MultiDimension Technology Co., Ltd. Page 1 of 5 Web: www.dowaytech.com/en , Email: info@dowaytech.com MDT USB2705A No.7 Guangdong Road, Zhangjiagang Free Trade Zone, Jiangsu, 215634, P.R. China Ver. 1.0a CALIBRA TION MAGNET CAL01 CAL01 – 800 GAUSS CALIBRATION MAGNET Features and Benefits /square6 High Field Magnetometer Calibration /square6 Calibrates Transverse and Axial Probes /square6 High Degree of Stability /square6 Accuracy Retained Between 0° C and 40° C /square6 Durable Nylon Housing /square6 Special Field Values Quoted on Request

Applications

/square6 USB2510x Magnetometer Calibration /square6 STEM Education /square6 Hobby and DIY General Description The CAL01 reference magnet provides a stable and co nvenient method to verify the accuracy and calibration of the USB 2510x series magnetometers. The CAL01 reference magnet is composed of two axially magnetized permanent magnet rings th at allow a probe to access the working region of the fixture from both axial and transverse orien tations, making the calibration fixture useful for both axial and transverse probe calibration. Unlik e conventional reference magnets that rely on placing a probe in a fixed position, which produce an error if the magnet hole is improperly sized or the probe is not shimmed correctly, the CAL01 is de signed so that magnet can be moved with respect to the probe in order to allow the user to find the positive and negative peak magnetic field values, which are used to compute the calibration f actors. When used in combination with MultiDimension’s graphical data logging user interf ace, this technique minimizes errors from probe misalignment and permits rapid accurate calibration of the USB2510x magnetometer probes. The plots below show the peak field values when moving an axial probe through the axial hole and when rotating the magnet around a transverse probe in the transverse hole. CAL01 IS NOT FOR USE WITH THE USB2705 MAGNETOMETER.

MultiDimension Technology Co., Ltd. Page 2 of 5 Web: www.dowaytech.com/en , Email: info@dowaytech.com MDT USB2705A No.7 Guangdong Road, Zhangjiagang Free Trade Zone, Jiangsu, 215634, P.R. China Ver. 1.0a Hardware Description The CAL01 reference magnet is composed of two axial ly magnetized permanent magnet rings that allow a probe to access the working region of the f ixture from both axial and transverse orientations, making the calibration fixture useful for both axial and transverse probe calibration. This forms a region of uniform magnetic field betwe en the rings, where provided the motion of the probe is confined to this region, provides a well defined peak magnetic field value. The housing of the magnets is composed of nylon, and it has guide holes for inserting the probe.

MultiDimension Technology Co., Ltd. Page 3 of 5 Web: www.dowaytech.com/en , Email: info@dowaytech.com MDT USB2705A No.7 Guangdong Road, Zhangjiagang Free Trade Zone, Jiangsu, 215634, P.R. China Ver. 1.0a Transverse Probe Calibration Procedure 1. Press the “calibrate to field” button. 2. Set the calibration field value to the magnetic field value indicated on the side of the calibration assembly 3. Make sure both sensitivity and offset checkboxes are selected 4. Insert transverse probe all the way into the transverse probe alignment hole. 5. Reset the chart in the calibration window 6. Rotate the magnet in order to sample the largest positive voltage and the largest negative voltage. The probe can be wiggled and the magnet r otated back and forth around the field extrema angles in order to sample the maximum and minimum values. 7. Save the calibration in either the flash for future use or just write the calibration to the sensor for immediate use only 8. Exit the calibration window 9. Enter the digitizer window and turn on the low pass filter (LPF) if desired. 10. If the offset has not been adequately corrected, then perform one of the offset calibration procedures Motion of the Magnet and Probe MultiDimension Technology Graphical Display

MultiDimension Technology Co., Ltd. Page 4 of 5 Web: www.dowaytech.com/en , Email: info@dowaytech.com MDT USB2705A No.7 Guangdong Road, Zhangjiagang Free Trade Zone, Jiangsu, 215634, P.R. China Ver. 1.0a Axial Probe Calibration Procedure 1. Press the “calibrate to field” button. 2. Set the calibration field value to the magnetic field value indicated on the side of the calibration assembly 3. Make sure both sensitivity and offset checkboxes are selected 4. Push the axial probe into the axial alignment hole to find the maximum positive (negative) voltage reading, you can wiggle the probe some and move in and out of the calibration assembly to verify the maximum value has been sampled. 5. Hold the probe steady and pull the calibration assembly away. 6. Flip the calibration magnet in order to reverse the magnetic field 7. Push the axial probe into the axial alignment hole to find the maximum negative (positive) voltage reading, you can wiggle the probe some and move in and out of the calibration assembly to verify the maximum value has been sampled. 8. Save the calibration in either the flash for future use or just write the calibration to the sensor for immediate use only 9. Exit the calibration window 10. Enter the digitizer window and turn on the low pass filter (LPF) if desired. 11. If the offset has not been adequately corrected, perform one of the offset calibration procedures Motion of the Magnet and Probe MultiDimension Technology Graphical Display

MultiDimension Technology Co., Ltd. Page 5 of 5 Web: www.dowaytech.com/en , Email: info@dowaytech.com MDT USB2705A No.7 Guangdong Road, Zhangjiagang Free Trade Zone, Jiangsu, 215634, P.R. China Ver. 1.0a Offset Calibration Procedure There are several methods that may be used.

  • Autozero the probe with the sensing axis aligned east-west
  • Autozero the probe inside a zero gauss chamber
  • Manual offset
  • Calibration method to permanently change the offset Calibration Method 1. Press the “calibrate to field” button. 2. Make sure the sensitivity checkbox is not selected 3. Make sure the offset checkbox is selected 4. Place the sensor in a zero gauss chamber OR align the sensing axis along the east-west direction 5. Reset the chart in the calibration window 6. Acquire several seconds of voltage vs time data, verifying that the background value is stable 7. Save the calibration in either the flash for future use or just write the calibration to the sensor for immediate use only 8. Exit the calibration window 9. Enter the digitizer window and turn on the low pass filter (LPF) if desired. MultiDimension Technology Graphical Display The information provided herein by MultiDimension T echnology Co., Ltd. (hereinafter MultiDimension) is believed to be accurate and reliable. Publication neither conveys nor implies any license under patent or other industrial or intellectual property rights. MultiDimension reserves the right to make changes to product specifications for the p urpose of improving product quality, reliability, a nd functionality. MultiDimension does not assume any liability arising out of the application and use of its products. MultiDimension’s custome rs using or selling this product for use in appliances, devices, or systems where malfunction c an reasonably be expected to result in personal inj ury do so at their own risk and agree to fully indemnify MultiDimension for any damages resulting from such applications. “MultiDimension”, “MultiDi mension Sensing the Future”, and “MDT” are registered trademarks of MultiDimension T echnology Co., Ltd.