48SV2 MSYSTEM | Alldatasheet
Document overview
- Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
- PDF pages: 6
Technical content
MODEL: 48SV2 48SV2 SPECIFICATIONS ES-9315 Rev.3 Page 1/6 https://www.m-system.co.jp/ Bargraph Indicators 48 Series BARGRAPH INDICATOR Functions & Features
- Small size bargraph indicator
- Displays a process variable in 51-segment bargraph LEDs
- Clear LEDs can be recognized from a distance
- Scale plate can be easily replaced
- Separable terminal block (Separable screwless spring terminal only)
- Displays upper and lower overranges.
- Zero & Span can be adjusted from the unit's front panel. 100 OR 18 (.71) (2.83) 100 (3.94) mm (inch) SPAN ZERO DOWN UP
ORDERING INFORMATION
- Code number: 48SV2-[1][2][3][4]-[5] Specify a code from below for each of [1] through [5]. (e.g. 48SV2-RV4WD-R)
- Special input range (For codes Z & 0)
- Bargraph scale (e.g. 0 – 100 %) (See ‘SCALE PLATE.’) [1] BAR LED COLOR R: Red Y: Amber G: Green B: Blue [2] MOUNTING DIRECTION V: Vertical H: Horizontal [3] INPUT Current A: 4 – 20 mA DC (Input resistance 10 Ω) B: 2 – 10 mA DC (Input resistance 20 Ω) C: 1 – 5 mA DC (Input resistance 39 Ω) D: 0 – 20 mA DC (Input resistance 10 Ω) E: 0 – 16 mA DC (Input resistance 12 Ω) F: 0 – 10 mA DC (Input resistance 20 Ω) G: 0 – 1 mA DC (Input resistance 200 Ω) H: 10 – 50 mA DC (Input resistance 5.1 Ω) Z: Specify current (See INPUT SPECIFICATIONS) Voltage 3: 0 – 1 V DC (Input resistance 1 MΩ min.) 4: 0 – 10 V DC (Input resistance 1 MΩ min.) 5: 0 – 5 V DC (Input resistance 1 MΩ min.) 6: 1 – 5 V DC (Input resistance 1 MΩ min.) 4W: -10 – +10 V DC (Input resistance 1 MΩ min.) 5W: -5 – +5 V DC (Input resistance 1 MΩ min.) 0: Specify voltage (See INPUT SPECIFICATIONS) [4] TERMINAL BLOCK S: Screwless spring terminal D: Separable screwless spring terminal [5] POWER INPUT DC Power T: 5 V DC (Operational voltage range 5 V ±10 %, ripple 10 %p-p max.) R: 24 V DC (Operational voltage range 24 V ±15 %, ripple 10 %p-p max.) SPARE PARTS
- Scale plate GENERAL SPECIFICATIONS Construction: Panel flush mounting Connection Terminal block "S": Screwless spring terminal Applicable wire size 0.13 to 1.5 mm2, stripped length 8 mm Terminal block "D": Separable screwless spring terminal Applicable wire size 0.5 to 1.5 mm2, stripped length 7 - 8 mm Housing material: Flame-resistant resin (gray) Isolation: Input to power Zero adjustment accuracy range: -10 – +10 % (front) Span adjustment accuracy range: 90 to 110 % (front) Scale plate: Flame resistant resin (white scale & characters on black base) Zero adj. indicator LED: Green LED turns on at zero adjustment Span adj. indicator LED: Green LED turns on at span adjustment. Read rate: 100 samples / sec.
MODEL: 48SV2 48SV2 SPECIFICATIONS ES-9315 Rev.3 Page 2/6 https://www.m-system.co.jp/ BARGRAPH DISPLAY LED: 55-segment LED (0 - 100% display: 51) 56 mm (2.2”) long, 3.5 mm (.14”) wide Display range: 0 - 100% (No scaling. The first or 55th segment turns on when lower/upper overrange occurs. The 2nd and 54th segments do not turn on.) Scale
- Numbers on the scale: 4 digits max. (Including decimal point and minus sign)
- Divisions: 50 max.
