47NLNT MSYSTEM | Alldatasheet

Document overview

  • Manufacturer or author: Provided By ALLDATASHEET.COM(FREE DATASHEET DOWNLOAD SITE)
  • PDF pages: 5

Technical content

MODEL: 47NLNT 47NLNT SPECIFICATIONS ES-9564 Rev.4 Page 1/5 https://www.m-system.co.jp/ Digital Panel Meters 47NL Series LOOP POWERED DIGITAL PANEL METER (4½ digit, LED display type, with terminal block) Functions & Features

  • 4 ½ digit loop powered digital panel meter
  • No external power source needed
  • Moving average function to suppress the display flickering
  • Scaling, forced zero, low-end cutout
  • Max. and Min. value display
  • Dedicated terminal block for input 48.0 (1.89) 96.0 (3.78) 48.5 (1.91)* Zro Spn D/P Tch Fnc Min Max H P L ˣ Scale/ ˢ Max/Min Shift Up mm (inch) * Panel thickness not included. MODEL: 47NLNT-[1][2]

ORDERING INFORMATION

  • Code number: 47NLNT-[1][2] Specify a code from below for each of [1] and [2]. (e.g. 47NLNT-R/Q)
  • Specify the specification for option code /Q (e.g. /SET) INPUT Current 4 – 20 mA DC [1] DISPLAY COLOR R: Red G: Green [2] OPTIONS blank: none /Q: With options (specify the specification) SPECIFICATIONS OF OPTION: Q EX-FACTORY SETTING /SET: Preset according to the Ordering Information Sheet (No. ESU-9564) GENERAL SPECIFICATIONS Construction: Panel mount type Degree of protection: IP66; Applicable to the front of the panel meter mounted according to the specified panel cutout. Connection: M3 screw terminals (torque 0.6 N·m) Solderless terminal: Refer to the drawing at the end of the section. Recommended manufacturer: Japan Solderless Terminal MFG.Co.Ltd, Nichifu Co.,ltd Applicable wire size: 0.25 to 1.65 mm2 (AWG 22 to 16) Screw terminal: Nickel-plated steel Housing material: Flame-resistant resin (gray) Setting: (Front button)
  • Scaled range
  • Moving average
  • Others (Refer to the instruction manual for details) Averaging: None or moving average Lockout setting: Prohibiting certain operations; protecting settings ■Recommended solderless terminal 6 (.24) max 3.3 (.13) max mm (inch) DISPLAY Display: 16 mm (.63) high, 4 ½ digits, 7-segment LED Display range: -19999 to 19999 Decimal point position: 10–1, 10–2, 10–3, 10–4 or none Zero indication: Higher-digit zeros are suppressed. Over-range indication: ‘-19999’ or ‘19999’ blinking for display values out of the display range. ‘S.ERR’ and ‘Min’ or ‘Max’ blinking when the input signal is out of the usable range. Function indicators: Zro, Spn, D/P, Tch, Fnc, Min, Max Display mode status and operation status, ON or blink (Display color is the same as display color code.) Engineering unit indication: Sticker label attached DC, AC, mV, V, kV, μA, mA, A, kA, mW, W, kW, var, kvar, Mvar, VA, Hz, Ω, kΩ, MΩ, cm, mm, m, m/sec, mm/min, cm/min, m/min, m/h, m/s2, inch, ℓ, ℓ/s, ℓ/min, ℓ/h, m3, m3/sec, m3/min, m3/h, Nm3/h, N·m, N/m2, g, kg, kg/h, N, kN, Pa, kPa, MPa, t, t/h, ℃, °F, %RH, J, kJ, MJ, rpm, sec, min, pH, %, ppm, etc.

MODEL: 47NLNT 47NLNT SPECIFICATIONS ES-9564 Rev.4 Page 2/5 https://www.m-system.co.jp/ INPUT SPECIFICATIONS ■ DC Current Current range: Approx. 3.7 to 23 mA DC Voltage drop: Approx. 6.3 V with 4 mA; approx. 6.5 V with 20 mA (Equivalent input impedance: Approx. 325 Ω) (There is voltage drop generated by using the unit. For the two-wire transmitter power supply, make sure that the operable voltage for the two-wire transmitter is ensured including the voltage drop by other devices and wiring resistance.) INSTALLATION Operating temperature: -10 to +55°C (14 to 131°F) Operating humidity: 10 to 90 %RH (non-condensing) Mounting: Screw mounting Weight: 115 g (0.25 lb) PERFORMANCE Accuracy Display: ±0.1 % ± 1 digit Temp. coefficient: ±0.015 %/°C (±0.008 %/°F) STANDARDS & APPROVALS EU conformity: EMC Directive EMI EN 61000-6-4 EMS EN 61000-6-2 RoHS Directive

MODEL: 47NLNT 47NLNT SPECIFICATIONS ES-9564 Rev.4 Page 3/5 https://www.m-system.co.jp/ EXTERNAL VIEW Max /Min Scale / Shift Up D/PSpnZro Tch Max Min Fnc (1) Main Display (2) Function Indicators (3) Max/Min Button (4) Button (7) Up Button (6) Shift Button (5) Scale/ Button ■ COMPONENT IDENTIFICATION No. COMPONENT FUNCTION (1) Main display Indicates present values, setting values and status of the unit. (2) Function indicators Indicates the status in each setting mode. (3) Max/Min button Used to switch the main display to show the present values, maximum values or minimum values. (4) Button Used to shift through setting items in each setting mode etc. (5) Scale/ button Used to move on to the scaling setting and other setting modes; or to shift through setting items in each set- ting mode. (6) Shift button Used to move on to the setting standby status of each setting mode or to shift through display digits in each setting item. (7) Up button Used to change setting values, to execute/cancel Forced Zero or to select setting values. Note: Refer to the operating manual for details on each function. EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch)

  • FRONT VIEW • SIDE VIEW ■ BODY
  • REAR VIEW Watertight packing Harness 4-M4 SCREW 8.0 (.31) deep approx. 30 (1 .18) 48.0 (1 .89) 1 2 3 4 5 6 7 8

MODEL: 47NLNT 47NLNT SPECIFICATIONS ES-9564 Rev.4 Page 4/5 https://www.m-system.co.jp/ 96.0 (3.78) 34 (1 .34) (.31) Max. 20.5 (.81) 6.4 (.25) (.08) 2–4.5 (.18) dia.

  • FRONT VIEW
  • SIDE VIEW ■ TERMINAL BLOCK
  • REAR VIEW 1 2 3 4 5 6 7 8 9 10 MOUNTING REQUIREMENTS unit: mm (inch)
  • Single Mounting • Clustered Mounting 4-4.5 (.18) dia. 20.5 (.81) (.67) min. 75 (2.95) Panel thickness: 1.0 to 3.2 mmPanel thickness: 1.0 to 3.2 mm

MODEL: 47NLNT 47NLNT SPECIFICATIONS ES-9564 Rev.4 Page 5/5 https://www.m-system.co.jp/ SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM Controllor, etc. 4 – 20mA DC INPUT ■ 4-WIRE CONNECTION EXAMPLE I/P Positioner, etc. Two-wire Transmitter – 2 INPUT Power Source Digital Computation A/D Converter Display / Setting Specifications are subject to change without notice.