March 2013, Rev. 1, 08/13 ©2013 Fluke Corporation. All rights reserved. Spec ifications are subject to change without notice. All product names are trademarks of their respective companies.

LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is two years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries, or to any product which, in Fluke's opinion, has been misused, altered, neglected, contaminated, or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with its functional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free or operate without interruption. Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available only if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one country is submitted for repair in another country. Fluke's warranty obligation is limited, at Fluke's option, to refund of the purchase price, free of charge repair, or replacement of a defective product which is returned to a Fluke authorized service center within the warranty period. To obtain warranty service, contact your nearest Fluke authorized service center to obtain return authorization information, then send the product to that service center, with a description of the difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination, alteration, accident, or abnormal condition of operation or handling, including overvoltage failures caused by use outside the product’s specified rating, or normal wear and tear of mechanical components, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, the product will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point). THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY. Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, the limitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not affect the validity or enforceability of any other provision. Fluke Corporation P.O. Box 9090 Everett, WA 98206-9090 U.S.A. Fluke Europe B.V. P.O. Box 1186

5602 BD Eindhoven

i Table of Contents Title Page

v List of Figures Figure Title Page

The 572-2 Infrared Thermometer (the Product) is for non-contact temperature measurement. The Product determines the surface temperature of an object by measuring the amount of infrared energy radiated by the object surface. The Product also uses a K-type thermocouple for contact temperature measurement. Contact Fluke To contact Fluke, call one of the following telephone numbers:

  • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
  • Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
  • Canada: 1-800-36-FLUKE (1-800-363-5853)
  • Europe: +31 402-675-200
  • Japan: +81-03-6714-3114
  • Singapore: +65-738-5566
  • Anywhere in the world: +1-425-446-5500 Or, visit Fluke's website at www.fluke.com. To register your product, visit http://register.fluke.com. To see, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.

A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test. Symbols used on the Product and in this manual are explained in Table 1. Laser safety markings are shown in Figure 1. Warning To prevent possible electrical shock, fire, eye damage, and personal injury:

  • Read all safety Information before you use the Product.
  • Use the Product only as specified, or the protection supplied by the Product can be compromised.
  • Do not use the Product around explosive gas, vapor, or in damp or wet environments.
  • Do not use the Product if it operates incorrectly.
  • See emissivity information for actual temperatures. Reflective objects result in lower than actual temperature measurements. These objects pose a burn hazard.
  • Do not look directly into the laser with optical tools (for example, binoculars, telescopes, microscopes). Optical tools can focus the laser and be dangerous to the eye.
  • Do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
  • Use the product only as specified, or hazardous laser radiation exposure can occur.
  • Do not use laser viewing glasses as laser protection glasses. Laser viewing glasses are used only for better visibility of the laser in bright light.
  • Do not open the Product. The laser beam is dangerous to eyes. Have the Product repaired only through an approved technical site.
  • Have an approved technician repair the product. Caution For safe operation and maintenance of the Product:
  • Remove the batteries if the Product is not used for an extended period of time, or if stored in temperatures above 50 °C. If the batteries are not removed, battery leakage can damage the Product.
  • Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
  • Have the Product repaired before use if the batteries leak.
  • Be sure that the battery polarity is correct to prevent battery leakage.
  • Do not short the battery terminals together.
  • Do not keep cells or batteries in a container where the terminals can be shorted.
  • Do not put battery cells and battery packs near heat or fire. Do not put in sunlight. Caution To avoid damage to the Product or the equipment under test, protect them from:
  • EMF (electro-magnetic fields) from arc welders, induction heaters, etc.
  • Static electricity
  • Thermal shock (caused by large or abrupt ambient temperature changes. For highest accuracy, allow 30 minutes for Thermometer to become stable before use).

Table 1. Symbols  Hazardous voltage. Risk of electrical shock.  Risk of danger. Important information. See manual.  Conforms to European Union directives. This product complies with the WEEE Directive (2002/96/EC) marking requirements. unsorted municipal waste. Go to Fluke’s website for recycling information.  Conforms to relevant South Korean EMC Standards. Conforms to China Metrology Certification.

