PXIE-SCP5100 NI | Alldatasheet

Document overview

  • PDF pages: 38

Technical content

Expandable PXI bundle based on PXIe-5105 Oscilloscope, 60 MHz, 12 bits, 8 Channels, 128 MB Specifications PXIe-1083 and PXIe-5105

Bundle P/N: 867010-01 Recommended Software Table of Contents This document combines the PDFs of this bundle together. The page numbers in the table of contents correspond to the page number of PDF the component’s documentation begins. Test Workflow P/N: 788509-35 Test Workflow is a bundle of select NI software featuring engineering-specific tools that help test professionals accomplish anything from their day-to-day work to overcoming their most challenging obstacles. Test Workflow includes:

  • LabVIEW - a graphical programming environment engineers use to develop automated research, validation, and production test systems.
  • InstrumentStudio- an application software that provides an integrated approach to interactive PXI measurements.
  • TestStand- a test executive software that accelerates system development and deployment for engineers in validation and production.
  • And more NI Software! PXIe-5105 (Oscilloscope) PXIe-1083 (5-Slot PXIe Chassis) Accessories:
  • SMB Female to Mini- Alligator Clips (x2)
  • Thunderbolt cable
  • Power cable, US

ains specifications for the PXIe-1083 chassis. Electrical The following section provides information about the PXIe-1083 AC input and DC output. AC Input Input rating 100 VAC to 240 VAC, 50 Hz/60 Hz, 6 A to 3 A Operating voltage range1 90 VAC to 264 VAC Nominal input frequency 50 Hz/60 Hz Operating frequency range1 47 Hz to 63 Hz Efficiency 78% typical Over-current protection Internal fuse in line Main power disconnect The AC power cable provides main power disconnect. Do not position the equipment so that it is difficult to disc onnect the power cord. The front-panel power switch causes the internal chassis power supply to provide DC power to the PXI Express backplane. Caution Disconnec t power cord to completely remove power. © National Instruments 3 PXIe-1083 Specifications

Maximum total available power for the PXIe-1083 is 293 W. wer available on +3.3 V and +5 V is 180 W. The maximum power available for each Thunderbolt port is 15 W (5 V/3 A). Table 1. Backplane Slot C urrent Capacity Note PCI V(I/O) pins in Hybrid P eripheral Slots are connected to +5 V. temperature considerations at 58 W.

Module cooling Forced air circulation (positive pressurization) through one 150 CFM fan Module slo t airflow direction Bottom of module to top of module Module intake Bottom of chassis Module exhaust Top, right side of chassis Slot cooling capacity 58 W; slot 6 supports 58 W cooling with high fan mode Power supply cooling Forced air circulation through integrated fans Power supply intake Front and left side chassis Power supply exhaust Rear of chassis Minimum chassis cooling clearances Above 44.45 mm (1.75 in.) Rear 44.45 mm (1.75 in.) Sides 44.45 mm (1.75 in.) Below Rack 44.45 mm (1.75 in.) Desktop 25.4 mm (1.00 in.) © National Instruments 5 PXIe-1083 Specifications

Maximum altitude 2,000 m (6,560 ft.), 800 mbar (at 25 °C ambient , high fan mode) Pollution Degree 2 Indoor use only. Operating Envir onment Ambient temperature range When all peripheral modules requir e ≤38 W cooling capacity per slot 0 °C to 50 °C (IEC 60068-2-1 and IEC 60068-2-2.)2 Meet s MIL-PRF-28800F Class 3 low temperature limit and high temperature limit. When any peripheral module requir es >38 W cooling capacity per slot 0 °C to 40 °C (IEC 60068-2-1 and IEC 60068-2-2.)2 Meet s MIL-PRF-28800F Class 3 low temperature limit and MIL- PRF-28800F Class 4 high temperature limit. Relative humidity range 20% to 80%, noncondensing Storage Environment Ambient temperature range –40 °C to 71 °C (IEC-60068-2-1 and IEC-60068-2-2.) [3] Meet s MIL- PRF-28800F Class 3 limits. Relative humidity range 10% to 95%, noncondensing ni.com6 PXIe-1083 Specifications

Operational shock 30 g peak, half-sine, 11 ms pulse (IEC-60068-2-27.)3 Meet s MIL- PRF-28800F Class 2 limits. Operational random vibration 5 t o 500 Hz, 0.3 grms Non-operating vibration 5 t o 500 Hz, 2.4 grms (IEC 60068-2-64.)3 Non-operating t est profile exceeds the requirements of MIL-PRF-28800F, Class 3. Acoustic Emissions Sound Pressur e Level (at Operator Position) (Tested in accordance with ISO 7779. Meets MIL-PRF-28800F requirements.)

