SAVE menu is displayed as follows. 1. In this function, the data is stored in the built-in memory or USB memory by CSV format. The data stored in USB memory can be used with a personal computer as a CSV file. “2650A_csvData” folder will automatically be created and the data will be stored in it as a file. 2. The object to be stored is selected with , or . SPECT: The current spectrum on the screen is stored. PARAM: The setting parameters are stored. BOTH: The spectrum and setting parameters are stored as one file. * The selected key will look pushed-in on the display. 3. When is pushed, SAVE is executed. The data is stored based on the file name displayed in the active area (refer to “4. Explanation of screen”). As for the file name, refer to “18.3 About File Name”. Up to 200 files can be stored in the built-in memory. Up to 1000 files can be stored in USB memory. 4. Pushing will return to the former menu.
18.3 About File Name
- The file name is attached to the file stored using this function as follows. LABEL-SP001.csv (1) (2) (3) (4) (1) This is the label entered from the label function. Refer to “24.1 Label function.” Note: Symbols and spaces are not recognized as a file name. (2) S : The object stored is the spectrum (SPECT). P: The object stored is the setting parameters (PARAM). SP: The object stored is both of spectrum and setting parameters (BOTH). (3) The consecutive three digit number from “000” is automatically attached if (1) and (2) are same. (4) This is the extension showing CSV format and is automatically attached. F1 F2 F3 F4 F5 F6 SPECT OBJECT PARAMM BOTH SAVE EXECUTE PRE SET RETURN F1 F2 F3
- If the label is not named, “2650A” will be the first part of the file name instead. For example: 2650A_SP001.csv 3. About file name at the time of storage When SAVE is selected, the file name is displayed in the active area following according to the above labeling convention. When SAVE is executed as it is, the data is stored with that file name. If the file exists in the storage device and is rotated left before SAVE is executed, those file names are displayed one by one. If SAVE is executed with an existing file name displayed, the data is overwritten on it.
18.4 Load Function
When is pushed, LOAD menu is displayed as follows. 1. When is rotated right or left, the stored file name is displayed one by one from the specified storage device in the active area as follows. 1) The storage device in which the file is stored. (USB: USB memory, MEM: built-in memory) 2) The file name selected. 3) Number attached to file The files in the specified folder of the storage device are sorted in alphabetical order, and this shows that number. 2. To search files, press , and the file will be displayed in order of the number attached to it in the active area. The search steps can be adjusted by pressing to search in steps of 10 files at a time or pressing to search in steps of 100 files at a time. Select it according to the number of stored files. 3. When is pushed, LOAD is executed. * When SPECT is selected, a spectrum waveform is loaded. The setting parameters of loaded spectrum are displayed in the setting parameters display area. [Refer to “4. Explanation of screen” for details.] F1 F2 F3 F4 F5 F6 RETURN File Search Step 10 100 LOAD EXECUTE DISP CLEAR 1) USB : LOAD 2) 2650A_SP012.csv 3) 12
- When PARAM is selected, setting parameters are loaded. * When the spectrum is loaded, the 2650A series enters HOLD state and the current spectrum disappears, replacing it with the loaded spectrum on main display. When is pushed right after, the loaded spectrum and a current spectrum are displayed together. 4. When is pushed, the loaded spectrum can be cleared on display.
18.5 Delete Function
When is pushed, DELETE menu is displayed as follows. 1. The file name to be deleted is selected from the specified device with , and the active area will indicate this name. As for the number of search steps, it is the same as “18.4 Load Function”. 2. When is pushed, DELETE is executed. The file displayed in the active area is deleted at this time. 3. Repeat step 1 and 2 above to delete more than one file.
18.6 Presetting (Initialization)
(For 2651, refer to “20. EMI test”) When is pushed, the setting parameters are reset to the preset values as shown below. Items Parameters Center frequency 1 GHz Frequency span 20 MHz Reference level 10 dBm Offset 0.0 dB Impedance 50 Ω Sweep time 30 ms Detection mode Sample mode (SMPL) RBW 100 kHz VBW 30 kHz Display scale 10 dB/div F1 F2 F3 F4 F5 F6 RETURN File Search Step 10 100 DELETE EXECUTE HOLD/RUN
- Measuring Function <MEAS> When is pushed, the following function menu is displayed. The measuring function can be selected as follows. * Once the measuring function is set, when is pushed, the function menu is directly returned to the last setting of the measuring function. In order to stop the measuring function or to select the other measuring function, push [F6] key (MEAS OFF), then the function menu is returned to the main menu of the measuring function. * Each of the four functions (Channel power, Adjacent channel leakage power, Occupied bandwidth and Frequency counter) and the marker function cannot be used at the same time. When is pushed while each of the four functions is selected, the measuring function will stop. Similarly, when the measuring function is selected while using the marker function, the marker function will stop. * Although 501 points can be displayed on the screen, the number of points calculated in the instrument is 1001.
19.1 Channel Power Measuremnt <CH POWER>
The total power in the specified frequency band is measured. Two modes, TOTAL and BAND, are available.
- TOTAL mode [ By pushing (MODE), TOTAL mode is selected.] The sum of power in the displayed spectrum specified by center frequency and frequency span is measured. F1 F2 F3 F4 F5 F6 FREQ COUNT E / F ANT M / F PROBE OBW CH POWER ACP MEAS MKR MEAS
- BAND mode [ By pushing (MODE), BAND mode is selected.] The sum of power in the band specified by band center and bandwidth is measured.
19.2 Adjacent Channel Leakage Power Measurement <ACP>
The adjacent channel leakage power is measured as the ratio of power in the range specified by offset frequency and bandwidth to carrier power. Both leakage power at the upper and lower side are measured. Furthermore, the method for measurement is selected out of three methods based on the classification of definition of carrier power; total power method, reference level method and in-band method.
- Mode selection and measurement [ By pushing (MODE), TOTAL, BAND or PEAK mode is selected.] * “ACP TOTAL”, “ACP BAND” or “ACP PK” is displayed in MEAS area on the screen. * “CP BAND” is displayed in MEAS area on the screen. * The measured value and the setting parameter are displayed at the lower part of the screen. 1. By (BAND ENTER) , the band center is set. 2. By (BAND WIDTH) , the band width is set. [Measuring Mode] [Measured Value] [Setting parameter] * “CP TOTAL” is displayed in MEAS area on the screen. * The measured value is displayed at the lower right corner on the screen. [Measuring Mode] [Measured Value]
- The measured value and the setting parameter are displayed at the lower part of the screen.
- Definition of Reference Carrier for each Mode 1. By (ACP OFFSET) , the offset frequency of adjacent channel is set. * The offset is from the center frequency of the reference carrier. 2. By (ACP WIDTH) , the band width of adjacent channel is set. 3. By (REFERENCE CENTER) , the center frequency of the reference carrier is set. * [F4] is only for TOTAL and BAND mode. 4. By (REFERENCE WIDTH) , the band width of reference carrier is set. * [F5] is only for BAND mode. [Measured value] [Setting parameter] [Measuring mode] TOTAL (total power method) This is based on the sum of power in the displayed spectrum. The center frequency of the reference carrier is set by [F4]. BAND (in-band method) This is based on the sum of power in the specified bandwidth. The center frequency of the reference carrier is set by [F4]. PEAK (reference level method) This is based on the power of peak on the screen. The center frequency of the reference carrier is automatically set to the peak on the screen.