hdp01.eps Figure 1. Safety Markings

Features

  • Dual laser sighting
  • Backlit display
  • Current Temperature plus MAX, MIN, DIF, AVG temperature displays
  • Two AA batteries
  • Hard case
  • 80PK-1 K-type thermocouple probe
  • Adjustable emissivity and predefined emissivity table
  • Infrared and thermocouple temperature display
  • Celsius or Fahrenheit temperature display
  • Tripod mount
  • Auto off
  • Standard miniature K-type thermocouple connector input
  • Printed 572-2 Getting Started
  • 572-2 Manuals CD
  • 12 or 24 hour clock
  • Last reading Hold (20 seconds)
  • Multi-language interface
  • High and low temperature alarm
  • Data storage and review
  • Trigger lock
  • USB 2.0 computer interface cable
  • FlukeView ® Forms Documenting Software
  • English
  • Spanish
  • French
  • German
  • Portuguese
  • Simplified Chinese To change the displayed language, refer to “Setup”. Menu Overview There are many settings that can be changed with the menu. Figure 2 shows the LCD and menu interface. Selecting the Menu softkey advances the menu to the next level. Table 2 is a top-level description of the menu. hdp02.eps

Figure 2. Menu Navigation Table 2. Top-Level Menu Description

1 Save Save reading to memory Menu Light Adjust backlight

2 Mem Review/delete memory

3 MnMx Enables Min/Max Menu Avg Enable Avg/Diff

6 Setup

Each menu item and function is explained in greater detail in the following sections.

To save readings: 1. Pull the trigger to take a measurement. 2. Release the trigger to stop taking the measurement. 3. Push the Save softkey to enter the Save menu. 4. Push the Yes softkey to save the reading. The reading is assigned a memory location and a time and date stamp. The reading includes:

  • IR temperature
  • Emissivity
  • Date/Time You can also push the Cancel softkey to abort saving the reading. Light The Product has a backlit display with two brightness levels. The Light softkey is used to adjust the backlight brightness. The backlight is on each time the trigger is pulled. To toggle the backlight brightness, push the Light softkey. The backlight can be disabled using the Setup menu. See “Setup” for more information. Memory The Product can store measurement records including time, date, emissivity, and measurement record numbers (see “Save” for more information). The Product can store 99 records. To access records stored in memory: 1. Push the Menu softkey until Mem shows as the left softkey function. 2. Push the Mem softkey to access the Memory menu. Saved readings can be read. Emissivity Menu The Emissivity menu includes a list of pre-defined materials and lists their typical emissivity values, see Table 3. For further information, see “Emissivity”. Note Default emissivity is 0.95. To access the Emissivity menu: 1. Push the Menu softkey until  (emissivity) shows as the right softkey function. 2. Push the  softkey.

No. softkey to manually enter the typical emissivity of a material.

  • If the Emissivity table is accessed, a list of materials and their suggested emissivity is shown. 1. Use the down arrow to navigate through the list. 2. Push the Enter softkey to choose the desired material.
  • To enter an emissivity value manually: 1. Push the No. softkey. 2. Use the down or up arrow softkey to change the entry. Hold down the up or down arrow softkey to increase the rate of change. 3. Push the Done softkey when finished to return to the main menu.

Table 3. Nominal Surface Emissivity Highlighted items may also be found in the emissivity table built into the Product. softkey to change between the measurements.

Min, Max, Avg, Differential The Product can measure minimum (MIN), maximum (MAX), average (AVG), or differential (Δ) temperatures each time a reading is taken. These values are not shown if a thermocouple is plugged into the Product. To turn on the Min Max mode: 1. Push the Menu softkey until MnMx (Min Max) shows as the left softkey function. 2. Push the MnMx softkey. The display shows the present reading, maximum and minimum readings, and the emissivity setting. To turn on the Avg/Dif mode: 1. Push the Menu softkey until Avg shows as the right softkey function. 2. Push the Avg softkey. The display shows the present reading, average reading, the differential reading between max and min (designated by Δ), and the emissivity setting. Alarm The Product has a programmable high and low alarm to designate high or low readings, depending on the thresholds entered. When the alarm level is reached, an alarm will sound and the display will flash orange and white. To set either the high or low alarm: 1. Push the Menu softkey until Alarm shows as the right softkey function. 2. Push the Alarm softkey to access the Alarm menu. 3. Push either the Hi or Lo softkey, depending on which alarm will be set. 4. Push the ON softkey to turn the alarm on. 5. Push the OFF softkey to turn the alarm off. 6. Use the Set softkey to access the Hi or Lo Alarm Set menu. 7. Use the down or up softkeys to change the alarm setting. 8. Once the necessary settings have been entered, push the Done softkey. Trigger Lock The Product trigger can be locked on for continuous measurement. To enable continuous measurement mode, hold down the trigger for a few seconds until the  symbol shows. To lock the trigger: 1. Push the Menu softkey until the lock symbol () shows as the left softkey function. 2. Push the  softkey to lock the trigger. The lock symbol shows on the display to designate a locked trigger. When the trigger is locked, the  softkey changes to . Push this softkey to unlock the trigger.

