Edition • 2023-07-21T16:36:13+02:00 Temperature measuring amplifier 2-channel IM12-TI02-2TCURTDR-2I-C0/24VDC The temperature transducer IM12- TI02-2TCURTDR-2I-C0/24VDC transmits temperature-dependent measured values gal- vanically isolated. The devices are suitable for operation in zone 2. The 2-channel device has two inputs for ther- mocouples acc. to IEC 60584, DIN 43710, GOST R 8.585-2001, low voltages (-150… +150 mV), RTDs acc. to IEC 60751, DIN 43760, GOST 6651-94 (2-, 3-wire) and resis- tors 0…5 kò (2-, 3-wire). Two 0/4 … 20 mA current outputs are available on the output side. The device is parameterized via FDT and IODD with a PC. The cold junction compen- sation can be set to internal, external, or to a constant value. The current outputs can be set to source or sink 0/4…20 mA. The inputs can be freely assigned to the outputs. In ac- cordance with the parameterization (E1, E2, E1 - E2 or E2 - E1), the input signals are pro- vided as a 0/4…20 mA standard current sig- nal. The devices have a green power LED (Pwr) and a red LED to indicate internal faults. For each input circuit there is a yellow and a red status LED. A fault in the input circuit leads to a flashing red LED according to NE44, an internal error to a steady red LED. The fault current can be adjusted to < 3.5 mA or > 21.5 mA. The device can be used in safety circuits up to SIL2 (high and low demand according to IEC 61508) and meets the requirements of the NE21. It is equipped with removable screw terminals. The device is equipped with removable screw terminals. ■ Input circuits monitored for wire-break and short-circuit ■ Parameterized via PC ■ Complete galvanic isolation ■ Removable screw terminals ■ ATEX use in Zone 2, cUL ■ SIL 2
Edition • 2023-07-21T16:36:13+02:00 Type IM12-TI02-2TCURTDR-2I-C0/24VDC ID 7580532 Nominal voltage 24 VDC Operating voltage 10…30 VDC Power consumption ð 2.7 W Power dissipation, typical ð 1.6 W Input circuits RTD Type DIN EN 60751 Pt50, Pt100, Pt 500, Pt1000 RTD Type DIN EN 43760 Ni50, Ni100, Ni500, Ni1000 RTD Type Gost 6651-94 Pt50, Pt100, Pt 500, Pt1000, CU50, Cu53, Cu100, CU500, CuZn100 TC Type DIN EN 60584 Type A, Type B, Type C, Type E, Type J, Type K, Type N, Type R, Type S, Type T TC Type DIN 43710 Type L TC Type Gost 8.585-2001Type A1, Type A2, Type A3, Type L, Type M Low voltage input -150…150 mV Resistance input 0…5000 ohms Reference temperature 23 °C Output circuits Output current 2 × source/sink (15…28 V) 0/4…20 mA Load resistance current output ð 0.8 kò Response characteristic Reference temperature 23 °C Measuring accuracy current output (including linearity, hysteresis and repeatability) ± 10 µA Temperature drift analog output 0.0025 %/K Accuracy, RTD input, 0…500 ohm ± 50 mò Temperature drift, RTD input, 0…500 ohm ± 5 mò/K Accuracy, RTD input, 500…5000 ohm ± 500 mò Temperature drift, RTD input, 500…5000 ohm ± 30 mò/K Measuring accuracy TC input (including linearity, hys- teresis and repeatability) ± 15 µV Temperature drift, TC input ± 3.2 µV/K Cold junction compensation error with cold junction compensation < 2 K Note With a 3-wire connection, the errors double Galvanic isolation Test voltage 2.5 kV RMS E1,E2-A1A,A2A 375 V peak value acc. to EN 60079-11 E1,E2 supply voltage 375 V peak value acc. to EN 60079-11 A1A supply voltage 300 V RMS acc. to EN 50178 and EN 61010-1 A2A supply voltage 300 V RMS acc. to EN 50178 and EN 61010-1 Important note For Ex-applications the values specified in the corre- sponding Ex certificates (ATEX, IECEx, UL, etc.) ap- ply. Important note If the device is used in applications to achieve func- tional safety according to IEC 61508, the safety manual must be used. Information in the data sheet are not valid for functional safety. Use in SIL safety circuits SIL 2 acc. to IEC 61508 Dimensions
Edition • 2023-07-21T16:36:13+02:00 Displays/Operating elements Operational readiness Green Switching state Yellow Error indication red Mechanical data Protection class IP20 Flammability class acc. to UL 94 V-0 Ambient temperature -25…+70 °C Storage temperature -40…+80 °C Dimensions 120 x 12.5 x 117 mm Weight 163 g Mounting instructions DIN rail (NS35) Housing material Polycarbonate/ABS Electrical connection Removable screw terminals, 2-pin Terminal cross-section 0.2…2.5 mm² (AWG: 24…14) Tightening torque 0.5 Nm Tightening torque 4.43 LBS-Inch Environmental conditions Operating height Up to 2000 m above sea level Pollution degree II Surge/Overvoltage cate- gory II (EN 61010-1) Standards used Voltage resistance and insulation EN 50178 EN 61010-1 EN 50155 GL VI-7-2 Shock EN 61373 class B EN 50155 GL VI-7-2 EN 60068-2-6 EN 60068-2-27 Temperature EN 60068-2-1 Ad EN 50155 GL VI-7-2 EN 60068-2-2 Bd EN 60068-2-1 Air humidity EN 60068-2-38 EMC EN 50155 GL VI-7-2 NE21 EN 61326-1 EN 61326-3-1 EN 61000-4-2 EN 61000-4-3 EN 61000-4-4 EN 61000-4-5 EN 61000-4-6 EN 61000-4-11 EN 61000-4-29 EN 55011 EN 55016 EN 50121-3-2 EN 61000-6-2
Edition • 2023-07-21T16:36:13+02:00 Accessories Type code Ident-No. Dimension drawing USB-2-IOL-0002 6825482 IO-Link Master with integrated USB port IOL-COM/3M 7525110 IO-Link communication line for connecting IO-Link devices to an IO-link master via a 3.5-mm jack plug IMX12-SC-2X-4BK 7580940 Screw terminals for IM(X)12 modules; included in delivery: 4 pcs. of 2-pin black terminals IMX12-CC-2X-4BK 7580942 Spring terminals for IM(X)12 modules; included in delivery: 4 pcs. black terminals, 2-pin IMX12-2-CJT 100003646