19.3 Occupied Bandwidth Measurement <OBW>
It is possible to measure the occupied frequency bandwidth defined as the width of points at N(%) of the total power (N% POWER), or as the width of points that are X(dB) lower than the peak level (XdB DOWN).
- N% POWER mode [ By pushing (MODE), N% POWER mode (N%) is selected.] The bandwidth of points at N(%) of the total power is measured.
- XdB DOWN mode [ By pushing (MODE), XdB DOWN mode (X dB) is selected.] The bandwidth of points that are X(dB) lower than the peak level is measured.
19.4 Electric Field Strength Measurement <E/F ANT>
Connecting a dipole antenna (option) to the RF input enables the measurement of the electric field strength. Other than the optional antenna is also available by creating the original compensation data. (Refer to “25.8 Writing of Original Compensation Data” for the details.) [Measuring mode] [Measured value] [Setting parameter] * “OBW XdB” is displayed in MEAS area on the screen. * The measured value is displayed at the lower right corner on the screen. 1. By (XdB) , the down level from the peak level is set. * Setting range : 0.1 to 99.0dB 2. By , the step size of XdB can be set to 1dB or 0.1dB. * “OBW N%” is displayed in MEAS area on the screen. * The measured value is displayed at the lower right corner on the screen. 1. By (RATIO %) , the percentage of the total power is set. * Setting range : 80.0 to 99.9% 2. By , the step size of RATIO % can be set to 1% or 0.1%. [Measuring mode] [Setting parameter] [Measured value]
[Specifications of dipole antenna (The antenna gain and VSWR are specified at the center of frequency range.)] * When M405 is used, the measurement errors will occur due to how to hold of instrument and influence of human body, because M405 is ¼ λ whip antenna. The error value will be several dB or more. In order to reduce the error, separate 2650A series from human body as much as possible.
- Mode Selection and Measurement By pushing (ANT), the antenna (M401/M402/M403/M404/M405/M406/USER) can be selected The measurement starts as soon as the antenna is selected. Items M401 M402 M403 M404 M405 M406 Type Sleeve Sleeve Sleeve Sleeve ¼ λ whip Sleeve Antenna gain >1dB >1dB >1dB >1dB >1dB >1dB Reference level setting range (except for the minimum value in screen shift) 93 to 143 dBμV/m 96 to 146 dBμV/m 99 to 149 dBμV/m 100 to 150 dBμV/m 87 to 137 dBμV/m 109 to 159 dBμV/m 2650A series * Turn the back to the signal source. Connect to RF INPUT. Antenna M401 M402 M403 M404 M405 M406 [Antenna connection] [Measurement environment]
- “E/F M40X” or “E/F USER” is displayed in MEAS area on the screen. * “USER” is the original compensation data which is created by the user. (Refer to “25.8 Writing of Original Compensation Data” for the details.) * The spectrum may exceed the area on the screen depending on the antenna gain compensation. The unit of amplitude axis automatically changes into [dBμV/m]. * The optimum center frequency and frequency span are automatically set according to the antenna. In addition, a spectrum other than the frequency range of the antenna is not displayed.
- Directivity of Antenna (reference data) * The following are the data when the antenna is connected to RF input directly with no obstacles around. Actually, the directivity changes due to the human body when handling the unit. * However, the data of M405 is the reference data which is included the influence of the human body. 90° 270° 180° E plane: X-Y axis (X direction=0°) M401 (900 MHz, E plane) Antenna gain vs. Frequency
M404 (2.4 GHz, E plane) Antenna gain vs. Frequency M403 (2 GHz, E plane) Antenna gain vs. Frequency 0.5 1.5 2.5 3.5 4.5 5.5 Frequency (GHz) ANT Gain (dBi) M402 (1.5 GHz, E plane) Antenna gain vs. Frequency 1.5 2.5 3.5 4.5 5.5 Frequency (GHz) ANT Gain (dBi)
19.5 Magnetic Field Strength Measurement <M/F PROBE>
The magnetic field distribution can be measured by using the magnetic field probe CP-2S (option). M406 (5.4 GHz, E plane) Antenna gain vs. Frequency 4.70 Frequency(GHz) 3.5 2.5 1.5 ANT Gain (dB) M405 (horizontal plane) Antenna gain vs. Frequency
“Specifications of magnetic field probe CP-2S” (Refer to the operating manual of CP-2S for the details.)
- Mode Selection and Measurement By pushing (PROBE), CP-2S or USER mode is selected. The measurement starts as soon as the probe is selected. * “M/F CP2S” or “M/F USER” is displayed in MEAS area on the screen. * “USER” is the original compensation data which is created by the user. (Refer to “25.8 Writing of Original Compensation Data” for the details.) The unit of amplitude axis automatically changes into [dBμA/m]. * A spectrum other than the frequency range of the probe is not displayed. Item Specifications Frequency range 10 MHz to 3 GHz Space resolution (-6 dB) approx. 0.25 mm (depending on objects) Reference level setting range: maximum 160 to 203 dBμA/m Reference level setting range: minimum (except for the minimum value in screen shift) 110 to 153 dBμA/m Measurement error approx. ±1 dB (probe simple substance) [Magnetic field probe connection] The tip of the magnetic field probe CP-2S is made of glass-ceramic board. Take care when handling the probe even though the strength of the glass-ceramic board is durable under normal operation.