38 W Profile

Auto fan (up to 30 °C ambient) 33.7 dBA High fan 50.8 dBA

58 W Profile

Auto fan (up to 30 °C ambient) 54.7 dBA High fan 55.3 dBA Sound Power Level © National Instruments 7 PXIe-1083 Specifications

Auto fan (up to 30 °C ambient) 44.9 dBA High fan 60.3 dBA Auto fan (up to 30 °C ambient) 63.4 dBA High fan 64.2 dBA Note The pro tection provided by the PXIe-1083 can be impaired if it is used in a manner not described in this document. Safety Compliance Standards This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: ■ IEC 61010-1, EN 61010-1 ■ UL 61010-1, CSA C22.2 No. 61010-1 Note For safety certifications, refer to the product label or the Product Certifications and Declarations section. EMC Guidelines This pr oduct was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic environment. ni.com8 PXIe-1083 Specifications

This product is intended for use in industrial locations. However, harmful interf erence may occur in some installations, when the product is connected to a peripheral device or test object, or if the product is used in residential areas. To minimize interference with radio and television reception and prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation. Furthermore, any changes or modifications to the product not expressly approved by NI could void your authority to operate it under your local regulatory rules. EMC Notices Refer to the following notices for cables, accessories, and prevention measures necessary to ensure the specified EMC performance. Notice For EMC declarations and certifications, and additional information, refer to the Product Certifications and Declarations section. Notice Chang es or modifications to the product not expressly approved by NI could void your authority to operate the product under your local regulatory rules. Notice Operate this product only with shielded cables and accessories. Electromagnetic Compatibility Standards This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: ■ EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity ■ EN 55011 (CISPR 11): Group 1, Class A emissions © National Instruments 9 PXIe-1083 Specifications

■ AS/NZS CISPR 11: Gr oup 1, Class A emissions Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medical equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note In Europe, Canada, Australia, and New Zealand (per CISPR 11) Class A equipment is intended for use in nonresidential locations. CE Compliance This product meets the essential requirements of applicable European Directives, as follo ws: ■ 2014/35/EU; Low-Voltage Directive (safety) ■ 2014/30/EU; Electromagnetic Compatibility Directive (EMC) ■ 2011/65/EU; Restriction of Hazardous Substances (RoHS) Product Certifications and Declarations Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, sear ch by model number, and click the appropriate link. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This pag e contains the environmental ni.com10 PXIe-1083 Specifications

regulations and directives with which NI complies, as well as other environmental information no t included in this document. EU and UK Customers

  • Waste Electrical and Electronic Equipment (WEEE)—At the end of the pr oduct life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/ environment/weee. 电子信息产品污染控制管理办法(中国 RoHS
  • 5 40e 中国 RoHS— NI 符合中国电子信息产品中限制使用某些有害物 质指令(RoHS) 。关于 NI 中国 RoHS 合规性信息,请登录 ni.com/environment/ rohs_china。(For information about China RoHS compliance, go to ni.com/ environment/rohs_china.) Backplane Size 3U-sized; 5 peripheral slots. Compliant with IEEE 1101.10 mechanical pack aging. PXI Express Specification compliant. Accepts both PXI Express and CompactPCI (PICMG 2.0 R 3.0) 3U modules. Backplane bare-board material UL 94 V -0 Recognized Backplane connectors Conforms to IEC 917 and IEC 1076-4-101, UL 94 V-0 rated System Synchronization Clocks

10 MHz Syst

em Reference Clock: PXI_CLK10 © National Instruments 11 PXIe-1083 Specifications

Maximum slot-to-slot skew 250 ps Accuracy ±25 ppm max (guaranteed over the operating temperature range) Maximum jitter 5 ps RMS phase-jitter (10 Hz–1 MHz range) Duty-factor 45% to 55% Unloaded signal swing 3.3 V ±0.3 V Note For other specifications, refer to the PXI-1 Hardware Specification.

