PXIE-SCP5101 NI | Alldatasheet

Document overview

  • Manufacturer or author: Provided By alldatasheet.com(free datasheet download site)
  • PDF pages: 36

Technical content

Expandable PXI bundle based on PXIe-5110 Oscilloscope, 100 MHz, 8 bits, 2 Channels, 512 MB Specifications PXIe-1083 and PXIe-5110

Module cooling Forced air circulation (positive pressurization) through one 150 CFM fan Module slo t airflow direction Bottom of module to top of module Module intake Bottom of chassis Module exhaust Top, right side of chassis Slot cooling capacity 58 W; slot 6 supports 58 W cooling with high fan mode Power supply cooling Forced air circulation through integrated fans Power supply intake Front and left side chassis Power supply exhaust Rear of chassis Minimum chassis cooling clearances Above 44.45 mm (1.75 in.) Rear 44.45 mm (1.75 in.) Sides 44.45 mm (1.75 in.) Below Rack 44.45 mm (1.75 in.) Desktop 25.4 mm (1.00 in.) © National Instruments 5 PXIe-1083 Specifications

Maximum altitude 2,000 m (6,560 ft.), 800 mbar (at 25 °C ambient , high fan mode) Pollution Degree 2 Indoor use only. Operating Envir onment Ambient temperature range When all peripheral modules requir e ≤38 W cooling capacity per slot 0 °C to 50 °C (IEC 60068-2-1 and IEC 60068-2-2.)2 Meet s MIL-PRF-28800F Class 3 low temperature limit and high temperature limit. When any peripheral module requir es >38 W cooling capacity per slot 0 °C to 40 °C (IEC 60068-2-1 and IEC 60068-2-2.)2 Meet s MIL-PRF-28800F Class 3 low temperature limit and MIL- PRF-28800F Class 4 high temperature limit. Relative humidity range 20% to 80%, noncondensing Storage Environment Ambient temperature range –40 °C to 71 °C (IEC-60068-2-1 and IEC-60068-2-2.) [3] Meet s MIL- PRF-28800F Class 3 limits. Relative humidity range 10% to 95%, noncondensing ni.com6 PXIe-1083 Specifications

Operational shock 30 g peak, half-sine, 11 ms pulse (IEC-60068-2-27.)3 Meet s MIL- PRF-28800F Class 2 limits. Operational random vibration 5 t o 500 Hz, 0.3 grms Non-operating vibration 5 t o 500 Hz, 2.4 grms (IEC 60068-2-64.)3 Non-operating t est profile exceeds the requirements of MIL-PRF-28800F, Class 3. Acoustic Emissions Sound Pressur e Level (at Operator Position) (Tested in accordance with ISO 7779. Meets MIL-PRF-28800F requirements.)

38 W Profile

Auto fan (up to 30 °C ambient) 33.7 dBA High fan 50.8 dBA

58 W Profile

Auto fan (up to 30 °C ambient) 54.7 dBA High fan 55.3 dBA Sound Power Level © National Instruments 7 PXIe-1083 Specifications

Figure 2. PXIe-1083 Chassis Dimensions (Side) Figure 3. PXIe-1083 Chassis Dimensions (Bottom) 1 The operating r ange is guaranteed by design.

2 This product meets the requirements of the environmental standards for electrical

equipment for measurement, control, and laboratory use.

3 This produc t meets the requirements of the environmental standards for electrical

equipment for measurement, control, and laboratory use. © 2022 National Instruments Corporation.

These specifications apply t o the PXIe-5110 with 64 MB and 512 MB of memory. Definitions Warranted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration. Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty. ■ Typical specifications describe the performance met by a majority of models. ■ Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing. ■ Measured specifications describe the measured performance of a representative model. Specifications are Nominal unless otherwise noted. Conditions Specifications are valid under the following conditions unless otherwise noted. ■ All vertical ranges, bandwidths, and bandwidth limiting filters ■ Sample rate set to 0.5 GS/s or 1 GS/s ■ Onboard sample clock locked to PXI_Clk100 reference clock ■ 15-minute warm-up time at ambient temperature ■ Chassis configured: [1] © National Instruments 3 PXIe-5110 Specifications

■ PXI E xpress chassis fan speed set to HIGH ■ Foam fan filters removed if present ■ Empty slots contain PXI chassis slot blockers and filler panels Warranted specifications are valid under the following conditions unless otherwise noted. ■ Ambient temperature range of 0 °C to 55 °C ■ Altitude ≤2,000 m ■ Calibration cycle maintained ■ Self-calibration run after: ■ Warm-up time has elapsed ■ Module has been power cycled ■ PC or controller has been restarted or wakes from sleep or hibernation modes ■ External calibration performed at 23 °C ±3 °C Typical specifications are valid under the following conditions unless otherwise noted. ■ Ambient temperature range of 0 °C to 55 °C ■ Altitude ≤2,000 m Vertical Analog Input Number of channels Two (simultaneously sampled) Input type Referenced single-ended Connectors BNC, ground referenced ni.com4 PXIe-5110 Specifications

