8846A FLUKE | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 12

Technical content

The Fluke 8845A and 8846A 6.5 digit precision multime- ters have the precision and versatility to handle your most demanding measurements, on the bench or in a system. These meters are both high performance and feature rich, yet also remarkably easy to use. These digital multimeters perform the functions you would expect to see in a multifunction DMM, including measuring volts, ohms, and amps, with performance that exceeds expec- tations. Basic V dc accuracy of up to 0.0024 %, 10 A current range, and a wide ohms range from 10 W to 1 GW with up to 10 μW resolution give you an unbeatable combination of measurement capability. You can also use the 8845A and 8846A to measure tempera- ture, capacitance, period, and frequency—the functions of a counter, capacitance meter, and thermometer are built in for unparalleled versatility. Extend the meters’ utility even more with their graphical display modes, including Trendplot paperless recorder mode, statis- tics and histograms—features you won’t find on other multimeters. Of course, these meters are also durable and dependable, features you expect from any Fluke meter. This unique combi- nation of features and perfor- mance makes the 8845A and 8846A an unbeatable value for a wide variety of applications, including manufacturing test, research and development, and service. Features at a glance

  • 6.5 digit resolution
  • Basic V dc accuracy of up to 0.0024 % (1 yr.)
  • Dual display
  • 100 μA to 10 A current range, with up to 100 pA resolution
  • Wide ohms ranges from 10 W to 1 GW with up to 10 μW resolution
  • 2 x 4 ohms 4-wire measurement technique
  • Both models measure frequency and period
  • 8846A also measures capacitance and temperature
  • USB memory drive port (8846A)
  • Fluke 45 and Agilent 34401A emulation
  • Graphical display
  • Trendplot™ paperless recorder mode, statistics, histogram
  • CAT I 1000 V, CAT II 600 V

8845A and 8846A Fluke Corporation 1 General Specifications Power Voltage Dimensions Display Vacuum Fluorescent Display, dot matrix Environment Temperature Relative Humidity (non-condensing) 28 °C to 40 °C <80 % 40 °C to 55 °C <50 % Altitude Safety Designed to comply with IEC 61010-1:2000-1, UL 61010-1A1, CAN/CSA-C22.2 No. 61010.1, CAT I 1000V/CAT II 600V EMC Designed to comply with IEC 61326-1:2000-11 (EMC) when used with shielded communications cables. This Meter has shown susceptibility to radiated frequencies greater than 1 V/m from 250 to 450 MHz. Triggering Memory (available separately, see “Accessories”) through front-panel USB port Math Functions Zero, dBm, dB, MX+B, Offset, DCV ratio and TrendPlot, Histogram, Statistics (min/max/average/standard deviation), and Limit Test

2 Fluke Corporation 8845A and 8846A

750 V ac (8845A), Diode, and 10 A ranges

RS-232C, DTE 9-pin, 1200 to 230400 baud (RS-232C to USB cable available to connect the Meter to a PC USB port. See Accessories) IEEE 488.2 LAN and “Ethernet 10/100 base T with DHCP (for IP_ADDRess) option” Warranty One year Electrical Specifications Accuracy specifications are valid for 6½ digit resolution mode after at least a 1-hour warm-up with Auto Zero enabled. 24-hour specifications are relative to calibration standards and assume a controlled electromagnetic environment per EN 61326-1:2000-11 DC Voltage Specifications 100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1 % minutes, add error: 0.0002 % range additional error + 5 μV. ZERO function for that range and NPLC setting. Voltage accuracy for the HI to LO Input (in ppm of the Input voltage), and Reference accuracy = DC Voltage accuracy for the HI to LO (Sense) Reference (in ppm of the Reference voltage). characteristics, and input signal changes. Input Characteristics Resolution Range Resolution 4½ Digits 5½ Digits 6½ Digits Input Impedance 100 mV 100.0000 mV 10 µV 1 µV 100 nV 10 M Ω or >10 GΩ [1] 1 V 1.000000 V 100 µV 10 µV 1 µV 10 M Ω or >10 GΩ [1] 10 V 10.00000 V 1 mV 100 µV 10 µV 10 M Ω or >10 GΩ [1] 100 V 100.0000 V 10 mV 1 mV 100 µV 10 M Ω ±1% 1000 V 1,000.000 V 100 mV 10 mV 1 mV 10 M Ω ±1% [1] Inputs beyond ±14 V are clamped through 200 kΩ typical. 10 MΩ is default input impedance. 8846A Accuracy Accuracy is given as ± (% measurement + % of range) Range 24 Hour (23 ±1 °C)