19.6 Frequency Counter <FREQ COUNT>
This function currently not supported. 20. EMI Test (2651)
20.1 Additional Function for EMI Test
Detection mode (Refer to “13.4 Setting of Detection mode”) By pushing , the detection mode can be set. POS SMPL NEG QP AV In the 2651, QP detection and AV detection are added for EMI measurement. POS (Positive Peak) : The spectrum of the maximum value between the sample points SMPL (Sample) : The spectrum of the momentary value between the sample points NEG (Negative Peak) : The spectrum of the minimum value between the sample points QP (Quasi Peak) : The spectrum of the quasi peak value between the sample points AV (Average) : The spectrum of the average value between the sample points As for QP detection, the following characteristics are chosen according to RBW settings. (based on CISPR16) RBW charging time constant dis-charging time constant meter time constant 9kHz 1ms 160ms 160ms 120kHz 1ms 550ms 100ms
- Preset (Initialization) (Refer to “18. Save / Load”) In the 2651, the radiated emission measurement and the conducted emission measurement are added to the preset as a default setting. By pushing , the menu is displayed NORMAL : The initial parameters of normal mode are set. EMI-C : The initial parameters of conducted emission measurement are set. EMI-R : The initial parameters of radiated emission measurement are set In the radiated mission measurement, “USER” is selected as an antenna. It is necessary to input the compensation data of antenna beforehand. F6 SWEEP F6 SAVE/LOAD
20.2 EMI Test
The QP detection is usually used in both of the radiated and conducted emission measurements, and the AV detection is usually used in the conducted emission measurement. The measurement time can be shortened by using them in the final measurement to the spectrums narrowed by the PK detection. Conducted Emission Measurement The conducted emission discharged through the power supply line is measured. In this measurement, LISN* (Line Impedance Stabilization Network) device is needed. The connection is shown in the figure below. Please refer to the manual of LISN for details of the connection and notes. 1. Turn on the power of the 2651 after the connection. 2. Set the initial parameters of conducted emission measurement by pushing . The setting parameters are as follows. Center Frequency : 25.5 MHz Frequency span : 50 MHz RBW : 9 kHz VBW : 1 MHz Sweep time : 3 sec Detection mode : Positive peak mode 3. Confirm whether the connection and the operation are correct by supplying the power to the EUT and measuring. 4. Set the number of times of the sweep in the MAX HOLD mode by operating 5. Search the frequency of disturbance noise that exceeds the AV or QP limits by using the marker function. 6. Set the center frequency to the point that exceeds the limits, the SPAN to 2 MHz and the SWEEP to 0.1 sec. And measure the frequency accurately. Power Supply EUT 2651 LISN * F6 SAVE/LOAD F2 F2 CALC
- Moreover, measure the frequency detected in the above with QP or AV detection, then the SPAN is 200 kHz and the SWEEP is 10 sec. Radiated Emission Measurement The electric field strength of disturbance noise that EUT radiates in the air is measured. 1. Set the compensation data of the antenna beforehand. Refer to “25.8 Writing original compensation data” for the details. 2. Turn on the power of the 2651 after connection. 3. Set the initial parameters of radiated emission measurement by pushing . The setting parameters are as follows. Center Frequency : 515 MHz Frequency span : 1 GHz RBW : 120 kHz VBW : 1 MHz Sweep time : 0.3 sec Detection mode : Positive peak mode 4. Confirm whether the connection and the operation are correct by supplying the power to the EUT and measuring. 5. Set the number of times of the sweep in the MAX HOLD mode by operating 6. Search the frequency of disturbance noise that exceeds the QP limits by using the marker function. 7. Set the center frequency to the point that exceeds the limits, the SPAN to 50MHz and the SWEEP to 30 msec. And measure the frequency accurately. 8. Moreover, measure the frequency detected in the above with QP detection, then the SPAN is 1MHz and the SWEEP is 30 sec. Antenna 2651 EUT F6 SAVE/LOAD F3 CALC
EMI standards (selected) Frequency Quasi Peak Average CISPR22 class A Limits of conducted disturbance at mains ports 0.15 MHz to 0.50 MHz
0.50 MHz to 30 MHz
Limits of conducted disturbance at mains ports 0.15 MHz to 0.50 MHz
0.50 MHz to 5 MHz
5 MHz to 30 MHz
Decreasing linearly with the logarithm of the frequency CISPR22 class A Limits of radiated disturbance
30 MHz to 230 MHz
230 MHz to 1000 MHz
1000 MHz to 3000 MHz
3000 MHz to 6000 MHz
--- --- 56 dBuV/m 60 dBuV/m Measurement distance 10 m CISPR22 class B Limits of radiated disturbance --- --- 50 dBuV/m 54 dBuV/m Measurement distance 10 m VCCI class A Limits of conducted disturbance at mains ports 0.15 MHz to 0.50 MHz Limits of conducted disturbance at mains ports 0.15 MHz to 0.50 MHz Decreasing linearly with the logarithm of the frequency. VCCI class A Limits of radiated disturbance --- --- Measurement distance 10 m Limits of radiated disturbance
30 MHz ~ 230 MHz
230 MHz ~ 1000 MHz
--- --- Measurement distance 10 m FCC part15 subpartB class A Limits of conducted disturbance at mains ports 0.15 MHz to 0.50 MHz FCC part15 subpartB class B Limits of conducted disturbance at mains ports 0.15 MHz to 0.50 MHz Decreasing linearly with the logarithm of the frequency. FCC part15 subpartB class A Limits of radiated disturbance
30 MHz to 88 MHz
88 MHz to 216 MHz
216 MHz to 960 MHz
39.1 dBuV/m 43.5 dBuV/m 46.4 dBuV/m 49.5 dBuV/m --- --- --- --- Measurement distance 10 m FCC part15 subpartB class B Limits of radiated disturbance 43.5 dBuV/m 46 dBuV/m 54 dBuV/m --- --- --- --- Measurement distance 3 m (Note: Information provided in this table serves as a reference only. B&K Precision is not liable for any damages due to any mistakes in the table.) 21. Screen Control <DSPL> When is pushed, the following function menu is displayed. DSPL
21.1 Setting Screen Display Color
When is pushed, the color of screen display can be selected. COLOR1 COLOR2 MONO
- COLOR1 : For normal use
- COLOR2 : For printing screen • MONO : Monochrome
21.2 ON/OFF Switching of LCD Backlight
Use button to toggle the backlight ON or OFF.
21.3 Adjusting Brightness of LCD Backlight
Use to set the brightness. It can be set in the range of 0 to 100. 22. Tracking Generator Mode (2652A) 22.1 Specification of T.G. Function Item Specification Frequency range 5 MHz to 3.3 GHz Output Level -10 dBm ± 1 dB @1 GHz(Fixed value) Output flatness ±1.5 dB Output impedance 50 Ω Output VSWR Less than 2.0 Output connector N(J) connector F1 F2 F3 F4 F5 F6 BRIGHT COLOR COLOR1 BACK LT ON
22.2 Description of I/O Connector
1) Input connector N (J) connector Input for an external signal Make sure that the total power of input signals does not exceed +27 dBm. 2) Output connector N (J) connector It is an output terminal of Tracking Generator. *Please set to “TG : OFF” to prevent the influence of the leak from the Tracking Generator in case of disuse of the T.G. function.. 22.3 ON/OFF Switching of T.G. Function When is pushed, the following function menu is display. 1. When is pushed, T.G. MODE is set. F1 F2 F3 F4 F5 F6 T.G. MODE MANUAL SWEEP AUTO ALL AUTO DET SMPL TRIG INT SWEEP
- Each time is pressed, T.G. output is alternately switched to ON or OFF.
22.4 Normalizing Function
NORM ON: The input level is made flat to the red line on the screen. 1. When is pushed, the same function menu as the previous item is displayed. 2. When is pushed, TG. MODE is set. 3. Each time is pushed, the normalizing function is alternately switched to ON or OFF. If the normalizing function is switched to ON, “NORM ON” is displayed on the screen. * If the setting is changed as follows, the normalizing function is automatically turned off.
- The span is expanded. · The center frequency is changed beyond the range normalized. ( When SPAN is changed from FULL SPAN according to center frequency, it is turned off.)
