Technical content
5] SCIENTIFIC VHEN MEASUREMENTS MATTER Overview The AP200is a complete, integrated CO, and H,0 atmospheric profile system. It measures carbon dioxide (CO,) and water vapour (H,0) concentration from up to eight intakes, which are normally spaced along the height of Benefits and Features ) Fully integrated system ) Requires only 13 W (average at 25°C and 12 Vdc) ) Contains a Campbell Scientific CR1000X Measurement and Control Datalogger Technical Description Measurements Primary: ) CO;, concentration at each intake ) H,0 concentration at each intake ) System diagnostic word Secondary: ) Sample flow rate Specifications System Enclosure Operating Temperature ~ -30°to +45°C SL(REmE AP200 CO2/H20 Atmospheric Profile System Complete, Integrated System Preprogrammed and low power a tower to give a vertical profile. The AP200 is often used in conjunction with an eddy covariance system to measure the storage term and give a more complete measure of the surface gas exchange. ) Automatically performs CO, zero and span ) Automated temperature and pressure control ) Data logger program included ) Sample cell pressure ) Sample cell temperature ) Other control variables Note:Contact a Campbell Scientific applications engineer from the micromet group for questions or updates regarding the datalogger program. Power Requirements 3.75 A (45 W) maximum power (cold startup) For comprehensive details, visit: www.campbellsci.eu/ap200
10 to 16 Vdc (voltage)
13 W (average power at 25°C)
12.5 W above +35°Cto 22.5 W at-30°C. 52.1x445x29.7 cm (205x 175 x117in) 159 kg (35 Ib) for AP200 base model 2729 (10 02) for CR1000KD 1kg (2.3 Ib) for LI-850 1549 (5.4 0z) for CFM100/ NL116 Dual-head diaphragm pump with a brushless dc motor Mounted in an insulated, temperature-controlled box inside system enclosure Pumping speed is automatically controlled to maintain the pump inlet pressure at the set point 90 liters per minute (LPM) 150t0 115.0kPa
80 W (turns on/off at 2°C)
~50 min (from -30° to +2°C)
0.7 W (turns on at 50°C and off
at45°Q) Mounted inside system enclosure Inlets Connections Mass Flow Sensor Heater Warm-up Time Fan Intake Assembly Filter Orifice Heater Mixing Volume Sample Connection Eight air sample inlets plus one inlet for zero, one inlet for CO, span, and one inlet for H,0 span 0.25-in Swagelok 0to 1.0 standard liters per minute (SLPM)
8.0 W (turns on/off at 5°C)
~20 min (from -30° to +4°C)
0.7 W (turns on at 45°C and off
at43°0) 1.0-in diameter, sintered stainless-steel disk filter, 10- micron pore size 2 kohm (0.07 W at 12 Vdc) 750mL 0.25 in. Swagelok Number of Connections for 3 (1 in, 2 out) Heater Cable Entry Seals Cable Diameter for Heater Cable Entry Seals Wire Diameter for Heater Cable Screw Terminals Wire Stripping Length for Heater Cable Screw Terminals Screw Tightening Torque for Heater Cable Screw Terminals Orifice Inside Diameter Dimensions Weight 281t066mm (0.11t00.261n)
261012 AWG
50mm (0.2in.)
0.4 Nem
0.178 mm (0.007 in) 31x125x19cm (12x5x 7.51n) 14kg (3.1 Ib) For comprehensive details, visit: www.campbellsci.eu/ap200 CAMPBELL 80 Hathern Road, Shepshed, LE129GX UK | +(0)1509828888 | sale@campbellsci.couk | www.campbellscieu AUSTRALIA | BRAZIL | CANADA | CHINA | COSTA RICA | FRANCE | GERMANY | INDIA | SOUTH AFRICA | SPAIN | THAILAND | UK | USA ©2020 Campbell Scientific, Ltd. | 01/15/2020