90 Days

(23 ±5 °C)

1 Year

(23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C

8845A and 8846A Fluke Corporation 3 8845A Accuracy Accuracy is given as ± (% measurement + % of range) Range 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C Additional Errors Digits NPLC Additional NPLC Noise Error 6½ 100 0 % of range 6½ 10 0 % of range 5½ 1 0.001 % of range 5½ .2 0.0025 % of range +12 μV 4½ 0.02 0.017 % of range +17 μV AC Voltage Specifications AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz, add an additional error of 0.1 % of range, and for 50 kHz to 100 kHz, add 0.13 % of range. volts-Hertz product (whichever is less) for any range. any range. AC Filter Bandwidth: Crest Factor Error (applies to non-sinusoidal waveforms only) Crest factor 2-3, 0.2 % of full scale Crest factor 3-4, 0.4 % of full scale Crest factor 4-5, 0.5 % of full scale Input Characteristics Resolution Range Resolution 4½ Digits 5½ Digits 6½ Digits Input Impedance 100 mV 100.0000 mV 10 µV 1 µV 100 nV 1 V 1.000000 V 100 µV 10 µV 1 µV 10 V 10.00000 V 1 mV 100 µV 10 µV 100 V 100.0000 V 10 mV 1 mV 100 µV 1000 V 1,000.000 V 100 mV 10 mV 1 mV

1 MΩ ±2 % shunted

by <100 pf 8846A Accuracy Accuracy is given as ± (% measurement + % of range) Range Frequency 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 100 mV 100 – 300 kHz [1] 1 V 100 – 300 kHz [1]

4 Fluke Corporation 8845A and 8846A

(23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 10 V 100 – 300 kHz [1] 100 V 100 –- 300 kHz [1] 3 – 5 Hz 1.0 + 0.015 1.0 + 0.0225 1.0 + 0.0225 0.1 + 0.00225 50 – 100 kHz [2] 1000 V 100 – 300 kHz [1][2] [1] Typically 30 % reading error at 1 MHz [2] 1000 Volt range is limited to 8 X 10 volt-Hertz 8845A Accuracy Accuracy is given as ± (% measurement + % of range) Range Frequency (Hz)

24 Hour

(23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 100 mV 100 – 300 kHz [1] 1 V 100 – 300 kHz [1] 10 V 100 – 300 kHz [1] 100 V 100 – 300 kHz [1] 50 – 100 kHz [2] 750 V 100 – 300 kHz [1] [2] [1] Typically 30 % reading error at 1 MHz [2] 750 Volt range is limited to 8 X 10 volt-Hertz

8845A and 8846A Fluke Corporation 5 Additional Low Frequency Errors Error is stated as % of reading. AC Filter Frequency 3HZ (slow) 20HZ (medium) 200HZ (fast) 10 – 20 Hz 0 0.25 – 20 – 40 Hz 0 0.02 – 40 – 100 Hz 0 0.01 0.55 100 – 200 Hz 0 0 0.2 200 Hz – 1 kHz 0 0 0.02 >1 kHz 0 0 0 Resistance Specifications are for 4-wire resistance function, 2 x 4-wire resistance, or 2-wire resistance with zero. If zero is not used, add 0.2 Ω for 2- wire resistance plus lead resistance, and add 20 mΩ for 2 x 4-wire resistance function. 100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1 % ZERO function for that range and NPLC setting. Input Characteristics Resolution Range Resolution 4½ Digits 5½ Digits 6½ Digits Source Current 10 Ω [1] 10.00000 Ω 1 m Ω 100 μΩ 10 μΩ 5 mA/13 V 100 Ω 100.0000 Ω 10 m Ω 1 m Ω 100 μΩ 1 mA/6 V 1 kΩ 1.000000 k Ω 100 m Ω 10 m Ω 1 m Ω 1 mA/6 V 10 kΩ 10.00000 k Ω 1 Ω 100 m Ω 10 m Ω 100 μA/6 V 100 kΩ 100.0000 k Ω 10 Ω 1 Ω 100 m Ω 100 μA/13 V 1 MΩ 1.000000 M Ω 100 Ω 10 Ω 1 Ω 10 μA/13 V 10 MΩ 10.00000 M Ω 1 k Ω 100 Ω 10 Ω 1 μA/13 V 100 MΩ 100.0000 M Ω 10 k Ω 1 k Ω 100 Ω 1 μA || 10 MΩ/10 V 1.0 GΩ [1]