Warning To prevent eye damage and personal injury, do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces. The laser of the Product is to be used for aiming purposes only. The laser turns off when the trigger is released. To enable or disable the laser: 1. Push the Menu softkey until Laser shows as the right softkey function. 2. Push the Laser softkey to enable or disable the laser.  shows on the display when the laser is enabled. Setup From the Setup menu, the backlight, time and date, and display language can be changed. Language To change the display language: 1. From the main menu, push the Menu softkey until Setup shows as the left softkey function. 2. Push the Setup softkey. 3. Use the down arrow softkey to move the indicator to Language. 4. Push the Enter softkey. 5. Use the down arrow to move the indicator to the desired language. 6. Push the Enter softkey to complete the language selection or push the Back softkey to return to the Setup menu. Backlight In normal use, the backlight is always on. Use this menu to change the backlight setting from on to off. Turn the backlight off to conserve battery power. 1. Push the Menu softkey until Setup shows as the left softkey function. 2. Push the Setup softkey. 3. Backlight has an indicator next to it. To enter the backlight menu, push the Enter softkey. 4. Push the OFF softkey to turn the backlight off, push the ON softkey to turn it on. 5. Push the Back softkey to return to the Setup menu. Time/Date To change the time on the Product: 1. Push the Menu softkey until Setup shows as the left softkey function. 2. Push the Setup softkey to enter the Setup menu. 3. Push the down arrow softkey to select Time/Date. 4. Push the Enter softkey. 5. Push the Time softkey.

  1. Push the necessary time format softkey ( 24hr or 12hr). 7. Use the up and down softkeys to change the selected hour to the correct hour. 8. Push the Next softkey to select the minutes. 9. Use the up and down softkeys to change the selection to the desired minute. 10. When in 12-hour mode, push the Next softkey to highlight the am/pm parameter. 11. Use the up and down softkey to change to am or pm. 12. Push the Done softkey when finished. The display reverts to the first page of the Time/Date menu. To change the date on the Product: 1. From the main menu, push the Menu softkey until Setup shows as the left softkey function. 2. Push the Setup softkey. 3. Push the down arrow softkey to select Time/Date. 4. Push the Enter softkey. 5. Push the Date softkey. 6. Select the date format: Day/Month/Year ( dmy) or Month/Day/Year (mdy). 7. Use the up and down softkeys to change the selected parameter. 8. Push the Next softkey and the arrow softkeys to select the month, date, or year parameters. 9. Use the up and down softkeys to change the selected parameter. 10. Push the Next softkey to move through each parameter. 11. Push the Done softkey when finished. The display reverts to the start of the Time/Date menu. Delete Data To delete stored data from the Product, from the main menu, push the menu softkey until Mem shows as the left softkey function. The last memory location shows on the display. To access the Delete menu, push the Delete softkey. From here, all of the Product memory or individual records can be deleted. Delete All Data To delete all of the records: 1. Push the All softkey. 2. At the confirmation screen, push the Yes softkey. Delete Individual Data Records To delete individual records: 1. Push the View softkey. 2. Use the down and up arrow softkeys to access the desired record. 3. Once the desired record is shown, push the Yes softkey to delete the record. 4. To cancel data deletion, pull the trigger.

spot size. The smaller the target, the closer you should be to it (see Figure 6). Figure 6. Field of View 0.95, the default setting for the Product. the tape or painted surface. probe is not inserted, release the trigger until HOLD shows on the display.

  • Record number
  • IR and probe temperature in °F or °C
  • Date/Time
  • Emissivity For more information, see “Save”. Data Download The stored data can be downloaded to a personal computer (PC) with the included USB cable and the included FlukeView Forms Documenting Software. See the FlukeView® Forms documentation for details. The USB input  is located on the top of the Product next to the external probe input  (see Figure 7). Note To prevent incorrect readings, do not do a temperature measurement of an earthed conductor while the Product is connected to a PC that is earthed by a three-phase grounding plug. 1 2 hdp07.eps

Figure 7. Thermocouple and USB Connections

the external contact probe to live electrical circuits. the Product via the probe input located on the top of the Product (see Figure 7). can be used simultaneously while the Product is taking non-contact measurements. computer and FlukeView Forms, the auto off feature is disabled. See Table 4 for solutions to possible problems when using the Product. Table 4. Troubleshooting