- The magnetic field strength measurement is selected.
- The AUTO tuning is executed. · The power supply is turned off. · The presetting is executed. * When the scale is 2 dB or 5 dB, the normalizing function does not operate correctly if the spectrum level is not displayed at a proper position on the screen. F1 F2 F3 F4 F5 F6 T.G. ON NORM ON SWEEP
- Storage/Print Screen Image <COPY> The following function menu is displayed when is pushed.
23.1 Selection of Image Area
When is pushed, the image area to be stored or printed can be selected. WHOLE : the whole screen image is stored or printed. SPECT : only the spectrum display area is stored or printed.
23.2 Print on Printer
When pushing after connecting the printer (option), the area selected in 23.1 is printed.
23.3 Storage into USB Memory
When pushing after connecting USB memory (removable storage) to this unit, the area selected in 23.1 is stored in USB memory in BMP (bit map) format. As for the file name, the number is automatically set and incremented in the form “2650A_001.bmp”. In the case when some characters are written in the label area, the file name will consist of it and the number put automatically such as “LABEL001.bmp”. The folder “2650A_IMG” is automatically generated and these files are stored inside. To change the color of the display before print, refer to “21. Screen Control”. F1 F2 F3 F4 F5 F6 AREA WHOLE PRINT EXECT Bmp → USB COPY MEM → USB EXECT COPY
※ Once you disconnect, then reconnect the same or a different USB flash memory drive, the 2650A series will not automatically recognize the new USB drive. After connecting the drive, you first must press the Save/Load, followed by F5 key (USB) in order for the 2650A to "see" the new drive, then proceed with copying data to the drive.
23.4 Transferring Internal Data to USB Memory in Lump Sum
When pushing after connecting USB memory (removable storage) to this unit, the data stored in the internal memory with SAVE/LOAD function are transferred to USB memory in a lump sum. During the transfer process, the screen on the lower right corner will display “Saving”. Wait until it disappears, indicating that transfer is complete, before continuing operation. The folder “2650A_SaveAll” is automatically generated and these files are stored inside. ※ Once you disconnect, then reconnect the same or a different USB flash memory drive, the 2650A series will not automatically recognize the new USB drive. After connecting the drive, you must first press the Save/Load key, followed by F5 key (USB) in order for the 2650A to "see" the new drive, then proceed as normal.
23.5 USB Printer (option)
The 2650A/2651 series enables a hard copy of the screen by connecting a USB printer (option) to USB A plug with the included USB cable.
- Turn on the power of the 2650A/2651 series unit first. After that, turn on the power of USB printer. Otherwise, it will not work correctly.
23.6 USB Memory
USB memory can be used as a removable storage. USB B plug Printer USB A plug
- Utility Function <UTIL> When is pushed, the following function menu is displayed. Label Function : Enters the label menu for customizing filename. Menu Off : Turns off menu and active area display. Buzzer Setting : Sets the condition of the buzzer sound. Clock Function Setting : Sets the built-in clock.
24.1 Label Function
When is pushed, the following function menu is displayed. 1. The kind of character to be written is selected with to . 012 : 0 to 9 (numeral) abc : a to z (lower case alphabet) ABC : A to Z (upper case alphabet) *Note: Certain special characters may not be recognized or valid as a filename. 2. A block cursor will be displayed in the label display area. By rotating , a number, small letter alphabet, capital letter alphabet or special character will be indicated on the cursor position. Turn the knob until the desired label character is displayed. 3. When is pushed, the cursor position will move right by one character. And the previous character is set. F1 F2 F3 F4 F5 F6 012 abc ABC @%* BACK SP ENTER F1 F2 F3 F4 F5 F6 LABEL MENU OFF BUZZER ALWAYS CLOCK CONFIG UTIL F1 UTIL F1 F4
- When is pushed, the cursor position will move left by one character. To change the character, turn the encoder. To delete, press F5 again. Characters previously entered can be deleted by consecutively pushing F5. 5. When finished, press to exit the label menu.
24.2 Menu Off
By pushing , the display of the function menu and the active area can be turned off temporarily.
24.3 Buzzer Setting
By pushing , the buzzer sound can be set to beep when keys are pressed or rotary encoder is turned. The following three options are selectable by pressing F3 until the desired option is displayed. OFF : The buzzer is off. ALARM : The buzzer will sound off when an incorrect setting is made. ALWAYS : The buzzer will sound whenever a key or the rotary encoder is operated. ※ When the battery voltage becomes low during battery operation, the buzzer will sound. (Even if OFF is selected, it will sound.)
24.4 Setting the Clock
When is pushed, the following function menu is displayed. The year, month, day and time are displayed below the screen. (Refer to “4. Explanation of Screen”) Set the year, month, day and time when 2650A/2651 series is used for the first time. The time information is updated even if the power is turned off. F1 F2 F3 F4 F5 F6 YEAR MONTH JUN DATE HOUR MIN ENTER F4 UTIL UTIL
- By rotating , the values for a selected parameter (choosen by pressing any button between F1 and F5) can be changed. 2. Used to set the year. Two last digits of the year are displayed. This can be set from 00 to 99. Used to set the month. This can be set from 1 to 12. Used to set the date. This can be set from 1 to 31. Used to set the hour. It is displayed in twenty-four hour format. Values can be set from 1 to 24. Used to set the minute. 3. By pushing , the values which are set from item 2. above are saved. The values are updated only when this key is pushed. * The clock function of the 2650A/2651 series is driven by a dedicated LSI, and the power is supplied by a built-in lithium battery.
- USB Device Function
25.1 Outline
The 2650A/2651 series is equipped with two kinds of USB connecters; A and B plugs. The A plug is for USB host and is connected to USB devices such as USB memory or printer. The B plug is for USB device and is connected to USB host such as a personal computer. The device function is explained in this chapter. As for the host function using USB A plug, refer to “23. Storage and print of screen image”.
25.2 How to Connect
When using the USB device, connect with the included USB cable as illustrated
25.3 Installation of Driver
An instrument specified USB driver is necessary in order to control the 2650A/2651 series with a PC via USB interface, and should be installed in the PC prior to using the software. The driver along with instructions can be downloaded from our website www.bkprecision.com. (Go to the Support section and select download). After the downloaded file is unzipped, the 2650A/2651 series is connected to a PC with the USB cable and then the power is turned on. The detection wizard for the USB driver starts on the PC screen the moment the power of the 2650A/2651 series is turned on, and install it following this wizard.
