Technical content

VER : A_3 update : 98.06.09 A C - D C M e d i c a l AEC60 SERIES 60 Watts http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 -1- ELECTRICAL SPECIFICATIONS Model No. AEC60-5S AEC60-9S AEC 60-12S AEC60-15S AEC60-24S AEC60-48S Max Output Wattage (W) 50W 60W 60W 60W 60W 60W Voltage 85-265 VAC or 100-370 VDC Frequency (Hz) 47-63 Hz Current (Full load) 2 A max. (115 VAC) / 1 A max. (230 VAC) Inrush Current (<2ms) 30 A max. (115 VAC) / 50 A max. (230 VAC) Input Leakage Current 0.5 mA max. Voltage (V.DC.) 5V 9V 12V 15V 24V 48V Voltage Accuracy ±2% Current (mA) max 10,000 6,660 5,000 4,000 2,500 1,250 Line Regulation (LL-HL) (typ.) ±1% Load Regulation (5-100%) (typ.) ±1% Minimum Load 1% Maximum Capacitive Load 470-23,000µF depending on model Ripple <0.2% Vout +40mV max (Vp-p) Noise <0.5% Vout +50mV max (Vp-p) Efficiency (at 230V) 82% 84% 86% 86% 86% 86% Trim ±10% Hold-up Time 10 ms min. Output Switching Frequency 100 kHz Over Power Protection Auto recovery Over Voltage Protection Zener diode clamp Protection Short Circuit Protection Auto recovery Input-Output (V.AC) 3000V Input-FG (V.AC) 1500V Isolation Output-FG . (V.AC) 500V Operating Temperature -40°C…+70°C (with derating) Storage Temperature -50°C…+85°C Temperature coefficient ±0.02%/°C Humidity 95% RH Environment MTBF >130,000 h @ 25°C (MIL-HDBK-217F) Weight 310 g Physical Cooling Method Free air convection Safety Agency Approvals CE, UL60950 EMI (Conducted & Radiated Emission) EN 55022 class B EMC EMS (Noise Immunity) EN 55024 All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated. KEY FEATURES „ Switching Power Module for PCB Mountable „ Universal Input: 85-265 VAC „ Single Output „ 5 VDC to 48 VDC Output „ -40 OC Starting „ 4000 VAC Input to Output Isolation „ CE, UL, Approval „ 2-Years Product Warranty

VER : A_3 update : 98.06.09 A C - D C M e d i c a l AEC60 SERIES 60 Watts http://www.archcorp.com.tw TEL: +886-2-26989508 FAX: +886-2-26981319 -2- Measured with 0.1μ 50V // 47μ 50V Ceramic Cap. Cross to output MECHANICAL DIMENSION ( Top View ) DERATING BLOCK DIAGRAM PIN# SINGLE

1 AC IN (N)

2 AC IN (L)

6 Trim