Table 1. Full-Scale (FS) Input Range and Vertical Offset Range

0.04 V ±5 V

0.1 V ±5 V

0.2 V ±5 V

0.4 V ±5 V

1 V ±5 V ±20 V

2 V ±5 V ±20 V

4 V ±5 V ±20 V

10 V ±2 V ±100 V

20 V — ±100 V

40 V — ±100 V

1 MΩ[2] |Peaks| ≤250 V DC

1 MΩ 100 MHz, typical

Figure 1. 50 Ω Full Bandwidth F requency Response, 1 GS/s, 1 Vpk-pk, Measured[6]

Table 2. Effectiv e Number of Bits (ENOB)[9]

20 MHz filter enabled Full bandwidth (Input Frequency <50 MHz)

Sample clock jitter[14] 2 ps RMS Phase-Locked Loop (PLL) Reference Clock Sources Internal Onboard clock (internal oscillator) External PXI_Clk100 (backplane connector) Duty cycle tolerance 45% to 55%, typical Triggers Supported triggers Reference (Stop) Trigger Ref erence (Arm) Trigger Start Trigger (Acquisition Arm) Advance Trigger Trigger types Edge Glitch Hys teresis Runt Width Window ni.com10 PXIe-5110 Specifications

Table 3. Analog Trigg er Time Resolution

Interpolator Status Time Resolution TIS Enabled TIS Disabled Disabled 1 ns 2 ns Trigger filters Low frequency (LF) reject 100 kHz High frequency (HF) reject 100 kHz Minimum threshold duration[15] Sample clock period Digital Trigger Sources PFI <0..3> (front panel HD-BNC connectors) PXI_T rig <0..7> (backplane connector) Time resolution PFI 2 ns PXI_Trig 8 ns Programmable Function Interface (PFI) Connectors PFI <0..3> (front panel HD-BNC connectors) Direction Bidirectional per channel As an input (trigger) Destinations Start Trigger (Acquisition Arm) ni.com12 PXIe-5110 Specifications

Reference (Stop) Trigger Ref erence (Arm) Trigger Advance Trigger Input impedance 49.9 kΩ VIH 2 V, typical VIL 0.8 V, typical Recommended input range 0 V to 3.3 V Maximum input overload +5 V tolerant Minimum pulse width 10 ns As an output (event) Sources Ready for Start St art Trigger (Acquisiton Arm) Ready for Reference Reference (Stop) Trigger End of Record Ready for Advance Advance Trigger Done (End of Acquisition) © National Instruments 13 PXIe-5110 Specifications

Output impedance 50 Ω Logic type 3.3 V CMOS Maximum current drive 12 mA Maximum frequency 50 MHz Minimum pulse width 10 ns Probe Compensation Connectors Probe compensation terminal Ground t erminal Output voltage[16] 0 V to 5 V Maximum overload voltage 25 V DC CableSense CableSense pulse voltage[17] 0.4 V CableSense pulse rise time[18] 4 ns Driver support for CableSense on the PXIe-5110 was first available in NI-SCOPE18.7. Related information ■ For more information about CableSense technology, refer to ni.com/ cablesense. ni.com14 PXIe-5110 Specifications

Table 4. Examples of Allocated Onboard Memory per Record, 512 MB Option

External calibration corrects the onboard references for gain and offset errors used in self-calibration and adjusts the compensation attenuator. All calibration constants are stored in nonvolatile memory. Self-Calibration Self-calibration is done on software command. The calibration corrects for gain, offset, interleaving spurs, and intermodule synchronization errors. Run self- calibration after the specified warm-up time has elapsed and any time the module is power cycled or the PC or controller is restarted or wakes from sleep or hibernation modes. Refer to the NI High-Speed Digitizers Help at ni.com/manuals for mor e information on when to self-calibrate the device. Calibration Specifications Interval for external calibration 2 years Warm-up time[21] 15 minutes Software Driver Software Driver support for this device was first available in NI-SCOPE18.6. NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibrate the PXIe-5110. NI-SCOPE provides application programming interfaces for many development environments. ni.com16 PXIe-5110 Specifications

NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments: ■ LabVIEW ■ LabWindows™/CVI™ ■ Measurement Studio ■ Microsoft Visual C/C++ ■ .NET (C# and VB.NET) Interactive Soft Front Panel and Configuration When you install NI-SCOPE on a 64-bit system, you can use InstrumentStudio to monitor, control, and record measurements from the PXIe-5110. InstrumentStudio is an application that allows you to perform interactive measurements on several different NI device types in a single application. Interactive control of the PXIe-5110 was first available via InstrumentStudio in NI-SCOPE18.6. InstrumentStudio is included on the NI-SCOPE media. NI Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5110. MAX is included on the driver media. Synchronization Channel-to-channel skew, between the channels of a PXIe-5110 50 Ω <100 ps

