©2019 Advanced Energy Industries, Inc. The Mikron® M310-HT is a portable blackbody calibration source utilizing a digital indicating temperature controller that may be set to any temperature between ambient 5 to 450°C (9 to 842°F). A precision platinum RTD temperature sensor is embedded in the blackbody emitter, providing high accuracy and repeatability. The temperature controller uses the industry standard PID algorithms to control the emitter temperature to within 0.3°C of the set points. The blackbody emitter mechanism uses a resistive heating device that provides a long life, short stabilization times, and stable temperature control. PRODUCT HIGHLIGHTS Excellent general purpose calibration High effective emissivity 1.00 @ 8 to 14 µm High accuracy, high resolution Excellent stability ±0.3°C per 8 hour period Manufactured and tested to meet rigid quality control standards Furnished with certificate of calibration traceable to NIST RS232 serial communication included TYPICAL APPLICATIONS Infrared temperature sensors Infrared thermal imaging systems Spectrographic analyzers Radiometers Flux meters MIKRON M310-HT Portable, compact low temperature blackbody calibration source with a large surface area. Ambient +5 to 450°C (+9 to 842°F). Temperature Range Ambient +5 to 450°C (+9 to 842°F) Measurement Uncertainty 0.25% of reading ±1°C Emissivity 1.00 effective emissivity @ 8 to 14 μm Heated Emitter Shape Flate plate Aperture Diameter 76 mm (3.00 in) Average Warm-Up Time < 30 min from ambient (to 400°C) AT A GLANCE
2 advancedenergy.com MIKRON M310HT Measurement Specifications Temperature Range Ambient +5 to 450°C (+9 to 842°F) Temperature Uncertainty 0.25% of reading ±1°C Display Accuracy vs. NIST Calibration See supplied NIST calibration report Temperature Resolution 0.1°C Stability 0.3°C per 8-hour period Source Non-Uniformity Approximately ±1°C @ 250°C or ±2°C @ 400°C Heated Cavity Shape Flat plate Exit Port Diameter 76 mm (3.00 in) Emissivity ε 1.00 effective emissivity @ 8 to 14 µm Standard Calibration Method Radiometric at 8 to 14 µm Temperature Sensor Precision platinum RTD Warm-up Time < 30 minutes from ambient to 400°C Slew Rate to 1°C Stability Approximately 6 min for a + 50°C change Slew Rate to 0.1°C Stability Approximately 10 min for a +50°C change Environmental Specifications Operating Ambient Temp 0 to 44°C (30 to 110°F) Cooling Fan cooled, air inlet on rear panel Operating Humidity 90% RH max, non-condensing Dimensions (H x W x D) 207.3 mm x 280.4 mm x 266 mm (8.2" x 11" x 10.5") Weight 5.6 kg (12.4 lbs) CE Certified Yes OVERVIEW TECHNICAL DATA Blackbody calibration sources are infrared radiators used for calibrating and verifying the output signals of infrared thermometers (pyrometers), thermal imaging systems, heat flux measurement systems, or spectrographic analysis systems. Advanced Energy supplies a unique selection of very precise calibration sources that are traceable to national standards. Quotations for custom designs and variations are available upon request. Mikron calibration sources have long been the gold standard to calibrate the instruments that keep your operations up and running. These blackbodies are superior because of the emissivity values, homogeneous emission areas, and a wide range of different sized apertures to adapt to the desired target area. In addition, fast heat-up times and high temperature stability are guaranteed. The quality of our calibration sources is guaranteed by tests, burn-in times, and radiometric calibrations. On most models, a certificate is provided to document the traceability to the international temperature scale ITS90 and NIST. 1 Accuracy calibration performed radiometrically, the uncertainty of emissivity and transfer standard are already included. 2 Provided stable AC mains voltage and minimum air flow across the exit port or emitter plate.
advancedenergy.com 3 MIKRON M310HT Communication and Electrical Specifications Remote Set Point Via RS232 Method of Control Digital PID controller Power Requirements 115 VAC ±10% 50 and 60 Hz 920 VA fused max, 600 W heater (230 VAC Optional) TECHNICAL DATA (CONTINUED) REFERENCE NUMBERS PN Description 14760-1-1-1-2-0-2 M310-HT, Ambient 5 to 450°C, 76 mm, 115 VAC @ 50 and 60 Hz, RS232, North American cable 14760-2-2-1-2-1-2 M310-HT, Ambient 5 to 450°C, 76 mm, 230 VAC @ 50 and 60 Hz, RS232, German cable
ENG-M310HT-235-01 10.19 DS TITLE For international contact information, visit advancedenergy .com. sales.support@aei.com +1 970 221 0108 ABOUT ADVANCED ENERGY Advanced Energy (AE) has devoted more than three decades to perfecting power for its global customers. AE designs and manufactures highly engineered, precision power conversion, measurement and control solutions for mission-critical applications and processes. AE’s power solutions enable customer innovation in complex semiconductor and industrial thin film plasma manufacturing processes, demanding high and low voltage applications, and temperature-critical thermal processes. With deep applications know-how and responsive service and support across the globe, AE builds collaborative partnerships to meet rapid technological developments, propel growth for its customers and power the future of technology. Specifications are subject to change without notice. Not responsible for errors or omissions. ©2019 Advanced Energy Industries, Inc. All rights reserved. Advanced Energy®, Mikron®, and AE® are U.S. trademarks of Advanced Energy Industries, Inc.