MODEL: 47NLR 47NLR SPECIFICATIONS ES-9565 Rev.3 Page 1/4 https://www.m-system.co.jp/ Digital Panel Meters 47NL Series RTD INPUT DIGITAL PANEL METER (4 digit, LED display type) Functions & Features

  • 4 digit RTD input digital panel meter
  • Moving average function to suppress the display flickering
  • Max. and Min. value display
  • Photo MOSFET relay alarm output 48.0 (1.89) 96.0 (3.78) 20.0 (.79) mm (inch) Zro Spn D/P Tch Fnc Min Max H P L Alarm/ Scale/ Max/Min Shift Up MODEL: 47NLR-[1][2]-R[3]

ORDERING INFORMATION

  • Code number: 47NLR-[1][2]-R[3] Specify a code from below for each of [1] through [3]. (e.g. 47NLR-4R-R/Q)
  • Specify the specification for option code /Q (e.g. /SET) [1] INPUT 1: JPt 100 (JIS'89) 3: Pt 100 (JIS'89) 4: Pt 100 (JIS'97, IEC) 5: Pt 50 Ω (JIS'81) 7: Pt 1000 [2] DISPLAY COLOR R: Red YR: Orange G: Green B: Blue W: White POWER INPUT DC Power R: 24 V DC (Operational voltage range 24 V ±10 %, ripple 10 %p-p max.) [3] OPTIONS blank: none /Q: With options (specify the specification) SPECIFICATIONS OF OPTION: Q EX-FACTORY SETTING /SET: Preset according to the Ordering Information Sheet (No. ESU-9565) GENERAL SPECIFICATIONS Construction: Panel mount type Degree of protection: IP66; Applicable to the front of the panel meter mounted according to the specified panel cutout. Only screw mounting conforms. Connection: Screwless spring terminal Applicable wire size: 0.2 to 0.5 mm2, stripped length 6 mm Housing material: Flame-resistant resin (gray) Isolation: Input to alarm output to power Burnout: Upscale standard; downscale optional by programming Setting: (Front button)
  • Alarm setpoint
  • Deadband (hysteresis)
  • Moving average
  • Others (Refer to the operating manual for details) Averaging: None or moving average Lockout setting: Prohibiting certain operations; protecting settings DISPLAY Display: 16 mm (.63) high, 4 digits, 7-segment LED Display range: -9999 to 9999 Minimum display/setting scale: 0.1°C or 1°F Decimal point position: 10–1 or none Zero indication: Higher-digit zeros are suppressed. Over-range indication: ‘S.ERR’ and ‘Min’ or ‘Max’ blinking when the input signal is out of the usable range. Burnout indication: ‘B.ERR’and function status 'Min' or 'Max' are displayed and blinking. Alarm status indication L indicator: Green turns on when the L alarm is tripped. H indicator: Red turns on when the H alarm is tripped. P indicator: Amber turns on when none of the other alarms is tripped. Only ‘P’ turns on when no-alarm is selected with alarm setpoint. Function indicators: Zro, Spn, D/P, Tch, Fnc, Min, Max Display mode status and operation status, amber ON or blink

MODEL: 47NLR 47NLR SPECIFICATIONS ES-9565 Rev.3 Page 2/4 https://www.m-system.co.jp/ Engineering unit indication: Sticker label attached DC, AC, mV, V, kV, μA, mA, A, kA, mW, W, kW, var, kvar, Mvar, VA, Hz, Ω, kΩ, MΩ, cm, mm, m, m/sec, mm/min, cm/min, m/min, m/h, m/s2, inch, ℓ, ℓ/s, ℓ/min, ℓ/h, m3, m3/sec, m3/min, m3/h, Nm3/h, N·m, N/m2, g, kg, kg/h, N, kN, Pa, kPa, MPa, t, t/h, ℃, °F, %RH, J, kJ, MJ, rpm, sec, min, pH, %, ppm, etc. INPUT SPECIFICATIONS ■ RTD: 3-wire RTDs Maximum leadwire resistance: 60 Ω per wire Sensing current: JPt 100 (JIS '89): 0.5mA Pt 100 (JIS '89): 0.5mA Pt 100 (JIS '97, IEC): 0.5mA Pt 50 Ω (JIS '81): 0.5mA Pt 1000: 0.05mA Temperature Range RTD CONFORMANCE RANGE USABLE RANGE JPt 100 (JIS ’89) °C -200 to +500 -230 to +530 °F -328 to +932 -382 to +986 Pt 100 (JIS ’89) °C -200 to +650 -230 to +680 °F -328 to +1202 -382 to +1256 Pt 100 (JIS ’97 , IEC) °C -200 to +850 -230 to +880 °F -328 to +1562 -382 to +1616 Pt 50Ω (JIS ’81) °C -200 to +649 -230 to +679 °F -328 to +1202 -382 to +1256 Pt 1000 °C -200 to +850 -230 to +880 °F -328 to +1562 -382 to +1616 OUTPUT SPECIFICATIONS ■ Alarm Output: Photo MOSFET Relay Rating: 26.4 V DC @ 100 mA (resistive load) ON resistance: ≤ 5 Ω INSTALLATION Power consumption