1 MΩ <100 ps

© National Instruments 17 PXIe-5110 Specifications

Synchronization with the NI-TClk API [22] NI-T Clk is an API that enables system synchronization of supported PXI modules in one or more PXI chassis, which you can use with the PXIe-5110 and NI-SCOPE. NI-TClk uses a shared Reference Clock and triggers to align the Sample Clocks of PXI modules and synchronize the distribution and reception of triggers. These signals are routed through the PXI chassis backplane without external cable connections between PXI modules in the same chassis. Module-to-module skew, between PXIe-5110 modules using NI-TClk [23] NI-TClk synchronization without manual adjustment [24] Skew, peak-to-peak [25] 200 ps Jitter, peak-to-peak [26] 120 ps NI-TClk synchronization with manual adjustment [24] Skew, average [25] 10 ps Jitter, peak-to-peak [26] 8 ps Sample Clock delay/adjustment resolution <1 ps Power Current draw +3.3 V DC 1.82 A +12 V DC 0.83 A ni.com18 PXIe-5110 Specifications

+3.3 V DC 6 W +12 V DC 10 W Total 16 W Total maximum power allowed 30 W Physical Dimensions 3U, one-slot, PXI Express/CompactPCI Express module 2.0 cm × 13.0 cm × 21.6 cm Weight 380 g (13.4 oz) Bus Interface Form factor Gen 1 x4 module Slot compatibility PXI Express or hybrid Environmental Characteristics Temperature Operating 0 °C to 55 °C © National Instruments 19 PXIe-5110 Specifications

Storage -40 °C to 71 °C Humidity Operating 10% to 90%, noncondensing Storage 5% to 95%, noncondensing Pollution Degree 2 Maximum altitude 4,600 m (at 25 °C ambient temperature) Shock and Vibration Operating vibration 5 Hz to 500 Hz, 0.3 g RMS Non-operating vibration 5 Hz to 500 Hz, 2.4 g RMS Operating shock 30 g, half-sine, 11 ms pulse Product Certifications and Declarations Ref er to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, sear ch by model number, and click the appropriate link.

1 For mor e information about cooling, refer to the Maintain Forced-Air Cooling

Note to Users available at ni.com/manuals. 2 Derat e above 250 kHz at 20 dB/dec until 2.5 MHz, then derate at 5 dB/dec. 3 Within ±5 °C of self-c alibration temperature. ni.com20 PXIe-5110 Specifications

4 Used t o calculate errors when the onboard temperature changes more than ±5 °C

from the self-calibration temperature.

5 Measur ed on one channel with test signal applied to another channel and the

same range setting on both channels. 6 Normalized t o 50 kHz. 7 Ex cludes ADC interleaving spurs. 8 Input frequencies <50 MHz, input r ange ≤4 Vpk-pk. -1 dBFS input signal. Includes second through fifth harmonics. 9 Input frequencies <50 MHz. -1 dBF S input signal corrected to FS. 1 kHz resolution bandwidth. 10 Applies to all filt er settings and input modes. Verified using a 50 Ω terminator connected to input. 11 Divide by n decimation from 500 MS/s. For more information on the sample clock and decimation, refer to the NI High-Speed Digitizers Help. 12 Single channel only. 13 Phase-lock ed to onboard clock. The default clock is PXI_Clk100. Refer to your chassis specifications for the timebase accuracy of PXI_Clk100. 14 Inte grated from 100 Hz to 10 MHz. Includes the effects of converter aperture uncertainty and the clock circuitry jitter. Excludes trigger jitter.

15 Data mus t exceed each corresponding trigger threshold for at least this minimum

duration to ensure analog triggering. 16 1 kHz, 50% duty cy cle square wave. 17 When measur ed with a high-impedance device. 18 When sourcing int o a 50 Ω cable or load. © National Instruments 21 PXIe-5110 Specifications

19 Onboar d memory is shared among all enabled channels. 20 For 512 MB op tion. You can exceed this value if you fetch records while acquiring data. For more information, refer to the Enable Records > Memory property in the NI High-Speed Digitizers Help at ni.com/manuals.

21 Warm-up time be gins after the chassis and either the controller or PC is powered

and NI-SCOPE is loaded. 22 NI-T Clk installs with NI-SCOPE.

23 Although you c an use NI-TClk to synchronize non-identical modules, these

specifications apply only to synchronizing identical modules. Specifications are valid under the following conditions: ■ All modules installed in the same PXI Express chassis ■ NI-TClk used to align the sample clocks of each module ■ All parameters set to identical values for each module ■ Self-calibration completed ■ Ambient temperature within ±1 °C of self-calibration For other configurations, including multi-chassis systems, contact NI Technical Support at ni.com/support.

24 Manual adjustment is the pr ocess of minimizing synchronization jitter and skew

by adjusting Trigger Clock (TClk) signals using the instrument driver.

25 Ske w is the misalignment between module timing across slots of a chassis and is

caused by clock and analog path delay differences.

26 Jitt er is the variation in module alignment that can be expected with each call to

NI-TClkSynchronize. ni.com22 PXIe-5110 Specifications © 2022 National Instruments Corporation.