25.4 Sample Program
Please contact us at www.bkprecision.com to request a sample code which demonstrates how to write a custom program using the commands in the following section. USB A plug USB B plug
25.5 Explanation of Command
- “CR(0D[HEX])+LF(0A[HEX])” is added to the end of every command. When the command is sent from PC, 2650A/2652A/2651 and 2658A return some responses, which include “OK” + CR + LF, “ERR” + CR + LF, or “(response to command)” + CR + LF. * By inputting “?” instead of “* *” for each command, the current setting parameters will be queried and a response of the value will be received by the PC. (Except for “ ---- Request” command and the command for inputting corrected data.) 1) Setting center frequency (* * * * * * *: Refer to “25.6 Input of frequency”) Example: FREQ1.235G Query: FREQ? 2) Request set marker Command : FREQSETMK * The center frequency is set according to the frequency of current marker position. 3) Setting frequency span Command : SPAN* * * * 2650A, 2652A, 2651: (* * * * = ZERO, 200K, 500K, 1M, 2M, 5M, 10M, 20M, 50M, 100M, 200M, 500M, 1G, 2G, FULL [unit : Hz]) 2658A : (* * * * = ZERO, 200K, 500K, 1M, 2M, 5M, 10M, 20M, 50M, 100M, 200M, 500M, 1G, 2G, 5G, FULL [unit : Hz]) Example: SPAN500K Query: SPAN? 4) Setting reference level Command : REF* * * (* * * = -60 to 10 [1 step, unit : dBm] ) Example: REF 5 Query: REF? 5) Setting reference level unit Command : UNIT * * * * (* * * * = DBM, DBUV, DBMV, DBV) Example: UNITDBUV Command Unit DBM dBm DBUV dBμV DBMV dBmV DBV dBV * When the unit is other than dBm, convert into dBm by using calculation formula in “9.8 Relation between reference level and ATT/AMP”.
6) Setting RBW Command : RBW* * * 2650A, 2652A, 2658A : ( * * * * = 3K, 10K, 30K, 100K, 300K, 1M, 3M, AUTO, ALL [ unit : Hz ]) 2651 : ( * * * * = 3K, 9K, 30K, 120K, 300K, 1M, 3M, AUTO, ALL [ unit : Hz ]) * ALL : The optimum RBW, VBW, and sweep time are set based on the frequency span. * AUTO : The optimum RBW is set based on the frequency span. Example: RBW3M Query: RBW? 7) Setting VBW Command : VBW* * * * ( * * * * = 100, 300, 1K, 3K, 10K, 30K, 100K, 300K, 1M, AUTO, ALL [ unit : Hz ] ) * ALL : The optimum RBW, VBW and sweep time are set based on the frequency span. * AUTO : The optimum RBW is set based on the frequency span. Example: VBW300 Query: VBW? 8) Selecting Measuring Function Command : MEAS* * * (* * * = CP, ACP, OBW, EF, MF, FC, OFF) Example: MEASOBW (To change measuring function when in a different measurement mode, set function OFF first by sending MEASOFF) Query: MEAS? 9) Request measuring result Command : MEASRES * Example of measuring result Channel power measurement ----- POW: - 25.5 dBm Adjacent channel power measurement ----- L: - 47.7 dBc U: - 48.3 dBc Occupied bandwidth measurement ----- C: 1.45 G W: 20.00 k Frequency counter ----- FC: 2400.0000 M * If the frequency counter is not installed, “Invalid for F/C” is responded. * If the level of spectrum is low and unmeasurable, “Non Signal” is responded. Command Measuring function CP Channel power measurement ACP Adjacent channel power measurement OBW Occupied bandwidth measurement EF Electric field strength measurement MF Magnetic field strength measurement FC Frequency counter (factory option) OFF OFF
10) Setting mode for channel power measurement Example: CPMODETOTAL Query: CPMODE? 11) Setting zone center frequency for channel power measurement ( * * * * * * *: = 0-500: screen position, center = 250) Example: CPCNTR300 Query: CPCNTR? 12) Setting zone width for channel power measurement ( * * * * * * *: = 0-500: screen position, center = 250) Example: CPWIDTH200 Query: CPWIDTH? 13) Setting mode for adjacent channel power measurement Command : ACPMODE* * * * * (* * * * * = TOTAL, BAND, PEAK) Example: ACPMODEPEAK Query: ACPMODE? 14) Setting band offset for adjacent channel power measurement ( * * * * * * *: = 0-500: screen position, center = 250) Example: ACPOFS100 Query: ACPOFS? 15) Setting bandwidth for adjacent channel power measurement ( * * * * * * *: = 0-500: screen position, center = 250) Example: ACPCHBW30 Query: ACPCHBW? 16) Setting reference band center frequency for adjacent channel power measurement Command Mode TOTAL TOTAL (total power method) BAND BAND (in-band method) PEAK PEAK (reference level method) Command Mode TOTAL The sum of power on the screen is measured. BAND The sum of power in the band specified is measured.
( * * * * * * *: = 0-500: screen position, center = 250) Example: ACPREF250 Query: ACPREF? 17) Setting reference bandwidth for adjacent channel power measurement ( * * * * * * *: = 0-500: screen position, center = 250) Example: ACPREFBW50 Query: ACPREFBW? 18) Setting mode for occupied bandwidth measurement Command : OBWMODE* * ( * * : N%, DB) Example: OBWMODEN% Query: OBWMODE? 19) Setting N% RATIO for occupied bandwidth measurement Command : OBWRATIO* * * Example: OBWRATIO85.6 Query: OBWRATIO? 20) Setting XdB DOWN for occupied bandwidth measurement Command : OBWDB* * * ( * * *: 0.1 to 99.0 [ 0.1 step , unit : dB] ) Example: OBWDB45.3 Query: OBWDB? 21) Setting antenna for electric field strength measurement Command : EFANT* * * * / M404 / M405 / M406 / USER) Example: EFANTM405 Query: EFANT? 22) Transfer user-compensation data for electric field strength measurement Command : EFUSER* * * * * (Refer to “25.8 Writing of Original Compensation Data” for the details.) Example: EFUSER300M:0.0DBI,350M:1.0DBI Command Mode N% N% POWER mode DB XdB DOWN mode Command Antenna M401 Setting data for M401 M402 Setting data for M402 M403 Setting data for M403 M404 Setting data for M404 M405 Setting data for M405 M406 Setting data for M406 USER Setting data for user’s original antenna