100 MHz Sys

tem Reference Clock: PXIe_CLK100 and PXIe_SYNC100 Maximum slot-to-slot skew 100 ps Accuracy ±25 ppm max (guaranteed over the operating temper ature range) Maximum jitter 3 ps RMS phase-jitter (10 Hz to 12 kHz range), 2 ps RMS phase-jit ter (12 kHz to 20 MHz range) Duty-factor for PXIe_CLK100 45% to 55% Absolute differential voltage (When terminated with a 50 Ω load to 1.30 V or Thévenin equivalent) 400 mV to 1000 mV Note For other specifications, refer to the PXI-5 PXI Express Hardware Specific ation. ni.com12 PXIe-1083 Specifications

talling the optional rack mount kits. Figure 1. PXIe-1083 Chassis Dimensions (Front)

Figure 2. PXIe-1083 Chassis Dimensions (Side) Figure 3. PXIe-1083 Chassis Dimensions (Bottom) 1 The operating r ange is guaranteed by design.

2 This product meets the requirements of the environmental standards for electrical

equipment for measurement, control, and laboratory use.

3 This produc t meets the requirements of the environmental standards for electrical

equipment for measurement, control, and laboratory use. © 2022 National Instruments Corporation.

anted specifications describe the performance of a model under stated operating conditions and are covered by the model warranty. Warranted specifications account for measurement uncertainties, temperature drift, and aging. Warranted specifications are ensured by design or verified during production and calibration. Characteristics describe values that are relevant to the use of the model under stated operating conditions but are not covered by the model warranty. ■ Typical specifications describe the performance met by a majority of models. ■ Nominal specifications describe an attribute that is based on design, conformance testing, or supplemental testing. ■ Measured specifications describe the measured performance of a representative model. Specifications are Typical unless otherwise noted. Conditions Specifications are valid under the following conditions unless otherwise noted. ■ All filter settings ■ All impedance selections ■ Sample clock set to 60 MS/s Warranted specifications are valid under the following conditions unless otherwise noted. ■ Temperature range of 0 °C to 55 °C © National Instruments 3 PXIe-5105 Specifications

■ The PXIe -5105 module is warmed up for 15 minutes at ambient temperature ■ Calibration cycle is maintained ■ The PXI Express chassis fan speed is set to HIGH, the foam fan filters are removed if present, and the empty slots contain PXI chassis slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users available at ni.com/manuals. ■ Ex ternal calibration is performed at 23 °C ± 3 °C Vertical Analog Input Number of channels Eight (simultaneously sampled) Input type Referenced single-ended Connectors SMB Impedance and Coupling Input impedance 50 Ω 50 Ω ±2%

1 MΩ 1 MΩ ±1% in parallel with a nominal capacitance of 50 pF

Input coupling AC[1], DC Voltage Levels Full-scale (FS) input range ni.com4 PXIe-5105 Specifications

1 MΩ only 30 V

1 MΩ |Peaks| ≤42 V

Table 1. DC Accur acy[2] Table 2. AC Amplitude Accuracy[3]

Table 3. Crosstalk[4]

0.2 V, 1 V, 6 V, and 30 V ≤-80 dB at 1 MHz

Table 4. Bandwidth (-3 dB)

0.2 V, 1 V, and 6 V 60 MHz

0.2 V, 1 V, 6 V, and 30 V 60 MHz

Figure 1. Frequency Response, 50 Ω, 1 Vpk -pk Input Range, Measured

1 MΩ Spectr

Table 9. Spurious-F ree Dynamic Range (SFDR)

0.2 V 70 dBc

1 V and 6 V 65 dBc

Table 10. Total Harmonic Distortion (THD)

0.05 V -72 dBc

0.2 V -75 dBc

1 V -65 dBc

6 V -68 dBc

Table 11. Signal to Noise and Dis tortion (SINAD)

0.05 V 50 dB

0.2 V 59 dB

1 V 61 dB

6 V 59 dB

1 MΩ Noise

Table 12. 1 MΩ RMS Noise[7]