1.000000 GΩ 100 k Ω 10 k Ω 1 k Ω 1 μA || 10 MΩ/10 V

[1] 8846A Only 8846A Accuracy Accuracy is given as ± (% measurement + % of range) Range 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 8845A Accuracy Accuracy is given as ± (% measurement + % of range) Range 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C

6 Fluke Corporation 8845A and 8846A

Digits NPLC Additional NPLC Noise Error 6½ 100 0 % of range 6½ 10 0 % of range 5½ 1 0.001 % of range 5½ 0.2 0.003 % of range ±7 mΩ 4½ 0.02 0.017 % of range ±15 mΩ DC Current 550 mA for 2 minutes on, 1 minute off. 100 dB for NPLC of 1 or greater with analog filter on and power line frequency ±0.1 % ZERO function for that range and NPLC setting. Input Characteristics Resolution Range Resolution 4½ Digits 5½ Digits 6½ Digits Shunt Resistance (Ohms) Burden Voltage 100 µA 100.0000 µA 10 nA 1 nA 100 pA 100 Ω <0.015 V 1 mA 1.000000 mA 100 nA 10 nA 1 nA 100 Ω <0.15 V 10 mA 10.00000 mA 1 μA 100 nA 10 nA 1 Ω <0.025 V 100 mA 100.0000 mA 10 µA 1 μA 100 nA 1 Ω <0.25 V 400 mA [3] 400.000 mA 100 µA 10 µA 1 μA 1 Ω <0.50 V 1 A [2] 1.000000 A 100 µA 10 µA 1 μA 0.01 Ω <0.05 V 3 A [1] 3.00000A 1 mA 100 µA 10 µA 0.01 Ω <0.15 V 10 A 10.00000 A 1 mA 100 µA 10 µA 0.01 Ω <0.5 V [1] Part of 10 A range. [2] Available on the front panel terminal only. [3] 400 mA available in software version 2.0 or greater only . 400 mA continuously; 550 mA for 2 minutes on, 1 minute off. Accuracy (8846A) Accuracy is given as ± (% measurement + % of range) Range 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 400 mA [3] 1 A [2] 3 A [1][2] 10 A [2] [1] Part of 10 A range [2] Available at front panel connectors only [3] 400 mA available in software version 2.0 or great er only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off.

8845A and 8846A Fluke Corporation 7 Accuracy (8845A) Accuracy is given as ± (% measurement + % of range) Range 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 400 mA [3] 1 A [2] 3 A [1][2] 10 A [2] [1] Part of 10 A range [2] Available at front panel connectors only [3] 400 mA available in software version 2.0 or great er only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off. Additional Current Errors Digits NPLC Additional NPLC Noise Error for 1 mA, 100 mA, 400 mA, 3A and 10A Additional NPLC Noise Error for 100 μA, 10 mA, 1A 6½ 100 0 % of range 0 % of range 6½ 10 0 % of range 0 % of range 5½ 1 0.001 % of range 0.01 % of range 5½ 0.2 0.011 % of range ±4 μA 0.11 % of range ±4 μA 4½ 0.02 0.04 % of range ±4 μA 0.28 % of range ±4 μA AC Current The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs from 1 % to 5 % of range, add an additional error of 0.1 % of range. mA for 2 minutes on, 1 minute off. AC Filter Bandwidth Crest Factor Error (applies to non-sinusoidal waveforms only) Crest factor 2-3, 0.2 % of full scale Crest factor 3-4, 0.4 % of full scale Crest factor 4-5, 0.5 % of full scale Input Characteristics Resolution Range Resolution 4½ Digits 5½ Digits 6½ Digits Shunt Resistance (Ohms) Burden Voltage 100 μA [1] 100.0000 μA 10 nA 1 nA 100 pA 100 Ω <0.015 V 1 mA [1] 1.000000 mA 100 nA 10 nA 1 nA 100 Ω <0.15 V 10 mA 10.00000 mA 1 μA 100 nA 10 nA 1 Ω <0.025 V 100 mA 100.0000 mA 10 µA 1 μA 100 nA 1 Ω <0.25 V 400 mA [4] 400.000 mA 100 µA 10 µA 1 μA 1 Ω <0.50 V 1 A [3] 1.000000 A 100 µA 10 µA 1 μA 0.01 Ω <0.05 V 3 A [2][3] 3.00000 A 1 mA 100 µA 10 μA 0.01 Ω <0.05 V 10 A [3] 10.00000 A 1 mA 100 µA 10 µA 0.01 Ω <0.5 V [1] 8846A Only [2] Part of 10 A range [3] Available at front panel connectors only [4] 400 mA available in software version 1.0.700.18 or great er only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400mA

