CS220 CAMPBELL | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 2

Technical content

5] SCIENTIFIC \\_/ WHEN MEASUREMENTS MATTER Overview The C5220 is a thermocouple that measures the temperature of a surface by direct contact. It typically monitors the temperature of a solar panel, but it can also monitor the Benefits and Features ) Well-suited for solar energy applications ) Thermocouple tolerance meets ASTM E230-ANSI MC 96.1 (reference junction at 0°C) Technical Description The C5220 uses a 24-AWG, twisted pair, shielded, fast response type E thermocouple. Type E thermocouples consist of a chromel wire and a constantan wire joined at a measurement junction. A voltage potential is generated when the measurement end of the thermocouple is at a different temperature than the reference end of the thermocouple. The magnitude of the voltage potential is related to the temperature difference. Therefore, temperature can be determined by measuring the differences in potential created at the junction of the two wires. A reference temperature measurement is required for thermocouple measurements. The temperature sensor built Specifications Sensor Type E thermocouple For comprehensive details, visit: www.campbellsci.eu/cs220 o ot ([ogtahi= €5220 Surface-Mount Type E Thermocouple Sensor Fast Response Ideal for monitoring solar panel temperatures temperature of other devices. This thermocouple easily interfaces with our dataloggers, and it is ideal for solar energy applications. ) Easy to install ) Compatible with most Campbell Scientific data loggers into many of our dataloggers' wiring panel typically provides this measurement. The C5220 consists of a stainless-steel overbraided cable and a thermocouple extension cable. Adhesive on one end of the overbraided cable secures the sensor to the solar panel. The mini connectors of the thermocouple extension and overbraided cables attach to each other. The pigtails of the extension cable connect to the datalogger. The C5220 can provide the temperature on the back of a solar panel, which is an important measurement in solar energy applications. The output of a solar panel is affected by its temperature. As the temperature of the solar panel increases, its output decreases. Measurement Description Back-of-module temperature

Operating Temperature Up to 260°C Accuracy Refer to the Thermocouple Range Measurementsection in the data logger manual. Thermocouple Type Chromel-Constantan Typical Output 60 uV/°C Response Time 0.15s Thermocouple Tolerances Meets ASTM E230-ANSI MC 96.1 Sensitivity +C Special Limits of 1.0°C or 0.4% Length of Stainless-Steel 1.0 m (3.3 ) (010 900°0). Braided Cable Reference junction at 0°C. Thermocouple Length 254cm (1.00in) Maximum Temperature of - Adheres up to 260°C. Thermocouple Width 191 cm 0751in) Adhesive Weight 2389 (84 0z) with 15.24-m (50- ft) cable For comprehensive details, visit: www.campbellsci.eu/cs220 CAMPBELL 80 Hathern Road, Shepshed, LE12 9GX UK | +(0)1509 828888 | sale@campbellsci.couk | www.campbellsci.eu SCIENTIFIC AUSTRALIA | BRAZIL | CANADA | CHINA | COSTA RICA | FRANCE | GERMANY | INDIA | SOUTH AFRICA | SPAIN | THAILAND | UK | USA ©2020 Campbell Scientific, Ltd. | 07/28/2020