Table 9. Spurious-F ree Dynamic Range (SFDR)[8]

0.2 V 72 dBc

1 V and 6 V 72 dBc

Table 10. Total Harmonic Distortion (THD)[8] Table 11. Signal to Noise and Dis tortion (SINAD)[8]

0.05 V 59 dB

0.2 V t

Figure 2. PXIe-5105 Dynamic Performance, 50 Ω, 1 Vpk -pk, with 24 MHz Filter

5 MHz, –1 dBFS input signal

Figure 1. PXIe-5105 Spectral Noise Density, 50 Ω, 0.05 Vpk -pk, with Anti-Alias Filter

30 M10 k 100 k 1 M 10 M

Table 12. 50 Ω RMS Noise[9]

24 MHz bandwidth filter disabled ≤500 ps

24 MHz bandwidth filter enabled ≤600 ps

Internal Onboard clock (internal VCXO)[11] External PFI 1 PXI S tar External frequency range 4 MHz to 65 MHz Exporting[12] Destination PFI 1 Maximum frequency 65 MHz Onboard Clock (Internal VCXO) Real-time sample rate range[13] 915.5 S/s to 60 MS/s Timebase frequency 60 MHz Timebase accuracy Not phase-locked to Reference clock ±25 ppm, warranted Phase-locked to Reference clock Equal to the Reference clock accuracy Sample clock delay range ±1 Sample clock period ni.com10 PXIe-5105 Specifications

Sample clock delay resolution <10 ps External Sample Clock Sources PFI 1 PXI S tar Frequency range[14] 4 MHz to 65 MHz Duty cycle tolerance 45% to 55% Phase-Locked Loop (PLL) Reference Clock Sources PXI_CLK10 (backplane connector) PFI 1 (front p anel SMB connector) Frequency range[15] 5 MHz to 20 MHz in 1 MHz increments Duty cycle tolerance 45% to 55% Exported Reference clock destination PFI 1 Triggers Ref erence (Stop) Trigger Supported trigger Reference (stop) trigger Trigger types Edge © National Instruments 11 PXIe-5105 Specifications

Trigger sources CH 0 to CH 7 PFI 1 PXI_T rig <0..6> Software Time resolution Sample clock timebase period Minimum rearm time[16] Internal Onboard clock 2.4 µs External Sample clock 144 × External clock period Holdoff From rearm time up to [(232 - 1) × Sample clock timebase period] Delay F rom 0 up to [(232 - 1) - Reques ted posttrigger samples] × (1/Actual sample rate), in seconds Related information ■ Refer to the NI High-Speed Digitizers Help for more information about which trigger sources are available for each trigger type. ni.com12 PXIe-5105 Specifications

Sources CH 0 to CH 7 (front panel SMB connectors) Trigger level range 100% FS Edge trigger sensitivity 2% FS Trigger jitter Sample clock timebase period Digital Trigger Trigger type Digital Sources PFI 1 (front panel SMB connector) PXI_TRIG <0..6> (b ackplane connector) Programmable Function Interface Connector PFI 1 (front panel SMB connector) Direction Bidirectional Coupling AC © National Instruments 13 PXIe-5105 Specifications

As a Sample Clock or Reference Clock Input voltage range Sine wave 0.65 Vpk-pk t o 2.8 V pk-pk (0 dBm to 13 dBm) Square wave 0.2 V pk-pk t o 2.8 V pk-pk Maximum input overload 7 Vrms with |Pe aks| ≤10 V Input impedance 50 Ω Coupling AC As an Input (Digital Trigger) Destinations Start trigger (acquisition arm) Ref erence (stop) trigger Arm Reference trigger Advance trigger Input impedance 150 kΩ, nominal VIH 2.0 V VIL 0.8 V ni.com14 PXIe-5105 Specifications

Maximum input overload -0.5 V, 5.5 V Maximum frequency 65 MHz As an Output Sources Start trigger (acquisition arm) Refer ence (stop) trigger End of record Done (end of acquisition) Sample clock timebase Reference clock Output impedance 50 Ω Logic type 3.3 V CMOS Maximum drive current ±24 mA Waveform Specifications Onboard memory size options [17] 128 MB Minimum record length 1 sample © National Instruments 15 PXIe-5105 Specifications

