LCS2 MSYSTEM | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 3
Technical content
MODEL: LCS2 LCS2 SPECIFICATIONS ES-1362 Rev.2 Page 1/3 http://www.m-system.co.jp/ Plug-in Signal Conditioners M-UNIT STRAIN GAUGE TRANSMITTER (remote sensing; isolated) Functions & Features
- Provides a DC output signal compatible with a bridge type strain gauge utilized in load cells, pressure transducers
- Six-wire remote sensing compensates for lead resistance variations
- Excitation adjustable from 2 V to 5 V
- Wide-range adjustment: 0 – 80 % for zero, 100 – 20 % for span
- High-density mounting Typical Applications
- Weighing system for tanks, hoppers, silos
- Weighing system using cranes
- Float level meter utilizing strain gauges (1.97) 136 (5.35) (3.15) mm (inch) MODEL: LCS2–[1][2]–[3][4]
ORDERING INFORMATION
- Code number: LCS2-[1][2]-[3][4] Specify a code from below for each of [1] through [4]. (e.g. LCS2-2A-B/E/K/Q)
- Special output range (For codes Z & 0)
- Specify the specification for option code /Q (e.g. /C01/S01) [1] INPUT STRAIN GAUGE 1: 1 mV/V 12: 1.25 mV/V 15: 1.5 mV/V 2: 2 mV/V 3: 3 mV/V 4: 4 mV/V 5: 5 mV/V 6: 10 mV/V 7: 20 mV/V 0: Specify (strain gauge and excitation) [2] OUTPUT Current A: 4 – 20 mA DC (Load resistance 750 Ω max.) B: 2 – 10 mA DC (Load resistance 1500 Ω max.) C: 1 – 5 mA DC (Load resistance 3000 Ω max.) D: 0 – 20 mA DC (Load resistance 750 Ω max.) E: 0 – 16 mA DC (Load resistance 900 Ω max.) F: 0 – 10 mA DC (Load resistance 1500 Ω max.) G: 0 – 1 mA DC (Load resistance 15 kΩ max.) Z: Specify current (See OUTPUT SPECIFICATIONS) Voltage 1: 0 – 10 mV DC (Load resistance 10 kΩ min.) 2: 0 – 100 mV DC (Load resistance 100 kΩ min.) 3: 0 – 1 V DC (Load resistance 100 Ω min.) 4: 0 – 10 V DC (Load resistance 1000 Ω min.) 5: 0 – 5 V DC (Load resistance 500 Ω min.) 6: 1 – 5 V DC (Load resistance 500 Ω min.) 4W: -10 – +10 V DC (Load resistance 2000 Ω min.) 5W: -5 – +5 V DC (Load resistance 1000 Ω min.) 0: Specify voltage (See OUTPUT SPECIFICATIONS) [3] POWER INPUT AC Power B: 100 V AC C: 110 V AC D: 115 V AC F: 120 V AC G: 200 V AC H: 220 V AC J: 240 V AC DC Power R: 24 V DC [4] OPTIONS (multiple selections) Input Signal Indicator blank: Without /E: With (0.0 - 100.0 % display) Response Time (0 – 90 %) blank: Standard (≤ 0.5 sec.) /K: Fast Response (Approx. 25 msec.) Other Options blank: none /Q: Option other than the above (specify the specification) SPECIFICATIONS OF OPTION: Q (multiple selections) COATING (For the detail, refer to M-System's web site.) /C01: Silicone coating /C02: Polyurethane coating /C03: Rubber coating TERMINAL SCREW MATERIAL /S01: Stainless steel
MODEL: LCS2 LCS2 SPECIFICATIONS ES-1362 Rev.2 Page 2/3 http://www.m-system.co.jp/ GENERAL SPECIFICATIONS Construction: Plug-in Connection: M3.5 screw terminals Screw terminal: Chromated steel (standard) or stainless steel Housing material: Flame-resistant resin (black) Isolation: Input to output to power Overrange output: Approx. -10 to +120 % at 1 – 5 V Excitation adjustment: 2 – 5 V (front) limited to ≤ 10 V when the remote sensing terminals are open. Zero adjustments (tare): 0 – 80 % (front) (Excitation voltage: factory default) Span adjustment: 100 – 20 % (front) (Excitation voltage: factory default) ■ DISPLAY (Input indicator) LCD digital display: 0.0 - 100.0 % (min. digit 0.1 %) (No scaling) INPUT SPECIFICATIONS Input: Bridge voltage from load cells Allowable leadwire resistance: ≤ Total Strain Gauge Resistance × 0.25
- Strain Gauge Rated output from strain gauge: 1 – 20 mV/V; Input to the transmitter must be over 3 mV.
- Excitation: 2 – 5 V adjustable (5 V standard) Maximum current: 65 mA OUTPUT SPECIFICATIONS ■ DC Current: 0 – 20 mA DC Minimum span: 1 mA Offset: Max. 1.5 times span Load resistance: Output drive 15 V max. ■ DC Voltage: -10 – +12 V DC Minimum span: 5 mV Offset: Max. 1.5 times span Load resistance: Output drive 10 mA max.; 5 mA for negative voltage output; at ≥ 0.5 V INSTALLATION Power input
- AC: Operational voltage range: rating ±10 %, 50/60 ± 2 Hz, approx. 5 VA
- DC: Operational voltage range: 24 V ±10 %, approx. 150 mA, ripple 10 %p-p max. Operating temperature: -5 to +60°C (23 to 140°F) Operating humidity: 30 to 90 %RH (non-condensing) Mounting: Surface or DIN rail Weight: 430 g (0.95 lb) PERFORMANCE in percentage of span Accuracy: ±0.1 % (input ≥ 3 mV) Display accuracy: ± (0.1 % of FS + 1 digit) (input ≥ 3 mV) Temp. coefficient: ±0.015 %/°C (±0.008 %/°F) (input ≥ 3 mV) Line voltage effect: ±0.1 % over voltage range Insulation resistance: ≥ 100 MΩ with 500 V DC Dielectric strength: 2000 V AC @1 minute for AC power
1000 V AC @1 minute for DC power
(input or output to power)
1000 V AC @1 minute (input to output)
2000 V AC @1 minute
(input or output or power to ground)
MODEL: LCS2 LCS2 SPECIFICATIONS ES-1362 Rev.2 Page 3/3 http://www.m-system.co.jp/ EXTERNAL VIEW Span Adj. Zero Adj. Excitation Adj. LCD Meter EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch) 5678 2111 10 80 (3.15) 80 (3.15) (.79) 40 (1 .57) 50 (1 .97) 7 .8 (.31) CLAMP (top & bottom) DIN RAIL 35mm wide2–4.5 (.18) dia. MTG HOLE 25 (.98) deep 11 –M3.5 SCREW
- When mounting, no extra space is needed between units. SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM Isolation Base Socket Output Driver OUTPUT Z S STRAIN GAUGE Close across the terminal 3 – 4 and 5 – 9 when the remote sensing is not used. * Option /E Low Drift DC Volt. Stabilizer Low Drift Amplifier EXC. S S Remote Sensing U(+) V(–) POWER Specifications are subject to change without notice.