8 Fluke Corporation 8845A and 8846A

Accuracy is given as ± (% measurement + % of range) Range Frequency (Hz) (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 100 μA 1 mA 10 mA 100 mA 400 mA [3] 1 A [2] 3 A [1][2] 10 A [2] [1] Part of 10 A range [2] Available only on front panel connectors [3] 400 m available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400mA; specification for current above 329 mA is typical. 8845A Accuracy Accuracy is given as ± (% measurement + % of range) Range Frequency (Hz) (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 10 mA 100 mA 400 mA [3] 1 A [2]

8845A and 8846A Fluke Corporation 9 Range Frequency (Hz) (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 3 A [1][2] 10 A [2] [1] Part of the 10 A range [2] Available only at front panel connectors [3] 400 m available in software version 1.0.700.18 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 minute off; maximum crest factor 3:1 at 400mA; specification for current above 329 mA is typical. Additional Low Frequency Errors Error is stated as % of reading. AC Filter Frequency 3HZ (slow) 20HZ (medium) 200HZ (fast) 10 – 20 Hz 0 0.25 – 20 – 40 Hz 0 0.02 – 40 – 100 Hz 0 0.01 0.55 100 – 200 Hz 0 0 0.2 200 Hz – 1 kHz 0 0 0.02 > 1 kHz 0 0 0 Frequency function. For the most accurate measurement, wait up to 1 second for the input blocking capacitor to settle. voltage, low-frequency signals. 8846A Accuracy Accuracy is given as ± % measurement Range Frequency 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 3 – 5 Hz 0.1 0.1 0.1 0.005 5 – 10 Hz 0.05 0.05 0.05 0.005 10 – 40 Hz 0.03 0.03 0.03 0.001 40 Hz – 300 kHz 0.006 0.01 0.01 0.001 100 mV to 1000 V [1][2] 300 kHz – 1 MHz 0.006 0.01 0.01 0.001 [1] Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10. [2] Limited to 8 X 10 volt-Hertz

