MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 1/8 https://www.m-system.co.jp/ Digital Panel Meters 47 Series THERMOCOUPLE INPUT DIGITAL PANEL METER (4-digit, LED display type) Functions & Features
- 4-digit thermocouple input digital panel meter
- 1/8 DIN size
- Moving average function to suppress the display flickering
- Max. and Min. value display
- IP66 front panel
- Separable terminal block
- Rear terminal cover for safety strapped to the meter Alar m/ Init/ Max/Min Shift Up Ini Alm D/P Adj Fnc Min Max HH H P L LL (1.89) mm (inch) 96 (3.78) 98.5 (3.88)
ORDERING INFORMATION
- Code number: 47LT-1[1][2][3]-[4][5] Specify a code from below for each of [1] through [5]. (e.g. 47LT-101G-M2/Q)
- Specify the specification for option code /Q (e.g. /C01/S01/SET) INPUT – Field-selectable 1: (PR), K (CA), E (CRC), J (IC), T (CC), B (RH), R, S, N [1] DC OUTPUT 0: Without Current A: 4 – 20 mA DC (Load resistance 550 Ω max.) D: 0 – 20 mA DC (Load resistance 550 Ω max.) Voltage 4: 0 – 10 V DC (Load resistance 10 kΩ min.) 5: 0 – 5 V DC (Load resistance 5000 Ω min.) 6: 1 – 5 V DC (Load resistance 5000 Ω min.) 4W: -10 – +10 V DC (Load resistance 10 kΩ min.) [2] ALARM OUTPUT 0: None 1: N.O. relay contact, 4 points 2: SPDT relay contact, 2 points [3] DISPLAY COLOR R: Red YR: Orange G: Green BG: Bluegreen B: Blue W: White [4] POWER INPUT AC Power M2: 100 – 240 V AC (Operational voltage range 85 – 264 V, 50/60 Hz) DC Power R: 24 V DC (Operational voltage range 24 V ±10 %, ripple 10 %p-p max.) P: 110 V DC (Operational voltage range 85 – 150 V, ripple 10 %p-p max.) [5] OPTIONS blank: none /Q: With options (specify the specification) SPECIFICATIONS OF OPTION: Q (multiple selections) COATING (For the detail, refer to M-System's web site.) Moving parts and indicators are not coated. /C01: Silicone coating /C02: Polyurethane coating /C03: Rubber coating TERMINAL SCREW MATERIAL /S01: Stainless steel EX-FACTORY SETTING /SET: Preset according to the Ordering Information Sheet (No. ESU-9512) GENERAL SPECIFICATIONS Construction: Panel flush mounting Degree of protection: IP66; applicable to the front of the panel meter mounted according to the specified panel cutout Connection: M3 separable screw terminal (torque 0.6 N·m) Screw terminal: Nickel-plated steel (standard) or stainless steel Housing material: Flame-resistant resin (gray) Isolation: Input to DC output to HH output or H output to L output or LL output to power Burnout: Upscale standard; downscale optional by programming Linearization: Standard Cold junction compensation: CJC sensor attached to the input terminals Setting: (Front button)
MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 2/8 https://www.m-system.co.jp/
- Input type
- Alarm setpoint
- Hysteresis (deadband)
- Others (Refer to the instruction manual for details) Read rate: 40 times/sec. (25 msec.) Averaging: None or moving average Lockout setting: Prohibiting certain operations; protecting settings DISPLAY Display: 16 mm (.63) high, 4 digits, 7-segment LED Display range: -1999 to 9999 Minimum display/setting scale: 1°C or 1°F (0.1°C for K and J, narrow span) Decimal point position: 10–1 or none (fixed depending upon the input type) Zero indication: Higher-digit zeros are suppressed. Over-range indication: ‘S.ERR’ and ‘Min’ or ‘Max’ blinking when the input signal is out of the usable range. Burnout indication: ‘B.ERR’ blinking Alarm status indication