CN400 OMEGA | Alldatasheet
Document overview
- Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
- PDF pages: 2
Technical content
Features
- 1/16 DIN Dual Display
- Single/Dual Universal Inputs
- Fast Speed up to 470 Samples/Sec
- Programmable by NFC/RFID
- Supports Heating/Cooling P.I.D
- Universal Power Supply Option
- UL Certified CN400 Series PID Process Controller The CN40x series includes 5 versions: Models with power supply 24..230 VAC/VDC ±15% 50/60 Hz – 6 Watt/VA CN401-11445 1 analogue input + 2 relays 2 A + 2 SSR + 2 D.I. + 1 analogue output V/mA CN401-11445-C4 1 analogue input + 2 relays 2 A + 2 SSR / D.I. + 1 analogue output V/mA + RS485 CN401-111445 1 analogue input + 3 relays 2 A + 2 SSR + 2 D.I. + 1 analogue output V/mA (1 D.O. + 1 D.I. when using CT) Model with power supply 24 VAC/VDC ±15% 50/60 Hz – 6 Watt/VA CN402-1114455- C4-DC 2 analogue input + 3 relays 2 A + 2 SSR + 2/4 D.I. + 2 analogue output V/mA + RS485 + CT (2 D.I./D.O when using CT) Model with power supply 115..230 VAC ±15% 50/60 Hz – 6 Watt/VA CN402- 1114455-C4 2 analogue input + 3 relays 2 A + 2 SSR + 2/4 D.I. + 2 analogue output V/mA + RS485 + CT (2 D.I./D.O when using CT)
1 or 2 - Configurable: 16 bit, selectable for TC type K, S, R, J, T, N, B (automatic compensation of cold junction) -25 to 85°C, ±0,2% F.S. ±1 digit F.S.), thermoresistances PT100, PT500, PT1000, Ni100, PTC1K, NTC10K (β 3435K), process signals 0 to 10 V (50000 points), 0/4 to 20mA (40000 points), 0 to 60 mV (25000 points), potentiometer 1 to 150 KΩ (50000 points) Sampling time: Programmable up to 2.1 ms (frequency up to 470 Hz) 2/4 digital inputs: Setpoint change, Hold, Run, Tuning launch, Start / Stop, Lock configuration 1 Current Transformer (C.T.) input: Selection C.T. 50 mA, 800 μs - 4096 points Inputs Outputs 2/3 relays: Relay 2 A - 250 V AC resistive change Software Features Control algorithms: proportional Tuning: Manual or automatic Data protection: Lock of control / alarm setpoint / Access to parameters by password Alarm modes: Absolute / Threshold, Band, High / Low deviation. Alarm with optional Manual reset. Loop Break Alarm Auto / Manual function: Output percentage command also with automatic change in case of sensor failure Double P.I.D: Heating / Cooling P.I.D Programmer function: Pre-programmed cycle / 3 steps Soft-Start: Rising gradient expressed as Degrees / Hour or fixed output percentage Open / Close logic: Open / Close logic for motorized valves Box: 48 x 48 (front panel) x 105 mm Power supply: 24 to 230 V AC / DC ±15% 50/60 Hz - galvanic isolation 2.5KV Consumption: 8 W Display: 4 digits 0.5” white + 4 digits 0.3” red Operating conditions: Temperature 0 to 45 °C, humidity 35 to 95 RH% MaterialBox: PC UL94V2 self-extinguishing, front panel: PC UL94V2 Weight: Approx. 185 g Sealing: IP65 (front panel) IP20 (box and terminal blocks) Quick set-up options: Memory Card, software LABSOFTVIEW, or EASY-UP APP: NFC Programming via APP Direct Link for Android smartphones
2 SSR:
1/2 analog: Selection 4 to 20 mA (40000 points ± 0,2% F.S.) or 0 to 10 V DC (40000 points ± 0,2% F.S.) for command or retransmis- sion PV / SPV Serial communication: RS485 Modbus RTU - Slave (4800 to 115200 baud code-T)