  • DC: 0.7 W max. Operating temperature: -10 to +55°C (14 to 131°F) Operating humidity: 10 to 90 %RH (non-condensing) Mounting: Magnet or screw mounting Weight: 85 g (3.0 oz) PERFORMANCE Accuracy Display: ±1℃ ± 1 digit (±2°F ± 1 digit) Temp. coefficient: ±0.015 %/°C (±0.008 %/°F) Response time: ≤ 0.5 sec. (alarm output: 0 – 100 % at 90 % setpoint) Burnout response: ≤ 10 sec. Line voltage effect: ±0.1 % over voltage range Insulation resistance: ≥ 100 MΩ with 500 V DC Dielectric strength: 1500 V AC @ 1 minute (input to alarm output to power to ground) STANDARDS & APPROVALS EU conformity: EMC Directive EMI EN 61000-6-4 EMS EN 61000-6-2 RoHS Directive

MODEL: 47NLR 47NLR SPECIFICATIONS ES-9565 Rev.3 Page 3/4 https://www.m-system.co.jp/ EXTERNAL VIEW Alarm /Max /Min Scale / Shift Up H P L D/PSpnZro Tch Max Min Fnc (4) Max/Min Button (7) Shift Button (8) Up Button (1) Main Display (3) Function Indicators (2) Alarm Indicators (6) Scale/ Button (5) Alarm/ Button ■ COMPONENT IDENTIFICATION No. COMPONENT FUNCTION (1) Main display Indicates present values, setting values and status of the unit. (2) Alarm indicators Indicates the comparison result between alarm setting values and present values. (3) Function indicators Indicates the status in each setting mode. (4) Max/Min button Used to switch the main display to show the present values, maximum values or minimum values. (5) Alarm/ button Used to check the alarm setpoints, to move on to the alarm and other setting modes; or to shift through set- ting items in each setting mode. (6) Scale/ button Used to move on to the initial and other setting mo des; or to shift through setting items in each sett ing mode. (7) Shift button Used to move on to the setting standby status of each setting mode or to shift through display digits in each settingitem. (8) Up button Used to change setting values or to select setting values. Note: Refer to the operating manual for details on each function. EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch) 96.0 (3.78) 48.0 (1 .89) 20.0(.79) ■ FRONT VIEW ■ REAR VIEW ■ SIDE VIEW

  • SCREW MOUNTING • MAGNET MOUNTING Watertight packing 20.0(.79) Magnet label sheet 4-M4 SCREW 8.0 (.31) deep 1 2 3 4 5 6 7 8

MODEL: 47NLR 47NLR SPECIFICATIONS ES-9565 Rev.3 Page 4/4 https://www.m-system.co.jp/ MOUNTING REQUIREMENTS unit: mm (inch) ■ PANEL MOUNTING ɹ (Conform to degree of protection IP66) Panel cutout must be such as M-System specified. Use watertight packing and unit fixing screws included in the product package. 4-4.5 (.18) dia. 20.5 (.81) (.67) 30.5 (1 .20) dia.

  • Single Mounting unit: mm (inch) ■ MAGNET MOUNTING ɹ (Not conform to degree of protection IP66) Panel cutout is not required. Use magnet sheet and stopper included in the product package.
  • Clustered Mounting unit: mm (inch)
  • Clustered Mounting unit: mm (inch) 96.0 (3.78) 48.0 (1.89) min. 120 (4.72) min. 100 (3.94) min.120 (4.72) min. 75 (2.95) Panel thickness: 1.0 to 3.2 mm SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM A/D Converter Display / Setting 7 POWER Ry Ry H COM L L OUTPUT H OUTPUT A B B INPUT Digital Computation Specifications are subject to change without notice.