Number of samples[18] Pretrigger Zero up to full record length Posttrigger Zero up to full record length Allocated onboard memory per rec ord[19] [(Record length in samples × 2 bytes/sample × number o f enabled channels) + 480] rounded up to the nearest 256 bytes Calibration Ex ternal Calibration External calibration calibrates the onboard references used in self-calibration and the external trigger levels. All calibration constants are stored in nonvolatile memory. Self-Calibration Self-calibration is done on software command. The calibration corrects for gain, offset, triggering, and timing errors for all input ranges. Calibration Specifications Interval for external calibration 2 years Warm-up time [20] 15 minutes Software Driver Software Driver support for this device was first available in NI-SCOPE14.1. ni.com16 PXIe-5105 Specifications

NI-SCOPE is an IVI-compliant driver that allows you to configure, control, and calibr ate the PXIe-5105. NI-SCOPE provides application programming interfaces for many development environments. Application Software NI-SCOPE provides programming interfaces, documentation, and examples for the following application development environments: ■ LabVIEW ■ LabWindows™/CVI™ ■ Measurement Studio ■ Microsoft Visual C/C++ ■ .NET (C# and VB.NET) Interactive Soft Front Panel and Configuration When you install NI-SCOPE on a 64-bit system, you can monitor, control, and record measurements from the PXIe-5105 using InstrumentStudio. InstrumentStudio is a software-based front panel application that allows you to perform interactive measurements on several different device types in a single program. Note InstrumentStudio is supported only on 64-bit systems. If you are using a 32-bit system, use the NI-SCOPE–specific soft front panel instead of InstrumentStudio. Interactive control of the PXIe-5105 was first available via InstrumentStudio in NI-SCOPE18.0 and via the NI-SCOPE SFP in NI-SCOPE14.1. InstrumentStudio and the NI-SCOPE SFP are included on the NI-SCOPE media. NI Measurement & Automation Explorer (MAX) also provides interactive configuration and test tools for the PXIe-5105. MAX is included on the driver media. © National Instruments 17 PXIe-5105 Specifications

an use the NI TClk synchronization method and the NI-TClk driver to align the Sample clocks on any number of supported devices, in one or more chassis. For more information about TClk synchronization, refer to the NI-TClk Synchronization Help, which is located within the NI High-Speed Digitizers Help. For other configurations, including multichassis systems, contact NI Technical Support at ni.com/support. Intermodule SMC Synchr onization Using NI-TClk for Identical Modules Specifications are valid for modules installed in one NI PXI-1042 chassis. These specifications do not apply to PCI modules. Specifications are valid under the following conditions: ■ All parameters are set to identical values for each SMC-based module. ■ Sample clock set to 60 MS/s. ■ All filters are disabled. Note Although you can use NI-TClk to synchronize non-identical modules, these specifications apply only to synchronizing identical modules. Skew [21] 500 ps Average skew after manual adjustment [22] <10 ps Sample clock adjustment resolution <10 ps Power Current draw +3.3 V DC 1.5 A ni.com18 PXIe-5105 Specifications

Figure 1. PXIe-5105 Table 13. PXIe-5105 Front Panel Connectors CH 0—CH 7 SMB jack Analog input connection; digitizes data and triggers acquisitions.

Maximum altitude 2,000 m (800 mbar) (at 25 °C ambient temperature) Pollution Degree 2 Indoor use only. Operating Envir onment Ambient temperature range 0 °C to 40 °C Relative humidity range 10% to 90%, noncondensing Storage Environment Ambient temperature range -40 °C to 71 °C Relative humidity range 5% to 95%, noncondensing Shock and Vibration Operating shock 30 g peak, half-sine, 11 ms pulse Random vibration Operating 5 Hz to 500 Hz, 0.3 grms Nonoperating 5 Hz to 500 Hz, 2.4 grms ni.com20 PXIe-5105 Specifications