- Engineering unit: Max. 4 characters or 2 double-width characters INPUT SPECIFICATIONS ■ DC Current: 0 – 50 mA DC; input resistor incorporated Minimum span: 1 mA When specifying a resistance value, choose from below. 5.1 Ω, 10 Ω, 12 Ω, 20 Ω, 39 Ω, 200 Ω ■ DC Voltage: -10 – +10 V DC Input resistance: ≥ 1 MΩ Minimum span: 0.1 V Offset: Max. 1.5 times span INSTALLATION Power consumption
- DC:
0.8 W or less @ 5 V rating
1.0 W or less @ 24 V rating
Operating temperature: -5 to +55°C (23 to 131°F) Operating humidity: 10 to 90 %RH (non-condensing) Mounting: Panel flush mounting Weight: 90 g (0.2 lb) PERFORMANCE in percentage of span Accuracy: ±2 % ±1 digit Response time: ≤ 0.5 sec. Insulation resistance: ≥ 100 MΩ with 500 V DC Dielectric strength: 2000 V AC @ 1 minute (input to power to ground) STANDARDS & APPROVALS EU conformity: EMC Directive EMI EN 61000-6-4 EMS EN 61000-6-2 RoHS Directive
MODEL: 48SV2 48SV2 SPECIFICATIONS ES-9315 Rev.3 Page 3/6 https://www.m-system.co.jp/ SCALE PLATE Minimum Divisions Maximum Divisions Bipolar Scale ■ WHAT MUST BE SPECIFIED WHEN ORDERING Following two methods can specify scale plate. a) Using ‘Scale Plate Designer’ Access ‘Design Scale Plate’ in the M-System web site (http://www.m-system.co.jp/english). Scale plate can be designed in this web site. By function below, it can be easy to create standard design or original design. [Design Automatically] Entering Minimum, Maximum, and Unit allows to create automatically a scale plate. Maximum created scale division number is ’50’. [Specify Division Interval] Division Interval can be specified according to the application. [Specify Division Number] It is available to create originally with scale division number, length of line, position, character size, font and detailed position. After designing is completed, register code is issued. Place the order with this code. Once scale plate is designed, it is recorded. The register code can be used any number of times. b) Specifying scale range and display unit when placing the order It is available to create by specifying scale range and display unit for right and left. Regarding design of scale plate such as division number, length of division number line, and character font, they are same as above [Design Automatically], M-System designs them. ■ DESIGNING BY ‘DESIGN AUTOMATIFICALLY’ How ‘Design Automatically’ creates scale design is described succinctly below. ■ TYPES OF DIVISIONS Five (5) types of divisions are used depending upon the scale span, which determined by the following equation: Scale Span = (Max. range value – Min. range value) 10n where n = integer (used to limit the calculated scale span to the minimum of 1.1, below 11.0.)