LL indicator: Green turns on when the LL alarm is tripped. L indicator: Green turns on when the L alarm is tripped. H indicator: Red turns on when the H alarm is tripped. HH indicator: Red turns on when the HH alarm is tripped. P indicator: Amber turns on when none of the other alarms is tripped. Note: Only ‘P’ turns on with no-alarm-output type. ‘LL’ or ‘HH’ does not turn on with dual-alarm-output type. All setpoints can be independently set either for Hi or Lo alarm trip. Function indicators: Ini, Alm, D/P, Adj, Fnc, Min, Max Display mode status and operation status, amber ON or blink Engineering unit indication: Sticker label attached DC, AC, mV, V, kV, μA, mA, A, kA, mW, W, kW, var, kvar, Mvar, VA, Hz, Ω, kΩ, MΩ, cm, mm, m, m/sec, mm/min, cm/min, m/min, m/h, m/s2, inch, ℓ, ℓ/s, ℓ/min, ℓ/h, m3, m3/sec, m3/min, m3/h, Nm3/h, N·m, N/m2, g, kg, kg/h, N, kN, Pa, kPa, MPa, t, t/h, ℃, °F, %RH, J, kJ, MJ, rpm, sec, min, min-1, pH, %, ppm, etc. INPUT SPECIFICATIONS Input resistance: 1 MΩ minimum Burnout sensing: ≤ 0.1 μA Default input type: K (CA) Temperature Range THERMOCOUPLE CONFORMANCE USABLE RANGE RANGE (PR) °C 0 to 1700 -30 to +1730 °F 32 to 3092 -22 to +3146 K (CA) °C -200 to +1300 -230 to +1330 °F -328 to +2372 -382 to +2426 K (CA), narrow span °C -100.0 to +450.0 -199 .9 to +500.0 E (CRC) °C 0 to 600 -30 to +630 °F 32 to 1112 -22 to +1166 J (IC) °C -100 to +850 -130 to +880 °F -148 to +1562 -202 to +1616 J (IC), narrow span °C -100.0 to +350.0 -150 .0 to +400.0 T (CC) °C -200 to +400 -230 to +430 °F -328 to +752 -382 to +806 B (RH) °C 400 to 1800 70 to 1830 °F 752 to 3272 158 to 3326 R °C 200 to 1700 -30 to +1730 °F 392 to 3092 -22 to +3146 S °C 0 to 1700 -30 to +1730 °F 32 to 3092 -22 to +3146 N °C -200 to +1300 -230 to +1330 °F -328 to +2372 -382 to +2426 OUTPUT SPECIFICATIONS ■ DC Output
- DC Current Operational range: -5 – +105 %
- DC Voltage Operational range: -5 – +105 % ■ Alarm Output: Relay contact Rated load: 250 V AC @ 3 A (cos ø = 1)
30 V DC @ 3 A (resistive load)
Maximum switching voltage: 250 V AC, 30 V DC Maximum switching power: 750 VA, 90 W (resistive load) Minimum load: 5 V DC @ 10 mA Mechanical life: ≥ 5 × 106 cycles (rate 180 cycles/min.) INSTALLATION Power consumption
- AC: Approx. 6.5VA
- DC: Approx. 3 W Operating temperature: -10 to +55°C (14 to 131°F) Operating humidity: 30 to 90 %RH (non-condensing) Mounting: Panel flush mounting Weight: 300 g (0.66 lb) PERFORMANCE Accuracy Display: ±1°C (or ±1.8°F) ±1 digit (PR, K, E, J, T) ±2°C (or ±3.6°F) ±1 digit (B, R, S, N)
MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 3/8 https://www.m-system.co.jp/ Output: ± 0.15 % (DC output = display + output) Cold junction compensation error: ±3°C at 25 ±10°C ±5.4°F at 77 ±18°F Temp. coefficient: ±0.015%/°C (±0.008%/°F) when the input range is set to the maximum span. Input resolution: Max. 16 bits Output resolution: Max. 13 bits Response time: ≤ 0.5 sec. (alarm output: 0 – 100 % at 90 % setpoint) ≤ 0.5 sec. (DC output: 0 – 90 %) Burnout response: ≤ 15 sec. Line voltage effect: ±0.1 % over voltage range Insulation resistance: ≥ 100 MΩ with 500 V DC Dielectric strength: 2000 V AC @ 1 minute (input to DC output to HH output or H output to L output or LL output to power to ground) STANDARDS & APPROVALS EU conformity: EMC Directive EMI EN 61000-6-4 EMS EN 61000-6-2 Low Voltage Directive EN 61010-1 Measurement Category II (alarm output) Installation Category II (power) Pollution degree 2 Input or DC output to alarm output to power: Reinforced insulation (300 V) Input to DC output: Basic insulation (300 V) RoHS Directive Protection against access to the terminal blocks: Finger protection (VDE 0660-514)
MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 4/8 https://www.m-system.co.jp/ EXTERNAL VIEW (4) Max/Min Button (5) Alarm/ Button (6) Init/ Button (7) Shift Button (8) Up Button (1) Main Display (2) Alarm Indicators (3) Function Indicators Alarm/ Init/Max/Min Shift Up Ini HH H P L LL Alm D/P Adj Fnc Min Max
- COMPONENT IDENTIFICATION No. COMPONENT FUNCTION (1) Main display Indicates present values, setting values and status of the unit. (2) Alarm indicators Indicate alarm status of the input signal. (3) Function indicators Indicate setting modes and status. (4) Max/Min button Used to switch the main display to show present values, maximum values or minimum values etc. (5) Alarm/ button Used to con/f_i rm alarm setting value and to move on to the alarm and other setting modes; or to shift through setting items in each setting mode. (6) Init/ button Used to move on to the initial and other setting modes; or to shift through setting items in each setting mode. (7) Shift button Used to move on to the setting standby status and shift through display digits in each setting mode. (8) Up button Used to choose setting value in each setting mode. Note: Refer to the operating manual for details on each function.
MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 5/8 https://www.m-system.co.jp/ EXTERNAL DIMENSIONS & TERMINAL ASSIGNMENTS unit: mm (inch) 96 (3.78) 48 (1.89) 103 (4.06) 95 (3.74) 2 (.08) 47 (1.85) 98.5 (3.88) 45 (1.77) 91.5 (3.60) 4–M3 SCREW ■ REAR VIEW
- No Alarm Output • Alarm Output ■ FRONT VIEW ■ SIDE VIEW ■ TOP VIEW 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 20–M3 SCREW TERMINAL 2–M3 SCREW 10–M3 SCREW TERMINAL 6.6 (.26)
MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 6/8 https://www.m-system.co.jp/ MOUNTING REQUIREMENTS unit: mm + 0.8 – 0 + 0.6 – 0 min. 120 min. 75 Panel thickness: 1.6 to 8.0 mm
MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 7/8 https://www.m-system.co.jp/ SCHEMATIC CIRCUITRY & CONNECTION DIAGRAM POWER U(+) V(–) LOAD Inductive Load (Coil) Varistor or CR Circuit
- DC Powered LOAD Inductive Load (Coil) Diode, Varistor or CR Circuit ■ Relay Protection
- AC Powered ■ ALARM OUTPUT CODE 1: N.O. contact, 4 points Ry Ry Ry Ry POWER U(+) Lc (COM ) Lb (NC) La (NO) Hc (COM ) Hb (NC) Ha (NO) V(–) ■ ALARM OUTPUT CODE 2: SPDT contact, 2 points Ry Ry Lc, LLc LLa La Hc, HHc Ha HHa LL OUTPUT L OUTPUT H OUTPUT L OUTPUT H OUTPUT HH OUTPUT POWER U(+) V(–) ■ ALARM OUTPUT CODE 0: No alarm output Digital Computation Display / Setting A/D Converter T/C comp. leadwire CJC SENSOR DC OUTPUT Isolation Output Driver Digital Computation Display / Setting A/D Converter T/C comp. leadwire CJC SENSOR DC OUTPUT Isolation Output Driver Digital Computation Display / Setting A/D Converter T/C comp. leadwire CJC SENSOR DC OUTPUT Isolation Output Driver *DO NOT loosen the screw of the terminal 3. Loosen only the terminals 1 – 2 when connecting extension wires. Note: The section enclosed by broken line is only with DC output option. U
MODEL: 47LT 47LT SPECIFICATIONS ES-9512 Rev.11 Page 8/8 https://www.m-system.co.jp/ Specifications are subject to change without notice.