Use the battery charge symbols to gauge the approximate level of charge left on the batteries. Note When in low battery mode, the Product does not store values. “Err” shows on the display if attempted.  Batteries are at 5 %. Before taking further readings, the batteries must be changed. Battery Replacement Warning To prevent possible electrical shock, fire, eye damage, and personal injury:

  • Do not open the Product. The laser beam is dangerous to eyes. Have the Product repaired only through an approved technical site.
  • Have an approved technician repair the Product. For safe operation and maintenance of the Product:
  • Repair the Product before use if the batteries leak.
  • Be sure that the battery polarity is correct to prevent battery leakage. Caution
  • Do not short the battery terminals together.
  • Do not keep cells or batteries in a container where the terminals can be shorted.
  • Do not put battery cells and battery packs near heat or fire. Do not put in sunlight. Note The batteries should be replaced in less than one minute after removal to avoid manually re-initializing the Product clock and date. To install or change the two AA batteries (see Figure 8): 1. Slide the locking tab on the battery door to the “ ” position and then open the handle. 2. Insert the batteries noting their correct polarity. 3. Close and lock the handle.

Figure 8. Battery Replacement Use soap and water on a damp sponge or soft cloth. See Table 5 for a list of replaceable parts. Table 5. Replaceable Parts

  • Soft Carrying Case (H6)
  • Calibration Certification
  • All K-type thermocouple probes with standard mini-connector. See Table 7 for more information. Recommended Temperature Probes Warning To avoid electrical shock or personal injury, do not connect the external contact probe to live electrical circuits. See Table 6 for a list of recommended temperature probes.

Table 6. Recommended Temperature Probes and air temperatures within ducts, vent temperatures. quick, accurate refrigerant temperatures. ducts, and vent temperatures. thermocouple to a pipe while keeping hands free. thermometer wells, vent temperatures, and for penetrating pipe insulation.

IR Temperature Range -30 °C to 900 °C (-22 °F to 1652 °F) IR Accuracy (Calibration geometry with ambient temperature 23 °C±2 °C) ≥0 °C: ± 1 °C or ± 1 % of reading, whichever is greater (≥32 °F: ± 2 °F or ± 1 % of reading, whichever is greater) IR Repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater. Display Resolution 0.1 °C / 0.1 °F Spectral Response 8 μm to 14 μm Response Time (95 %) <500 ms K-Type Thermocouple Input Temperature Range -270 °C to 1372 °C (-454 °F to 2501 °F) K-Type Thermocouple Input Accuracy (with ambient temperature 23 °C±2 °C) ≥-40 °C: ±1 % or 1 °C, whichever is greater ≥-40 °F: ±1 % or 2 °F, whichever is greater K-Type Thermocouple Repeatability ±0.5 % of reading or ±0.5 °C (±1 °F), whichever is greater K-Type Thermocouple Resolution 0.1 °C/0.1 °F Optical Distance:Spot (90 % energy) 60:1(calculated at 90 % energy) Laser sighting Offset dual laser, output <1 mW Minimum spot size 19 mm Operational Emissivity Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials (see Table 3 for list of materials and value) Probe input KTC (Should display simultaneous probe and IR temperature) Data storage 99 points Communication USB 2.0 Operating Altitude 2000 meters Storage Altitude 12,000 meters Relative Humidity 10 % to 90 % RH non-condensing up to 30 °C (86 °F) Operating Temperature 0 °C to 50 °C (32 °F to 122 °F) Storage Temperature -20 °C to 60 °C (-4 °F to 140 °F) Vibration 2.5 G, IEC 60068-2-6 Weight 0.322 kg (0.7099 lb) Power 2 AA Batteries Battery Life 8 hours with laser and backlight on; 100 hours with laser and backlight off, at 100 % duty cycle (thermometer continuously on) Standards and Agency Approval EMC: meets IEC61326-1: Portable Safety Compliance: IEC 60825-1, Class 2 Implement Standard: Q/SXAV 16

Applies to use in Korea only. Class A equipment; (Industrial Broadcasting & Communication Equipment)[1] [1] This product meets the requirement for industrial (Class A) electromagnetic wave equipment and seller or user should take notice of it. This equipment is intended t use in business environment and is not to be used in homes. K-Type Thermocouple Bead Probe Specifications Measurement Range -40 °C to 260 °C (-40 °F to 500 °F) Accuracy ±1.1 °C (±2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F). Typically within 1.1 °C (2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F) Cable Length 1 m (40 in) K-type thermocouple cable with standard miniature thermocouple connector and bead termination