23) Setting probe for magnetic field strength measurement Command : MFPROBE* * * * ( * * * *: CP2S / USER) Example: MFPROBECP2S Query: MFPROBE? 24) Transfer user-compensation data for magnetic field strength measurement Command : MFUSER* * * * * (Refer to “25.8 Writing of Original Compensation Data” for the details.) Example: MFUSER300M:0.0DBI,350M:1.0DBI 25) Start and Stop calculation function Command : CALC* * * (* * *: OFF / MAX / MIN / AVE / OVR) Example: CALCMIN Query: CALC? 26) Setting number of times to sweep for MAX HOLD Command : MAXNO* * * * ( * * * * : 2 / 4 / 8 / 16 / 32 / 64 / 128 / 256 / 512 / 1024 / 0) * 0 = unlimited) Example: MAXNO1024 Query: MAXNO? 27) Setting number of times to sweep for MIN HOLD Command : MINNO* * * * ( * * * *: 2 / 4 / 8 / 16 / 32 / 64 / 128 / 256 / 512 / 1024 / 0) * 0 = unlimited) Example: MINNO512 Query: MINNO? 28) Setting number of times to sweep for AVERAGE Command : AVENO* * * * ( * * * *: 2 / 4 / 8 / 16 / 32 / 64 / 128 / 256 / 512 / 1024 / 0) * 0 = unlimited) Example: AVENO16 Query: AVENO? 29) Setting number of times to sweep for OVERWRITE Command : OVWNO* * * * ( * * * *: 2 / 4 / 8 / 16 / 32 / 64 / 128 / 256 / 512 / 1024 / 0) * 0 = unlimited) Example: OVWNO128 Query: OVWNO? Command Calculation OFF OFF MAX MAX HOLD MIN MIN HOLD AVR AVERAGE OVR OVER WRITE
30) Setting display scale Command : SCALE* * Example: SCALE10 Query: SCALE? 31) Setting sweep time Command : SWEEP* * * * Example: SWEEP0.1S Query: SWEEP? 32) Setting detection mode Command : DET* * * ( * * *: POS / NEG / SMP ) Example: DETPOS Query: DET? 33) Setting trigger source Command : TRG* * * ( * * *: INT / EXT) Example: TRGEXT Query: TRG? 34) Request AUTOTUNE Command : AUTO * The response is returned after tuning. 35) Hold/Run function Command : HOLD Command : RUN 36) Request marker information Command : MKRRES * Example of returned data : 1.42G -15dBm Query: MKRRES? Command Sweep time 10M 10ms 30M 30ms 0.1S 0.1s 0.3S 0.3s 1S 1s Command Sweep time 3S 3s 10S 10s 30S 30s AUTO AUTO ALL ALL AUTO Command Display scale 2 2dB/div 5 5dB/div 10 10dB/div Command Trigger source INT Internal EXT External Command Detection mode POS Positive peak mode NEG Negative peak mode SMP Sample mode QP QP mode (only for MSA438E) AVG AVG mode (only for MSA438E)
37) Setting marker mode ( * * * * *: NORM / DELTA) Example: MKRDELTA Query: MKR? 38) Setting marker position by frequency The position of active marker is set by frequency. ( * * * * * * *: Refer to “25.6 Input of Frequency”) Query: NORMMKR? 39) Setting marker position by number of points on horizontal axis The position of active marker is set by number of points on horizontal axis. Command : MKRPOSI* * * ( * * *: 0 to 500 ; 0 = far left, 500 = far right) Example: MKRPOSI40 40) Setting peak search mode Command : PEAK* * * * ( * * * * : NORM / ZONE) Example: PEAKZONE Query: PEAK? 41) Request peak search Command : PKSEARCH* * / 07 / 08 / 09 / 10 / 11) Example: PKSEARCH09 42) Setting zone center frequency for peak search ( * * * * * * *: Refer to “25.6 Input of Frequency”) Example: PKCNTR1.345G Query: PKCNTR? 43) Setting zone width for peak search ( * * * * * * *: Refer to “25.6 Input of Frequency”) Example: PKWIDTH256M Query: PKWIDTH? Command Marker mode NORM Normal marker DELTA Delta marker Command Destination of marker
01 The maximum peak on the screen
02 The 2nd peak on the screen
11 The 11th peak on the screen
44) Setting unit of marker Command : CONV* * * ( * * *: DBM / W / DBV / V / DBUVM / VM) Example: CONVDBUVM Query: CONV? 45) Request print on printer Command : PRTSIW USB printer connected to USB A plug prints. 46) Request for transfer of spectrum Command : SRS* * * * Example: SRSCURR *The number of the commands is given to the saved file. *Refer to “25.7 Transfer of Spectrum Data” for the returned data. 47) Request for transfer of spectrum in USB memory Command : SRSU* * * * Example: SRSU111 * The number of the command is given to the saved file. * Refer to “25.7 Transfer of Spectrum Data” for the returned data. 48) Request for transfer of spectrum data of 1001 points Command : SRSF * Refer to “25.7 Transfer of Spectrum Data” for the returned data. 49) Request preset Command : PRESET 50) Setting remote control Command : REMOTE* * * ( * * *: ON / OFF) Example: REMOTEON Query: REMOTE? Command Unit of marker DBM dBm W W DBV dBV V V DBUVM dBμV/m VM V/m Command Transferred spectrum CURR Current spectrum
000 Spectrum data of 000
199 Spectrum data of 199
Command Transferred spectrum
000 Spectrum data of 000 in USB memory
199 Spectrum data of 199 in USB memory
ON Operation by function key and encoder cannot be done. Control the unit with USB commands. OFF Operation by function key and encoder can be done. USB commands are available
- When the remote control is ON, “REMOTE” is displayed in the operating information area on the screen. ( Refer to “4. Explanation of Screen” for the details.) 51) Single sweep Command : CAPT * It sweeps only once and will be in a HOLD state. 52) Setting offset level Command : OFFSET* * * * * (* * * * *: -50.0 to 50.0 [0.1 step, unit: dB] ) Example: OFFSET-25.6 Query: OFFSET? 53) Setting input impedance Command : IMP* * (* * : 50 / 75) Example: IMP75 Query: IMP? * When the input impedance is selected, the offset level is set as described above. 54) Clearing spectrum data and parameter Command : MCLR* * * Example: MCLRWALL * For internal memory only. 55) Setting character for label Command : LBL* * * ( * * *: The number of characters is 16 or less) Example: LBLsample_spectrum Query: LBL? 56) Setting clock ( * * * * * * * * : aabbccdd ; aa: two last digits of year, bb: month (01 to 12), cc: hour (00 to 23), dd: minute (00 to 59) Example: CLC0902111103 Query: CLC? Command Offset level 50 Offset level is set to 0dB. 75 Offset level is set to 5.7dB. Command Cleared data ALL All of spectrum data
57) Setting tracking generator Command : TG* * * ( * * *: ON / OFF) Example: TGON Query: TG? 58) Setting normalizing function Command : NORM* * * ( * * *: ON / OFF) Example: NORMOFF Query: NORM? 59) Save/Load function Command : DEV*** ( * * *: MEN/USB) 60) Request On/Off of Offset Level Command: OFFSETOF? *Returned data: ON or OFF. 61) Request firmware version Command: VER
25.6 Input of Frequency
The frequency is input as follows. * * * * * * * : 0.0k to 999.9k (0.1 step, unit: Hz) : 0.0M to 999.9M (0.1 step, unit: Hz) : 0.0000G to 3.3G (0.0001 step, unit: Hz) ---- 2650A, 2652A, 2651 : 0.0000G to 8.5G (0.0001 step, unit: Hz) ---- 2658A * The offset frequency and the zone width can be input only in the range decided by the center frequency and frequency span. If the value out of the range is input, an error occurs. * The offset frequency and the zone width change when the frequency span is changed.
25.7 Transfer of Spectrum Data
The spectrum data outputs as a numeric string separated by commas “,”. [., ., ……, .] The unit is dBm, and it is available up to two decimal places.
