Technical content
Datasheet sections
Read the NI-9232 Safety, Environmental, and Regulatory Information and complete the software and hardware installation procedures in your chassis documentation. Safety Guidelines Caution Observe all instructions and cautions in the user documentation. Using the product in a manner not specified can damage the product and compromise the built-in safety protection. Attention Suivez toutes les instructions et respectez toutes les mises en garde de la documentation d'utilisation. L'utilisation du produit de toute autre façon que celle spécifiée risque de l'endommager et de compromettre la protection de sécurité intégrée. Safety Voltages Connect only voltages that are within the following limits. Isolation Channel-to-channel None Channel-to-earth ground Continuous 60 V DC, Measurement Category I Withstand 1,000 V RMS, verified by a 5 s dielectric withstand test Safety Guidelines for Hazardous Locations The NI-9232 is suitable for use in Class I, Division 2, Groups A, B, C, D, T4 hazardous locations; Class I, Zone 2, AEx nA IIC T4 Gc and Ex nA IIC T4 Gc hazardous locations; © National Instruments 3 NI-9232 Getting Started
and nonhazardous locations only. Follow these guidelines if you are installing the NI-9232 in a potentially explosive environment. Not following these guidelines may result in serious injury or death. Caution Do not disconnect I/O-side wires or connectors unless power has been switched off or the area is known to be nonhazardous. Caution Do not remove modules unless power has been switched off or the area is known to be nonhazardous. Caution Substitution of components may impair suitability for Class I, Division 2, or Zone 2. Caution The system must be installed in an enclosure certified for the intended hazardous (classified) location, having a tool secured cover/door, where a minimum protection of at least IP54 is provided. Caution For Division 2 and Zone 2 applications, connected signals must be within the following limits. Capacitance 0.2 µF maximum Special Conditions for Hazardous Locations Use in Europe and Internationally The NI-9232 has been evaluated as Ex nA IIC T4 Gc equipment under DEMKO 12ATEX 1202658X and is IECEx UL 14.0089X certified. Each NI-9232 is marked II 3G and is suitable for use in Zone 2 hazardous locations, in ambient temperatures of -40 °C ≤ Ta ≤ 70 °C. If you are using the NI-9232 in Gas Group IIC hazardous locations, you must use the device in an NI chassis that has been evaluated as Ex nC IIC T4, Ex IIC T4, Ex nA IIC T4, or Ex nL IIC T4 equipment. ni.com4 NI-9232 Getting Started
Caution Transient protection shall be provided that is set at a level not exceeding 140% of the peak rated voltage value of 85 V at the supply terminals to the equipment. Caution The system shall only be used in an area of not more than Pollution Degree 2, as defined in IEC/EN 60664-1. Caution The system shall be mounted in an ATEX/IECEx-certified enclosure with a minimum ingress protection rating of at least IP54 as defined in IEC/EN 60079-15. Caution The enclosure must have a door or cover accessible only by the use of a tool. Electromagnetic Compatibility Guidelines This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic environment. This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a peripheral device or test object, or if the product is used in residential or commercial areas. To minimize interference with radio and television reception and prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation. Furthermore, any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules. Notice To ensure the specified EMC performance, operate this product only with shielded cables and accessories. Do not use unshielded cables or © National Instruments 5 NI-9232 Getting Started
accessories unless they are installed in a shielded enclosure with properly designed and shielded input/output ports and connected to the product using a shielded cable. If unshielded cables or accessories are not properly installed and shielded, the EMC specifications for the product are no longer guaranteed. Caution Electrostatic Discharge (ESD) can damage the NI-9232. To prevent damage, use industry-standard ESD prevention measures during installation, maintenance, and operation. Caution To ensure the specified EMC performance for the NI-9232 with screw terminal, you must install clamp-on ferrite beads (part number 782802-01) in accordance with the product installation instructions. Refer to the NI-9232 product page on ni.com for purchasing information about clamp-on ferrite beads. Notice To ensure the specified EMC performance, the length of all I/O cables must be no longer than 3 m (10 ft). Cable Requirements for EMC Compliance Select and install cables for the NI-9232 in accordance with the following requirements: ■ Connect the cable shield to the chassis ground (grounding screw of the chassis). ■ For the NI-9232 with screw terminal, install a clamp-on ferrite bead (part number 782802-01) on the input cable for each channel that you are connecting to on the NI-9232. ■ For the NI-9232 with screw terminal, clamp-on ferrite beads must be installed on the cable as close to the module as possible. Placing the ferrite elsewhere on the cable noticeably impairs its effectiveness. ni.com6 NI-9232 Getting Started
Figure 1. Cable Connections for EMC Compliance model number, and click the appropriate link. ensure that the desired EMC performance is attained.
Figure 2. NI-9232 Pinout path and the signal ground reference. Table 1. Signal Descriptions You can connect ground-referenced or floating signal sources to the NI-9232.
Visit ni.com/support to find support resources including documentation, downloads, and troubleshooting and application development self-help such as tutorials and examples. Visit ni.com/services to learn about NI service offerings such as calibration options, repair, and replacement. Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI. NI corporate headquarters is located at 11500 N Mopac Expwy, Austin, TX, 78759-3504, USA. ni.com10 NI-9232 Getting Started © 2022 National Instruments Corporation.