10 Fluke Corporation 8845A and 8846A

Accuracy is given as ± % measurement Range Frequency 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Compensation/ °C Outside 18 to 28 °C 3 – 5 Hz 0.1 0.1 0.1 0.005 5 – 10 Hz 0.05 0.05 0.05 0.005 10 – 40 Hz 0.03 0.03 0.03 0.001 100 mV to 750 V [1][2] 40 Hz – 300 kHz 0.006 0.01 0.01 0.001 [1] Input >100 mV. For 10 – 100 mV, multiply percent measurement error by 10. [2] Limited to 8 X 10 volt-Hertz Gate Time vs. Resolution Gate Time Resolution 0.01 5½ 0.1 6½ 1.0 6½ Additional Low Frequency Errors Error stated as percent of measurement for inputs >100 mV. For 10 – 100 mV, multiply percent by 10. Resolution Frequency 6½ 5½ 4½ 3 – 5 Hz 0 0.12 0.12 5 – 10 Hz 0 0.17 0.17 10 – 40 Hz 0 0.2 0.2 40 – 100 Hz 0 0.06 0.21 100 – 300 Hz 0 0.03 0.21 300 Hz – 1 kHz 0 0.01 0.07 > 1 kHz 0 0 0.02 Capacitance (8846A Only) Accuracy is stated as ±(% of measurement + % of range) Range Resolution 1 Year Accuracy [1] (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C 1 nF 1 pF 2% ± 2.5 % 0.05 + 0.05 10 nF 10 pF 1% ± 0.5 % 0.05 + 0.01 100 nF 100 pF 1% ± 0.5 % 0.01 + 0.01 1 µF 1 nF 1% ± 0.5 % 0.01 + 0.01 10 µF 10 nF 1% ± 0.5 % 0.01 + 0.01 100 µF 100 nF 1% ± 0.5 % 0.01 + 0.01 1 mF 1 µF 1% ± 0.5 % 0.01 + 0.01 10 mF 10 µF 1% ± 0.5 % 0.01 + 0.01 100 mF 100 µF 4% ± 0.2 % 0.05 + 0.05 [1] Stated accuracy is attained when Zero function is used. Temperature (8846A only) Accuracy is stated as ± °C and is based on a Platinum RT100 (DIN IEC 751, 385 type) RTD with less than 10 ohms lead resistance. The accuracy listed in the table below are valid only when using the 4-wire RTD measurement function. Specifications do not include probe accuracy, which must be added. Accuracy Range Resolution 90 Days (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C -200 °C 0.001 °C 0.06 0.09 0.0025 -100 °C 0.001 °C 0.05 0.08 0.002 0 °C 0.001 °C 0.04 0.06 0.002 100 °C 0.001 °C 0.05 0.08 0.002 300 °C 0.001 °C 0.1 0.12 0.002 600 °C 0.001 °C 0.18 0.22 0.002

Digits NPLC Additional NPLC Noise Error C° 0 001 ½ 6 C° 0 01 ½ 6 C° 30.0 1 ½ 5 C° 21.0 2.0 ½ 5 4 ½ 0.02 0.6 °C Continuity Accuracy is given as ± (% measurements + % of range) Range 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C Diode Test Accuracy is given as ± (% measurements + % of range) Range 24 Hour (23 ±1 °C) (23 ±5 °C) (23 ±5 °C) Temperature Coefficient/ °C Outside 18 to 28 °C Measurement Rates (IEEE488[4]) Measurements/Second [1] Function Digits Setting Integration Time

60 Hz (50 Hz) 8845A 8846A

6½ 10 NPLC 167 (200) ms 6 (5) 6 (5) 5½ 1 NPLC 16.7 (20) ms 60 (50) 60 (50) 5½ 0.2 NPLC 3.3 ms 270 270 DC Volts, DC Current, and Resistance 4½ 0.02 NPLC 500 us 995 995 6½ 3 Hz 0.47 0.47 6½ 20 Hz 1.64 1.64 AC Voltage and AC Current [2] 6½ 200 Hz [3] 4.5 4.5 6½ 1 s 1 1 5½ 100 ms 9.8 9.8 Frequency and Period 4½ 10 ms 80 80

2 AN 2/1 6 ecnaticapaC

[1] Typical measurement rates with auto-zero off, delay = 0, display off, auto range off and math off. [2] Maximum measurement rates for 0.01 % of ac step. When dc input varies, additional settling delay is required. [3] For remote operation or external trigger using default settling delay [4] Speeds available in OutG SW 1.0.700.18 or higher. Note that the measurements rates for RS232 can vary depending on the baud rate chosen. If the baud rate selected is 115,200, the maximum measurement rate is 711 measurement/s. The LAN bus has a maximum measurement rate of 963 measurement/s. Fluke Corporation PO Box 9090, Everett, WA 98206 U.S.A. Fluke Europe B.V. PO Box 1186, 5602 BD Eindhoven, The Netherlands Fluke. Keeping your world up and running.® For more information call: In the U.S.A. (800) 443-5853 or Fax (425) 446-5116 In Europe/M-East/Africa +31 (0) 40 2675 200 or Fax +31 (0) 40 2675 222 In Canada (800)-36-FLUKE or Fax (905) 890-6866 From other countries +1 (425) 446-5500 or Fax +1 (425) 446-5116 Web access: http://www.fluke.com ©2006-2008 Fluke Corporation. Specifications subject to change without notice. Printed in U.S.A. 2/2008 2748044 D-EN-N Rev B Pub_ID: 11152-eng Rev 02