- Description String Explanation Example SPECT The data which follows the string "SPECT" is the spectrum data. SPECT * *, * *, · · · It is the spectrum data. This string consists of all data points separated by a comma “,” after each data value, and the string can have a total of ten data points per line. For example, transferring spectrum data of 1001 points require a total of 101 lines. -110.12,
25.8 Writing of Original Compensation Data
When antennas or probes other than the option is used in electric field strength measurement or magnetic field strength measurement, it is necessary to write each original compensation data into the unit. Please write the data into the unit according to the following description. There are two methods for writing. Either use the 2650A PC Software or write a custom program using interface commands. 1) Preparation items * Windows PC (with USB interface) [The writing can be done with 2650A/2651 series unit only.] * 2650A PC software (Refer to 3) below) 2) Example of writing data As an example, the compensation data (antenna gain) of antenna M405 and the compensation data (compensation coefficient) of magnetic field probe CP-2S are shown below. * Compensation data (antenna gain) of antenna M405 Frequency 300MHz 350MHz 400MHz 450MHz 500MHz * Compensation data (compensation coefficient) of magnetic field probe CP-2S Frequency 10MHz 100MHz 1GHz 2GHz 3GHz Compensation * Data of ten points or less can be written. 0Hz is not valid as a data point. 3) Using 2650A PC software for writing 1. Create the text file of compensation data.
Create a new text file, and open it with a text editor. Write the frequency and compensation data in the following format. * Format “frequency”: ”compensation data”, “frequency”: “compensation data”, · · · Example of M405 * Write the unit with a capital letter as it is case sensitive. G (GHz) is also available. 2. Write the text file into the unit with 2650A PC software. Connect the personal computer to 2650A series with the included USB cable and turn on the power of the 2650A/2651 series. From the upper menu of the PC software, select [File] → [Write E/F User Data] in case of electric field strength measurement, or select [File] → [Write M/F User Data] in case of magnetic field strength measurement. When the text file created beforehand is selected, the data is written. * Install the specified USB driver in the personal computer prior to using the PC software. (Refer to “25.3 Installation of Driver”) 4) Method of using original program for writing 1. Prepare the USB communication software. Connect the personal computer to the 2650A/2651 series with the included USB cable and turn on the power of the 2650A/2651 series. 2. Write the data into the unit. Transfer the data in the following format with the USB communication software. * Format In case of electric field strength measurement; EFUSER “frequency”: ”compensation data”, “frequency”: “compensation data”, · · · In case of magnetic field strength measurement; MFUSER “frequency”: ”compensation data”, “frequency”: “compensation data”, · · · Example of CP-2S * Write the unit with a capital letter as it is case sensitive. G (GHz) is also available. 3. When the writing ends correctly, “OK” is returned. 5) How to use Set the measuring function of the 2650A/2651 series to electric field strength measurement mode or magnetic field strength measurement mode. By pushing [MEAS] → [E/F ANT], electric field strength measurement mode is set. By pushing [MEAS] → [M/F PROBE], magnetic field strength measurement mode is set. Select “USER” from antennas or probes on the display by pushing [F1]. Then, the measurement by the written data will begin. * The written data remains even if power is turned off. * The measuring mode returns to a usual measurement mode when turning off power. 6) About antenna gain
In the following context, the antenna gain means an absolute gain. When the antenna gain is a relative gain, it can change into an absolute gain by adding +2.15 dB. Absolute gain [dBi] = Relative gain [dBd] + 2.15 dB As a reference, the conversion formula to electric field strength is indicated below: E = √(480π2 x Pa) / (Ga x λ2)) E: Electric field strength [V/m] Pa: Received electric power [W] Ga: Antenna gain [times] = 10(antenna gain [dBi] ÷ 10) λ : Wavelength [m] = (3 × 108) ÷ frequency [Hz] 26. Option Dipole antenna M401 to M406 Connector : N (P) Refer to “19.4 Electric Field Strength Measurement” for the details. Magnetic field probe CP-2S. Refer to “19.5 Magnetic Field Strength Measurement” for the details. USB printer With AC adaptor and one roll of paper Refer to “23. Storage / Print of Screen Image” for the details. Roll printer (10 rolls) For USB printer Specification Frequency range : 10 MHz to 3 GHz Space resolution : approx.0.25 mm (depending on objects) Dimensions : outside 12 φ×135 mm probe tip 2 mm(W)×1 mm(T) Connectors : SMA(P) Specification Printing method : Thermal line dot method Paper : 80 mm width thermal paper Power source : internal : AA-sized alkaline battery (4 pcs) external : 7.5 VDC/3 A (dedicated AC adapter) Dimensions : 134(W)×60(H)×180(D) mm Weight : approx.450 g (mainframe only)
- Basic Performance Test (2650A/2651/2652A/2658A) To keep the quality of the unit, regular performance testing is recommended. This section describes a method and specification of basic performance testing. If a problem is found in the results of basic performance testing, or formal testing is required,please contact B&K Precision.
27.1 Frequency Characteristics
Adjust the output level of the spectrum analyzer calibration unit (thereafter, “calibration unit”) so that the displayed power value is -15dBm at each frequency for this unit, and measure the absolute value with a receiver for calibration (microwave power meter, etc.). Setting of 2650A/2652A/2651/2658A Specifications Measurement value Judgment Center frequency Frequency span RBW *1) 50 kHz 200 kHz 10 kHz Within Reference±2.6 dB±1dot 100 kHz 200 kHz 30 kHz Within Reference±2.6 dB±1dot 1 MHz 2 MHz 100 kHz Within Reference±2.6 dB±1dot 10 MHz 10 MHz 3 MHz Within Reference±2.0 dB±1dot
100 MHz 10 MHz 3 MHz Reference
1 GHz 10 MHz 3 MHz Within Reference±1.0 dB±1dot 2 GHz 10 MHz 3 MHz Within Reference±1.0 dB±1dot 3.3 GHz 10 MHz 3 MHz Within Reference±1.0 dB±1dot 6.2 GHz *2) 10 MHz 3 MHz Within Reference±1.0 dB±1dot 8.5 GHz *2) 10 MHz 3 MHz Within Reference±1.0 dB±1dot *1) RBW switching error is included at RBW other than 3 MHz. *2) 2658A only
- Setting of 2650A/2652A/2651/2658A · Setting of calibration unit Reference level : -15 dBm Frequency : Same as a center frequency of VBW : 1 MHz 2650A/2652A/2651/2658A. Sweep time : 1 s Detection mode : SMPL Output power : Adjust the power indication of 2650A/2652A/2651/2658A to -15 dBm. Display scale : 2 dB/div Spectrum analyzer calibration unit Receiver for calibration 2650A/2651 series [Connection diagram]
27.2 Accuracy of Reference Level
Adjust the output level of the calibration unit so that the displayed value of this unit is the 0th div from the top, and calibrate the absolute value with the receiver for calibration (microwave power meter, etc.). Setting of 2650A/2652A/2651/2658A Specifications Measurement value Judgment Reference level +10 dBm within ±1.4 dB±1dot 0 dBm within ±1.4 dB±1dot -10 dBm within ±1.4 dB±1dot -15 dBm within ±0.8 dB±1dot -20 dBm within ±1.4 dB±1dot -30 dBm within ±1.4 dB±1dot -40 dBm within ±1.4 dB±1dot * Input attenuator switching error is included at the reference level other than -15 dBm.