Compliance and Certifications Safe ty Compliance Standards This product is designed to meet the requirements of the following electrical equipment safety standards for measurement, control, and laboratory use: ■ IEC 61010-1, EN 61010-1 ■ UL 61010-1, CSA C22.2 No. 61010-1 Note For safety certifications, refer to the product label or the Product Certifications and Declarations section. Elec tromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement, control, and laboratory use: ■ EN 61326-1 (IEC 61326-1): Class A emissions; Basic immunity ■ EN 55011 (CISPR 11): Group 1, Class A emissions ■ EN 55022 (CISPR 22): Class A emissions ■ EN 55024 (CISPR 24): Immunity ■ AS/NZS CISPR 11: Group 1, Class A emissions ■ AS/NZS CISPR 22: Class A emissions ■ FCC 47 CFR Part 15B: Class A emissions ■ ICES-001: Class A emissions Note In the United States (per FCC 47 CFR), Class A equipment is intended for use in commercial, light-industrial, and heavy-industrial locations. In Europe, Canada, Australia, and New Zealand (per CISPR 11), Class A equipment is intended for use only in heavy-industrial locations. © National Instruments 21 PXIe-5105 Specifications

Note Group 1 equipment (per CISPR 11) is any industrial, scientific, or medic al equipment that does not intentionally generate radio frequency energy for the treatment of material or inspection/analysis purposes. Note For EMC declarations, certifications, and additional information, refer to the Product Certifications and Declarations section. Pr oduct Certifications and Declarations Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for NI products, visit ni.com/product-certifications, sear ch by model number, and click the appropriate link. Environmental Management NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers. For additional environmental information, refer to the Engineering a Healthy Planet web page at ni.com/environment. This pag e contains the environmental regulations and directives with which NI complies, as well as other environmental information not included in this document. EU and UK Customers Waste Electrical and Electronic Equipment (WEEE)—At the end of the pr oduct life cycle, all NI products must be disposed of according to local laws and regulations. For more information about how to recycle NI products in your region, visit ni.com/ environment/weee. ni.com22 PXIe-5105 Specifications

电子信息产品污染控制管理办法(中国 RoHS )

  • 5 40e 中国 RoHS— NI 符合中国电子信息产品中限制使用某些有害物 质指令(RoHS) 。关于 NI 中国 RoHS 合规性信息,请登录 ni.com/environment/ rohs_china。(For information about China RoHS compliance, go to ni.com/ environment/rohs_china.) 1 AC c oupling available on 1 MΩ input only. 2 Within ±5 °C of self-c alibration temperature.

3 For a 50 kHz signal with amplitude 90% of full-sc ale input range measured within

±5 °C of self-calibration temperature.

4 Measur ed from one channel to another channel, with same range settings on both

channels. 5 AC c oupling available on 1 MΩ input only. 6 -1 dBFS input signal. Includes the sec ond through the fifth harmonics. 24 MHz bandwidth filter enabled. 7 Verified using a 50 Ω t erminator connected to input. 8 -1 dBFS input signal. Includes the sec ond through the fifth harmonics. 24 MHz bandwidth filter enabled. 9 Verified using a 50 Ω t erminator connected to input. 10 10 MHz sine input signal.

11 Internal Sample clock is lock ed to the Reference clock or derived from the

onboard VCXO. 12 You c annot export a decimated Sample clock signal. © National Instruments 23 PXIe-5105 Specifications

13 Divide by n decimation used for all rates less than 60 MS/s. For more information about the Sample clock and decimation, refer to the NI High-Speed Digitizers Help. 14 Divide by n decimation available where 1 ≤ n ≤ 65,535. For more information about the Sample clock and decimation, refer to the NI High-Speed Digitizers Help. 15 Default of 10 MHz. The PLL R eference clock frequency must be accurate to ±50 ppm. 16 Holdoff set t o 0. Onboard Sample clock at maximum rate. 17 Onboar d memory is shared between all enabled channels. 18 Single-r ecord and multirecord acquisitions. 19 The maximum number of rec ords is 100,000. 20 Warm-up time be gins after the NI-SCOPE driver is loaded. Unless manually disabled, the NI-SCOPE driver automatically loads with the operating system and enables the module. 21 Caused by clock and analog p ath delay differences. No manual adjustment performed.

22 For mor e information about manual adjustment, refer to the Synchronization

Repeatability Optimization topic in the NI-TClk Synchronization Help. ni.com24 PXIe-5105 Specifications © 2022 National Instruments Corporation.