- Type 1: 1.1 ≤ Scale Span < 1.3 Number of divisions: 22 to 25.9 Scale: Starts at 0, increments in 0.02 / 0.2 / 2 / 20 / 200. Min. and max. values are indicated. 4 digits including negative sign and decimal point. Division lines: Long, Short, Medium, Short, Long (4 division lines repeating)
- Type 2: 1.3 ≤ Scale Span < 2.0 Number of divisions: 26 to 39.9 Scale: Starts at 0, increments in 0.03 / 0.3 / 3 / 30 / 300. Min. and max. values are indicated. 4 digits including negative sign and decimal point. Division lines: Long, Short, Medium, Short, Medium, Short, Long (6 divisions repeating) Minimum Divisions Maximum Divisions Bipolar Scale 0.2 0.4 0.6 0.8 1.2 1.29 600 400 200 -200 -400 -600 0.6 0.3 0.0 -0.3 -0.6 -0.8 0.8 0.3 0.6 0.9 1.2 1.5 1.8 1.99 120 130
MODEL: 48SV2 48SV2 SPECIFICATIONS ES-9315 Rev.3 Page 4/6 https://www.m-system.co.jp/
- Type 3: 2.0 Scale Span < 2.6 Number of divisions: 32 to 41.9 Scale: Starts at 0, increments in 0.05 / 0.5 / 5 / 50 / 500. Min. and max. values are indicated. 4 digits including negative sign and decimal point. Division lines: Long, Short, Medium, Short, Medium, Short, Medium, Short, Long (8 divisions repeating)
- Type 5: 5.5 Scale Span < 11.0 Number of divisions: 22 to 43.9 Scale: Starts at 0, increments in 0.01 / 0.1 / 1 / 10 / 100 / 1000. Min. and max. values are indi-cated. 4 digits including negative sign and decimal point. Division lines: Long, Medium, Medium, Medium, Long (5 divisions repeating)
- Type 4: 2.6 Scale Span < 5.5 Number of divisions: 26 to 43.9 Scale: Starts at 0, increments in 0.05 / 0.5 / 5 / 50 / 500. Min. and max. values are indicated. 4 digits including negative sign and decimal point. Division lines: Long, Medium, Medium, Medium, Long (4 divisions repeating) Minimum Divisions Maximum Divisions Bipolar Scale Minimum Divisions Maximum Divisions Bipolar Scale Minimum Divisions Maximum Divisions Bipolar Scale [Example] : Bargraph range 0 to 300 cm (Type 4) Scale range : 0 – 300 Scale unit (bargraph) : cm 120 100 -50 -100 -120 0.5 1.5 2.59 0.5 0.4 0.3 0.2 0.1 -0.1 -0.2 -0.3 -0.4 -0.5 100 200 300 400 500 550 10.9 250 200 150 100 -50 -100 -150 -200 -250 0.5 1.5 2.5 3.5 4.5 5.49 100 150 200 260 300 250 200 150 100 cm OR ZERO SPAN DOWN UP A1234 EXTERNAL VIEW Span adj. Indicator LED Adjustment button Adjustment button Zero adj. Indicator LED Span adj. Indicator LED ■VERTICAL MOUNTING ■HORIZONTAL MOUNTING 100 OR 80 60 40 20 0 %ZERO SPAN DOWN UP A1234 100 OR 0 % ZEROSPAN DOWN UP A1234
MODEL: 48SV2 48SV2 SPECIFICATIONS ES-9315 Rev.3 Page 5/6 https://www.m-system.co.jp/ EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm 100 OR 72 ±0.5 18 ±0.5 69 ±0.5 17.5±0.5 ■ SCREWLESS SPRING TERMINAL ZERO SPAN DOWN UP Approx.1696 ±0.54±0.5 69 ±0.5 17.5±0.5 ■ SEPARABLE SCREWLESS SPRING TERMINAL 100 OR 0 % 72 ±0.5 18 ±0.5 ZERO SPAN DOWN UP
MODEL: 48SV2 48SV2 SPECIFICATIONS ES-9315 Rev.3 Page 6/6 https://www.m-system.co.jp/ PANEL CUTOUT unit: mm Panel thickness: 0.5 - 2.0 Panel thickness: 0.5 - 2.0 Panel thickness: 0.5 - 2.0 Panel thickness: 0.5 - 2.0 Note 1. When mounting vertically, leave 3 cm or more of space above and below the unit. 1 cm or more of space to right adn left the unit. Note 2. When mounting horizontally, leave 3 cm or more of space above and below the unit. 2.5 cm or more of space to left and right of the unit.
- Clustered .PVOUJOH 69.5 + 1 – 0.2 69.5 + 1 – 0.2 28 min. + 0 – 0.2 + 0 – 0.2 18 + 0 – 0.2
- 4JOHMF.PVOUJOH
- Clustered Mounting • Single Mounting 69.5 97 min. + 1 – 0.2 69.5 + 1 – 0.2 – 0.2 69.5 + 1 – 0.2 – 0.2
- VERTICAL MOUNTING
- HORIZONTAL MOUNTING SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM Low Drift Amplifier Digital Computation Circuit Display Circuit INPUT * Input shunt resistor incorporated for current input. POWER Specifications are subject to change without notice.