- Setting of 2650A/2652A/2651/2658A · Setting of calibration unit Center frequency : 100 MHz Frequency : 100 MHz Frequency span : 10 MHz Output power : Adjust it so that the indicated RBW : 3 MHz value of 2650A/2652A/2651/2658A VBW : 1 MHz is at the 0th div from the top. Sweep time : 1 s Detection mode : SMPL Display scale : 2 dB/div
27.3 Display Accuracy of Center Frequency
Measure the frequency with the peak search function of 2650A/2652A/2651/2658A. Setting of 2650A/2652A/2651/2658A Specifications Measurement value Center frequency Judgment Frequency span Center frequency Frequency span RBW
100 MHz 200 kHz 3 kHz within ±50 kHz±1dot
±RBW x 6%
100 MHz 10 MHz 30 kHz
100 MHz 20 MHz 100 kHz
within ±360 kHz±1dot ±RBW x 20%
100 MHz 200 MHz 100 kHz
1 GHz 20 MHz 100 kHz
2 GHz 20 MHz 100 kHz
3.3 GHz *1 20 MHz 100 kHz
6.1 GHz *2 20 MHz 100 kHz
8.5 GHz *2 20 MHz 100 kHz
*1 2650A/2652A/2651 only *2 2658A only
- Setting of 2650A/2652A/2651/2658A · Setting of calibration unit Reference level : -15 dBm Frequency : Same as a center frequency of 2650A/2652A/2651/2658A. VBW : AUTO Sweep time : 1 s Output power : -15 dBm Detection mode : SMPL * However, calibrate the signal generator in advance. Display scale : 10 dB/div
27.4 Display Accuracy of Frequency Span
Adjust the frequency of the calibration equipment so that the peaks are at the positions of f1 and f9, and measure the frequencies of f1 and f9. Calculate from f1 and f9 the display accuracy of the frequency span. *1 2650A/2652A/2651 only *2 2658A only
- Setting of 2650A/2652A/2651/2658A · Setting of calibration unit Reference level : -15 dBm Frequency : Adjust it to the positions of f1 VBW : AUTO and f9. Sweep time : 1 s Output power : -15 dBm Detection mode : SMPL Display scale : 10 dB/div
27.5 Linearity of Amplitude Axis
Adjust the level of the calibration unit so that the peak is at the top of the amplitude axis (0th div), and regard the point set at that time as the reference. Gradually lower the output, starting from the reference, and measure the amplitude value of 2650A/2652A/2651/2658A. Setting of 2650A/2652A/2651/2658A Output of calibration unit Specifications Measurement value Judgment Display scales 10 dB/div XdBm (adjust it to the 0th div) Reference(-15 dBm) X-10 dBm Within -25 dBm±0.8 dB±1dot 5 dB/div XdBm (adjust it to the 0th div) Reference(-15 dBm) X-5 dBm Within -20 dBm±0.4 dB±1dot 2 dB/div XdBm (adjust it to the 0th div) Reference(-15 dBm) X-2 dB Within -17 dBm±0.2 dB±1dot
- Setting of 2650A/2652A/2651/2658A · Setting of calibration unit Setting of 2650A/2652A/2651/2658A Specifications Measurement value Measurement value (f9 - f1) Judgment Frequency span Center Frequency RBW 200 kHz 100 MHz 3 kHz within ±180 kHz ×3%±1dot
10 MHz 100 MHz 100 kHz within ±8 MHz ×3%±1dot
20 MHz 100 MHz 300 kHz within ±18 MHz ×3%±1dot
200 MHz 100 MHz 3 MHz within ±180 MHz ×3%±1dot
500 MHz 1 GHz 3 MHz within ±400 MHz ×3%±1dot
2 GHz 1 GHz 3 MHz within ±1.8 GHz ×3%±1dot FULL(3.3 GHz)*1 1.65 GHz 3 MHz within ±2.64GHz ×3%±1dot 2 GHz *2 4.8 GHz 3 MHz within ±1.8 GHz ×3%±1dot 2 GHz *2 7.4 GHz 3 MHz within ±1.8 GHz ×3%±1dot FULL(8.5 GHz)*2 4.25 GHz 3 MHz within ±7.65 GHz ×3%±1dot * f1: 1st div from the left on the spectrum display f9: 9th div from the left on the spectrum display
Center frequency : 100 MHz Frequency : 100 MHz Reference level : -15 dBm Frequency span : 10 MHz RBW : 3 MHz VBW : 1 MHz Sweep time : 1 s Detection mode SMPL
Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the product in the original packaging with proof of purchase to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Non-Warranty Service: Please go the support and service section on our website www.bkprecision.com to obtain a RMA #. Return the product in the original packaging to the address below. Clearly state on the RMA the performance problem and return any leads, probes, connectors and accessories that you are using with the device. Customers not on an open account must include payment in the form of a money order or credit card. For the most current repair charges please refer to the service and support section on our website. Return all merchandise to B&K Precision Corp. with pre-paid shipping. The flat-rate repair charge for Non-Warranty Service does not include return shipping. Return shipping to locations in North America is included for Warranty Service. For overnight shipments and non-North American shipping fees please contact B&K Precision Corp. B&K Precision Corp. Yorba Linda, CA 92887 www.bkprecision.com 714-921-9095 Include with the returned instrument your complete return shipping address, contact name, phone number and description of problem. LIMITED THREE-YEAR WARRANTY B&K Precision Corp. warrants to the original purchaser that its products and the component parts thereof, will be free from defects in workmanship and materials for a period of three years from date of purchase. B&K Precision Corp. will, without charge, repair or replace, at its option, defective product or component parts. Returned product must be accompanied by proof of the purchase date in the form of a sales receipt. To obtain warranty coverage in the U.S.A., this product must be registered by completing a warranty registration form on our website www.bkprecision.com within fifteen (15) days of purchase. Exclusions: This warranty does not apply in the event of misuse or abuse of the product or as a result of unauthorized alterations or repairs. The warranty is void if the serial number is altered, defaced or removed. B&K Precision Corp. shall not be liable for any consequential damages, including without limitation damages resulting from loss of use. Some states do not allow limitations of incidental or consequential damages. So the above limitation or exclusion may not apply to you. This warranty gives you specific rights and you may have other rights, which vary from state-to-state. B&K Precision Corp. Yorba Linda, CA 92887 www.bkprecision.com 714-921-9095
Yorba Linda, CA 92887 www.bkprecision.com © 2009 B&K Precision Corp. Printed in Japan