GC52 Differential Pressure Transmitter Data Sheet Data Sheet All specifications are subject to change without notice. All sales subject to standard terms and conditions. Ashcroft® and Trust the Shield® are trademarks of Ashcroft Inc. The following non-Ashcroft trademarks are the property of their respective owners: Viton™. For more information, see Ashcroft Brands & Trademarks ©2022 Ashcroft Inc. GC-52_transmitter_ds_RevC_10-19-22 ashcroft.com | 1.800.328.8258 | Contact Us
FEATURES
Bright backlit 4-digit LCD display Scaling function allows display to indicate user defined physical units Internal “push-button” configurability allows quick range changes “Loop Check” function allows unit to output 4-20 mA without applying pressure Linear flow output and totalization via square root extraction TYPICAL USES Pressurized & non-pressurized tank levels Flow measurement (liquid/gas) Pollution monitoring equipment KEY BENEFITS
- Up to 8X smaller than a conventional process transmitter
- Compact, robust design
- Easily rotatable display, 90° increments
- Robust NEMA 4 (IP66) aluminum die cast housing PERFORMANCE SPECIFICATIONS Reference Temperature: 73 °F (23 °C) Accuracy: ±0.50% of span (URL) Includes the effects of linearity, hysteresis, and repeatability Stability: ±0.25% of span (URL)/year Output Resolution: 0.1% of span (URL) Model: 4 digit, 10 mm LCD with LED backlight ENVIRONMENTAL SPECIFICATIONS Temperature Effects: (–10 °C to 60 °C) ±0.03% FS/C° Temperature Limits: Storage: Operating: Compensated: 5 °F to 150 °F (–15 °C to 65 °C) 14 °F to 140 °F (–10 °C to 60 °C) 14 °F to 140 °F (–10 °C to 60 °C) FUNCTIONAL SPECIFICATIONS Static (Line) Pressure Pressure Range:
4 IWC to ≤400 IWC
Proof: 300 psi Burst: 800 psi Single Side (Differential) Pressure Range: ≤8 IWC, ±4 IWC ≥20 IWC, ±8 IWC Proof: 30 psid 100 psid Burst: 130 psid 130 psid Static (Line) Pressure Effects Pressure Range: ≥20 IWC, ±8 IWC
8 IWC, ±4 IWC
4 IWC
Effect: ±0.3% Range/100 psi (URL) ±0.7% Range/100 psi (URL) ±1.5% Range/100 psi (URL) Response Time: 100 ms (user adjustable) Vibration: 5 g’s 150 Hz Shock Effect: 10 g’s 16 ms ELECTRICAL SPECIFICATIONS Output Signal: 4-20 mA (2 Wire) Supply Voltage: 12-32 Vdc Rangeablility/ Adjustment: Zero –10% to 110% Span Span –10% to 110% Span (Accuracy and output resolution based upon full scale (URL) value) Insulation Resistance: 50 Vdc (>100 MΩ) EMC Compliance: EMC Directive 2004/108/EC IEC/EN 61326-1: 2006 (EMI Class A/ EMS Table 2) IEC/EN 61326-2-3: 2006 ((Annex BB (Pressure Transducer)) GC52 Pressure Transmitter
GC52 Differential Pressure Transmitter Data Sheet Data Sheet All specifications are subject to change without notice. All sales subject to standard terms and conditions. Ashcroft® and Trust the Shield® are trademarks of Ashcroft Inc. The following non-Ashcroft trademarks are the property of their respective owners: Viton™. For more information, see Ashcroft Brands & Trademarks ©2022 Ashcroft Inc. GC-52_transmitter_ds_RevC_10-19-22 ashcroft.com | 1.800.328.8258 | Contact Us PHYSICAL SPECIFICATIONS Weight: Approx. 1.0 lb. Environmental Rating: IP66/NEMA 4 Mounting: Mounting bracket included Pressure Connection: 1/4 NPT Female Electrical Connection (External): ½ NPT Female Conduit Cable Gland (Cable diameters 0.35˝ to 0.47˝) 1,000 750 500 250 03 22010 1,020 Load Limitations 4-20mA Output Only 12 24 545 OPERATING REGION Loop Resistance (V) LOOP SUPPLY VOLTAGE Vmin = 12V+[0.022A*RL] *Includes a 10% safety factor R L = RS + RW RL = Loop Resistance (ohms) RS = Sense Resistance (ohms) RW = Wire Resistance (ohms) WETTED MATERIAL Diaphragm Process Connection Media Compatibility
316 SS, Viton™ &
Fluids and gases compatible with 316 SS, Viton™ and Alumina Ceramic NON-WETTED Enclosure Aluminum, epoxy coated OPTIONAL FEATURES Calibration Report: 9 point traceable calibration certificate Pressure Connection: 54 mm conversion joint for 3-way manifold LOAD LIMITATIONS 4-20 mA OUTPUT ONLY ORDERING CODE Example: GC52 7 F02 42 CG 4IWL – XRH Model GC52 - Rangeable wet/wet differential pressure transmitter GC52 Accuracy 7 - ±0.50% of span 7 Pressure Fitting F02 - 1/4 NPT Female F02 Output Signal 42 - 4-20 mA Output signal 42 Electrical Connection CG - Cable gland CG CD - ½ NPT Female conduit Pressure Range Bidirectional 4IWL - ±4 IWC 4IWL 8IWL - ±8 IWC 20IWL - ±20 IWC 40IWL - ±40 IWC 80IWL - ±80 IWC 200IWL - ±200 IWC Differential 4IW - 0-4 IWD 8IW - 0-8 IWD 20IW - 0-20 IWD 40IW - 0-40 IWD 80IW - 0-80 IWD 200IW - 0-200 IWD 400IW - 0-400 IWD Option (if including an option(s) must include an “X’’) X__ RH - 9 pt. Traceable calibration certificate RH
GC52 Differential Pressure Transmitter Data Sheet Data Sheet All specifications are subject to change without notice. All sales subject to standard terms and conditions. Ashcroft® and Trust the Shield® are trademarks of Ashcroft Inc. The following non-Ashcroft trademarks are the property of their respective owners: Viton™. For more information, see Ashcroft Brands & Trademarks ©2022 Ashcroft Inc. GC-52_transmitter_ds_RevC_10-19-22 ashcroft.com | 1.800.328.8258 | Contact Us Dimension Drawings Dimensions in inches 1.0 4.49 3.62 2.91 2.28 4.72 5.12 1.97 2.36 0.69 2 x Ø 0.06 Drain 4 x Ø 0.22 Panel Mounting Holes Bracket t = 0.08 Ø 2.56 0.422.07 Drain Outlet Equalizing Valve High Pressure Port NPT FemaleLow Pressure Port NPT Female 0.82 1.42 Electrical Connection PG 13.5 threaded housing, factory installed options include cable gland or ½ NPTF conduit connection inH2O GC52 3.37 0.43 0.91 1.14 0.12 (2 locations) (4 locations) 1.62 2.13 2.41 1.24 -20UNF (4 locations) 2.17 DIMENSIONS 54mm Conversion Joint Body Part Number 141A444-01